1 2 3 4 Optimization of automated gas sample collection and IRMS analysis of δ 13C of CO2 in air Matthias J. Zeeman1 , Roland A. Werner1 , Werner Eugster1 , Rolf T. W. Siegwolf2 , Günther Wehrle2 , Joachim Mohn3 , Nina Buchmann1 1 Institute 8 of Plant Sciences, ETH Zurich, Universitaetsstrasse 2, CH–8092 Zurich, Switzerland of Atmospheric Chemistry, Paul Scherrer Institute, Bachstrasse 1, CH–5232 Villingen, Switzerland 3 Laboratory for Air Pollution & Environmental Technology, Empa, Überlandstrasse 129, CH–8600 Dübendorf, Switzerland 9 Abstract 5 6 7 10 11 12 13 14 15 16 17 2 Laboratory The application of 13 C/12 C in ecosystem–scale tracer models for CO2 in air requires accurate measurements of mixing ratios and stable isotope ratios of CO2 . To increase measurement reliability and data intercomparability as well as to shorten analysis times, we have improved an existing field sampling setup with portable air sampling units and developed a laboratory setup for analysis of δ 13 C of CO2 in air by isotope ratio mass spectrometry (IRMS). The changes consist of (a) optimization of sample and standard gas flow paths, (b) additional software configuration and (c) automation of liquid nitrogen refilling for the cryogenic trap. We achieved a precision better than 0.1 and an accuracy of 0.11±0.04 for δ 13 C of CO2 in air and unattended operation of measurement sequences up to 12 hours. 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 The interest in the global atmospheric carbon cycle has intensified as a response to reported trends in global climate change. These trends are primarily related to atmospheric increases in greenhouse gas concentrations. 1 On the global average, carbon dioxide (CO2 ) plays the most important role and thus ecosystem oriented research has particularly focused on CO2 . The potential use of the stable isotope ratios of CO2 (e.g. 13 C/12 C, 18 O/16 O) in ecosystem–scale atmosphere–biosphere process studies has often been highlighted and is believed to be a powerful tool for carbon cycle studies, in particular to disentangle ecosystem flux components.e.g. 2–7 It is commonly used to quantify mixing contributions from sources with differing isotopic compositions. 8,9 However, this requires accurate measurements of both CO2 mixing ratios and isotopic composition in order to be useable in ecosystem–scale tracer model approaches. 4,10,11 On local (species 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 ∗ Correspondence to ecosystem) scales this can be quite a challenge; CO2 mixing ratios and isotopic composition in the air close to the vegetation are known to fluctuate strongly, i.e. on short time scales (seconds to hours), especially under less turbulent atmospheric conditions due to accumulation of CO2 . Moreover, with conventional flask sampling the measurement strategy is mostly limited to discrete sampling, and typically these samples need to be transferred to a distant laboratory for analysis by an Isotope Ratio Mass Spectrometer, so the insight into ecosystem processes is hampered by technical and logistical constraints. In this paper, we aim to optimize and extensively test air sampling and analysis of stable carbon and oxygen isotope ratios in atmospheric CO2 for stable isotope studies at the ecosystem level. The setup described here has been successfully used for grassland ecosystem studies in Switzerland and intercomparisons of stable iso- to: MJ Zeeman, ETH Zurich, Institute of Plant Sciences, Universitaetsstrasse 2, CH–8092 Zurich, Switzerland, Email: matthias.zeeman@ipw.agrl.ethz.ch, Phone: +41 44 632 81 96, Fax: +41 44 632 11 53 1 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 tope ratio instrumentation (e.g. a comparison of a quantum cascade laser based absorption spectrometer, a field-deployable Fourier transform infrared spectrometer and an Isotope Ratio Mass Spectrometer).e.g. 12–14 The basic considerations for the chosen measurement approach haves been (a) the collection of samples at multiple locations for (b) sample measurements by laboratory based high precision Isotope Ratio Mass Spectrometer. The most important implication of this approach is that the conditions (e.g. temperature, pressure) might be different between location of sample collection and the laboratory. Thus, gas samples might be contaminated during the storage period between sampling and analysis, which is especially likely if samples are collected at higher altitudes under reduced ambient pressure. 15–18 For time series analysis, e.g. to understand diurnal cycles or effects of weather events, samples or series of samples are repetitively collected at equally spaced time intervals. If a Keeling plot approach (inverse [CO2 ] related to isotope δ –value) is used, the accuracy of y–axis intercepts is directly related to the precision and accuracy of the measurements.cf. 4 Thus, the analysis must be as accurate and precise as possible, deviations should be on the order of 0.1 for δ 13 C at most. To achieve our aims, we have substantially improved existing gas sampling equipment previously described by Theis et al. 19 and developed a new Isotope Ratio Mass Spectrometer setup, programming and measurement routines for δ 13 C of CO2 in air. An overview of this improved setup is shown in Figure 1 for both field and laboratory setup. An important part of these improvements was to optimize the automation of the operations during sampling and isotope ratio analysis to allow for accurate timings and increased reproducibility. Thus, our objectives were to 1) apply digital communication protocols between the sampling unit and the control computer to store status information from the sampling unit in order to eliminated the potential error of sample misidenfication.cf. 20 2) We wanted to increase precision and reliability of the IRMS measurements for CO2 in air samples and optimize sample preparation steps. 3) We wanted to reduce the time required per IRMS analysis of a CO2 in air sample to increase throughput in the laboratory and reduce storage times of the samples. 109 Methodology 110 Field setup 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141 142 143 144 145 146 147 148 149 150 151 152 153 154 155 156 2 Three devices are used in our field setup (Fig. 1), consisting of an home-built air inlet selection unit, followed by an InfraRed Gas Analyzer (IRGA) for CO2 mixing ratios (model LI-840, LI-COR, Lincoln, Nebraska, USA) and a sample manifold at the end. This sample manifold is a modified and improved version of the device termed Automated Sampler of Air (ASA) by Theis et al. 19 . It contains 33 glass flasks sample containers connected to three multiport Valco-valves (EMTMA2ST12MWE, VICI, Schenkon, Switzerland) allowing independent filling of each individual sample container with sample air. We continue to use the abbreviation “ASA” to refer to the portable air sampling unit described here, because its key components (the Valvo-valves) and its function as sample manifold have not changed with respect to the Theis et al. 19 version, despite the modifications described here. During field deployment, a single inlet is selected from a series of continuously purged air inlets (Synflex™Type 1300, formerly known as Dekabon™, Gembloux SA/NV, Belgium; ID 4 mm, ≈ 1 L min−1 ). After a particle filter (Gelman, LI-COR), a T-split diverts the airflow (a) to the IRGA and a subsequent small pump (DC12/8FK, Fürgut GmbH, Germany) inside the inlet selection unit, and (b) to an ASA sample inlet. The flow through the IRGA is kept at a continuous rate of 0.9 L min−1 (Fig. 1), within the manufacturer supplied specifications for the IRGA. Once inside the ASA (Fig. 2), the sample air is pushed by a pump, diverted on activation of a solenoid valve (EVT307-5D0-02F-Q, SMC, Weisslingen, Switzerland) through a drying column containing magnesium perchlorate (Fluka, Switzerland) and is filtered (SS-4FW-2, Swagelok, USA) before being pushed further through 300 mL glass flasks (Ernst Keller & Co AG, Basel, Switzerland) or 10 mL stainless steel loops (SL10KSTP, VICI, Schenkon, Switzerland) connected to the multiport Valco-valves with ≈ 0.9 L min−1 . At one of the 12 positions of each Valco-valve a short stainless steel capillary is used as low volume bypass to allow aligning multiple Valco-valves in series, typically three or four per ASA. To create an over-pressure 157 158 159 160 161 162 163 164 165 166 167 168 169 170 171 172 173 174 175 176 177 178 179 180 181 182 183 184 185 186 187 188 189 190 191 192 193 194 195 196 197 198 199 200 201 202 203 204 205 206 of at least 50 kPa in the sample containers, we changed the original Theis et al. 19 design and the position of the Teflon membrane pump (N811KDC, KNF, Germany) in combination with a poppet check valve (SS-6C-MM-1, Swagelok, USA) and an adjustable flow meter (V-100, Vögtlin, Switzerland) at the exit. By having pressurized sample containers, the chance of contamination during transport and laboratory analysis is minimized. In the laboratory (Zurich, 400 m above sea level, a.s.l.), the pressure excess (pressure above ambient) directly after the pump inside the ASA (Fig. 2) was typically found to be ≈ 90 kPa and ≈ 50 kPa before the adjustable flow meter. This is sufficient for the collection of samples at alpine locations (e.g. >2000 m a.s.l.), though higher pressures could be reached at the expense of flow rates. Although multiple ASAs can be used in series for sampling by using the bypass position of the solenoid valves, we have chosen a parallel setup utilizing a flow split with poppet check valves (SS6CA-MM-3, Swagelok, USA) to prevent any backflow from ASAs with inactive pumps or from the open connections after removal of one ASA. The position of the IRGA (Fig. 1) parallel to the ASA has shown no discernible different results. As an advantage over a sequential setup our parallel version allows to continue the concentration measurements of the sample gas even if the ASAs are disconnected or inactive. The IRGA can alternatively be positioned directly before the ASA to directly analyze the gas that subsequently flows through the sample containers of the ASA. The inlet selection unit and ASAs are configured and operated by a field computer via RS232 serial communication lines connected to digital controllers (C-Control I STATION 2.0, Conrad Electronics GmbH, Germany), programmed to operate the rotation valves, solenoid valve, pump, and digital flow meter, and to provide status information for later use in post-processing of CO2 concentration and stable isotope data. The keypad and LCD display of the digital controllers are used to confirm correct operation of the devices before and during field deployment. The serial communication, data storage and post processing are handled by scripts written in Perl language. After having completed an in situ field sampling sequence, the ASAs with up to 33 or 44 gas sample containers per ASA are transported back to 208 the laboratory for subsequent (same day) isotope ratio analysis. 209 Laboratory setup 207 210 211 212 213 214 215 216 217 218 219 220 221 222 223 224 225 226 227 228 229 230 231 232 233 234 235 236 237 238 239 240 241 242 243 244 245 246 247 248 249 250 251 252 253 254 3 The precise determination of δ 13 C and δ 18 O values in CO2 of large numbers of air samples implies a precise and reproducible sampling technique as well as an automated and easy-to-use coupling of the sample containers (glass flasks or steel loops) to the Isotope Ratio Mass Spectrometer (Delta plus XP, Finnigan MAT, Bremen, Germany). A series of multiposition valves are used for the flow path of the sample preparation (Fig. 3). A 6-position dead-end path Valco-valve (ASD6MWE,VICI, Schenkon, Switzerland) and a 4-port 2-position Valco-valve (AC4UWE, VICI, Schenkon, Switzerland) allow the alignment of up to four independent reference air gas bottles (laboratory air gas cylinder with different CO2 mixing ratios and δ -values) or helium, using the same sample preparation path as the gas sampled with the ASA allowing referencing according to the Identical Treatment (IT) principle. 21 A feed capillary delivers pure He to the ASA (Fig. 3, valves 1 and 2), allowing a pressure build-up in the glass flasks that flushes the sample gas at a rate of about 5 mL min−1 through a water trap (Nafion dryer) to the cryogenic focus trap where condensable gases (mainly CO2 and N2 O) are cryogenically trapped. After diverting the non-condensable gases to a vent (Fig. 3, valve 4), the cryogenically trapped sample is thawed and subsequently flushed by He into the Gas Chromatograph column (Poraplot Q 25 m × 320 nm i.d., Varian, Walnut Creek, USA, held at 24 ◦ C) to allow separation of CO2 from N2 O and is subsequently led to the Isotope Ratio Mass Spectrometer for analysis (Fig. 3, valve 5). The trapping efficiency was checked beforehand with an IRGA (LI-840, LI-COR, Lincoln, Nebraska, USA) behind the frozen cryogenic trap. In contrast to Theis et al. 19 , who used a Precon (Finnigan MAT) hooked up to the Isotope Ratio Mass Spectrometer, we modified the Gasbench II system (Finnigan MAT) to directly interface with individual ASA units. This modification of the Gasbench (Fig. 3, bottom panel) comprises the replacement of the “Gas Chromatograph”-type split 22 by a ConFloIIIlike split 23 and the replacement of the stainless 255 256 257 258 259 260 261 262 263 264 265 266 267 268 269 270 271 272 273 274 275 276 277 278 279 280 281 282 283 284 285 286 287 288 289 290 291 292 293 294 295 296 297 298 299 300 301 302 303 steel sample loop with a home-built cryogenic focus trap (1/16" stainless steel capillary filled with Ni-wire) at the 8-port valve inside the Gasbench, which is configured to operate as 6-port valve (Fig. 3, valve 4). A second 4-port 2-position Valcovalve (AC4UWE, VICI, Schenkon, Switzerland) inside the sample preparation path (Fig. 3, valve 3) operates as a vent to release the pressure excess inside the sample containers and allows for high flow purging of the sample preparation path (10 mL min−1 ) and the Isotope Ratio Mass Spectrometer flow path (17 mL min−1 ) with pure He. In our system without Precon this would otherwise not have been possible. Without the posibility to flush with high flow (50 mL min−1 ) as shown by Theis et al. 19 , our measurement time would have been 835 s. With the help of the pressure vent and related high He flows we are able to reduce the analysis time per sample to 610 s, a period comparable to Theis et al. 19 . The cryogenic trap and all valves in the Gasbench, the external referencing unit, the rotary valve systems of the ASA and an automated liquid nitrogen refill procedure are computer controlled by modified Isodat script language (ISL) scripts, available in the vendor supplied ISODAT NT software package (Ver. 2.0 SP2.63, Finnigan MAT). To avoid overloading of the cryogenic trap with sample gas of high CO2 concentration (> 1000 µ mol mol−1 ) and to circumvent a possible non-linearity of the Gasbench and Isotope Ratio Mass Spectrometer combination with signal strength a , the signal strength of each sample is adjusted to be close to that of the Isotope Ratio Mass Spectrometer reference by changing the cryogenic trapping period depending on the sample concentration. We first tested the relation between the sample CO2 concentration and the cryogenic trapping period empirically for each of the different sample container volumes and tube lengths, determined the best fit (Fig. 4) and tested the results with known dilutions of a CO2 in air mixture of a known stable isotope composition. Furthermore, to improve stable conditions for the cryogenic trap, the liquid nitrogen (LN2 ) level in the Dewar was kept within the 95–100% range, either by manual refill or automated refill. To facilitate partly unattended operation (12 hours) and thus a certain level of autonomy of this setup, we installed a LN2 rea The effect should be < 0.06V −1 337 fill system that consists of a balance measuring the weight loss of the evaporating LN2 from the Dewar in combination with timing signals received from the Isotope Ratio Mass Spectrometer (Fig. 5) and that is supplied by a 30 L LN2 tank (up to 48 hours operation). Empirically, de-icing of the Dewar is required every 12 hours. As a last step, the ISODAT NT software configuration had to be adjusted to our modified setup, in particular the measurement timing schedule (Fig. 6). First, the sample preparation steps of flushing the capillaries with He and sample gas through the input lines before the gasbench, which is conventionally done between two measurements using a “pre-script”, is scheduled during and in parallel with the IRMS analysis for the next measurement. A second important modification that helped to save time was achieved by switching to high He flow rates while purging the input line capillaries before and in the Gasbench as described above. Third, changes to the ISODAT NT software configuration for the Gasbench allowed for variably timed operations parallel to the measurement of the reference standards (multitasking) within the “chromatography” part of the measurement time schedule. Thus, the cryogenic trapping period could be varied within the chromatogram sequence without disturbing other tasks (e.g. the measurement of the reference standards) and was no longer required to be executed before the chromatogram (e.g. in a pre-script). The combination of the described modifications allowed to shorten the IRMS measurement time per sample and increased the number of samples analyzed per day. 338 Isotope ratio analysis 304 305 306 307 308 309 310 311 312 313 314 315 316 317 318 319 320 321 322 323 324 325 326 327 328 329 330 331 332 333 334 335 336 The carbon and oxygen isotopic composition of the CO2 is expressed as the relative difference of its isotope abundance ratio relative to that of an international standard. This difference, usually expressed in per mill, is defined as ¸ (13 C/12 C)Sample − 1 · 103 δ C[]V–PDB = 13 12 ( C/ C)V–PDB · 13 · δ 18 O[]V–PDB–CO2 = (1) ¸ (18 O/16 O)Sample − 1 · 103 (18 O/16 O)V–PDB–CO2 (2) for δ 13 C for the reference standards, according to the vendor supplied instruction manual. 4 339 340 341 342 343 344 345 346 347 348 349 350 351 352 353 354 355 356 357 358 359 360 361 Post-run off-line calculation and drift correction for assigning the final δ 13 C and δ 18 O values on the V-PDB (Vienna PeeDee Belemnite) and V-PDB-CO2 scaleb were done following the IT principle as described by Werner and Brand 21 . The δ 13 C and δ 18 O values of the laboratory air standards (Zurich CO2 -in-air standards) were determined at the Max-Planck-Institut für Biogeochemie (MPI-BGC, Jena, Germany) according to Werner et al. 25 . The accurate assignment of the corresponding δ -values on the V-PDB and the VPDB-CO2 scale was performed in Jena by measuring the Zurich CO2 -in-air standards versus the “Jena-Reference AirSet” (J-RAS) as standard reference material (SRM).e.g. 24 Any isotope ratio data presented in this article are reported in [] deviation from V–PDB and V–PDB–CO2 for 13 C and 18 O, respectively. Typically several measurements of a laboratory reference standard are placed at the beginning and at the end of each measurement series, for post-calculation of corrections. Quality control (QC) standards are used to evaluate this correction procedure. 383 384 385 386 387 388 389 390 391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 362 363 364 365 366 367 368 369 370 371 372 373 374 375 376 377 378 379 380 381 382 Results and discussion After ensuring linearity, we tested the effects of the variable cryogenic trapping period of our ASA– Gasbench–Isotope Ratio Mass Spectrometer setup on δ 13 C measurements for a range of CO2 concentrations and for different sample containers used (glass or metal). Furthermore, we tested the performance (precision, accuracy) of the δ 13 C measurements for typical use of the described ASA– Gasbench–Isotope Ratio Mass Spectrometer setup. Linearity tests with gases of different CO2 in air or He mixing ratios have shown a strong relationship between the IRMS peak amplitude and the offset between the δ 13 C (or δ 18 O) of CO2 in a sample and its δ 13 C (or δ 18 O) reference value (Fig. 7). For δ 13 C, this offset increases strongly with lower relative peak amplitude. We suppose the origin of this effect is the signal to noise ratio of the analysis. Thus, to ensure measurement intercomparability, it is required to correct for this effect, e.g. by optimizing the peak amplitudes of the 407 408 409 410 411 412 413 414 415 416 417 418 419 420 421 422 423 424 425 426 427 428 b The samples to a limited range close to the amplitudes of the Isotope Ratio Mass Spectrometer reference gas. This optimization in effect means that the cryogenic trap should freeze the same amount of CO2 for each sample, independent of sample container and capillary volumes. Due to differences in pressure build-up as function of container volume, the relationships between CO2 concentration of the sample and the trapping period required for a peak amplitude close to the reference have to be determined empirically (cf. Fig. 4, Table 1) and were thus tested by analyzing a broad range of dilutions of a CO2 in air mixture with CO2 free air (Fig. 8, top panel). The resulting relative amplitudes for this range of diluted samples (Fig. 8, bottom panel) are between 85 and 125%, well within the typical variability of peak amplitudes (cf. Fig. 7), reflecting the quality of the chosen fit function and the inaccuracy caused by the low time resolution of the variable trapping period (Fig. 4). This last aspect is mostly defined by the 1 s time resolution of the Isotope Ratio Mass Spectrometer chromatogram procedure, for which the inaccuracy increases at higher CO2 concentrations. For example, CO2 delivered by steel capillary for the concentration range [355,390] µ mol mol−1 and [1365,1502] µ mol mol−1 are represented by a 34 s and 20 s trapping period, respectively. If very high concentrations (> 5000 µ mol mol−1 ) are expected, the relative peak amplitude of the samples could be allowed to be > 100%. However, a new empirical relation would need to be determined for lower flow rates of sample through the cryogenic trap or the Isotope Ratio Mass Spectrometer software for the variable trapping period (see Appendix) would need to be changed to use a time resolution shorter than 1 s. In any case, application of the empirical relations of trapping period and sample concentration requires not only that the concentration needs to be known prior to IRMS analysis, but also that the CO2 concentration data from the IRGA (Fig. 1) needs to be collected and processed prior to the laboratory analysis. Since we modified the Isotope Ratio Mass Spectrometer setup substantially, we tested the performance (i.e. precision and accuracy) by mea- virtual non-existing standard V-PDB is defined by adopting a δ 13 C value of +1.95 and a δ 18 O value of –2.2 for NBS 19 exactly. Via assigning these δ -values the hypothetical mineral V-PDB or rather the CO2 produced from it would be the standard for δ 13 C and δ 18 O values. The term V-PDB-CO2 refers to the oxygen isotopic composition of the CO2 evolved from the mineral by reaction with water-free H3 PO4 at 298 K. For details, see e.g. Ghosh et al. 24 5 429 430 431 432 433 434 435 436 437 438 439 440 441 442 443 444 445 446 447 448 449 450 451 452 453 454 455 456 457 458 459 460 461 462 463 464 465 466 467 468 469 470 471 472 473 474 475 476 477 478 suring δ 13 C (and δ 18 O) of a laboratory reference standard and QC standards that passed through the sample flow path (Fig. 9) or were sampled by the ASA beforehand (Fig. 10). The overall precision of δ 13 C measurements was determined to be <0.08 (σ ) for samples with standards stored in glass flasks inside an ASA (N=33), <0.11 (σ ) for samples with standards stored in stainless steel loops inside an ASA (N=44) and <0.06 (σ ) for directly supplied standards (N=5), over the course of several measurement campaigns between February 2006 and March 2008. The decrease in precision with increasing sample numbers (N=5,33 or 44) suggests that reference standards must be included in the sample sequence more frequently to correct for possible drift effects. The slight difference between the QC standard I and II (Fig. 9) can be explained by methods to determine the respective reference value. For the QC standard I the reference value was determined by the ETH IsoLab based on an average difference (N=5) to the laboratory reference standard. For the second QC standard on the other hand, the reference value was determined by an internationally acknowledged laboratory against several international standards with high precision. In general, based on periodic measurements of standards (mostly QC Standard I) using the ASAs, the overall accuracy was determined to be 0.11±0.04 (σ ), reflecting measurements with the ASA–Isotope Ratio Mass Spectrometer setup during one year, i.e. March 2007 to March 2008. For δ 13 C, we did not find an effect of the surface properties (e.g. volume, surface:volume, surface material) of the stainless steel loop versus the glass flask sample containers on the precision and accuracy of stored samples. However, for δ 18 O, the used stainless steel loop containers appeared inadequate and would require extensive pre-treatment, such as the removal of residual water from the surfaces that can otherwise exchange oxygen atoms in an equilibrium reaction with CO2 and thus potentially change the stable isotope composition of the sample. A treatment with long periods of dry air flushing in combination with heating, as suggested by Gemery et al. 16 , makes steel loops far less practical for δ 18 O measurements than their counterpart, i.e. glass flask as sample containers. We evaluated the reliability of the filling procedure of the ASAs and the influence of transport of samples from the field to the 504 laboratory by filling the same ASA in the field and sebsequently in the laboratory with the same standard gas. The resulting δ 13 C and δ 18 O measurements showed no significant difference (∆δ 13 C=0.04, ∆δ 18 O=0.05, N=5) between the respective filling locations (Sophia Etzold, Institute of Plant Sciences, ETH Zurich, Switzerland, unpublished data). For studies using the Keeling plot approach, e.g. to determine the signature of the respiration source via a statistical regression approach, the optimization of peak amplitudes provides a clear and essential improvement for intercomparability of δ 13 C measurements. 8 Small (systematic) errors in δ 13 C values would lead to much larger uncertainty in the determination of the respiration signature (the intercept of the Keeling plot regression line). 4,11 Our results show that the system described here not only can provide precision of at least 0.1 with an accuracy of 0.11±0.04 (σ ) for δ 13 C, but also allows unattended operation in the field and retain a measurement time per sample of 610 s. This clearly fullfills the quality criteria necessary to perform gradient measurements of stable isotope ratios of CO2 in air for the study of atmosphere–biosphere interactions. 505 Acknowledgments 479 480 481 482 483 484 485 486 487 488 489 490 491 492 493 494 495 496 497 498 499 500 501 502 503 515 Peter Plüss (ETH) and Patrick Flütsch (ETH) are kindly acknowledged for their extensive technical support. We would like to thank Matthias Saurer (PSI), Willi A. Brand (MPI-BGC), Michael Rothe (MPI-BGC) and Sophia Etzold (ETH) for their advice and helpful discussions. Our work has also benefited from discussions with Peter Weigel and Andreas Hilkert (Thermo Fischer). This work has been supported by the Swiss National Science Foundation (SNF), grant 200021-105949. 516 References 506 507 508 509 510 511 512 513 514 517 518 519 520 521 522 6 1. IPCC. Climate Change 2007: The Physical Science Basis. Contribution of Working Group I Contribution to the Fourth Assessment Report of the Intergovernmental Panel on Climate Change. Cambridge University Press 2007. ISBN 0521705967. 523 524 2. Yakir D, Wang XF. Nature 1996; 380: 515– 517. 564 565 566 525 526 3. Bowling DR, Tans PP, Monson RK. Global Change Biology 2001; 7: 127–145. 567 568 527 528 529 530 531 4. Pataki DE, Ehleringer JR, Flanagan LB, Yakir D, Bowling DR, Still CJ, Buchmann N, Kaplan JO, Berry JA. Global Biogeochem. Cycles 2003; 17: 1022. DOI: 10.1029/2001GB001850. 569 570 571 572 573 532 533 534 535 536 5. Ogee J, Peylin P, Ciais P, Bariac T, Brunet Y, Berbigier P, Roche C, Richard P, Bardoux G, Bonnefond JM. Global Biogeochem. Cycles 2003; 17: 1070. DOI: 10.1029/2002GB001995. 574 575 576 577 578 537 538 6. Knohl A, Buchmann N. Global Biogeochem. Cycles 2005; 19: GB4008. 579 580 539 540 541 542 543 544 545 546 547 548 549 550 551 552 553 554 555 7. Bowling DR, Pataki DE, Randerson JT. New Phytologist 2008; 178: 24–40. DOI: 10.1111/j.1469-8137.2007.02342.x. 581 8. Keeling CD. Geochim. Cosmochim. Acta 1958; 13: 322–334. 584 9. Keeling CD. Geochim. Cosmochim. Acta 1961; 24: 277–298. 586 10. Ogee J, Peylin P, Cuntz M, Bariac T, Brunet Y, Berbigier P, Richard P, Ciais P. Global Biogeochem. Cycles 2004; 18: GB2019. DOI: 10.1029/2003GB002166. 588 11. Zobitz JM, Keener JP, Schnyder H, Bowling DR. Agric. For. Meteorol. 2006; 136: 56–75. 12. Tuzson B, Zeeman MJ, Zahniser MS, Emmenegger L. Infrared Physics & Technology 2008; 51 : 198–206. DOI: 10.1016/j.infrared.2007.05.006. 582 583 585 587 589 590 591 592 593 594 595 596 597 598 556 557 558 559 560 561 562 563 13. Mohn J, Zeeman MJ, Werner RA, Eugster W, Emmenegger L. Isot. Environ. Health Stud. 2008; (in press). 14. Tuzson B, Mohn J, Zeeman MJ, Werner RA, Eugster W, Zahniser MS, Nelson DD, McManus JB, Emmenegger L. Applied Physics B. Lasers and Optics 2008; 92 : 451–458. DOI: 10.1007/s00340-008-3085-4. 7 15. Rothe M, Jordan A, Brand WA. In: World Meteorological Organization Global Watch World Meteorological Organization, 2005; 161: 64–70. 16. Gemery PA, Trolier M, White JWC. J. Geophys. Res.-Atmospheres 1996; 101: 14415– 14420. 17. Sturm P, Leuenberger M, Sirignano C, Neubert REM, Meijer HAJ, Langenfelds R, Brand WA, Tohjima Y. Journal Of Geophysical Research-Atmospheres 2004; 109: D04309. 18. Knohl A, Werner RA, Geilmann H, Brand WA. Rapid Communications In Mass Spectrometrys 2004; 18: 1663–1665. DOI: 10.1002/rcm.1528. 19. Theis DE, Saurer M, Blum H, Frossard E, Siegwolf RTW. Rapid Commun. Mass Spectrom. 2004; 18: 2106–2112. DOI: 10.1002/rcm.1596. 20. Clark MJR, Whitfield PH. Water Resources Bulletin 1994; 30: 1063–1079. 21. Werner RA, Brand WA. Rapid Commun. Mass Spectrom. 2001; 15: 501–519. DOI: 10.1002/rcm.258. 22. Merritt DA, Brand WA, Hayes JM. Org. Geochem. 1994; 21: 573–583. DOI: doi:10.1016/0146-6380(94)90003-5. 23. Werner RA, Bruch BA, Brand WA. Rapid Commun. Mass Spectrom. 1999; 13: 1237– 1241. 24. Ghosh P, Patecki M, Rothe M, Brand WA. Rapid Commun. Mass Spectrom. 2005; 19: 1097–1119. 25. Werner RA, Rothe M, Brand WA. Rapid Commun. Mass Spectrom. 2001; 15: 2152–2167. 599 Appendix 669 vice. The execution of the script code (or Actionscript device) within the acquisition method was scheduled each second during a specific time window as shown in Figure 6. Contrary to the ISL delay command for delayed operations, the proposed solution does not interfere with the timing of other tasks during the acquisition and proved to be a reliable and effective way of performing time variable tasks parallel to tasks with fixed timing during the IRMS chromatogram part of a measurement. Figure 4 shows different regression fits. Based on this, we have used different fitting functions for different concentration ranges, or applied a look-up-table approach using if statements. We were not able to program the power function fits (y = a · xb ) with decimal values of “b” in the ISL script, and have relied on Taylor or logarithmic functions instead. 670 FIGURE CAPTIONS 652 653 600 601 602 The ISL script code used for multitasking the concentration dependend activation of the cryogenic trap is described in the following example. 603 1 604 2 605 3 606 4 607 5 608 6 609 7 610 8 611 9 612 10 613 11 614 12 615 13 616 14 617 618 15 619 16 620 17 621 622 18 623 624 19 625 20 626 21 627 22 628 23 629 24 630 25 631 632 633 634 635 636 637 638 639 640 641 1 643 2 644 3 655 656 657 658 include " lib \ s t d i s l . i s l "; include " lib \ Instrument . i s l "; i n c l u d e " l i b \ GasBench_lib . i s l " ; f u n c t i o n T r a p T i m e rF u n ( number A) { number B = ( −10.393 * l o g (A) ) −96; r e t u r n B; } main ( ) { number nA = _GetSequenceNumber ( " Concentration " ,60) ; number nB = c a l l T r a p T i m e r F u n ( nA ) ; number nC = 150 − nB ; number nD = _ R e f G e t P r o f i l e N u m b e r ( " ASA " ," Start " ,0) ; number nE = a b s ( _ G e t T i c k C o u n t ( ) − nD ) ; 659 660 661 662 663 664 665 666 667 668 671 672 i f ( nE > ( nC * 1 0 0 0 ) ) { _ S e t ( " Gas Bench / V a l c o " ,LOAD) ; _ U s e r I n f o ( " CryoTrap a c t i v e ! " , 0 , 0 ) ; } } The sample concentration is read from the sequence table and converted to a trapping period via an empirically fitted function (code line 14–15) as shown in Figure 4, from which a target start time in seconds is calculated (16). The elapsed time is calculated (17–18) and compared to the target start time in order to decide if the trap should be set to an activated state or not (20–24). For correct calculation of the elapsed time, the start time of the chromatogram process was stored from the internal millisecond counter with the command code 642 645 s c r i p t TrapTimer 654 numbers nF = ( _ G e t T i c k C o u n t ( ) ) ; _ R e g S e t P r o f i l e N u m b e r ( "ASA" , " S t a r t " , nF ) ; c a l l Startchromatogram () ; 673 674 675 676 677 678 679 680 681 682 683 684 685 686 687 688 689 690 691 692 646 647 648 649 650 651 within the Acquisition ISL script used for this method. Please note that the included lines (1–2) are added just before the chromatogram is started (3). For effective use of the script in the ISODAT NT acquisition software, it was incorporated into the Gasbench configuration as an ActionScript de- 693 694 695 696 697 8 Figure 1: Overview of improved setup, from sampling air in the field (top) to measurement by Isotope Ratio Mass Spectrometer (bottom). Communication connections and sample gas flow paths are indicated by broken lines and thick lines, respectively. After field operation, the ASA is transported to the lab and interfaced with the Isotope Ratio Mass Spectrometer. Figure 2: Flow diagram of the ASA showing the difference between field (top) and laboratory (bottom) operation. The ASA is shown in field sampling mode: the flow is diverted by the solenoid valve, the air sample is dried, filtered and pushed through the Valvo-valve (shown here in bypass loop position). During sampling, an adjustable flow regulator (set to ≈ 0.9 L min−1 ) and a poppet check valve (one-way, opening at >7 kPa) help create a pressure excess in the glass flask or stainless steel loop sample containers. Per ASA, three or four Valvo-valves with sample containers are connected in series, adding up to a maximum total of 33 or 44 samples per unit. Two manual threeway valves are used to switch between field and laboratory setup. Figure 3: Flow diagram of the laboratory setup. Shown here is the situation for flow of He (valve 1) with sample air from the ASA (2) to the Gasbench 698 699 700 701 702 703 704 705 706 707 708 709 710 711 (3) and subsequent cryogenic focus trap (4), while at the same time the Gas Chromatograph and the inlet of the Isotope Ratio Mass Spectrometer are flushed with He (4 and 5). See text for details. Figure 4: Empirical relations between sample CO2 concentration and trapping periods required for peak amplitudes that are equal to the Isotope Ratio Mass Spectrometer reference gas, for samples delivered to the Isotope Ratio Mass Spectrometer from three different types of containers (glass flasks, steel loops, steel capillary). For glass flask sample containers, two fit functions are shown. For fit parameters and further details, see Table 1 and Appendix. 738 739 740 741 742 743 744 745 746 747 748 749 750 712 713 714 715 716 717 718 719 720 721 722 723 724 725 726 727 728 729 730 731 732 733 734 Figure 5: Decision diagram for the automated liquid nitrogen (LN2 ) refilling system used for the cryogenic trap. Activation of the solenoid valve for LN2 flow into the Dewar depends on fulfillment of the conditions for the weight of the cryogenic trap Dewar (A), the temperature of the balance (B) and a signal from the Isotope Ratio Mass Spectrometer during a specific period of the IRMS measurement protocol (C) and a waiting period after filling. Figure 6: The timeline of the Isotope Ratio Mass Spectrometer measurement protocol implemented in the ISODAT NT software. The flow path of the input line and Gasbench is flushed with sample air (S) and the cryogenic trap is lowered into liquid nitrogen before sample air is let into the trap to be frozen (FS) for a period that is variable and derived from the sample CO2 mixing ratio. After the trap is raised to thaw, the cryogenically focused content is carried by He through the Gas Chromatograph to the Isotope Ratio Mass Spectrometer for analysis. Meanwhile, the sample inlet tubes are flushed with pure He (He) or the next sample (Snext ) with higher flow rate (+). 751 752 753 754 755 756 757 758 759 760 761 762 763 764 765 766 767 768 769 770 771 772 735 736 737 δ 13 C Figure 7: Linearity performance of analysis using the modified Gasbench, expressed as the deviation from the δ 13 C reference value against the 773 774 775 9 relative peak amplitude (A) in the chromatogram. See text for details. Figure 8: Application of concentration-dependent variable cryogenic trapping periods in the Gasbench for measurements of δ 13 C. In the top panel, the deviation from the δ 13 C reference value is shown for different dilutions of a CO2 in air mixture with a constant δ 13 C value. In the bottom panel, the deviation from the δ 13 C reference value is shown for the corresponding relative peak amplitude in the chromatogram. The SD for the δ 13 C measurements is 0.04 (N=13). Figure 9: Deviation between measured δ 13 C (δSample ) and reference δ 13 C (δRef ) for a laboratory reference standard and two quality control (QC) standards that were directly suplied through steel capillaries, at different dates. The deviations of QC standard I (top) and QC standard II (bottom) are on average –0.02±0.04 and 0.06±0.05, respectively. The maximum observed SD for each fivesample series of reference standard and QC standards (I and II) is 0.05. Figure 10: The deviation between measured δ 13 C (δSample ) and reference δ 13 C (δRef ) for a laboratory reference standard and a quality control (QC) standard that was sampled beforehand in glass flasks (3 min of ≈ 1 L min−1 flushing into an ASA). The maximum observed SD for each five-sample series of reference standard is 0.05 and 0.07 for δ 13 C and δ 18 O, respectively. For QC standard I, the average deviations from their reference value are 0.08±0.04 and –0.19±0.05 for δ 13 C and δ 18 O, respectively. The QC standard I is the same as shown in the top panel of Figure 9. Table 1: Fit parameters for the relationships between sample CO2 concentrations and trapping periods for different types of sample containers. See text and Figure 4 for details. 776 FIGURES AND TABLES create a pressure excess in the glass flask or stainless steel loop sample containers. Per ASA, three or four Valvo-valves with sample containers are connected in series, adding up to a maximum total of 33 or 44 samples per unit. Two manual threeway valves are used to switch between field and laboratory setup. Field setup Air inlet selection unit Pump COMcontroller 0.9 L min Particle filter -1 sample Sample sample manifold manifold manifold (ASA) 0.9 L min-1 Gas analyzer COMcontroller Field PC Transport to the lab Lab setup Hesample & Ref. manifold valve COMcontroller Gasbench Sample sample sample manifold manifold manifold (ASA) COMcontroller Cryogenic focus trap LN2 dewar COMcontroller 777 Reference and sample preparation Mass Spectrometer He COMcontroller He Vent Vent 4 5 He (Dilution off) Cryogenic focus trap Split Mass Spectrometer Figure 3: Flow diagram of the laboratory setup. Shown here is the situation for flow of He (valve 1) with sample air from the ASA (2) to the Gasbench (3) and subsequent cryogenic focus trap (4), while at the same time the Gas Chromatograph and the inlet of the Isotope Ratio Mass Spectrometer are flushed with He (4 and 5). See text for details. Adjustable flow meter Poppet check valve 100 ● ● Digital flow meter Entrapment period [s] Field Gas Chromatograph Water trap 779 Lab Particle filter 12-pos. Valco-valve with 11 sample volumes He, high flow He Pump Drying column Water trap A 3 He + Sample Sample manifold (ASA) Gasbench II, modified Out Solenoid valve He Lab PC A Vent 2 1 Ref 1 Figure 1: Overview of improved setup, from sampling air in the field (top) to measurement by Isotope Ratio Mass Spectrometer (bottom). Communication connections and sample gas flow paths are indicated by broken lines and thick lines, respectively. After field operation, the ASA is transported to the lab and interfaced with the Isotope Ratio Mass Spectrometer. Sample in Ref 2 Ref 3 ● 80 ● 60 ● Glass flasks (ASA); y = a1·x1+a2·x2+a3·x3+b Glass flasks (ASA), alt. fit; y = a·xb Steel loops (ASA); y = a·ln(x)+b Steel capillary (Ref. inlet); y = a·ln(x)+b ● ● ● 40 ● ● ● ● ● ● ● ● 40 ● ● 20 20 0 300 778 Sample + He to GasBenchMass Spectrometer 780 80 60 ● ● 0 He in 100 500 700 900 1100 1300 1500 CO2 concentration [µ µmol mol−1] Figure 4: Empirical relations between sample CO2 concentration and trapping periods required for peak amplitudes that are equal to the Isotope Ratio Mass Spectrometer reference gas, for samples delivered to the Isotope Ratio Mass Spectrometer from three different types of containers (glass flasks, steel loops, steel capillary). For glass flask sample containers, two fit functions are shown. For fit parameters and further details, see Table 1 and Appendix. Figure 2: Flow diagram of the ASA showing the difference between field (top) and laboratory (bottom) operation. The ASA is shown in field sampling mode: the flow is diverted by the solenoid valve, the air sample is dried, filtered and pushed through the Valvo-valve (shown here in bypass loop position). During sampling, an adjustable flow regulator (set to ≈ 0.9 L min−1 ) and a poppet check valve (one-way, opening at >7 kPa) help 10 Dewar for the cryogenic trap A) Balance 781 m < "tare" B) Temperature of the balance T > -20°C C) IRMS Signal during acquisition Solenoid COMController valve LN2 Figure 5: Decision diagram for the automated liquid nitrogen (LN2 ) refilling system used for the cryogenic trap. Activation of the solenoid valve for LN2 flow into the Dewar depends on fulfillment of the conditions for the weight of the cryogenic trap Dewar (A), the temperature of the balance (B) and a signal from the Isotope Ratio Mass Spectrometer during a specific period of the IRMS measurement protocol (C) and a waiting period after filling. 782 Figure 6: The timeline of the Isotope Ratio Mass Spectrometer measurement protocol implemented in the ISODAT NT software. The flow path of the input line and Gasbench is flushed with sample air (S) and the cryogenic trap is lowered into liquid nitrogen before sample air is let into the trap to be frozen (FS) for a period that is variable and derived from the sample CO2 mixing ratio. After the trap is raised to thaw, the cryogenically focused content is carried by He through the Gas Chromatograph to the Isotope Ratio Mass Spectrometer for analysis. Meanwhile, the sample inlet tubes are flushed with pure He (He) or the next sample (Snext ) with higher flow rate (+). 11 (ASample / ARef) × 100 [%] 0 50 100 200 400 02:00 O 0.4 δSample − δRef [‰] 0.2 0.2 ● ●● ●● ● ● ● ●● ● ● ● ● ● ● ● ● ● ● ● ● ● ●● ● ● ●● ●● ● CO2 in air (400 µmol mol−1) ● ● ● ●● CO2 in He (4000 µmol mol−1) 0.0 −0.2 −0.4 13 0.4 C 0.2 0.0 δSample − δRef [‰] 18 0.4 0.0 −0.2 −0.4 −1 ● ● ●●● ● ● ● ●●●● ●● ●● ● ● ●● ●● ●●●● ● ●● ●● ● ● ● ● ●●● ● CO2 in air (400 µmol mol ) CO2 in He (4000 µmol mol−1) 0.4 0.2 0.2 −0.2 ●●● ●● 0 50 200 Figure 7: Linearity performance of δ 13 C analysis using the modified Gasbench, expressed as the deviation from the δ 13 C reference value against the relative peak amplitude (A) in the chromatogram. See text for details. C ●● ● ●● δSample − δRef [‰] −1 CO2 concentration [µ µmol mol ] 900 1200 1500 1800 δSample − δRef [‰] 13 C 0.2 0.0 −0.2 C 0.2 786 0.0 0 50 100 150 200 (ASample / ARef) *100 [%] Figure 8: Application of concentration-dependent variable cryogenic trapping periods in the Gasbench for measurements of δ 13 C. In the top panel, the deviation from the δ 13 C reference value is shown for different dilutions of a CO2 in air mixture with a constant δ 13 C value. In the bottom panel, the deviation from the δ 13 C reference value is shown for the corresponding relative peak amplitude in the chromatogram. The SD for the δ 13 C measurements is 0.04 (N=13). ● ●●●● 00:00 ●●●● ● ● ●●● ● ●●● ●● Reference standard 04:00 −0.2 0.2 0.0 −0.2 08:00 18:00 20:00 22:00 00:00 C 0.4 0.2 ● ● ● ●●● ● ●● ●●●●●●●● ●●●●● ●●●●● ●●●●● ● 0.0 −0.2 −0.4 0.4 0.2 0.0 −0.2 2008−03−12 ● QC standard I Reference standard 18 ● QC standard I Reference standard O −0.4 0.4 0.2 0.0 ● ●● ●● ● ● ● ● ● ● ●● ●●●●● ●● ●●●●●● ● ●●● ●● ● −0.2 2008−03−12 −0.4 18:00 20:00 22:00 00:00 Time of day [h] Figure 10: The deviation between measured δ 13 C (δSample ) and reference δ 13 C (δRef ) for a laboratory reference standard and a quality control (QC) standard that was sampled beforehand in glass flasks (3 min of ≈ 1 L min−1 flushing into an ASA). The maximum observed SD for each five-sample series of reference standard is 0.05 and 0.07 for δ 13 C and δ 18 O, respectively. For QC standard I, the average deviations from their reference value are 0.08±0.04 and –0.19±0.05 for δ 13 C and δ 18 O, respectively. The QC standard I is the same as shown in the top panel of Figure 9. −0.2 2007−07−18 ● ●●●● ● QC standard II 20:00 16:00 0.0 Reference standard 13 0.0 −0.4 13 0.2 0.2 −0.2 −0.2 2007−07−18 −0.2 784 0.2 0.0 ●●● ● ● ●● ●●● ●●●● ● 2008−03−13 16:00 600 ● QC standard I Time of day [h] 0.4 300 0.0 Figure 9: Deviation between measured δ 13 C (δSample ) and reference δ 13 C (δRef ) for a laboratory reference standard and two quality control (QC) standards that were directly suplied through steel capillaries, at different dates. The deviations of QC standard I (top) and QC standard II (bottom) are on average –0.02±0.04 and 0.06±0.05, respectively. The maximum observed SD for each fivesample series of reference standard and QC standards (I and II) is 0.05. 400 (ASample / ARef) × 100 [%] 0 ●●●●● 13 785 100 ● ● ●●● 2008−02−26 −0.2 783 18:00 0.2 ●●●● ● 16:00 0.0 −0.2 14:00 C 0.0 0.2 10:00 13 0.0 −0.2 06:00 787 12 Table 1: Fit parameters for the relationships between sample CO2 concentrations and trapping periods for different types of sample containers. See text and Figure 4 for details. Fit function Fit parameters 1 2 3 Glass flasks (in ASA) y = a1 · x + a2 · x + a3 · x + b a1 = −3.71 · 10−1 , a2 = 3.536 · 10−4 , a3 = −1.215 · 10−7 , b = 174 a = −15.37, b = 144 Steel loops (in ASA) y = a · ln(x) + b Steel capillary (Ref. & QC standards) y = a · ln(x) + b a = −10.39, b = 96 13
© Copyright 2024