System Components for Sample Gas Sampling Data Sheet 10/235.10 EN November 2006 A gas analysis system usually consists of two groups of devices: • The gas analyzer measures the concentrations of the sample gas components and converts them into an electrical signal. • Supplementary devices sample the sample gas from the process, condition it and feed it in the gas analysis system. Contents In most cases, the sample gas sampled from the process cannot be processed by the gas analyzer without further conditioning. Too high a dust content, temperature and dew point as well as too high or too low pressure, and not least interference components in the sample gas can impair the operational capability of the gas analyzer and falsify the measuring result. Supplementary devices such as the sampling probe, the sample gas line, the sample gas cooler and pumps and filters ensure that the sample gas inlet conditions of the connected gas analyzers are satisfied and faultless measuring results are obtained independently of the process sequence and the local conditions. Modular Sample Gas Probe System FE2 Filter Unit Annular Heating Element Heating Sleeve PFE2 Filter Unit PFE3 Filter Unit Selection Table for Probe Tube Types and Probes Type 40, Type 40 W Probe Tubes Type 42 Probe Tube (heated) Type 60 Probe Tube (Cement Flue Gas Probe) Probe 2 Probe 11 Ceramic Filter Process Pressure Reducers Heated Sample Gas Lines CGWB 13 Heated Sample Gas Line TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines Electronic Temperature Controller and Switching Relay Page 2 3 4 5 6 11 15 16 17 18 19 20 21 22 24 26 29 35 Modular Sample Gas Probe System Easily adaptable to different sampling conditions thanks to modular construction Components: • • • • FE2, PFE2, PFE3 filter units Probe tube types 40, 42, 60 Annular heating element Heating sleeve Probe tubes and filter unit heatable Connections for test gas and back purging Filter with high retention rate and long service life Easy filter replacement Usable in emission measuring systems conforming to 13th BImSchV and 17th BImSchV (German Federal Immissions Protection Law) 2 System Components for Sample Gas Sampling 10/23-5.10 EN November 2006 FE2 Filter Unit Use The FE2 filter unit is a component of the modular gas sampling probe system. It is used for filtering gases containing dust. Process Gas Sampling Conditions Pressure pabs = 50 to 600 kPa Temperature max. 200 °C (coated max. 180 °C) Flow Rate 30 to 500 l/h, relative to 100 kPa and 0 °C Pressure Drop approx. 0.6 hPa at 100 l/h Dust Content max. 3 g/m3 (application recommendation); max. 40 g/m3 with probe tube type 40 (see page 16) with purge-back device; max. 2000 g/m3 with probe tube type 60 (see page 18) Filter Characteristics Connection Ports Filter Element Surface filter Sample Gas / Test Gas G 1/4 internal thread to DIN ISO 228/1 Filter Fineness 0.3 µm Compressed Air (2 Ports) G 1/2 internal thread to DIN ISO 228/1 Retention Rate 99.98 % Probe Tube G 3/4 internal thread to DIN ISO 228/1 Seal Integrity 10–4 hPa l/s Compressed Air Supply for Purge Back Dead Volume 280 ml (See PFE2 Filter Unit, page 6) Materials of the Gas-carrying Parts Instrument Air Free of dust, water and oil Pressure max. 600 kPa Casing and Connections Special steel Mat. No. 1.4571 Flange Special steel Mat. No. 1.4541 Dimensional Drawing (Dimensions in mm) Filter Aluminum oxide surface filter 105 14 130 160 65 Coating Thermoset plastic, resistant up to 180 °C Installation Flange DN 65, PN 6, Form B to DIN 2573, flange gasket A; positioning of holes: 0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°, 360° 280 40...65 Mounting Flange connect or bolt to the probe tube Weight approx. 8.0 kg 10/23-5.10 EN November 2006 Ordering Information Catalog No. Housing and flange SS, filter fineness 0.3 µm 23042-4-0730709 Housing and flange SS, coated, 0.3 µm 23042-4-0730718 System Components for Sample Gas Sampling 3 Annular Heating Element Use For heating the FE2 filter unit, which is a component of the modular gas sampling probe system (see page 3). Heating is necessary if condensation is likely from moist gases and there is consequently risk of the filter becoming clogged and corroded. Data Power Supply 115 VAC or 230 VAC, 50 to 60 Hz Heater Output 200 W, 100 W or 50 W – depending on setting (see below) The heating resistors are factory-set for heat output of 100 W Temperature of Heater (unregulated) approx. 120 °C to 180 ° C (measured at ambient temperature of 20 °C, still air) Ambient Temperature 0 to 55 °C (still air) Degree of Protection of Housing IP50 Materials Housing: aluminum alloy, heating surface: brass Dimensional Drawing (dimensions in mm) Weight approx. 1.7 kg Electrical Connections Terminals via M20 cable gland. The electrical supply conductor must be heat resistant (≥ 200 °C). Installation Two half shells are placed round the FE2 filter unit to be heated and closed with a wing screw. Option Regulated Heating with 1 x Pt 100 Ordering Information Heating Circuit Ring heater 230 V, 50/60 Hz Ring heater 115 V, 50/60 Hz Ring heater with Pt 100, 230 V, 50/60 Hz Ring heater with Pt 100, 115 V, 50/60 Hz 230 / 115 V AC 200 W 4 230 / 115 V AC 100 W Catalog No. 23029-5-7865312 23029-5-8008813 23029-5-7865316 23029-5-7865317 230 / 115 V AC 50 W System Components for Sample Gas Sampling 10/23-5.10 EN November 2006 Heating Sleeve Regulated Heating Sleeve for FE2 Filter Unit Heater Micanite Holding Temperature max. 350 °C Temperature Control max. 200 °C (Pt 100) Power Supply 115 VAC or 230 VAC Power Consumption 250 VA Connection Terminal box with M16 cable gland, degree of protection IP54 Supply Cable 3 x 1 mm2 with silicone sheath, length 1 m Ambient Temperature –20 to +45 °C Relative Humidity ≤ 75 % yearly mean; infrequent slight condensation permissible Weight approx. 1 kg Essential Accessories Ordering Information Heated sleeve, 230 V, 50/60 Hz, 250 VA Heated sleeve, 115 V, 50/60 Hz, 250 VA Isolation for heated sleeve Catalog No. 23043-5-8018410 23043-5-8018411 23043-5-8018412 – Insulating Mat – Temperature Controller (see page 35) 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 5 PFE2 Filter Unit Use The PFE2 filter unit consists of the FE2 filter unit with a regulated heating sleeve, installed in a special-steel protective case. It is used to filter dust-containing gases and is a component of the modular gas sampling probe system. When used for gases with high dust content, the PFE2 can optionally be fitted with valves for automatic cleaning of the filter element in the FE2 filter unit. FE2 Filter Unit Technical data see page 3 Protective Case Dimensions 450 x 450 x 400 mm (H x W x D) Material Special steel, Mat. No. 1.4301 Degree of Protection of Housing IP54 Ambient Temperature –20 to +45 °C Control Valve Solenoid Valve 3/2-way solenoid valve Weight approx. 18 kg Pressure Range 600 kPa (6 bar) Options – Annular heating element, unregulated (see page 4) – Heating sleeve, regulated (Pt 100), with insulating mat (see page 5) Connections G 1/8 internal thread DIN ISO 228/1 Material Brass Pneumatic Valve Mode of Operation Closed in normal position Diaphragm Valve 2/2-way straight-way valve Ambient Temperature 0 to +100 °C Operating Pressure / Control Pressure 600 kPa (6 bar) Electrical Connections Standard connector Nominal Diameter DN 12 Connections G 1/2 internal thread DIN ISO 228/1 Control air: G 1/8 internal thread DIN ISO 228/1 Material Brass Nominal Diameter DN 2.0 Power Supply 115 VAC or 230 VAC, 50 to 60 Hz Power Consumption 8 VA Mode of Operation Closed in the normal position Ambient Temperature 0 to +100 °C 6 System Components for Sample Gas Sampling 10/23-5.10 EN November 2006 80 165 ø11,5 ø6,5 450 50 80 M4 ø19,5 65 210 M4 75 180 155 110 2x M32 3x M25 2x M20 System Components for Sample Gas Sampling 400 300 170 160 Take holes for cable glands on the left into account Terminal box with cable glands Use as required 450 90 155 M4 M4 450 80 Flange plate with pipe clamp 2 1/2" RST 37-2 hot galvanized DIN 3568 for mounting at probe mounting adapter Elongated holes for M8 safety collar bolts Compressed air Purging Bulkhead fitting 10/12x1 mm Test gas connection Bulkhead fitting 4/6x1 mm Control air Bulkhead fitting 4/6x1 mm Sample gas connection Cover swivels upwards Electrical connections 550 300 10/23-5.10 EN November 2006 155 75 PFE2 Filter Unit Probe Protective Case Dimensional Drawing (dimensions in mm) 17 0 7 440 PFE2 Filter Unit Mounting of Probe Protective Case Minimum distance xmin of the mounting flange from the wall depending on mounting angle α: α 10° 15° 20° 25° 30° 35° xmin / mm 229 248 268 287 307 324 X α Ordering Information Catalog No. 23028-0↓ Design Without protection box with protection box, IP54 (450 x 450 x 400 mm) with protection box, IP54, anti-freeze heating with protection box, IP54, ambient temp. max. 65 °C ↓ ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 2 3 4 Filter housing and flange (equal FE2) Housing and flange SS, with filter element <0.3 µm (filtration fineness) Housing and flange SS, coated, with filter element <0.3 µm (filtration fineness) 7 8 Heating Without with heating sleeve and insulation (max. 200 °C) with ring heater, unregulated 1 2 3 Control / Monitoring Without with 1x Pt 100 with 2x Pt 100 1 2 3 Back purging unit (without compressed air unit) Without back purging with back purging, 1-stage and SS mounting flange plate with back purging, 2-stage and SS mounting flange plate w/o back p., w test/emerg. purg. valve, unheated, w/o connection adapter (ACF-NT) w back p., 1-st., with test/emerg. purg. v., heat. shut-off valve, SS conn. ad. (ACF-NT) w back p., 2-st., with test/emerg. purg. v., heat. shut-off valve, SS conn. ad. (ACF-NT) with back purging, 2-stage, heated shut-off valve, SS connection adapter (FID) 0 1 2 3 4 5 6 Power supply 230 V, 50...60 Hz 115 V, 50...60 Hz 1 2 ↓ Catalog No. 23028-0- 8 ↓ System Components for Sample Gas Sampling ↓ ↓ ↓ ↓ ↓ 10/23-5.10 EN November 2006 10/23-5.10 EN November 2006 Solenoid Valve Aeration -Y4 Diaphragm Valve Cleaning Filter Surface and Probe Tube -Y3.2 Pilot Operation Valve Cleaning Filter Surface and Probe Tube -Y3.1 Junction Box Protection Box System Components for Sample Gas Sampling Power Supply Heated Check Valve -Y5 Filter Unit B D C Heated Sample Line -E2 Pt100 Connection A Check Valve Control Air Inlet Bulk Head Union 6 mm Calibration Gas Inlet Bulk Head Union 6 mm Instrument Air Inlet Bulk Head Union 12 mm Diaphragm Valve Puls-Instrumentenluft -Y1.2 Pilot Operation Valve Puls-Instrument Air -Y1.1 Diaphragm Valve Cleaning Filter -Y2.2 Pilot Operation Valve Cleaning Filter -Y2.1 A: Calibration Gas Inlet G 1/4" -> 6 mm Tube Fitting D: Sample Outlet G 1/4" -> 6 mm or 8 mm Tube Fitting C: Back Pressure Purging Filter Surface and Probe Tube Inlet G 1/2" -> 12 mm Tube Fitting B: Back Pressure Purging Filter Inlet G 1/2" -> 12 mm Tube Fitting PFE2 Filter Unit Pneumatics Connections for 2-Stage Back-Purging, Test/Emerg. Purging Valve and Heated Shutoff Valve 9 PFE2 Filter Unit Circuit and Terminal Diagram Protection Box External Junction Box Analyzer Cabinet -X1 J L 1 N 2 PE 3 3x1,5 mm² -X1 3x2,5 mm² Heating Sleeve Filter Unit Heated Sample Line -E1 J Power Supply Heating Sleave Filter Unit, 230 or 115 VAC, 50...60 Hz, 250 W, required fusing 6 A, controller required ! L 4 J 5 -E2 6 J N 7 PE 8 3x1,5 mm² L 9 10 J -E3.1 21.10.2005 Power Supply Sample Line, 230 VAC: < 35 m -> 1~, > 35...60 m -> 3~, 115 VAC: < 15 m -> 1~, > 15...40 m -> 3~, 50...60 Hz, 90 W/m, in case of a 3~ sample line connect L2 -> Termination 5, L3 -> Termination 6, required fusing 16 A, controller required ! J N -E3.2 Power Supply Heater Probe Tube, 230 VAC, 50...60 Hz, max. 800 W, required fusing 6 A, in case of 115 VAC power supply wire the heating elements -E3.1 and -E3.2 parallel. controller required ! 11 PE 12 -X1 3x1,5 mm² -B01 set @ 30°C 3x1,5 mm² L 13 N 14 PE 15 -E4 Frost Protection Heated Probe Tube Power Supply Frost Protection, 230 or 115 VAC, 50...60 Hz, 300 W, required fusing 6 A -X1 L 16 N 17 PE 18 19 -E5.2 Heated Check Valve -E5.1 Power Supply PTC-Elements and Powerpack -G02 / Heated Solenoid Valve in Sample Outlet, 230 or 115 VAC, 50...60 Hz, 250 W, required fusing 6 A 20 21 -T01 4x1,5 mm² 115 V~ 24 V= 1,3 A -X2 3x1,5 mm² L 1 -Y3.1 N 2 PE Power Supply Pilot Operation Valve -Y3.1, Cleaning Filter Surface and Probe Tube, 230 or 115 VAC, 50...60 Hz, 8 W required fusing 1 A 3 ... L 4 -Y1.1 N 5 PE Power Supply Pilot Operation Valve -Y1.1, Puls-Instrument Air, 230 or 115 VAC, 50...60 Hz, 8 W required fusing 1 A 6 ... L 7 -Y2.1 N 8 ... 9 PE Power Supply Pilot Operation Valve -Y2.1, Cleaning Filter, 230 or 115 VAC, 50...60 Hz, 8 W required fusing 1 A L 10 -Y4 N 11 12 PE Power Supply Pilot Operation Valve -Y4, Aeration, 230 or 115 VAC, 50...60 Hz, 8 W required fusing 1 A 3 -Y5 2 -X2 1 L 13 A1 A2 -K01 N 14 15 PE 3x1,5 mm² Power Supply Relais -K05/ Check Valve -Y5, 230 or 115 VAC, 50...60 Hz, required fusing 1 A 3...18 x 0,14 mm² -X3 2x0,14 mm² 1 2x0,14 mm² 2 Resistance Thermometer PT100, Filter Unit Sample Outlet, 3-Wire-Connection 3 4 5 Resistance Thermometer PT100, Sample Line, 3-Wire-Connection 6 7 8 9 2x 0,14 mm² 10 11 12 13 14 15 16 17 18 2x 0,14 mm² 2x0,14 mm² 2x0,14 mm² 10 Resistance Thermometer PT100, Probe Tube, 3-Wire-Connection System Components for Sample Gas Sampling Resistance Thermometer PT100, Filter Unit Sample Outlet (Safety Wiring), 3Wire-Connection Resistance Thermometer PT100, Sample Line (Safety Wiring), 3-Wire-Connection Resistance Thermometer PT100, Probe Tube (Safety Wiring), 3-Wire-Connection 10/23-5.10 EN November 2006 PFE3 Filter Unit Design The PFE3 filter unit consists of a modified FE2 filter unit with a self-regulating heater. It is installed in a stainless steel protective case. Process Gas Sampling Conditions Pressure pabs = 50 to 600 kPa Temperature max. 200 °C Flow Rate 30 to 500 l/h, relative to 100 kPa and 0 °C Pressure Drop approx. 0.6 hPa at 100 l/h Dust Content max. 3 g/m3 (application recommendation) Filter Characteristics Filter Element Surface filter Installation Filter Fineness 0.3 µm Flange DN 65, PN 6, Form B to DIN 2527; positioning of holes: 45°, 135°, 225°, 315° Seal Integrity 10–4 hPa l/s Mounting Flange connect or bolt to the probe tube Dead Volume 280 ml Dimensions 50 / 20 x 135 mm Connection Ports Heater Test Gas Bulkhead fitting 6 mm (option) Sample Gas Pipe fitting 6 mm Type Self-regulating heater (400 W) Purge Gas Bulkhead fittings 18 mm (option) Temperature 180 °C (ambient temperature 20 °C) Probe Tube G 3/4 internal thread to DIN ISO 228/1 Temperature Alarm approx. 150 °C Power Supply 115 VAC or 230 VAC, 50 to 60 Hz Protective Case Options – Annular heating element (see page 4) – Heating sleeve (see page 5) Dimensions 270 x 203 x 330 mm (H x W x D) Material Special steel Mat. No. 1.4301 Materials of the Gas-carrying Parts Degree of Protection of Case IP54 Casing, Flange and Connections Special steel Mat. No. 1.4571 Weight approx. 17 kg incl. FE2 filter unit Seals FPM Ambient Temperature –20 to +45 °C 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 11 12 266 Cleaning Filter Surface and Probe Tube, Bulk Head Union 18 mm M20x1,5 System Components for Sample Gas Sampling 124 3x M20x1,5 2x M12x1,5 Junction Box -X1 IP66 200 Cleaning Filter, Bulk Head Union 18 mm Tube, CU 15x1 mm Tube, SS 316Ti, 6x1 mm Calibration Gas Bulk Head Union 6 mm Tube, CU 15x1 mm 330 58 PFE3 Filter Unit Dimensional Drawing, Gas Connections (dimensions in mm) 10/23-5.10 EN November 2006 130 270 80 PFE3 Filter Unit Circuit and Terminal Diagram Protection Box External Junction Box Analyser Cabinet -X1 Heated Aluminium Forms Filter Unit L 1 J N Power Supply Heated Aluminium Form Filter Unit, 230 or 115 VAC, 50...60 Hz, 250 W, required fusing 6 A 2 PE 3x1,5 mm² PE -E1.2 -E1.1 -E1.4 -E1.3 set @ 140°C -X1 3x2,5 mm² Heated Sample Line L 3 J N -E2 4 PE PE Power Supply Sample Line, 230 or 115 VAC, 50...60 Hz, 90 W/m, max. 3600 W / 5700 W, required fusing 16 A / 25 A, controller required ! -X1 2x0,14 mm² 5 6 2x0,14 mm² Resistance Thermometer PT100, Filter Unit Sample Outlet, 3-Wire-Connection 7 8 9 Resistance Thermometer PT100, Sample Line, 3-Wire-Connection 10 11 Temperature Alarm, Surface heated Aluminium Forms 12 13 2x0,2 mm² 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 13 PFE3 Filter Unit Ordering Information Catalog No. 23046-0↓ Design with protection box IP54 (330 x 203 x 270 mm) ↓ ↓ ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 Filter housing and flange in stainless steel Filter unit < 0.3 µm (filtration fineness), without purge connectors Filter unit < 0.3 µm, coated, without purge connectors Filter unit FE2, < 0.3 µm, with purge connectors, 2-stage Filter unit FE2, coated, < 0.3 µm, with purge connectors, 2-stage 4 5 6 7 Heating with self regulated heating and temperature alarm (115 / 230 VAC) with ring heater, unregulated with heated sleeve (max. 190 °C, Pt 100 required) 1 2 3 Controlling / Monitoring without with 1 x Pt 100 (only for heated sleeve) with 2 x Pt 100 (only for heated sleeve) 3 4 5 Connectors for back purging (without compressed air unit) without with bulkhead unions for back purging, 1-stage with bulkhead unions for back purging, 2-stage 1 2 3 Power supply 230 VAC, 50…60 Hz 115 VAC, 50…60 Hz 1 2 Test gas / emergency purging valve without with, unheated, bulkhead union for connecting 1 2 ↓ Catalog No. 23046-0- 14 ↓ 0 ↓ System Components for Sample Gas Sampling ↓ ↓ ↓ ↓ ↓ ↓ 10/23-5.10 EN November 2006 Probe tube type 60 R water cooled 3) Probe tube type 60 S water cooled 3) Probe system H water cooled 3) Probe 2 partly heated Probe 2 heated Probe tube type 42 SS heated Probe tube type 40 silicone carbide Probe tube type 40 SS 1.4893 Probe tube type 40 SS 1.4571 coated Probe tube type 60 water cooled 3) < 130°C < 180°C < 300°C 180–450°C 180–900°C < 1300°C < 1400°C Dust content < 50 mg/m3 < 1000 mg/m3 < 3 g/m3 > 3 g/m3 1) 40–100 g/m3 < 2000 g/m3 Water vapor < 65 °C dew point > 65 °C Acidic aerosols (acid dew pt. 130 °C) Pressure < 2 bar < 5 bar Technical data see … p. 16 Probe tube type 40 W SS 1.4571 Probe tube type 40 SS 1.4571 Selection Table for Probe Tube Types and Probes 1) 1) 1) 1) 1) 1) 1) 1) 1) 1) p. 18 p. 18 p. 18 p. 18 p. 19 p. 19 Temperature 2) 1) p. 16 1) p. 16 p. 16 1) p. 16 1) p. 17 1) Back purging required 2) with regulated heating sleeve (see page 5) 3) “ACK Cement Analysis System” data sheet (Document No. 10/23-8.40 EN) Lead-Through Tube Installation Brickwork with metal sheeting 1 2 5 3 Isolated sheet-metal channel 1 2 Brickwork 5 3 1 4 2 5 3 α Flow direction of process gas 1 Lead-through tube 2 Mounting flange DN 65, PN 6, Form B to DIN 2573 3 Gasket 4 Welded-on rectangular block 5 Sampling probe flange Minimum distance xmin of the mounting flange from the wall depending on mounting angle α: α xmin / mm 10° 133 15° 138 20° 143 25° 147 30° 151 35° 153 Dimensions in mm 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 15 Type 40, Type 40 W Probe Tubes Use Dimensional Drawing (dimensions in mm) For sampling hot dust-containing gases in conjunction with the FE2 / PFE2 filter unit (see page 3 or page 6) Flow Rate max. 300 l/h Pressure Drop approx. 10 hPa at 30 to 90 l/h G3/4 L1 max. 450 °C max. 180 °C max. 900 °C max. 300 °C max. 1.300 °C 16 Special Steel Material No. 1.4893 (max. 900 °C) 16,8 ø Temperature SS Mat. No. 1.4571 SS Mat. No. 1.4571 (coated) SS Mat. No. 1.4893 SS Mat. No. 1.4571 (Type 40 W) Silicone carbide (max. length 1.000 mm) 17,2 ø pabs = 50 to 500 kPa (0.5 to 5 bar) 21,4 ø Pressure 12,6 ø Process Gas Sampling Conditions Special Steel Material No. 1.4571 (max. 450 °C) G3/4 L1 Dust Content max. 3 g/m3 (application recommendation); max. 40 g/m3 with purge-back device in conjunction with FE2 / PFE2 filter unit (not for coated probe tube) 16 Silicon-Carbide (max. 1300 °C) Mounting Angle any; 8° to 90° downwards from the horizontal is recommended. It is imperative to install the silicon carbide probe tube vertically. 34 ø 15 ø G 3/4 external thread DIN ISO 228/1 Protective Tube 23 ø SiC Installation G3/4 105 L1 16 Ordering Information Catalog No. 23036-0↓ Version Stainless steel (1.4571), up to 500 °C Stainless steel (1.4571), coated, up to 180 °C Silicon carbide up to 1300 °C, max. 1000 mm Stainless steel (1.4893), up to 900°C Stainless steel (1.4571), up to 300 °C, heated (Type 40 W for cement) 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 2 4 5 6 Nominal length 500 mm 1000 mm 1500 mm 2000 mm 2500 mm 3500 mm (only for heated version) 4000 mm (only for heated version) 4500 mm (only for heated version) 1 2 3 4 5 6 7 8 ↓ Catalog No. 23046-0- 16 ↓ System Components for Sample Gas Sampling ↓ 10/23-5.10 EN November 2006 Type 42 Probe Tube (heated) Use Probe Tube Heater • Low gas temperature • High moisture content in gas • Corrosive components in the process gas Heat Output nominal length 1.000 mm: 2 x 200 W nominal length 1.500 mm: 2 x 300 W nominal length 2.000 mm: 2 x 400 W Typical places of use are: • Sampling points in wet desulphurization plants (crude gas, clean gas, part-cleaned clean gas) • Waste incineration plants • Auxiliary boilers and stack emission measurements Nominal Voltage 115 VAC or 230 VAC, 50 to 60 Hz Control Control temperature max. 200 °C Process Gas Sampling Conditions Temperature Sensor Pt 100 Pressure pabs = 200 kPa (2 bar) Continuous Operation at 200 °C flue gas temperature 450 °C Temperature < 200 °C Ordering Information Flow Rate max. 250 l/h Catalog No. 23041-4-8018392 23041-4-8018393 23041-4-8018394 Length 1000 mm Length 1500 mm Length 2000 mm Flow Velocity max. 12 m/s (> 1000 mm max. 7 m/s) Dust Content max. 3 g/m3 (application recommendation); max. 40 g/m3 with purge-back device in conjunction with FE2 / PFE2 filter unit (see page 3 or page 6) T90 Time approx. 0.4 s at 250 l/h Dimensional Drawing (dimensions in mm) and Terminal Diagram 40 17 160 21 1000/1500/2000 mm 0 Heater 120 40 x 2 25 x 1,5 13 Pt100 123456 789 Bridge: Line 230 V Terminals 3+4 Line 115 V Terminals 2+3/4+5 Length M20 Wattage (W) 400 600 800 Weight (kg) 8 10 12 (mm) 1000 1500 2000 Line Sensor 90 Temperature Controller Connection 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 17 Type 60 Probe Tube (Cement Flue Gas Probe) Use Cooling Water Supply For sampling hot cement flue gas with high dust content for CO, CO2, NO and SO2 analysis in “dry” cement production process (e.g. recuperator furnaces, furnaces with grate coolers), at rotary kiln inlet and/or after precalcination. Quality Drinking water quality with approx. 30 % anti-freeze if required Outlet Temperature min. 50 °C (or > dew point of the cement flue gas), max. 85 °C Process Gas Sampling Conditions Pressure pabs = max. 400 kPa (4 bar) Pressure pabs = max. 200 kPa (2 bar) Quantity Circulated approx. 3 m3/h (closed system) Temperature max. 1200 °C for length = 3500 mm max. 1250 °C for length = 3000 mm max. 1300 °C for length = 2500 mm max. 1350 °C for length = 2000 mm max. 1400 °C for length = 1500 mm Type 60 S Probe Tube Model with reinforced protection tube and flange DN 150, PN 16 Flow Rate max. 250 l/h Ordering Information Dust Content max. 2000 g/m3 with periodical compressed air cleaning in conjunction with FE2 filter unit (see page 3) Length 1500 mm Length 2000 mm Length 2500 mm Length 3000 mm Length 3500 mm Length 1500 mm, reinforced protection tube Length 2000 mm, reinforced protection tube Length 2500 mm, reinforced protection tube Length 3000 mm, reinforced protection tube Length 3500 mm, reinforced protection tube Type 60 S, length 1500 mm Type 60 S, length 2000 mm Type 60 S, length 2500 mm Type 60 S, length 3000 mm Type 60 S, length 3500 mm Type H, length 1500 mm Type H, length 2000 mm Type H, length 2500 mm Type H, length 3000 mm Type H, length 3500 mm Dead Time (Tt) for Length = 3500 mm approx. 40 s at 60 l/h, approx. 9 s at 250 l/h 90% Time (T90) for Length = 3500 mm approx. 82 s at 60 l/h, approx. 17 s at 250 l/h Pressure Drop approx. 10 hPa Materials of the Gas-carrying Parts Special steel, Material No. 1.4541 Mounting A flanged probe connector (internal diameter 200 mm) should be installed by the customer. The mounting flange is essential for installing the probe tube. The connections should be aligned according to the local situation. Catalog No. 23038-4-0856831 23038-4-0856832 23038-4-0856833 23038-4-0856834 23038-4-0856835 23038-4-8018478 23038-4-8018479 23038-4-8018480 23038-4-8018481 23038-4-8018482 23038-4-0730642 23038-4-0730643 23038-4-0730644 23038-4-0730645 23038-4-0730646 23020-4-0730656 23020-4-0730647 23020-4-0730648 23020-4-0730649 23020-4-0730650 Dimensional Drawing (dimensions in mm) ~60 3 2 A ø60,3 ~70 1 35 ø89 A 90 L2 L3 3 L1 18 System Components for Sample Gas Sampling 1 Sample gas inlet ports 2 Thickened shank of probe tube 3 Cooling water connections L1 Probe tube length L2 Length of thickened shank: 800 mm for L1 ≤ 2000 mm 1300 mm for L1 ≥ 2500 mm L3 200 mm with reinforced protection tube 10/23-5.10 EN November 2006 Probe 2 Use Wherever large volumes of gas are sampled, gas pressure is raised and there is a high moisture or dust content in the gas. Electrical Connections Harting connector, IP55 Probe Tube Heating Process Gas Sampling Conditions Pressure pabs = max. 400 kPa (4 bar) Heat Output (unregulated) Nominal length 1200 mm: 650 W partially heated (for heating of the wall bushing): 255 W Nominal Voltage 230 VAC, 50 to 60 Hz (115 VAC on inquiry) Temperature max. 450 °C (probe tube partly heated) max. 130 °C (probe tube heated Protective Case (Option) Flow Rate max. 2000 l/h Dust Content max. 20 g/m3 (application recommendation), max. 100 g/m3 with purging Dimensions 270 x 203 x 330 mm (H x W x D) Material Special steel Mat. No. 1.4301 Purging Pressure max. 600 kPa (6 bar) Degree of Protection of Case IP54 Heating Temperature of Inlet Filter ≥ 200 °C (at process gas temperature of 20 °C) Weight approx. 7 kg 90% Time (T90) < 2 s at 250 l/h Ambient Temperature –20 to +45 °C Filter Fineness 0.3 µm Dimensional Drawing and Connection Diagram Z Material of the Gas-carrying Parts Support Tube, Heating Rod, Connecting Pipe Special steel, Material No. 1.4571 650 1 Inlet Filter Aluminum oxide α Seals Ceramic fiber 1 80 Z Flange Special steel, Material No. 1.4581 1=N 2 = L1 4 260 Mounting Flange DN 65, PN 6, Form B, to DIN 2573 position of holes: 45°, 135°, 225°, 315° 1 Filter mounting 4 Harting connector 2 Gas sampling probe tube 3 Sample gas outlet or test gas / purge air inlet Mounting Angle any; 10° to 35° downwards from the horizontal recommended (take direction of flow into account), see dim. drawing Ordering Information Sample Gas Outlet G 1/4 internal thread DIN ISO 228/1 Test Gas / Purge Air Inlet G 3/8 internal thread DIN ISO 228/1 10/23-5.10 EN November 2006 Catalog No. 1200 mm, 230 V, 50/60 Hz, 650 W 23022-4-0768301 1200 mm, 230 V, 50/60 Hz, 255 W, part. heat.23022-4-0768305 Prot. box, insulat., IP54, (330x203x270 mm) 23022-4-0730719 System Components for Sample Gas Sampling 19 Probe 11 Probe 11 Dimensional Drawing (dimensions in mm) Use in interior air, garage and tunnel monitoring, including corrosive atmospheres. The materials used ensure that with the gases normally occurring no adsorption occurs. Technical Data Sampling Pressure Atmospheric pressure Gas Temperature –25 to +50 °C, dry Sample Gas Flow Rate max. 70 l/h Dust Content max. 1 mg/m3 Pressure Drop with New Filter ≤ 10 hPa (mbar) at 70 l/h Service Life of Filter (estimated) filter constant (2500) d (days) = flow rate (l/h) x dust concentration (mg/m3) Ordering Information Sample Gas Connection Port Bolted connection for flexible pipe with internal diameter 4 mm and external diameter 7 mm Room air probe 11 Catalog No. 23021-4-0859585 Materials Polycarbonate (probe head), aluminum oxide (filter) Transport and Storage Temperature –25 to +70 °C Mounting On wall or suspended from ceiling 20 System Components for Sample Gas Sampling 10/23-5.10 EN November 2006 Ceramic Filter Use For filtering dust from gases in the gas conditioning systems of analyzer systems when the gas sampling probe does not have a filter. The ceramic filter is positioned upstream the other devices in the sample gas conditioning chain. The ceramic filter can be heated with the annular heating element if condensation from the sample gas is likely. Technical Data Flow Rate (Air) max. 300 l/h Gas Pressure pabs = 20 kPa to 1.7 MPa (0.2 to 17 bar) at 20 °C filter temperature, max. 200 kPa (2 bar) at 180 °C Pressure Drop 0.1 to 0.85 hPa at 30 to 250 l/h (air) Dynamic Characteristics at Flow rate (air) 30 l/h 3.7 s Dead time Tt 11.5 s 90% time T90 60 l/h 2.3 s 5.5 s 90 l/h 1.7 s 3.9 s Filter Fineness 0.3 µm Dimensional Drawing (dimensions in mm) Construction, Mounting Materials of the Gas-carrying Parts Housing: steel, St 37 K or special steel, Material No. 1.4571; Filter: aluminum oxide; Seals: FPM, Cu, Al2O3 fiber incl. T-handle (not shown) Gas Connection Ports G 1/4 internal thread DIN ISO 228/1 or pipe connectors for pipe of internal diameter 4 mm Filter Surface approx. 212 cm2 Weight approx. 2.1 kg Filter Heating Annular heating element, see page 4 Mounting On wall with angle bracket Ordering Information Ceramic filter, steel version Ceramic filter, stainless steel version 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling Catalog No. 23031-5-8018177 23031-5-8018178 21 Process Pressure Reducers Use If the pressure of the process gas is too high for the analyzer, a pressure reducer is required to lower the pressure. The type of pressure reducer used is determined by the properties of the process gas and the required sample gas pressure. Technical Data Casing: Brass Casing: Brass Casing: Special steel Casing: Special steel Single-stage with one manometer (outlet) max. 1.5 MPa (15 bar) 0.8 to 10 kPa (8 to 100 mbar) Single-stage without manometer max. 3.5 MPa (35 bar) 10 to 70 kPa (0.1 to 0.7 bar) 80 kPa (0.8 bar) – 25 to 250 l/h Special steel EPDM Special steel Single-stage with two manometers (inlet / outlet) max. 3.5 MPa (35 bar) 10 to 70 kPa (0.1 to 0.7 bar) 80 kPa (0.8 bar) 0 to 4 MPa (0 to 40 bar) 0 to 150 kPa (0 to 1.5 bar) 25 to 250 l/h Special steel EPDM Special steel approx. 2.35 kg max. 50 °C approx. 2.35 kg max. 50 °C 23422-4-0730699 23422-4-0730700 Weight Gas Temperature and Ambient Temperature approx. 1.25 kg max. 50 °C Single-stage with two manometers (inlet / outlet) max. 3.5 MPa (35 bar) 10 to 40 kPa (0.1 to 0.4 bar) 50 kPa (0.5 bar) 0 to 4 MPa (0 to 40 bar) 0 to 60 kPa (0 to 0.6 bar) 25 to 250 l/h Buna rubber Buna rubber Special steel; gascarrying parts: brass approx. 2.65 kg max. 50 °C Catalog No. 23035-5-5864958 23035-5-5864951 Construction Inlet Pressure Outlet Pressure (adjustable) Safety Valve Set to – Manometer Display Inlet Pressure Manometer Display Outlet Pressure Flow Rate Material Diaphragm Material Gasket Material Housing – 0 to 250 hPa (0 to 250 mbar) 25 to 250 l/h Buna rubber Buna rubber Brass – Supplied: Pressure reducer with angle bracket 22 System Components for Sample Gas Sampling 10/23-5.10 EN November 2006 Process Pressure Reducers Dimensional Drawings (dimensions in mm) Pressure reducer single-stage with one manometer (outlet) (Catalog No. 23035-5-5864958) Pressure reducer single-stage without manometer (Catalog No. 23422-4-0730699) x. 30 appro Ø4 79 able) ° (vari 105 GasTech approx. 94 (depending on handwheel position) Pressure reducer single-stage with two manometers (inlet / outlet) (Catalog No. 23035-5-5864951) Pressure reducer single-stage with two manometers (inlet / outlet) (Catalog No. 23422-4-0730700) 118 appro ble) (varia 148 Ø4 ° x. 30 GasTech approx. 94 (depending on handwheel position) 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 23 Heated Sample Gas Lines Heated Sample Gas Line CGWB 13 Parallel strip heater (holding temperature max. 200 °C) Pt 100 resistance thermometer for regulation Self-regulating anti-frost heater (max. 30 °C) Self-regulating heater (option, holding temp. max. 100 °C) Heated sample gas line TBL 01-S Fixed resistor heat. conductor (holding temp. max. 200 °C) Pt 100 resistance thermometer for regulation Self-regulating heater (option, holding temp. max. 100 °C) Heated sample gas line TBL 01-C Self-regulating heater (holding temp. max. 100 / 120 °C) Integral pipe-end connectors for cement analysis systems Heated sample gas line TBL 01-E Competitive version Pipe ends with polyamide hard caps Selection Table for Heated Sample Gas Lines Heater Self-regulating strip heater (30 °C) Self-regulating strip heater (100 °C) Self-regulating strip heater (120 °C) Fixed resistor heat. cond. (200 °C) Parallel strip heater (200 °C) Outer covering Options Line can be shortened Small bending radius Gas connections Electrical connections Maximum line length Applications CO, O2, CO2 CO, O2, CO2 SO2, NOx, CnHm Cement analysis systems 24 CGWB 13 TBL 01-S TBL 01-C TBL 01-E x x – – x Polyvinyl chloride (PVC) – x – x – Polyamide (PA) – x x – – Metal or PVC – – – x – Polyamide (PA) x – Various options Heater and Pt 100 fixed 130 m – x Various options Heater and Pt 100 fixed 100 m – x Various options Heater and Pt 100 fixed 70 m – x PTFE Heater and Pt 100 fixed 40 m Anti-frost heater – – Holding temp. 100 °C Holding temp. 100 °C – Holding temp. 200 °C Holding temp. 200 °C – – – Holding temp. ≤ 120 °C System Components for Sample Gas Sampling – – Holding temp. 200 °C – 10/23-5.10 EN November 2006 Heated Sample Gas Lines Integral Pipe-end Connectors Flange Silicone cap Metal cap POM cap and shrunk-on hose Hard cap 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 25 CGWB 13 Heated Sample Gas Line Technical Data Frost protected line Outer Covering Medium-carrying Pipe Integral Pipe-end Connectors PVC PTFE Various options (see ordering information, page 28) Heater self-regulating strip heater Power Supply 230 V or 115 V, 50 to 60 Hz Heat Output 16 W/m Making / holding current Current Consumption = 65 / 25 mA Holding Temperature see diagram Ambient Temperature –30 to +60 °C Outer Diameter 30 mm Flexible Sample Gas Pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm Maximum Line Length 130 m Line Duct M42 Minimum Bending Radius 300 mm Weight approx. 1 kg/m Connect. Lead for Heater (fixed)3 x 2.5 mm2; length 5 m Connect. Lead for Pt 100 (fixed) – Self-regulating strip heater Regulated parallel strip heater PVC PTFE Various options (see ordering information, page 28) self-regulating strip heater 230 V or 115 V, 50 to 60 Hz 30 W/m 0.15 A/m PVC PTFE Various options (see ordering information, page 28) CPD heating element, regulated 230 V or 115 V, 50 to 60 Hz 60 W/m 0.3 A/m see diagram –30 to +60 °C 43 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 90 m M42 300 mm approx. 1 kg/m 3 x 2.5 mm2; length 5 m – 200 °C –30 to +60 °C 40 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 60 m M42 300 mm approx. 1 kg/m 3 x 2.5 mm2; length 5 m LIYCI 2 x 0.14 mm2; length 5 m Holding Temperature Diagrams Medium pipe temperature °C Frost protected line, 6 x 4 x 1 mm PTFE, 16 W/m, self-regulating 50 40 30 20 - 30 - 20 - 10 0 + 10 + 20 + 30 + 40 Ambient temperature °C Medium pipe temperature °C Self-regulating strip heater, 6 x 4 x 1 mm PTFE, 30 W/m 100 90 80 70 - 30 - 20 - 10 0 + 10 + 20 + 30 + 40 Ambient temperature °C 26 System Components for Sample Gas Sampling 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 54 46 63 Replacable tube nipple, 316ss, 6x1 or 8x1 mm 300 25 Sample take-off side Sample take-off side Length by customer’s request max. length depending on MCB or switching load of the controller respectively Length by customer’s request max. length depending on MCB POM cap Shrinking hose POM cap Shrinking hose Pt100 Cable 2 x 0.14 mm² Power Supply Cable 3 x 1.5...2.5 mm² 300 Analyzer side Power Supply Cable 3 x 1.5...2.5 mm² 300 Analyzer side CGWB 13 Self-regulating strip heater 49 42 42 10/23-5.10 EN November 2006 46 46 Process tube overhang, SS-nipple 6x1 or 8x1 mm CGWB 13 Heated Sample Gas Line Dimensional Drawings (dimensions in mm) CGWB 13 Regulated parallel strip heater 27 CGWB 13 Heated Sample Gas Line Ordering Information Catalog No. 23281-0↓ Version Parallel strip heater 60 W/m, line can be shortened, regulated (Pt 100), 200 °C Self-regulating 16 W/m, 30 °C (antifreeze) Self-regulating 30 W/m, 100 °C ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 2 3 7 Integral pipe-end connectors Probe: Metal cap for PFE2/3 – Analysis: Plastic cap with shrunk-on hose Probe: Metal cap for PFE2/3 – Analysis: Flange Both sides with POM plastic cap (for self-regulated versions only) Probe: PTFE cap for PFE2/3 in conf. set – Analysis: POM plastic cap Probe: Metal cap for PFE2/3 in conf. set – Analysis: Plastic cap with shrunk-on hose 2 3 4 5 6 Gas conducting line PTFE tube 6x4x1 mm PTFE tube 8x6x1 mm 1 2 Power supply 230 V, 50 to 60 Hz 115 V, 50 to 60 Hz 1 2 Controlling and monitoring without Pt 100 (for self-regulated versions only) with 1 x Pt 100 (fixed built-in) 0 1 ↓ Catalog No. 23281-0- 28 ↓ 0 ↓ System Components for Sample Gas Sampling ↓ ↓ ↓ ↓ 10/23-5.10 EN November 2006 TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines Technical Data TBL 01-S Self-regulating strip heater Outer Covering Medium-carrying Pipe Integral Pipe End Connectors Power Supply Heat Output Holding Temperature Ambient Temperature Outer Diameter Maximum Line Length Line Duct Minimum Bending Radius Weight Electrical Connection for Heater and Pt 100 Outer Covering Medium-carrying Pipe Integral Pipe End Connectors Power Supply Heat Output Holding Temperature Ambient Temperature Outer Diameter Maximum Line Length Line Duct Minimum Bending Radius Weight Electrical Connection for Heater and Pt 100 10/23-5.10 EN November 2006 TBL 01-S Fixed resistor heating conductor Polyamide corrugated tubing Polyamide corrugated tubing PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm Probe: metal cap, analyzer: flange or hard caps on both sides (pipe projecting 300 mm) or silicon caps on both sides (probe: 6/4 stainless steel connections, analyzer: pipe projecting 300 mm) 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz 30 W/m 90 W/m see diagram on page 26 200 °C –30 to +60 °C –30 to +60 °C 43 mm 43 mm 100 m 40 m M48 M48 200 mm 200 mm approx. 1 kg/m approx. 1 kg/m fixed, 3 x 2.5 mm2, length 5 m fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m TBL 01-C Self-regulating strip heater TBL 01-E Fixed resistor heating conductor Metal corrugated tubing (for ACK) or PVC outer covering (for ACMP) PTFE pipe 6 x 4 x 1 mm Probe: metal cap – analyzer: silicon cap (mandatory for ACK) or POM plastic caps on both sides (mandatory for ACMP) or probe: PTFE cap – analyzer: POM plastic cap (mandatory for ACMP 120°C) 230 V or 115 V, 50 to 60 Hz 30 W/m or 47 W/m 100 °C or 120 °C –30 to +60 °C 43 mm 70 m M48 200 mm approx. 1 kg/m fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m Polyamide corrugated tubing PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm Polyamide hard caps 230 V or 115 V, 50 to 60 Hz 90 W/m 200 °C –30 to +60 °C 43 mm 40 m M48 200 mm approx. 1 kg/m fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m System Components for Sample Gas Sampling 29 50 System Components for Sample Gas Sampling 50 56 42 Ø 6 mm 300 Replacable tube nipple, 316ss, 6x1 or 8x1 mm 60 25 8 6 Power supply Cable 3 x 1.5...2.5 mm² 81 PA corrugated hose Analyzer side Power supply Cable 3 x 1.5...2.5 mm² Hard cap PA corrugated hose Length by customer’s request max. length depending on MCB or switching load of the controller respectively Pt100 Cable 2 x 0.14 mm² Length by customer’s request max. length depending on MCB 350 Hard cap 81 24 25 Sample take-off side 300 Process tube overhang 54 350 42 42 Sample take-off side 63 30 49 TBL 01-S, Pipe end connectors with hard caps 49 Analyzer side TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines Dimensional Drawings (dimensions in mm) TBL 01-S, Pipe end connectors with flange and metal cap 10/23-5.10 EN November 2006 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 50 300 Process tube overhang Hard cap 2 x M16 Cable Gland Power supply Cable 3 x 1.5...2.5 mm² Length by customer’s request max. length depending on MCB or switching load of the controller respectively PA corrugated hose Pt100 Cable 2 x 0.14 mm² Power cupply Cable 3 x 1.5...2.5 mm² Length by customer’s request max. length depending on MCB or switching load of the controller respectively 42 Hard cap 350 81 54 49 6 300 56 42 Ø 6 mm Process tube overhang Sample take-off side 24 25 Sample take-off side TBL 01-S, Pipe end connectors with metal caps for ACF-NT Analyzer side Replaceable tube nipple, 316ss, 8x1 mm PA corrugated hose Pt100 Cable 2 x 0.14 mm² 63 50 60 Analyzer side TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines Dimensional Drawings (dimensions in mm) TBL 01-E 31 42 54 300 Process tube overhang Replacable tube nipple, 316ss, 6x1 mm 63 49 25 Sample take-off side 81 45 Length by customer’s request max. length depending on MCB PVC Jacket Length by customer’s request max. length depending on MCB or switching load of the controller respectively System Components for Sample Gas Sampling 300 Pt100 Cable 2 x 0.14 mm² Power Supply Cable 3 x 1.5...2.5 mm² 300 Analyzer side Pt100 Cable 2 x 0,14 mm² Power supply Cable 3 x 1,5...2,5 mm² POM cap Shrinking hose Silicone cap Process tube overhang Analyzer side TBL 01-C, Pipe end connectors for ACK 46 Metall corrugated hose 51 32 46 Sample take-off side TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines Dimensional Drawings (dimensions in mm) TBL 01-C, Pipe end connectors for ACMP 10/23-5.10 EN November 2006 42 TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines TBL 01-S Ordering Information Catalog No. 23283-0↓ Version Fixed resistor heating strip, 90 W/m, Holding temperature 200 °C Self-regulated heating strip, 30 W/m, Holding temperature 100 °C Fixed resistor heating strip for ACF-NT, 90 W/m, Holding temperature 200 °C ↓ ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 2 4 Gas conducting line with PTFE tube 6x4x1 mm with PTFE tube 8x6x1 mm (mandatory for ACF-NT) 1 3 Material of sheathing with plastic hose of PA (outside and indoors installation) 1 Connections Metal caps w. 6/4 SS connectors on both sides (mandatory for ACF-NT) Probe: Metal cap / Analyzer: Flange, 6/4 SS connectors on both sides Plastic hard caps on both sides (tube 300 mm surplus, self-regulated only) Silicone caps both sides / probe 6/4 SS conn. / Analyzer: tube 300 mm surplus A B C D Control / Monitoring without Pt 100 (self-regulated version only) with 1 x Pt 100 0 1 Power supply 230 V, 50 to 60 Hz, for self-regulating strip heater (not for ACF-NT) 115 V, 50 to 60 Hz, for self-regulating strip heater (not for ACF-NT) 230 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m 115 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m 230 V, 50 to 60 Hz, for fixed resistor heat. cond. 90 W/m (for ACF-NT, length ≤ 35 m) 115 V, 50 to 60 Hz, for fixed resistor heat. cond. 90 W/m (for ACF-NT, length ≤ 15 m) 400 V / 230 V, 50 to 60 Hz, for fixed res. heat. cond. 90 W/m (ACF-NT, length 35-60 m) 200 V / 115 V, 50 to 60 Hz, for fixed res. heat. cond. 90 W/m (ACF-NT, length 15-40 m) 1 2 3 4 5 6 7 8 ↓ Catalog No. 23283-0- 10/23-5.10 EN November 2006 ↓ 0 ↓ System Components for Sample Gas Sampling ↓ ↓ ↓ ↓ ↓ 33 TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines TBL 01-C Ordering Information Catalog No. 23285-0↓ Design Self-regulating strip heater, 30 W/m, 100 °C (ACK) Self-regulating strip heater, 30 W/m, 100 °C (ACMP coal bin) Self-regulating strip heater, 47 W/m, 120 °C (ACMP pre-heater) ↓ ↓ ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 2 3 4 Gas conducting line with PTFE tube 6x4x1 mm 1 Material of sheathing with metal corrugated tubing for ACK (outdoor and indoor installation) with PVC sheathing for ACMP (outdoor and indoor installation) 1 2 Connections Probe: Metal cap, Analyzer: Silicone cap (mandatory for ACK) + shortening kit POM plastic caps on both sides (mandatory for ACMP ) + shortening kit Probe: PTFE caps, Analyzer: POM caps (mandatory for ACMP 120°C ) + shorten. kit 1 2 3 Control / Monitoring with 1 x Pt 100 1 Power supply 230 V, 50 to 60 Hz 115 V, 50 to 60 Hz 1 2 ↓ ↓ Catalog No. 23285-0- ↓ ↓ ↓ ↓ TBL 01-E Ordering Information Fixed resistor heating conductor 90 W/m, PTFE 6x4x1, PA caps, 230 V Fixed resistor heating conductor 90 W/m, PTFE 8x6x1, PA caps, 230 V Fixed resistor heating conductor 90 W/m, PTFE 6x4x1, PA caps, 115 V Fixed resistor heating conductor 90 W/m, PTFE 8x6x1, PA caps, 115 V 34 Catalog No. 23284-4-0730695 23284-4-0730696 23284-4-0730697 23284-4-0730698 System Components for Sample Gas Sampling 10/23-5.10 EN November 2006 Electronic Temperature Controller and Switching Relay Temperature Controller Connection Diagram 48 x 48 mm, C-rail housing Weight 140 g Catalog No. 23345-5-8328907 parameters not set 23345-5-8328907P parameters set for filter unit 23345-5-8328907L parameters set for sample gas line 23345-5-8328907C parameters set for self-regulating heater for ACMP L1 N 342 343 341 142 143 Alarm 1 Temperature max. 230V /3A 0/12V 81 83 max. 230V /3A Alarm 2 Limiter 242 243 80 112 111 113 Logic Outputs Switching Relay Load Voltage not used Dimensions + 110 to 240 VAC +10 / –15 %, 48 to 63 Hz + 2 x relay 3 A, 250 VAC Power Supply Binary Output 2 Outputs GND Measurement Input Pt 100 Binary Output 1 –200 to +850 °C 110...240 VAC -15/+10% 48...63 Hz Control Range Urms = 24 to 265 V 4+ 3- Max. Load Current Irms = 30 A (ambient temperature = 25 °C) Weight 200 g Catalog No. 23345-5-8328731 M2A ϑ 1 L N 2 Last For load current > 2 A use coupling relay L 10/23-5.10 EN November 2006 System Components for Sample Gas Sampling 35 The Company’s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice. www.abb.com Printed in the Fed. Rep. of Germany (11.06) © ABB 2006 ABB Automation GmbH Analytical Stierstaedter Strasse 5 60488 Frankfurt am Main Germany Hotline: +49 69 793040 Fax: +49 69 79304566 EMail: analyticalmkt.deapr@de.abb.com 10/235.10 EN Rev. 6 ABB has Sales & Customer Support expertise in over 100 countries worldwide.
© Copyright 2024