Keys to Getting a Good Fingerstick and Capillary Blood Sample Doing a proper fingerstick is essential to getting good results on any device for Point-of-Care testing MAXIMIZE GOOD BLOOD CIRCULATION • Warm the hand by washing in warm water or holding it in a heating pad or hand warmer for a couple of minutes. • Massage the finger with a downward motion several times before performing the fingerstick. • Lower the hand to below the heart level when collecting the drop of blood. USE THE RECOMMENDED LANCET DEVICE We recommend a 21 gauge, single use lancet device. If you are still having trouble, you can try using a 19 gauge lancet. CHOOSE THE RIGHT SPOT Choose a spot that is on the bottom side of the tip of one of the center fingers of either hand. (You should check to be sure that the blood will easily drop onto the Test Strip before you perform the fingerstick) TRY USING A MICROSAFE CAPILLARY COLLECTION TUBE A capillary collection tube can help draw the correct amount of blood, and can make it easier to apply the sample to the Test Strip. COLLECT AND APPLY AN ADEQUATE SAMPLE TO THE STRIP Release pressure and repeat, if necessary, to collect a Quickly express blood down toward the large hanging drop of blood. Immediately apply one finger by gently squeezing across the entire drop to the sample well on the strip. Do not add a finger, at the last joint, not the fingertip. second drop after the first drop is applied. Do not "milk" the finger. • Healthcare professionals should always wear gloves. • The meter should be on a level surface, and not moved or touched during testing. • If you need to repeat the test, perform a new fingerstick. HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 0200189 Rev F Training Materials TRAINING-IN-A-BINDER Contents Chapter 1 – Instructor and Site Preparation…………… 2 Chapter 2 – Description of the Meter and Strip……...… 4 Chapter 3 – How to Change the Meter Setup………..… 8 Chapter 4 – How to Perform a Test…………….……..…10 Chapter 5 – Troubleshooting………………….……...… 14 Chapter 6 – Finishing the Training………………...…… 16 Chapter 7 – Forms…………………………………......…17 Chapter 8 – Your Site Specific Information………..…… 18 Contact Information Technical Support and Customer Service Toll Free: 1.877.436.6444 Technical Support: 24 hours/day, 7 days/week Customer Service: 7am-5pm pst, Monday-Friday HemoSense Inc. 651 River Oaks Parkway San Jose, CA 95134 USA phone: fax: website: email: 408.719.1393 408.719.1184 www.hemosense.com service@hemosense.com ©2005 HemoSense Inc. All rights reserved. Made and printed in the USA. PN 0200263 RevC 1 Chapter 1 – Instructor and Site Preparation Preparation for Training Before training in person or over the telephone, ensure that the site is prepared. This section covers setting up the class, training materials and supplies. Call to set up an appointment for training and review the Class Guidelines below. After setting up the appointment, FAX or email the Training Letter to the customer. The Training Letter contains a checklist of the Class Guidelines outlined below. Request the site have the trainees view the video and review the other training material prior to the training. Setting up the Class: Training Area: Ensure the site has sufficient space for the trainer, trainees, training materials and supplies. There should be enough space for participants to take notes and to practice with the INRatio meter. A flat, stable work surface is required to run the meter. The temperature of the room should be between 50°F and 95°F. Scheduling: The suggested time for the training is 1 hour. It is recommended to schedule 1.5 hours to allow for adequate practice of the fingerstick technique and for questions. Factor in additional time if review of site policies and procedures is required. Class Size: The recommended class size is four to six participants. Training Materials: HemoSense Training Materials: INRatio Professional Instruction Video: The video runs approximately 10 minutes and provides step by step instructions for setting up the meter, performing a fingerstick, running a test and cleaning the meter. INRatio Professional User’s Guide: The User’s Guide is provided with each INRatio Kit. It is the primary reference for users of the device. Inserts: Separate package inserts are provided with the strips, lancing devices and micro capillary tubes. Note: for the Unistick®2 Extra single use lancing device, the instructions for use are printed on the box. INRatio Quick Reference Guide: A Quick Reference Guide is provided with every kit which instructs the user in a number of easy steps how to obtain a test result. Refer to the User’s Guide, Package Inserts and Quick Reference Guide frequently throughout the training to familiarize the user with where additional information may be found. 2 Other Training Materials: This training binder may be customized further by adding training materials e.g. overheads, handouts, additional information, site policies etc. Extra tabs are provided to make this easier. Please indicate the institution’s name or logo on this extra material to distinguish it from the HemoSense provided materials. Check with the site being trained if they would like to review their Policies and Procedures prior to beginning the product training. Ordering Additional Training Materials: To order additional copies of the training material please contact HemoSense Customer Service at 1-877-463-6444. Supplies: Make sure that the site has received their system before performing the training. In addition, if the training is being held in person at the site, take with you demo meters plus strips and other supplies. Allow for one demo meter per two trainees. The following is the list of supplies that should be available for the training: - INRatio Meter - INRatio Test Strips - Lancing Device and Lancets - Micro-capillary Tubes - Gauze and Alcohol Swabs - Latex Gloves - Biohazard Disposal Container - INRatio Professional User’s Guide - Package Inserts for Strips, Lancets and Micro-capillary Tubes - Quick Reference Guide General Instructions 1. When performing training follow the structure of the training material in this binder. 2. Convey the information outlined in each section and demonstrate where in the product labeling (User’s Guide and or Package Inserts) the information can be found. Use examples and analogies as appropriate. 3. The N symbol indicates information that should be shown or explained to the trainees. 4. The symbol indicates information that should be stressed and reemphasized multiple times during the training. 5. As you complete each section ask the questions at the end of the section and assure yourself that the trainees are familiar with the answers. Ask other questions as appropriate. 6. Complete your copy of the check list as you go to indicate you have covered all the relevant sections. Fill in the name of the institution and the date of the training. Sign the checklist and keep it for your records. i 3 Chapter 2 – Description of the Meter and Strip Understanding the System This section covers descriptions of the meter, strip, operating conditions and safety. -Begin the class by viewing the video if the trainees have not already done so. -Review the components of the kit and the features of the meter and strip. -Review the operating and storage conditions for the meter and strip. N Show trainees where more information can be found in the User’s Guide sections “Principals of Operation”. Professional Kit Components Describe each of the components of the kit, Power Supply, Quick Reference Card, User’s Guide, Lancets, Training Video, and Warranty Reply Card. The Meter 1. Instruct the trainees to turn the Meter so that the back faces them. Point out the Power Jack and the Data Port. i Explain that the Power Jack is only to be used with the HemoSense Power Supply provided with the kit. 2. Instruct the trainees to turn the meter upside-down. Point out the Serial Number, the Battery Door Release, the Battery Door and the telephone number for Technical Support. Explain how to open the Battery Door to install the batteries. Specify the number and type of batteries to be used. N Show trainees where information on Installing Batteries can be found in the User’s Guide. 3. Point out the important features of the Meter, the Display, the OK, MENU and UP/DOWN Buttons, the Cleaning Door, the Test Strip Guide and the Sample Light. 4. Point out that the meter requires no maintenance other than routine cleaning. Explain and demonstrate how to perform cleaning. Include discussion of site cleaning policy as required e.g. after every test, once per day, once per week. N Show trainees where information on Maintenance and Cleaning can be found in the User’s Guide. i Caution that the meter should not be immersed in liquid and that abrasive cleaner should not be used. The menu button: a) May be used to turn on the meter. b) Accesses the meter menu. c) Accesses the memory functions of the meter. d) All of the above. 4 The Strip 1. Point out the important features of the Strip, the Electrodes, the Test Areas and the Sample Well. Explain where the blood is applied and how it runs up the track to the test area. 2. Point out the two QC channels and explain that each time a test is run with the patients’ sample a normal and a therapeutic QC is automatically run at the same time. Explain that the results can be recorded and that because of these on-strip controls there is no need for external liquid controls. Point out that the ranges for QC for each strip lot can be found on the HemoSense web site (See Chapter 8). 3. Explain that when QC values are displayed without an error message, the meter and strip are working properly and the testing was done using acceptable techniques. Explain that when a QC error is displayed the controls are out of range and a series of Troubleshooting steps will need to be taken. Troubleshooting will be discussed in a separate section. N Show trainees where this information can be found in the User’s Guide section “Principals of Operation” and “Quality Control”. 1. How many channels are on the test strip? 2. How many controls are on the test strip? Operating and Storage Conditions Review the proper operating and storage conditions for the meter and the strip. Point to where these conditions may be found in the Strip Package insert and in the User’s Guide. Explain that to ensure the accuracy of the test results, the system should be operated in proper conditions and the strip should be stored carefully in the appropriate environmental conditions. Review all of the operating and storage conditions in the following sections. i Stress the importance of this information and emphasize the fact that the results may not be accurate if these conditions are not adhered to. Meter Operating Conditions Instruct trainees on the following: Do: • Perform testing at Room Temperature between 50°F and 95°F. • Perform testing at a Relative Humidity between 10% and 95%, without condensation. • Make sure the meter is at room temperature before use. • Test with the meter on a level surface that is free of vibration. • Transport the meter in the original container or another secure container. Do not: • Do not perform testing below 50°F or above 95°F. 5 • • • • • Do not perform testing where there is condensation e.g. a steamy bathroom Do not bring the meter to the patient’s finger. Bring the finger to the meter. If this is not possible use a micro-capillary tube. Do not perform testing where the surface is uneven such as a bed or a person’s lap. Do not perform testing on a vibrating surface such as in a car or an airplane. Do not drop the meter or treat it roughly by carrying it around unprotected in a briefcase or a purse. N Show trainees where this information can be found in the User’s Guide sections “Introduction”, “Principals of Operation” and “Operational Precautions and Limitations”. To run a test you can hold the meter in your hand while applying the sample as long as you set it down within 10 seconds. True or False? Strip Storage Conditions Determine what the site policy is for storing strips. Depending on the policy (some sites do not want to give the option of refrigeration) emphasize the preferred storage method. Instruct trainees on the following: Do: • Store the strips at Room Temperature, below 90°F, until the expiration date. • If desired, store the strips refrigerated, 35°F to 45°F, until the expiration date. • If refrigerated, allow the sealed pouch to come to room temperature for 5 minutes before opening for testing. • Store the strips in their original container. • Use the test strip within 10 minutes of opening the foil pouch. Do not: • Do not leave the strips in conditions above 90°F such as your car in the summer time or on a window ledge in the sun. • Do not freeze the strips • Do not use the strips after the expiration date printed on the pouch. Point out where the expiration date may be found on the box and on the pouch. N Show trainees where this information can be found in the Strip Package Insert section “Storage and Handling”. 1. How long can test strips be stored: a) Until the expiration date on the box. b) For six months. c) For three months or until the expiration date on the box, whichever comes first. d) Indefinitely. 2. How long can the strip be out of the pouch before performing a test? 6 Safety N Show trainees where this information can be found in the User’s Guide section “Hazards”. i Important Points to Emphasize and Repeat ? Always perform testing on a flat, level surface. ? Always store and use the strips under the recommended conditions. 7 Chapter 3 – How to Change the Meter Setup Main Menu Instruct the trainees to turn the meter on by pressing any button. Point out the Self Test Screen. Note this screen only appears for X seconds and is then replaced by the Insert Strip Screen. Explain that pressing the Menu Button at this point will take you to the main menu. Point out that there are a number of options, Review Memory, Print Memory, Change Setup, Turn Meter Off and Run Test. Spend some time reviewing each of these options and explaining the choices available. N Show the trainees where they can find this information in the User’s Guide section “Main Menu Functions”. Review Memory Explain that the meter automatically saves the test results in memory and that this option allows review of past test results. Point out that the meter can store up to 60 time and date stamped results at any one time and that when the memory is full, the first result will be lost when a new one is recorded. N Explain that pages 23 and 24 of the User’s Guide show the screens that will appear in the Review Memory menu. Print Memory Explain that this option allows the memory to be printed and that a printer is available for purchase from HemoSense or their distributors. Explain also that the “Printer Operating Instructions” should be followed for set up and operation for the printer. Change Setup Explain how to set the various options in the Change Setup menu, Time and Date, Units, Target, Printer, Language. Point out that if the site requires QC to be recorded then the units should be set to the INR+PT+QC. Otherwise the INR option is appropriate for most situations. Ask the site which units are required to be displayed per their policy and emphasize this screen. i Stress the importance of recording the correct units in the log sheet. Recording a PT seconds value of 12 as an INR value can lead to inappropriate treatment decisions. N Explain that pages 25 to 39 of the User’s Guide show the screens that will appear in the Review Memory menu. Explain that these pages walk through the steps to change each of the setup options. 8 Turn Meter Off Explain how this option allows the meter to be turned off. Point out that the meter will automatically turn itself off after 10 minutes of inactivity in the Main Menu in order to save batteries. To turn off the INRatio meter: a) Wait 10 minutes, and it will turn off by itself. b) Press the menu key, scroll down to "TURN METER OFF" and press the OK button. c) Since there is no on/off button, it must be unplugged. d) Both a. and b. Run Test Explain that this option will return the user to the testing menu to the Insert Strip screen. i Important Points to Emphasize and Repeat ? Check that the units being recorded are correct. 9 Chapter 4 – How to Perform a Test Performing a Test This section covers the steps for performing a test including preparations, performing a fingerstick, applying the sample, reading and recording the results. After walking through the steps, instruct the trainees to break into pairs and perform a fingerstick on each other. Observe the fingerstick technique and provide instruction throughout the process Preparing to do a Test Instruct the trainees on the steps necessary to prepare for the test. 1. If the strip has been refrigerated allow it to come to room temperature for 5 minutes before opening the pouch. (Remember site policy on storage. If room temperature storage is required, ignore this step). 2. Prepare the lancing device and set aside until ready to do the test. Note: Instruct the trainees how to prepare the lancing device at this point. i Stress the importance of using the recommended lancing device with 21 gauge lancet (Unistick®2 Extra for single use or auto-Lancet™ Device for multiple use). Point out that a lancing device/lancet for a blood glucose meter will not produce enough blood for an INRatio test. N Show the trainees where they can find instructions for use in the Lancing Device Package Insert. 3. Place the meter on a flat, stable surface like a table and turn it on using any button. Note: Inform the trainees about the self-test and software version on the Self-Test screen. 4. Wait for the Insert Strip screen to appear. 5. Open the pouch, remove the test strip and insert it into the meter. The Strip Code screen will appear. Note: Instruct the trainees on how to insert the strip correctly using the strip guides and making sure the words on the strip are readable by the user. Point out the green sample light in the sample well. i Stress the importance of using the strip within 10 minutes of opening the pouch. 6. Compare the strip code on the pouch with the one displayed on the screen. Note: Instruct the trainees on how to change the strip code. i Stress the importance of entering the correct strip code. Point out that the results may be inaccurate if the code in the meter does not match the code assigned to the strips. 10 7. Wait for the Apply Sample screen. Note: Point out that the system is now ready to do a test. Explain that there is a 5 minute window from this point to do the Fingerstick and apply the sample before the meter times out. Explain also that there is a 10 sec window at the end of the 5 minute period within which pressing any button will reinstate the Apply Sample screen. N Show the trainees where to find these steps in the User’s Guide in the section “Performing a Test”. 1. If the strip code displayed on the meter DOES NOT match the strip code number (the code in bold print on the outside of the strip pouch), you should: a) Discard that strip and find one that matches. b) Use the strip anyway. c) Correct the code on the meter display by using the "up/down" arrow key. d) Call HemoSense Technical Support. 2. If the test strip is inserted upside down: a) It will break the meter. b) The meter will not recognize that a strip has been inserted. c) The meter will run a test anyway, and the results will be correct. d) The meter will run a test anyway, but the results may be erroneous. Performing a Fingerstick Review the steps necessary to obtain an adequate sample from a fingerstick. A minimum sample size of 15uL is required in a large hanging drop. Use examples to describe a large hanging drop or demonstrate on yourself and show them what the drop should look like. N Show the trainees where to find these steps in the User’s Guide in the section “Collecting a Fingerstick Sample”. Hand out the sheet “Tips for Obtaining a Good Fingerstick Sample” and use it to walk through the steps. 1. Increase the circulation to the hand. Note: Instruct the trainees how to increase circulation using the tip sheet and the User’s Guide. 2. Select a site on the finger, clean with alcohol and dry thoroughly. Note: Instruct the trainees on how to choose a site, clean and dry using the instructions in the tip sheet and the User’s Guide. i Stress the importance of drying the site after wiping with alcohol. Point out that if not dried correctly, a large hanging drop will not form and the blood will smear on the 11 finger. The site may sting when the fingerstick is performed if there is still alcohol on the finger. i Stress the importance of using the recommended Alcohol Swabs i.e. alcohol only. Using swabs with lanolin or other additives may interfere with the test results. 3. Puncture the finger with the lancet following the instructions for the lancet being used. 4. Apply gentle, continuous pressure until a large hanging drop forms. Note: Instruct the trainees on how and where to apply pressure to obtain a clean, adequate drop of blood following the instructions in the tip sheet and the User’s Guide. i Instruct the trainees not to apply strong repetitive pressure i.e. not to milk the finger. Point out that this may cause an inaccurate result. Applying a Fingerstick Sample Instruct the trainees on how to properly apply the large hanging drop to the sample well. 1. Apply the sample to the sample well on the strip over the green light. i Stress the importance of applying the sample immediately after performing the fingerstick. Point out that the PT test result is based on timing and that the blood starts to clot as soon as it is exposed to the air. Any delay in applying the sample will lead to an inaccurate result. i Instruct the trainees that not hitting the target will result in the sample smearing on top of the strip which will lead to insufficient sample in the well. i Stress that it is very important not to move the meter after the sample has been applied. Point out that doing so may give an inaccurate result. 2. When the meter beeps and displays the Testing Sample screen, the test has begun. i Stress the importance of applying sufficient blood in order to get an accurate result. A good way to check this is to look at the sample well after the test has started to see if there is some blood left in the sample well. i Stress that it is very important not to apply a second drop of blood. Applying a second drop of blood may lead to an inaccurate result. If there is not enough sample, begin the test again with a new fingerstick and a new strip. 3. When the meter beeps again the test is complete and the result is displayed in the units chosen in the setup. 4. Remove the strip. Discard the strip and lancet in a biohazard, puncture proof container. 5. Record the result as appropriate (depending on your institution’s policy) 12 N Show the trainees where to find these steps in the User’s Guide in the section “Collecting a Fingerstick Sample” and “Performing a Test” and in the tip sheet “Tips for Obtaining a Good Fingerstick Sample”. 1. The amount of blood that should be applied to a test strip is: a) A large "hanging" drop of blood. b) At least 15 µL of blood. c) Enough blood to cover the sample well. d) All of the above. 2. If the drop of blood is not big enough you can just add more blood. True or False? Applying a Sample using a Micro-capillary Tube Instruct the trainees on the use of micro-capillary tubes for applying a fingerstick sample to the strip. Some sites may not have Micro-capillary tubes at the time of phone training. In this case walk through the Package Insert. Point out that this techniques should be used in situations where the sample cannot be applied directly to the application site e.g. bed ridden patient or patient with dexterity problems. i Stress the importance of not squeezing the bulb on the Micro-capillary Tube while collecting the sample. This can lead to bubbles in the sample which may lead to an inaccurate result. i Stress again that it is very important not to move the meter after the sample has been applied. Point out that doing so may give an inaccurate result. N Show the trainees where to find this information in the Micro-capillary Package Insert. i Important Points to Emphasize and Repeat ? The importance of good fingerstick technique ? Never move the meter once the strip is inserted ? Enter the correct calibration code 13 Chapter 5 – Troubleshooting Troubleshooting Walk through this section with the trainees making sure to cover all the Error Messages, Abnormal Results and Limitations of the System Error Messages Walk the trainees through the Troubleshooting section of the User’s Guide. Hand out the sheet “Error Message Troubleshooting Guide”. Explain each Error Screen and describe, using examples what action should be taken. N Show the trainees where to find the “Error Screen” descriptions and the suggested “Action to Take” in the Troubleshooting section of the User’s Guide (pages 46 to 49) and on the Error Code Summary in this training binder. If a message "Error QC 2L", indicating that the on-board controls are out of range, appears on the display: a) Call HemoSense Technical Support immediately. b) Rerun the test. If you get a result the second time, the system is working properly. c) Discard the rest of the box of test strips. Abnormal Results Walk the trainees through the Results section of the Test Strip Package Insert. Explain what the normal and therapeutic ranges are and what to do in the event of an unusual result. Point out that the HemoSense recommendation is to repeat the test if a value is unusually low or high. If the result still does not match the clinical situation, test by another method. i Stress the importance of responding appropriately to an unusually high or low result. Instruct the trainees to familiarize themselves with their site policy on confirmed high and low results and act accordingly. Limitations of the System Walk the trainees through the Operational Precautions and Limitations section of the User’s Guide and the Limitations section of the Test Strip Package Insert. Be sure to cover all of the following: 1. Use only fresh capillary whole blood. No plasma or anticoagulated whole blood. No blood from a venous draw. 2. Minimum 15uL of blood required. 3. The system is sensitive to heparin and low molecular weight heparin. 4. Do not use outside of 30 to 55% hematocrit. 14 N Show the trainees where to find this information in the Test Strip Package Insert and in the User’s Guide page 41. 1. The INRatio system can be used with: a) Capillary whole blood. b) Venous whole blood. c) Plasma. d) All of the above. Warranty Inform the trainees that the warranty period is one year from the date of purchase. N Show the trainees where to find this information in the User’s Guide page 50. i Important Points to Emphasize and Repeat ? Be aware of policies and procedures for an abnormally high or low result ? Be aware of the importance of the limitations of the system Troubleshooting Tip Sheets Provided by HemoSense in this Section 1. Keys to Getting a Good Fingerstick and Capillary Blood Sample 2. Error Message Troubleshooting Guide 3. Printer Troubleshooting Guide 15 Chapter 6 – Finishing the Training Review The purpose of the review is to reinforce and repeat points or steps in the operation of the system that are key to obtaining an accurate result and to interpreting that result correctly. Review these key points and answer any remaining questions: 1. 2. 3. 4. Fingerstick technique Storage of supplies Changing the calibration code Product limitations Competency Quiz The competency quiz determines whether trainees have understood the key concepts. Use this exercise to determine if there are any remaining areas of confusion. Training Certification Fill out the Training Certificate, sign and give to the participants. Retain a copy for your own records. 16 Chapter 7 – Forms Forms Provided by HemoSense Competency Quiz Competency Quiz Answer Key Training Certificate Example Log Sheet 17 Chapter 8 – Your Site Specific Information Feel free to insert your own site documents here. 18 Printer Trouble Shooting Guide Problem: Solution: Green light is flashing Printer is out of labels. Install a new roll of labels. If labels are installed, look for jammed or damaged labels, and reload the labels. The green light can also flash if the wrong labels are used. Problem: Solution: Label jam Open the cover. Cut the label roll between the back feed slot and the roll. Push the form feed button to advance the labels through the printer. Cut the label strip through so there will be a flat edge on the end of the strip. Reload the labels again. Problem: Solution: Printer will not print Make sure the printer cable is properly connected. Make sure the printer is turned on. The green light will be on. Check the power connection at the wall and at the printer. Plug the printer into another wall outlet. Problem: Solution: Self Test to verify that the printer is printing properly Load a roll of labels. Make sure the power cord is connected. Turn the printer off. Hold the form feed button down and as you are holding it down, turn the printer on. When the printer starts printing, release both buttons simultaneously. The printer will print out a few lines and at the bottom of it, it will say "ready". Problem: Solution: Paper label roll does not fit in the printer The right hand plate on the spindle has a triangle on one side. This side should be facing the paper roll. Snap the plate on tightly. If the plate is not in this position the spindle will not fit in the slot properly. Problem: Solution: Print quality is poor Print head may be temporarily misaligned. Unload and reload the labels and try printing again. Problem: Solution: Printer prints gibberish Can happen if there is a power surge. Unplug and plug the printer again. Problem: Solution: Printing is faint This is a thermal printer which does not have any ink to change. Reload the labels, clean the rollers and reload the labels after the rollers have been allowed to air dry. Recommended Labels: Dymo Labelwriter White Shipping Labels, 2-1/8" x 4", 220 labels per roll Available at office supply stores. 2-1/4" x 4" or 2-5/16" x 4" will also work. HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500261 Rev A INRatio PT/INR Monitoring System: Correlations with Plasma-based Laboratory and Whole Blood-based Point-of-Care PT Measurement Systems Introduction Clinical Laboratory Standards Institute (CLSI), formerly known as NCCLS, guidelines for specimen collection, transportation, storage and processing to minimize the affect of pre -analytical variables on the lab sample. A minimu m of 20 samples is recommended. The INRatio PT/INR Monitoring System was developed for point-of-care (POC) testing, specifically to improve patient management for healthcare providers by using a simple finger-stick sampling of capillary whole blood and providing an immediate result. To facilitate a smooth transition to a new device such as the INRatio Monitor, comparisons to the current measurement system are sometimes performed. This paper provides examples of completed correlations comparing the INRatio system to plasma based lab systems and other POC whole blood based devices. Factors to Consider As described in previous INRatio Technical Bulletins [#101–105] there are many factors to consider when comparing PT results between different systems, including but not limited to, factors that affect warfarin metabolism, factors that affect the PT test itself, such as sample type, reagent source/sensitivity, and instrument measurement principle (Table 1). It is important to note that, as mentioned in Table 1, the INR calculation itself explains reasons for some differences: ISI INR = (Patient PT/ MNPT) Purpose The following correlations are provided as examples only of previous studies between INRatio and existing PT systems. These comparisons reinforce the strong correlation found between INRatio and other systems, including POC (whole blood) and lab-based (plasma samples), as well as to reference laboratories, and demonstrate acceptable performance of INRatio. (MNPT= Mean Normal PT, ISI=International Sensitivity Index) The ISI is an exponent, so any error in the ISI can magnify errors in the calculation (hence, a low ISI due to a highly sensitive reagent will minimize errors). In addition, results exceeding 5.0 INR generally have reduced trueness, precision and linearity both in POC and lab based PT testing and should be interpreted accordingly.1 Procedure The evaluations were performed using the HemoSense Evaluations Guidelines, which recommend adherence to Table 1: Factors that affect PT/INR results when comparing different systems. Factors that affect warfarin mechanism or metabolism • Illness (e.g. congestive heart failure, hyper- & hypo-thyroidism, hepatic failure) • Drugs • Diet (foods containing vitamin K and alcohol) Factors that affect the PT test itself can cause differences when comparing systems • Lack of comparability of INR between reagents when used at onset of warfarin therapy • Loss of accuracy and precision when using reagents with high ISI (low sensitivity thromboplastin) • Inherent variability with different instrumentation • Lack of reliability of ISI provided by manufacturer • Incorrect calculation of INR from use of inappropriate control plasmas used to determine the Mean Normal PT (MNPT) Factors when comparing capillary whole blood labbased plasma system • Pretest variables (especially for plasma samples) • Contaminated line (Heparin) • Antiphosp holipid antibodies • Hematocrit Incorrect ISI of thromboplastin • Incorrect calibration • Change in ISI over time • Poor distribution of warfarin samples • Incorrect choice of reference INRatio Monitor Characteristics The INRatio Monitor determines the PT by measuring the change in impedance of the sample as clotting occurs. INRatio test strips use a recombinant thromboplastin reagent with a low ISI, as re-commended by ACCP, CAP and WHO.2 Each lot of INRatio Test Strips is calibrated to a plasma based lab system traceable to the WHO standard. Lot to lot variability is adjusted for by entering a test strip calibration code with each new strip lot. In the comparisons provided here, detailed reagent information (source and ISI value of comparative reagent) was not provided by the site. 1 Data Interpretation In general terms, the “r” value may be interpreted as a measure of scatter in the data. A factor influencing the “r” value is the method of clot detection employed by the system. The closer the “r” value is to 1.0 the better the agreement between the individual points. Comparing systems with different clot detection methods may give rise to higher levels of scatter even when comparing lab systems (see fig. 9). The slope of the regression line may be interpreted as a reflection of a systematic bias between the systems. A factor influencing the slope is the calibration of the system. All systems are calibrated differently, a fact reflected in the slopes of the lab to lab comparisons (see fig. 7, 8 and 9). The closer the slope is to 1.0, the less the bias between the systems. Comparisons between whole blood POC systems: POC systems with the same sample type i.e. whole blood fingerstick show very good agreement even though the measurement principals are quite different. Fig. 1 Fig. 2 CoaguCheck vs. INRatio ProTime vs. INRatio y = 1.0141x + 0.0323, r = 0.94 y = 1.0441x - 0.0099, r = 0.94 5 7 6 4 INRatio INRatio 5 3 4 3 2 2 1 1 0 0 0 1 2 3 4 0 5 1 2 3 4 5 6 7 ITC Protime Coagucheck Comparisons between whole blood POC and lab-based plasma systems: Even with the differences in sample types added to the mix, the following comparisons between INRatio and several laboratory-based plasma samples continue to show good correlations ranging from 0.93 to 0.99. Few (if any) clinical discrepancies that would result in different dosing decisions are apparent. In each case the evaluating site implemented the INRatio system due to acceptable agreement coupled with the additional features of the INRatio system. Fig. 3 Fig. 4 AMAX vs. INRatio MLA vs INRatio y = 1.0457x + 0.1999, r = 0.96 6 y = 0.9759x - 0.0438, r = 0.976 5 5 4 INRatio INRatio 4 3 3 2 2 1 1 0 0 1 2 3 4 5 0 6 0 Amax 1 2 3 MLA 2 4 5 Fig. 5 Fig. 6 Sysmex vs. INRatio ACL vs. INRatio y = 1.0809x + 0.104, r = 0.95 6 5 INRatio 5 4 INRatio y = 0.9889x - 0.3382, r = 0.96 6 4 3 3 2 2 1 1 0 0 0 1 2 3 4 5 0 6 1 2 3 4 5 6 ACL 1000 Sysmex 1500 Commentary: Variability between laboratory analyzers is recognized and reported by others To understand the variability among PT methodologies better, three comparisons are provided in Figures 7, 8 and 9 between lab systems commonly used: MLA, Sysmex, STA and AMAX. In addition, according to data provided by Dr. Alan Jacobson, observed variation in INR can be high even between laboratory and reference laboratory analyzers (Fig 10). Plasma samples tested on 18 different instrument/reagent combinations demonstrated that as much as 2.5 INR differences can be observed for a single sample run on multiple systems. Fig. 7 Fig. 8 MLA vs Sysmex MLA vs. STA y = 0.8419x + 0.1663, r = 0.99 6 y = 0.7277x + 0.6668, r = 0.96 7 6 5 Sysmex 1500 5 STA 4 3 4 3 2 2 1 1 0 0 0 1 2 3 4 5 0 6 1 Fig. 9 4 5 6 Instrument/Reagent Combinations & Observed Variation in INR y = 0.7245x + 0.3779, r = 0.92 7 5.5 6 5 5 4.5 4 INR AMAX 3 Fig. 10 MLA vs AMAX 3 4 3.5 2 3 1 2.5 0 0.00 2 MLA 1800 INR MLA 1800 INR 2 1.00 2.00 3.00 4.00 5.00 6.00 7.00 1.5 MLA 1800 INR Sample 3 7 Conclusions Transitioning to a new coagulation measurement system for PT is not difficult but has several points to consider due to the many variables that exist across all PT/INR testing modalities. In the context of this variability, the correlations presented here show strong performance for INRatio versus commonly used lab systems. The frequently asked question “Will it match?” should be carefully considered. However, in the end consistency of test method is the key to the management of patients on oral anticoagulant therapy. The INRatio provides that consistency together with an easy to learn and easy to use system.4 Measurement Principles of Comparison Systems INRatio detects clot formation by a change in the electrical impedance of the blood sample that occurs when fibrinogen is converted to fibrin. ProTime Microcoagulation System pumps the sample back and forth until a clot forms. The clot is detected optically as the motion of the blood decreases with clotting. CoaguChek uses alternating magnetic fields in the device which cause iron particles on the strip to move within the sample; the endpoint is reached when the blood clot stops the iron particles from moving. Plasma based systems generally detect clot formation via mechanical (AMAX) or optical means (ACL, MLA, STA, Sysmex). References 1 NCCLS Point of Care monitoring of anticoagulation therapy; approved guideline, NCCLS document H49-A Vol. 24 No. 23 (TSBN 1-56238540-2). NCCLS, 940 West Valley Road, Suite 1400, Wayne, PA, 19087-1898 USA, 2004. 2 The Seventh ACCP Conference on Anticoagulation Therapy: Evidence-Based Guidelines, Supplement to Chest, Volume 126, No. 3 (Supplement). September 2004: 2085-2095. ACCP: American College of Chest Physicians, CAP: College of American Pathologists, WHO: World Health Organization. 3 Unpublished Abstract, Alan Jacobson, Loma Linda VA 4 Self management of oral anticoagulation with the INRatio system: impact of a structured teaching program on patient’s knowledge of medical background and procedures; European Journal of Cardiovascular Prevention and Rehabilitation 2004, 11:000–000; Heinz Völler, Clemens Dovifat, Johannes Glatz, Heinrich Körtke, Uwe Taborski and Karl Wegscheider. Sysmex is manufactured by and a trademark of Sysmex Corporation. STA is manufactured by and a trademark of Diagnostica Stago Inc. ACL is manufactured by and a trademark of Instrumentation Laboratory/Beckman Coulter Inc. AMAX is manufactured by and a trademark of Trinity Biotech. MLA is manufactured by and a trademark of Medical Laboratory Automation, Inc. CoaguChek is manufactured by and a trademark of Roche Diagnostics. ProTime is manufactured by and a trademark of ITC. HemoSense, Inc. 651 River Oaks Parkway San Jose, CA 95134 USA 1-877-436-6444 toll free (408) 719-1393 tel (408) 719-1184 fax www.hemosense.com ©2005 HemoSense, Inc. All rights reserved. PN 5500285 Rev A 4 inratio professional TrainingCertificate Understanding the System Meter description Strip description Meter operating conditions Strip storage conditions Safety Main Menu Performing a Test Menu functions Preparing to do a test Performing a fingerstick Applying a fingerstick sample Applying a sample using a micro-capillary tube Troubleshooting Error messages Abnormal results Limitations of the system Warranty print name of trainer signature of trainer date print name of trainee & employee id signature of trainee date institution & location or unit PN 5500265 REV A Error Message Troubleshooting Guide The INRatio meter can report abnormal operating conditions by displaying error or warning messages appropriate to the condition on the meter display. The following provides a list of the error messages generated by the INRatio meter with a description of the conditions which cause the message to be displayed. Some errors allow the user to recover from the error condition whereas others are non-recoverable. In most cases the user can recover from the error condition by repeating the test. The following table describes both recoverable and non-recoverable error message. Recovery procedures are described for each recoverable error. Error Screen • What’s the problem? What to do? The sample was applied during the warm up period before the “Apply Sample” screen appeared. • • A used strip was inserted. • Be sure to use a new strip for every test. • Not enough blood reached the test area. The possible causes for this are: o Not enough blood applied to the sample well. o The strip channel is blocked. • Begin the process again with a new test strip and perform another fingerstick. Make sure to apply a large hanging drop of blood. Consult the Fingerstick Tip Sheet if you are having problems obtaining a good sample. Do not try to add more blood from the first fingerstick. • • • • • • Only certain combinations of numbers and letters are recognized by the monitor. The code entered was not recognized by the monitor. • The monitor had a data processing problem while calculating the results from some or all of the three channels. This may be caused by the following: o A problem with the individual test. o Not enough blood was applied to the sample well. • The error appears as a three digit number. Each digit represents one of the channels on the strip and may appear as 1, 2, 3, or 4. Example: ERROR 411 • • • • Begin the process again with a new test strip. Be sure to wait until the “Apply Sample” screen appears before applying the sample. Re-enter the correct code at the “Strip Code” prompt. If the message repeats, call HemoSense Technical Support. Begin the process again with a new test strip and perform another fingerstick. Make sure to apply a large hanging drop of blood. Consult the Fingerstick Tip Sheet if you are having problems obtaining a good sample. Do not try to add more blood from the first fingerstick. If the error message repeats, call HemoSense Technical Support. HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500264 Rev C 1 Error Screen What’s the problem? What to do? • An invalid date value was entered. • Check the date and enter the correct day, month and year. • The monitor was turned on with a test strip in the strip holder. • If the strip was a used strip remove it and insert a new strip when the “Insert Strip” screen appears. If the strip was a new strip, remove it and reinsert it when the “Insert Strip” screen appears. Do not use a new strip that has been out of the pouch for longer than 10 minutes. If the message recurs, call HemoSense Technical Support. • • • • The test result is outside the range of the monitor. • • Repeat the test. If the message repeats, contact your physician. • An invalid target range has been entered e.g. the lower end of the target range entered is higher than the upper range • Re-enter the correct upper and lower values and continue with the setup. • The temperature of the room is less than 65oF (18oC) and the monitor cannot heat up within the allowed time. • Move the monitor to a warmer location and wait until it has warmed up before performing the test. • The temperature of the room is higher than 95 oF (35oC) and the monitor cannot cool down. • Move the monitor to a cooler location and wait until it has cooled down before performing the test. HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500264 Rev C 2 Error Screen What’s the problem? • • One or both of the on-board controls are out of range due to one of the following causes: o The strip has been exposed to excessive hear or humidity. o The user technique is poor. If the monitors units are set to INR or INR+PT, the Error may look like any of the following: QC 1H, QC 1L, QC 2H, QC 2L, QC 1+2 What to do? • • • • • • No INR value is displayed because one or both of the on-board controls are out of range due to one of the following reasons: o The strip has been exposed to excessive hear or humidity. o The user technique is poor. QC values are displayed because the monitors units are set to INR+PT+QC • • • • • • The printer option has been selected and one of the following may have happened: o There is no printer connected to the meter. o The printer is not functioning properly. There is no paper in the printer. • • • Repeat the test with a new test strip and a fresh fingerstick paying careful attention to proper technique. If the message repeats, it may be an indication that the strips have deteriorated. Perform the following checks before proceeding: 1) Is the strip outside the expiration date on the pouch? 2) Was the strip stored outside the recommended storage conditions e.g. in a hot place? 3) Are the pouches of the remaining unopened strips damaged? If any of these conditions occurred, do not use the strips for further testing. Call HemoSense Technical Support if the message repeats. Repeat the test with a new test strip and a fresh fingerstick paying careful attention to proper technique. If the message repeats, it may be an indication that the strips have deteriorated. Perform the following checks before proceeding: 4) Is the strip outside the expiration date on the pouch? 5) Was the strip stored outside the recommended storage conditions e.g. in a hot place? 6) Are the pouches of the remaining unopened strips damaged? If any of these conditions occurred, do not use the strips for further testing. Call HemoSense Technical Support if the message repeats. Recheck all the connections to make sure the printer is connected. If the error message repeats, call HemoSense Technical Support. Insert paper into the printer and try to print again. HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500264 Rev C 3 Error Screen Error 01, Error 02, Error 03, Error 10, Error 11, Error 12, Error 13, Error 14, Error 15 and Error 20. What’s the problem? The batteries are running low. • • Continue with the present test. Replace the battery as soon as possible. • The batteries are dead and the meter will not operate. • • Replace the batteries or Plug the unit into a wall outlet. • If you see any of these errors it means there is a problem with the monitor These errors cause the meter to shut down: o If the OK button is pressed o If the message has been displayed for 60 sec. An error has been detected during the monitor self test process which causes the green light in the application area to flash • If a non-recoverable error is displayed, make note of the error number and contact HemoSense Technical Support. • If a non-recoverable error is displayed, make note of the error number and contact HemoSense Technical Support. • These errors are non-recoverable Flashing green light in the application area This is a non-recoverable error What to do? • • HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500264 Rev C 4 INRatio Example Log Sheet Test Results Log • Quality Control Check • Daily Cleaning/Maintenance Log Instrument Serial #________________ Patient/Resident Strip Lot #_______________________ Test Results PT INR QC OK? Yes No Strip Code______________________ Date/Time By QC Results QC1 QC2 Daily Cleaning/Maintenance Log (To document cleaning, place initials in box under date) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500263 Rev A 31 INRatio Competency Quiz Answer Key 1. (d) 2. False 3. Three 4. Two 5. (a) 6. 10 min 7. (d) 8. (c) 9. (b) 10. (d) 11. False 12. (b) 13. (a) HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500262 Rev B INRatio Competency Quiz 1. The menu button: a. May be used to turn on the meter. b. Accesses the meter menu. c. Accesses the memory functions of the meter. d. All of the above 2. To run a test you can hold the meter in your hand while applying the sample as long as you set it down within 10 seconds. True or False? 3. How many channels are on the test strip? _________ 4. How many controls are on the test strip? __________ 5. How long can test strips be stored: a. Until the expiration date on the box. b. For six months. c. For three months or until the expiration date on the box, whichever comes first. d. Indefinitely. 6. How long can the strip be out of the pouch before performing a test? __________ 7. To turn off the INRatio meter: a. Wait 10 minutes, and it will turn off by itself. b. Press the menu key, scroll down to "turn meter off" and press the OK button. c. Since there is no on/off button, it must be unplugged. d. Both a) and b). 8. If the strip code displayed on the meter DOES NOT match the strip code number (the code in bold print on the outside of the strip pouch), you should: a. Discard that strip and find one that matches. b. Use the strip anyway. c. Correct the code on the meter display by using the "up/down" arrow key. d. Call HemoSense Technical Support 9. If the test strip is inserted upside down: a. It will break the meter. b. The meter will not recognize that a strip has been inserted. c. The meter will run a test anyway, and the results will be correct. d. The meter will run a test anyway, but the results may be erroneous 10. The amount of blood that should be applied to a test strip is: a. A large "hanging" drop of blood. b. At least 15 µL of blood. c. Enough blood to cover the sample well. d. All of the above 11. If the drop of blood is not big enough you can just add more blood. True or False? 12. If a message "Error QC 2L", indicating that the on-board controls are out of range, appears on the display: a. Call HemoSense Technical Support immediately. b. Rerun the test. If you get a result the second time, the system is working properly. c. Discard the rest of the box of test strips. 13. The INRatio system can be used with: a. Capillary whole blood. b. Venous whole blood. c. Plasma. d. All of the above. HemoSense, Inc. 651 River Oaks Parkway, San Jose, CA 95134 1-877-436-6444 toll free (408)719-1393 tel (408)719-1184 fax www.hemosense.com © 2005 HemoSense, Inc. All rights reserved. PN 5500262 Rev B Prothrombin Time/INR Measurement in the Physician’s Office: Improved Quality of Care Steven Cohen MD, Cobb Medical Group, Austell GA 30106 In today’s health care environment, both insurance companies and Medicare are monitoring physician performance. The cost of providing care, the degree to which a physician’s patients achieve certain measured outcomes (HbA1c levels in diabetics, percentage of patients with coronary artery disease on beta blockers, the frequency of mammograms and PAP smears in women, etc.) and the level of patient satisfaction with a physician’s treatment are tracked. In some areas,1 insurance companies and employers provide bonuses to physicians who meet defined benchmarks (e.g., number of patients with Hemoglobin A1c under 7.0, etc.). In an ongoing pilot project, Medicare pays physicians and hospitals bonuses if certain quality benchmarks are met.2 Physicians today have to meet the challenge of helping patients achieve a defined outcome in a cost-effective fashion. At the same time, the combination of increasing overhead and flat or decreasing reimbursement from payors means that physicians cannot simply hire more employees to achieve these outcomes. It is necessary to “work smarter,” not just “work harder,” to maintain a thriving practice in today’s healthcare environment. The medical literature has demonstrated that point of care testing (defined as performing a test while the patient is present, as opposed to sending the test to a reference lab and receiving the test results at a later time) is helpful in treating conditions as varied as bacterial vaginosis,3 diabetes,4 respiratory syncitial virus,5 and anticoagulant therapy.6 The ability to perform a PT/INR in-house and obtain results at the time of a patient’s visit can help physicians increase patient satisfaction while maintaining the same level of clinical outcome (defined as the percentage of time that a PT/INR is within the therapeutic range).7 In the United States an estimated 250,000 people are hospitalized annually with thromboembolic events8 and 2.2 million people are affected by atrial fibrillation.9 The incidence of complications in patients treated with warfarin ranges from 0.22.2/100 patient-years.10 This emphasizes the importance of monitoring the PT/INR and keeping it within the therapeutic range, to minimize both the risk of complications from the underlying condition and bleeding complications from warfarin. The Hemosense INRatio® system is a CLIA waived test that can be performed by the nursing staff to provide PT/INR results before the patient leaves the clinic. Combining a nurse visit with the testing gives the staff an opportunity to check medication compliance, monitor for the presence of complications, provide ongoing patient education on the proper use of warfarin and monitor the use of medications or herbal treatments which can interact with warfarin11 (see Appendix 1). One valuable aspect of point of care testing is the elimination of “phone tag” in this age of voice mail and answering machines. If the specimen is sent to the laboratory, the cost of notifying the patient of the results by phone can be estimated at $4.25/test based on the following assumptions: 1. Staff salary of $14/hour with an additional 25% cost of benefits, for a total hourly rate of $17.50. 2. Two calls made by clinical staff before contacting a patient with the results (5 minutes/call to pull the chart, call the patient, leave a message or explain the results and to refile the chart). 1 3. Five minutes of staff time for the operator to answer the call from the patient requesting the results, take a message, and attach it to the chart for the clinical staff to respond to the phone call. Providing the results to the patient at the time of testing would eliminate all but 3 minutes of the time above (estimated time for nurse to explain results to patient) for a savings of $3.40/test. This translates into an annual savings of $2448 for a physician who performs 3 tests/day. Additionally, the patient has a better chance of understanding and properly following physician instructions. A nurse or medical assistant may use a structured questionnaire (see Appendix 1) to help monitor the use of warfarin. This can fulfill the criteria for a 99211 visit, which are as follows:12 1. The E/M service must be provided face to face. 2. The patient must be an established patient in the practice. 3. The visit must be medically necessary. 4. The physician must have previously seen the patient so that the services of the staff are incident to the treatment plan established by the physician. The physician does not have to be in the same room where the patient is being seen, but must be in the suite. The documentation should clearly indicate that the physician is periodically participating in the care of the patient. 5. The documentation must identify the date of service as well as the identity of the staff providing the care. The American College of Physicians July/August 2000 newsletter reports that ICD-9 code v58.61, long-term (current) use of anticoagulants should be the primary code used for ordering the prothrombin time. The diagnosis that prompted the use of warfarin (e.g., atrial fibrillation 427.31, pulmonary embolism and infarction 415.10) can be used as the secondary diagnosis.13 An example of a structured questionnaire that may be administered by the nurse is listed in Appendix 1. Appendix 2 contains an example of a flow sheet for continuing warfarin treatment. 2 References 1 Colwell J. Pay-for-performance takes off in California. ACP Observer, Jan-Feb 2005. http://www.acponline.org/journals/news/jan05/pfp.htm. 2 CMS Office of Public Affairs, Fact Release, Jan 31, 2005. http://www.cms.hhs.gov/media/press/release.asp?Counte r=1343. 3 West B, et al. Evaluation of a new rapid diagnostic kit (FemExam) for bacterial vaginosis in patients with vaginal discharge syndrome in The Gambia. Sex Transm Dis. 2003 Jun;30(6):483-9. 4 Feinberg WM, Blackshear J, Laupacis A, Hart RG. Prevalence, age distribution, and gender of patients with atrial fibrillation: analysis and implications. Arch Intern Med 1995;155:469-73. 5 Mackie PL, McCormick EM, Williams C. Evaluation of Binax NOW RSV as an acute point-of-care screening test in a paedriatric accident and emergency unit. Commun Dis Public Health 2004 Dec;7;328-30. 6 Caffee AE, Teichman PG. Improving anticoagulation management at the point of care. Fam Pract Manag 2002 Feb;9:35-37. 7 Shiach CR, et al. Reliability of point-of-care Prothrombin time testing in a community clinic: a randomized crossover comparison with hospital laboratory testing. Br J Haematol 2002 Nov;119:370-75. 8 Goldhaber SZ. Pulmonary embolism. N Engl J Med 1998 Jul;339:93-104. 9 Maisel WH. Atrial fibrillation in heart failure: epidemiology, pathophysiology, and rationale for treatment. Am J Cardiol. 2003 Mar 20;91(6A):2D-8D. 10 Braunwald, E. Heart Disease: A Textbook of Cardiovascular Medicine, 6th ed, p 1699-1700, 2001. 11 Price CP. Medical and economic outcomes of point-ofcare testing. Clin Chem Lab Med 2002 Mar;40: 246-51. 12 Audit and Reimbursement Department, University of Missouri Health Care, July 2002 Newsletter. http://www.muhealth.org/~UPReimbursement/julynews0 2.htm 13 Baker B. ACP-ASIM Observer, July 2000. http://www.acponline.org/journals/news/julaug00/billingguides.htm APPENDIX 1 Patient name______________________________________ DOB________________ Date of Visit_____________ PT/INR Assessment - CLARIFY ANY YES ANSWERS YES ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ NO ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ ____ Change in meds since last visit Recent antibiotic use? Use of OTC pain relievers? Nsaids, aspirin, tylenol? Use of OTC meds for stomach, Zantac, Tagamet, antacids etc? Vitamin use? Change in intake of green herbal teas or green leafy vegetables? Change in alcohol intake (beer, wine, liquor)? Alcohol intake in last 48 hours? Any recent illness? Any herbal medications? Specify which ones. (May circle herbs below) Recently stopped any herbal therapy? Any extra doses of warfarin since last visit? Any missed doses of warfarin? Travel plans? Planned dental visit in near future? Planned surgery in near future? Prescription for warfarin needed? Any bleeding? Any bruising? Chest pain/leg pain or swelling/dyspnea/headache? Hospitalization since last visit? Herbs reported to affect anticoagulation (grouped by mechanism) Anticoagulants: Coumarin constituents: Alfalfa, Angelica, Aniseed, Arnica, Asafetida, Boldo, Capsicum, Celery, Chamomile(german), Chamomile (roman), Dandelion, Fenugreek, Focus, Horse-Chestnut, Nettle, Passionflower, Prickly Ash (northern or southern), Quassia, Red Clover Inhibition of platelet aggregation or activity: Bilberry, Clove, Cat’s Claw, Feverfew, Ginger, Gingko, Liquorice Salicylate constituents: Meadowsweet, Poplar, Willow Peroxidase stimulates synthesis of arachidonic acid metabolites: Horseradish Procoagulants: Coagulant: Agrimony Heparin antagonist: Golden Seal Lectin agglutinating activity: Mistletoe Coagulant in vivo: Yarrow Barnes, J. et. Al. Herbal Medicines: A guide for healthcare professionals. Pharmaceutical Press, London, 2nd edition, 2002. Staff Signature______________________________________ 3 APPENDIX 2 ANTICOAGULATION FLOW SHEET Patient name______________________________________ DOB________________ MR#_____________ Diagnosis for anticoagulation________________________ ICDM code_____________ Date warfarin began_______________ Date to finish_______________ / permanent INR Goal_____________________ Date PT INR Warfarin Dose Dose Change Next PT ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ _____________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ ______________ _______ _______ ____________ ____________ ___________ _____________________________ HemoSense Inc. 651 River Oaks Parkway San Jose, CA 95134 USA Tel: (408) 719-1393 Fax: (408) 719-1184 Email: moreinfo@hemosense.com Website: www.hemosense.com This publication sponsored by HemoSense, Inc. P/N 5500294 RevA © 2005 HemoSense, Inc. All rights reserved. Printed in the United States of America. HemoSense and INRatio are trademarks of HemoSense, Inc. in the United States of America and other countries. Staff Signature
© Copyright 2024