Quasar 900 User's and Maintenance Manual Open-Path Gas Detection System

Quasar 900
Open-Path Gas Detection System
User's and Maintenance Manual
Factory Mutual
Class I Div. 1 Groups B, C, D
Class II Div. 1 Groups E, F, G
ATEX, IECEx
Ex II 2 (2) GD,
Ex d e ib [ib Gb] IIB+H2 T4 Gb
Ex tb IIIC T135°C Db
Document Ref: TM 888200, Rev. I, January 2014
218 Little Falls Rd., Cedar Grove, NJ 07009 USA;
Phone: +1 (973) 239 8398
Fax: +1 (973) 239 7614
Web-Site: www.spectrex.net ; Email: spectrex@spectrex.net
SafEyeTM Quasar 900 Gas Detector
ii
Legal Notice
The SafEye monitoring system described in this document is the property of Spectrex, Inc.
No part of the hardware, software or documentation may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any language or computer
language, in any form or by any means, without prior written permission of Spectrex, Inc.
While great efforts have been made to assure the accuracy and clarity of this document,
Spectrex, Inc. assumes no liability resulting from any omissions in this document, or from
misuse of the information obtained herein. The information in this document has been carefully
checked and is believed to be entirely reliable with all of the necessary information included.
Spectrex Inc. reserves the right to make changes to any products described herein to improve
reliability, function, or design, and reserves the right to revise this document and make
changes from time to time in content hereof with no obligation to notify any persons of
revisions or changes. Spectrex, Inc. does not assume any liability arising out of the application
or any use of any product or circuit described herein; neither does it convey license under its
patent rights or the rights of others.
Warning: This manual should be carefully read by all
individuals who have or will have responsibility for using,
maintaining or servicing the product.
The Source and Detector are not field-repairable due to the
meticulous alignment and calibration of the sensors and the
respective circuits. Do not attempt to modify or repair the
internal circuits or change their settings, as this will impair
the system's performance and void the Spectrex, Inc.
Product warranty.
.
Legal Notice
iii
SafEyeTM Quasar 900 Gas Detector
Release History
Rev
Date
Revision History
Prepared by
Approved by
F
February
First Release
Ian Buchanan
Eric Zinn
G
June 2013
Second release
Ian Buchanan
Eric Zinn
H
August 2013
Third Release
Ian Buchanan
Eric Zinn
I
January 2014
Fourth Release
Ian Buchanan
Eric Zinn
iv
Release History
About this Guide
This manual describes the Quasar 900 Open-Path Gas Detection System
and its features and provides instructions on the installation, operation and
maintenance.
This guide includes the following chapters and appendices:

Chapter 1, Scope, provides a general introduction and overview of the
product and the guide, with a brief description of its content.

Chapter 2, Technical Description, describes the Detector’s theory of
operation.

Chapter 3, Operation Mode, describes the Detector’s operation
modes, user interface and indications.

Chapter 4, Technical Specifications, describes the Detector’s
electrical, mechanical and environmental specifications.

Chapter 5, Installation Instructions, describes how to install the
Detector, including wiring and mode settings.

Chapter 6, Operating Instructions, describes the operating
instructions and power-up procedures.

Chapter 7, Maintenance Instructions, contains the maintenance and
support procedures.

Chapter 8, Troubleshooting, describes the solutions to problems that
may arise with the Detector.

Appendix A, Wiring Option Configurations, provides wiring diagrams
for installation.

Appendix B, Accessories, provides a list of accessories available for
the Quasar 900 Open-Path Gas Detection System.

Appendix C, SIL-2 Features, details the special conditions to comply
with the requirements of EN 61508 for SIL-2.
About this Guide
v
SafEyeTM Quasar 900 Gas Detector
Abbreviations and Acronyms
vi
Abbreviation
Meaning
ATEX
Atmosphere Explosives
AWG
American Wire Gauge
BIT
Built In Test
EMC
Electromagnetic Compatibility
EOL
End of Line
FOV
Field of View
HART
Highway Addressable Remote Transducercommunication protocol
IAD
Immune at Any Distance
IECEx
International Electrotechnical Commission Explosion
IPA
Isopropyl Alcohol
IR
Infrared
JP5
Jet Fuel
Latching
Refers to relays remaining in the ON state even after
the ON condition has been removed
LED
Light Emitting Diode
LPG
Liquefied Petroleum Gas
mA
MilliAmps (0.001 amps)
MODBUS
Master-slave messaging structure
N.C.
Normally Closed
N.O.
Normally Open
N/A
Not Applicable
NFPA
National Fire Protection Association
NPT
National Pipe Thread
SIL
Safety Integrity Level
UNC
Unified Coarse Thread
VAC
Volts Alternating Current
Abbreviations and Acronyms
Table of Contents
Quasar 900 Open-Path Gas Detection System
User's and Maintenance Manual ....................................................................... i
Legal Notice .................................................................................................... iii
Release History ............................................................................................... iv
About this Guide.............................................................................................. v
Abbreviations and Acronyms ............................................................................. vi
1
Scope ......................................................................................................... 1
1.1
2
Product Overview ................................................................................... 1
Technical Description ................................................................................ 3
2.1
Features ............................................................................................... 3
2.2
Applications........................................................................................... 4
2.3
Principle of Operation ............................................................................. 4
2.3.1
Definitions of Terms ......................................................................... 4
2.3.2
Spectral Finger Print ......................................................................... 4
2.3.3
Optical Path .................................................................................... 5
2.3.4
Microprocessor Based ....................................................................... 5
2.3.5
Gas Sensitivity ................................................................................ 5
2.3.6
Gas Calibration ................................................................................ 5
2.3.7
Flash Source ................................................................................... 5
2.3.8
Heated Optics .................................................................................. 6
2.3.9
HART Protocol ................................................................................. 6
2.3.10
Handheld Unit ................................................................................. 6
2.3.11
Modbus RS-485 ............................................................................... 7
2.3.12
Tilt Mount ....................................................................................... 7
2.4
Product Certification ............................................................................... 7
2.4.1
ATEX, IECEx .................................................................................... 7
2.4.2
FM / FMC ........................................................................................ 8
2.4.3
SIL-2.............................................................................................. 8
2.4.4
Functional Test ................................................................................ 8
2.5
Models and Types .................................................................................. 8
2.6
Description ............................................................................................ 9
2.6.1
Flash Source Unit ............................................................................ 10
2.6.2
Detector Unit .................................................................................. 11
Table of Contents
vii
SafEyeTM Quasar 900 Gas Detector
3
Operating Modes...................................................................................... 13
3.1
Operational Modes ................................................................................13
3.1.1
Normal Mode .................................................................................. 13
3.1.2
Maintenance Call Mode (3mA Output) ................................................ 13
3.1.3
Fault Mode ..................................................................................... 14
3.1.4
Zero Calibration Mode (1mA Output) ................................................. 14
3.2
Visual Indicators ...................................................................................15
3.3
Output Signals ......................................................................................15
3.3.1
0-20mA Current Output ................................................................... 15
3.3.2
RS-485 Interface ............................................................................ 16
3.4
4
System Setup .......................................................................................16
3.4.1
Detection Function Programming ...................................................... 16
3.4.2
Detection Setup Function ................................................................. 17
3.4.3
Detector Default Setup .................................................................... 18
Technical Specifications ........................................................................... 19
4.1
General Specifications ...........................................................................19
4.2
Electrical Specifications ..........................................................................20
4.2.1
Power Consumption ........................................................................ 20
4.2.2
Electrical input protection ................................................................ 20
4.2.3
Electrical outputs ............................................................................ 20
4.3
Mechanical Specifications .......................................................................21
4.4
Environmental Specifications ..................................................................22
5
4.4.1
High Temperature ........................................................................... 22
4.4.2
Low Temperature ............................................................................ 22
4.4.3
Humidity ........................................................................................ 22
4.4.4
Salt and Fog ................................................................................... 22
4.4.5
Water and Dust .............................................................................. 22
4.4.6
Shock and Vibration ........................................................................ 22
4.4.7
Electromagnetic Compatibility (EMC) ................................................. 23
Installation Instructions .......................................................................... 25
5.1
Introduction .........................................................................................25
5.2
General Considerations ..........................................................................25
viii
5.2.1
Personnel ....................................................................................... 25
5.2.2
Tools Required................................................................................ 25
5.2.3
Site Requirements .......................................................................... 26
Table of Contents
TM 888200 Rev (I), January 2014
6
5.2.4
The Source and Detector ................................................................. 26
5.2.5
Guidance Tips for Gas Detector Locations........................................... 26
5.3
Preparations for Installation ...................................................................26
5.4
Certification Instructions ........................................................................28
5.5
Conduit/Cable Installation ......................................................................29
5.6
Detector/Source Mounting ......................................................................29
5.6.1
Tilt Kit P/N 888270.......................................................................... 29
5.6.2
Detector/Source Installation ............................................................. 29
5.7
Detector Wiring ....................................................................................30
5.8
Detector Terminal Wiring .......................................................................34
5.9
Flash Source Wiring...............................................................................34
5.9.1
Wiring ........................................................................................... 34
5.9.2
Terminal Wiring .............................................................................. 35
Operating Instructions ............................................................................ 37
6.1
SafEye Operation ..................................................................................37
6.2
Alignment of Unit ..................................................................................37
6.3
Powering up the System ........................................................................38
6.4
Safety Precautions ................................................................................39
6.5
Signal Verification .................................................................................39
6.5.1
7
8
Signal Values Limitation ................................................................... 39
6.6
Zero Calibration ....................................................................................40
6.7
Functional Check of Unit ........................................................................41
Maintenance Instructions ........................................................................ 43
7.1
General Maintenance .............................................................................43
7.2
Periodic Maintenance .............................................................................43
7.2.1
Routine Optical Surface Cleaning ...................................................... 44
7.2.2
Signal Verification ........................................................................... 44
7.2.3
Function Check of Unit ..................................................................... 44
Troubleshooting ...................................................................................... 45
Appendices .................................................................................................... 47
A
Wiring Configurations .............................................................................. 49
A.1
RS-485 Communication Network .............................................................53
Table of Contents
ix
SafEyeTM Quasar 900 Gas Detector
B
C
Accessories .............................................................................................. 55
B.1
Tilt Mount P/N 888270 ...........................................................................55
B.2
Pole Mount (U-Bolt 5 inch) P/N 799225 ....................................................55
B.3
Commissioning Kit P/N 888247 ...............................................................55
B.4
RS485 Hand-Held Diagnostic Unit P/N 799810 ..........................................55
B.5
HART Hand-Held Diagnostic Unit P/N 888810 ...........................................56
B.6
HART Hand-Held Harness Kit P/N 888815 ................................................56
B.7
USB/RS485 Harness Converter Kit P/N 794079-8 ......................................56
B.8
Mini Laptop Kit P/N 777820 ....................................................................56
B.9
Sunshade P/N 888263 ...........................................................................56
SIL-2 Features ......................................................................................... 57
C.1
Safety Relevant Parameters for the Quasar 900 Open Path Gas Detector .....57
C.2
General Conditions for Safe Use ..............................................................58
Technical Support ......................................................................................... 59
x
Table of Contents
TM 888200 Rev (I), January 2014
List of Figures
Figure 1: P/N Definition of Quasar 900 .................................................................. 9
Figure 2: Flash Source ....................................................................................... 10
Figure 3: Detector ............................................................................................. 11
Figure 4: Tilt Mount ........................................................................................... 31
Figure 5: Detector and Tilt Mount Assembly .......................................................... 32
Figure 6: Detector with Cover Removed ............................................................... 33
Figure 7: Source with Cover Removed ................................................................. 35
Figure 8: Magnetic Mode Selector ........................................................................ 40
Figure 9: Detector Wiring Terminal ...................................................................... 49
Figure 10: Source Wiring Terminal ...................................................................... 50
Figure 11: 0-20mA Sink 4-Wire........................................................................... 51
Figure 12: 0-20mA Non-isolated Sink 3-Wire ........................................................ 51
Figure 13: 0-20mA Source 3-Wire ....................................................................... 52
Figure 14: RS-485 Networking for Wiring Option 3 ................................................ 53
List of Figures
xi
SafEyeTM Quasar 900 Gas Detector
List of Tables
Table 1: Gas Concentrations Measurement Terms .................................................. 4
Table 2: Model Numbers and Installation Distances ................................................ 8
Table 3: Detector LED Indications ....................................................................... 15
Table 4: Source LED Indications .......................................................................... 15
Table 5: Standard (default) 0-20mA Current for the Gas Channel ............................ 16
Table 6: Detector Default Set Up ......................................................................... 18
Table 7: Source Default Set Up ........................................................................... 18
Table 8: Detection Distance Range ...................................................................... 19
Table 9: Detector and Source Maximum Power Consumption .................................. 20
Table 10: Tilt Mount Kit...................................................................................... 29
Table 11: Wiring Options .................................................................................... 34
Table 12: Flash Source Wiring Options ................................................................. 35
Table 13: Maintenance Channels Limit Values ....................................................... 39
Table 14: Troubleshooting .................................................................................. 45
xii
List of Tables
1
Scope
➣ In this chapter…
Product Overview
1.1
page 1
Product Overview
The SafEye Quasar 900 IR Open-Path Gas Detector employs an advanced
Xenon Flash Source and integrated electronics package, both housed in
improved, Stainless steel housings to provide high quality and performance,
fast response, line of sight gas monitoring. This high quality is backed by 3
years warranty for the complete SafEye system and 10 years for the Xenon
Flash source bulb.
Quasar 900 detects ambient combustible gases over a path length of up to
660ft. (200 m), even in harsh environments where dust, fog, rain, snow or
vibration can cause a high reduction of signal. The SafEye Quasar 900 can
maintain operation in up to 90% signal obscuration and ±0.5 degree of
misalignment.
Quasar 900 is manufactured only in stainless steel with heated optical
window to improve performance in icing, snow, and condensation
conditions. The programmable functions are available through a RS 485 or
HART port used with Host software supplied by Spectrex and a standard PC
or an I.S handheld unit.
The Quasar Source and Detector unit enclosures are ATEX and IECEx
approved Exd flameproof with an integral segregated rear, Exe terminal
compartment which avoids exposure of the sensors and electronics to
surrounding environment. The Detector also has a plug interface for
connection to handheld PC or HART handheld, which is intrinsically safe.
Hence the combined approval - Ex d e ib [ib Gb] IIB+H2 T4 Gb, Ex tb IIIC
T135°C Db IP66.
This manual consists of a full description of the Detector and its features. It
contains instructions on the installation, operation and maintenance.

To use the WinHost software to change the required functions and for
maintenance, refer to Manual TM 888050 for instructions

To use the I.S Handheld Unit to change the required functions and for
maintenance, refer to Manual TM 799060 for instructions.

To use HART Protocol to change the required functions and for
maintenance refer to Manual TM 888030. The HART can be connected on
the 0-20mA line or through the IS port for setting
Product Overview
1
SafEyeTM Quasar 900 Gas Detector
2
Product Overview
2
Technical Description
➣ In this chapter…
2.1
page 3
Applications
page 4
Principle of Operation
page 4
Product Certification
page 7
Models and Types
page 8
Description
page 9
Features

Long Range Gas Detection up to 660ft (200m)

Simultaneous detection of C1-C8 flammable gases

High Sensitivity and fast response to Hydrocarbon gases

Heated optics to improve performance in icing, condensation and snow
conditions

Continuous operation in extreme and harsh environmental

Conditions

Solar blind and immune to industrial environments

Withstands extreme vibration conditions

Standard 0-20 mA Output

HART Protocol: Communication Protocol

RS-485 Output Modbus compatible for PC Communication Network for a
maximum of 247 systems

Simple one person installation, alignment, and calibration

ATEX & IECEx per EX II 2 GD, Ex d e ib [ib Gb] IIB+H2 T4 Gb, Ex tb IIIC
T135°C Db IP66

TUV approved per SIL2 requirements

FM / FMC approved per: Class I Div. 1 Group B, C and D; Class II/III
Div. 1 Group E, F and G

Functional approved per FM 6325

Functional tested by FM per EN60079-29-4

Programmable configuration via the handheld unit

Fast connection to I.S. approved handheld diagnostic/calibration unit

Features
Features
(3mA) “maintenance call”

Warranty 3 years for the complete SafEye system

Warranty 10 years for Xenon Flash bulb
3
SafEyeTM Quasar 900 Gas Detector
2.2
Applications
The Quasar 900 system may be used to monitor flammable gas
concentration in various applications, such as:
2.3

Petrochemical, pharmaceutical and other chemical storage and
production areas.

Flammable and toxic chemical storage sites and hazardous waste
disposal areas.

Refineries, oil platforms, pipelines, refueling stations and fuel storage
facilities.

Hazardous loading docks, transportation depots and shipping
warehouses

Engine rooms

Compressor and pumping stations

Test cells

LNG-LPG Systems

Offshore Floating Production Storage and Shipping vessels (FPSO), and
fixed oil rigs.
Principle of Operation
The Quasar system detects gases through a dual spectral range monitoring,
analyzing the absorption of radiation caused by gases in the atmosphere
and comparing it (ratio) to background atmospheric absorption.
2.3.1
Definitions of Terms
The following list defines gas concentrations measurement terms that are
used in this manual:
Table 1: Gas Concentrations Measurement Terms
2.3.2
Term
Description
LEL
Lower Explosive Limit: The minimum concentration of a
substance (gas/vapor) in air mixture that can be ignited.
This mixture is different for every gas/vapor, measured in
% of LEL.
LEL.m
Integral of Concentration in LEL units (1 LEL = 100% LEL)
and the operation distance in meters (m).
Spectral Finger Print
Each hazardous material is detected at a specific wavelength selected
according to its specific spectral absorption or "finger print". The detection
process involves two separate filters, one transmitting radiation that is
absorbed by a particular gas and one that is not sensitive to it.
4
Applications
TM 888200 Rev (I), January 2014
2.3.3
Optical Path
The presence of hazardous airborne vapors, gases, or aerosols in a
monitored area is detected when the defined substance crosses/enters the
optical path between the radiation source unit and the detector.
Hazardous gases /vapors present in the atmosphere cause absorption of the
radiation pulse in the optical path between the radiating source and the
detector unit at some specific wavelengths. This causes a change in the
signal intensity received by the detector, which is translated into an output
related to the detector’s measuring scale.
The system analyzes the defined open path at the spectral bands specific to
the materials being monitored. The Automatic Gain Control (AGC) unit
compensates for environmental disturbances such as fog, rain, etc., through
a constant comparison with its dual spectral beam.
2.3.4
Microprocessor Based
The incoming signals are analyzed by the built in microprocessor.
A sophisticated mathematical algorithm calculates between the various
functions of the detected signal thresholds. Statistics, ratio algorithms, data
communications, diagnostics and other functions are performed.
2.3.5
Gas Sensitivity
The SafEye Quasar 900 Model uses wavelengths around 2.3µ spectral band
to measure air flammability potential between the source and detector. At
this wavelength, all hydrocarbon materials have an absorption peak. This
enables the detector to achieve both regular sensitivity of 0-5 LEL.m.
The Quasar 900 detects hydrocarbon gases including methane, ethylene,
propane, ethane, butane, and so on.
2.3.6
Gas Calibration
The Quasar 900 has three calibrations that can be changed by function
Setup:

Gas 1 – 100% methane

Gas 2 – 100% propane

Gas 3 – 100% ethylene
The full scale of methane and propane is 5 LEL.m.
The full scale of ethylene is 8 LEL.m.
2.3.7
Flash Source
The Xenon Flash Source was originally introduced in the first SafEye
development and was designed to overcome false alarms, which were
experienced by early generations of the open path system. The new SafEye
Quasar 900 employs the latest generation of flash bulbs to provide even
more power and extended operation life (10 years).
Principle of Operation
5
SafEyeTM Quasar 900 Gas Detector
2.3.8
Heated Optics
SafEye Quasar includes heated optics for the Detector and source. The
heater increases the temperature of the optical surface by 5-8°F (3-5°C)
above the ambient temperature to improve performance in icing,
condensation and snow conditions. The heated optics is configured to
automatically operate when the change in temperature requires the heating
(default).
However, the heated optics can be defined as one of the following modes:
1
Not operated (not an option on source unit)
2
On continuously
3
Automatic, per temperature change (default)
See System Setup, page 16.
When operated ‘per temperature change’, the user can define the start
temperature below which the window will be heated (default 41°F (5°C)).
This temperature can be defined between 32°F (0°C) to 122°F (50°C). The
heating will stop when the temperature is 27°F (15°C) above the start
temperature.
2.3.9
HART Protocol
The Quasar 900 uses the HART Protocol.
HART Communication is a bi-directional industrial field communication
protocol used to communicate between intelligent field instruments and host
systems. HART is the global standard for smart instrumentation and the
majority of smart field devices installed in plants worldwide are HARTenabled.
HART technology is easy to use and very reliable.
Through the HART connection, the SafEye is able to perform:

Detector set-up

Detector troubleshooting

Detector health and status
For more details, refer to the HART Manual TM 888030.
The HART communication can be connected on the 0-20mA line or through
the IS connection with a standard Handheld loaded with our host software
and a special harness.
2.3.10
Handheld Unit
The I.S.-approved handheld diagnostics unit (Part no. 799810) is available
to make installation and maintenance easier. This is an all-on-one diagnostic
/ calibration / interrogation plug-in unit that allows for one-person
installation and maintenance.
6
Principle of Operation
TM 888200 Rev (I), January 2014
The handheld unit can be used:
1
For on-site function programming and set up changes to the Detector.
2
During installation, the handheld unit will display all the detector’s
parameters and confirm that the installation has been completed
successfully. It is also required to perform the necessary Zero
calibration function
3
For Maintenance and Trouble Shooting – The handheld unit will provide
recommendation of maintenance action to overcome and optimize the
Detector’s performance.
For more details, refer to Manual TM 799060.
2.3.11
Modbus RS-485
For more advanced communications, the Quasar 900 has a RS-485 Modbuscompatible output that provides data communication from a network (up to
247 Detectors) to a host computer or universal controller for central
monitoring. This feature enables easy maintenance, local and remote
diagnostic tools.
2.3.12
Tilt Mount
The new design stainless steel tilt mount provides a smaller installation
footprint that can conform to limited space constraints, while the sturdy
construction maintains alignment even in constant vibration. The improved
‘X’ and ‘Y’ axis worm-gear adjustments provide quick and easy alignment for
installation and maintenance procedures.
2.4
Product Certification
The open path Quasar 900 is approved for the following certification:
2.4.1

ATEX, IECEx, page 7

FM / FMC, page 8

SIL-2, page 8

Functional Test, page 8
ATEX, IECEx
The Quasar 900 is ATEX approved per SIRA 12ATEX1212X and IECEx per
IECEx SIR 12.0086X per:

ATEX Ex II 2 GD
Ex d e ib [ib Gb] IIB+H2 T4 Gb
Ex tb IIIC T135°C Db IP66

T Ambient –55°C to +65°C
This product is suitable to use in hazardous zones 1 and 2 with IIB+H2
group vapors present.
Product Certification
7
SafEyeTM Quasar 900 Gas Detector
2.4.2
FM / FMC
The Quasar 900 is approved to FM / FMC Explosion Proof per:

Class I, Div. 1 Group B, C and D, T6 -58°F (-50°C) ≤ Ta ≤ 149°F (65°C)

Dust Ignition Proof – Class II/III Div. 1, Group E, F and G.

Ingress Protection – IP66 & IP68, NEMA 250 Type 6P.
IP68 is rated for 2 meter depth for 45 minutes.
2.4.3
SIL-2
The Quasar 900 Is TUV approved for SIL2 requirements per IEC 61508.
The alert condition according to SIL-2 can be implemented by alert signal
via 0-20mA current loop.
For more details and guidelines on configuring, installing, operating and
servicing, see SIL-2 Features, page 57, and TUV report no. 968/EZ
619.00/13.
2.4.4
Functional Test
The Quasar 900 was functional approved per FM 6325
The Quasar 900 was functional tested by FM per EN60079-29-4
2.5
Models and Types
The Quasar 900 includes 4 models with the same detector and different
source. That gives the ability to get detection at distances of 7-200m.
Table 2: Model Numbers and Installation Distances
8
Model
No.
Detector
Source
Min.
Installation
Distance
Max.
Installation
Distance
901
QR-X-11X
QT-X-11X
23 ft (7 m)
66 ft. (20 m)
902
QR-X-11X
QT-X-21X
50 ft (15 m)
132 ft. (40 m)
903
QR-X-11X
QT-X-31X
115 ft (35 m)
330 ft. (100m)
904
QR-X-11X
QT-X-41X
265 ft (80 m)
660 ft. (200 m)
Models and Types
TM 888200 Rev (I), January 2014
Figure 1: P/N Definition of Quasar 900
2.6
Description
The SafEye comprises two main units:

The Flash Infrared Source (Transmitter)

The Infrared Detector (Receiver)
Quasar 900 detects gases over an open path transmitted from the Flash
Source to the Detector.
Description
9
SafEyeTM Quasar 900 Gas Detector
2.6.1
Flash Source Unit
The Flash Source unit emits IR radiation pulses at the rate of two pulses per
second. The pulse width (5-10µsec) is very powerful. The front of the
source SafEye has a lens that collimates the IR beam for maximum
intensity. The front window is heated to improve performance in icing,
condensation and snow conditions.
There are four source types:

For Short Range –901 – source P/N QT-X-11X

For Medium Range 1- 901 – source P/N QT-X-21X

For Medium Range 2- 903 – source P/N QT-X-31X

For Long Range - 904 – source P/N QT-X-41X
1
Front Window Section
6
Back Cover
2
Label
7
Earth Terminal
3
Main Housing
8
Front Window
4
Mounting Plate
10
Cable Inlet
5
Junction Box
11
Indication Led
Figure 2: Flash Source
The sources for models 901, 902 and 903 are the same electronically and
optically. The only difference is in the apparatus
The source for model 904 has different optics with different XENON lamp
source.
10
Description
TM 888200 Rev (I), January 2014
2.6.2
Detector Unit
The Detector receives the transmitted pulsed radiation signals from the
Flash Source. The signals are then amplified and fed into an analog to digital
signal converter to be processed by the internal microprocessor. When the
signals drop below a prescribed level, the internal microprocessor will
compensate for them. This will allow the signal to be maintained even in
severe weather conditions. The data is sent to the output interface section.
The front window of the Detector is heated to improve performance in icing,
condensation and snow conditions.
There is one detector type that is suitable for all Quasar models 901, 902,
903 and 904 – P/N QR-X-11X.
1
Front Window Section
7
Earth Terminal
2
Label
8
Front Window
3
Main Housing
9
Handheld Fast Connection
4
Mounting Plate
10
Cable Inlet
5
Junction Box
11
Indication Led
6
Back Cover
Figure 3: Detector
Description
11
SafEyeTM Quasar 900 Gas Detector
12
Description
3
Operating Modes
➣ In this chapter…
3.1
Operational Modes
page 13
Visual Indicators
page 15
Output Signals
page 15
System Setup
page 16
Operational Modes
The Quasar 900 has four operational modes:
3.1.1

Normal Mode, page 13

Maintenance Call Mode (3mA Output), page 13

Fault Mode, page 14

Zero Calibration Mode (1mA Output), page 14
Normal Mode
This mode is used for gas detection. In normal mode, the following statuses
are possible:

Normal - Signal received from gas detection is at safe levels.

Warning* - Gases have been detected at warning levels.

Alarm* - Gases have been detected at alarm levels.
Note: For the standard 0-20mA output, the Warning and Alarm levels are
not relevant. The user chooses these alarm levels at the controller. The
output Detector will be 4mA at zero reading and 20mA for full-scale reading.
Warning and Alarm states are relevant where the stepped 0-20mA output is
used and will energize the warning and alarm relays.
If the RS485 output is used, the Detector changes its status from ‘N’ to ‘W’
at Warning level and to ‘A’ at Alarm.
3.1.2
Maintenance Call Mode (3mA Output)
This mode indicates low signal or low signal ratio that may be caused by a
dirty window; misalignment; poor source or that one of the Detectors
parameters is at the ‘limit’ value.
The Detector continues to function and reads any gas present, but provides
a (3mA) pre-warning signal that a maintenance procedure is required.
Operational Modes
13
SafEyeTM Quasar 900 Gas Detector
3.1.3
Fault Mode
In the Fault status, there are three fault types:

Fault 1 (2mA Output) – Non critical
If this occurs, it is due to blockage, very low signal or in the case of
partial obscuration or full beam block and detection is no longer
possible. The Detector’s proper operation can be restored (Auto reset)
during operation if the condition causing the problem is removed or
resolved. This mode will occur after a delay of 60 sec. from the moment
of the fault. This delay is important to eliminate momentary obscuration
due to passing through the beam.

Fault 2 (2.5mA) - Misalignment
This occurs due to poor alignment.

Low Voltage Fault (1mA Output)
In this case, detection is disabled due to low voltage being supplied to
the Detector. The Detector returns to proper operation only when proper
voltage level restored.

Fault 2 (1mA Output) – Critical
In this case, the detection is disabled due to electrical / software
operational failure or central device (memory, processor) fault. Such
fault causes the Detector to cease operation. In case of fault in 0-20mA
loop, the output will be 0mA.
3.1.4
Zero Calibration Mode (1mA Output)
This mode zeros the base level from which the Detector will read gas.
It should only be performed when there are:

No combustible gases present

Clear path between the Flash Source and Detector

Clear weather conditions.
Zero calibration must be done after installation, re-alignment and window
cleaning, using the handheld unit (or Host software on PC).
14
Operational Modes
TM 888200 Rev (I), January 2014
3.2
Visual Indicators
One 3-color LED indicator is located in the back side of the detector/source
and can be seen through the back cover window. See Figure 2 and
Figure 3, item 11.
The detector statuses are listed in Table 3.
Table 3: Detector LED Indications
Detector Status
LED Color
LED Mode
Fault
Yellow
4Hz – flashing
Alignment / Standby
Yellow
1Hz – flashing
Zero Calibration
Yellow
Constant
Normal
Green
1Hz – flashing
Warning
Red
2Hz – flashing
Alarm
Red
Steady
The source statuses are listed in Table 4.
Table 4: Source LED Indications
Source Status
3.3
LED Color
LED Mode
Fault
Yellow
4Hz – flashing
Normal
Green
1Hz – flashing
Output Signals
The SafEye system provides the following outputs:
3.3.1

0-20mA Current Output, page 15

RS-485 Interface, page 16
0-20mA Current Output
The 0-20mA output provides the detector status measurement showing a
continuous reading of exact gas concentration.
The 0-20mA functions as current sink, but it can be configured as Source
(see Appendix A).
The maximum permitted load resistance for the 0-20mA output is 600Ω.
Visual Indicators
15
SafEyeTM Quasar 900 Gas Detector
Table 5: Standard (default) 0-20mA Current for the Gas Channel
Current Reading
3.3.2
Status and Description
0mA +0.3mA
Fault in 0-20mA loop
1mA ±0.3mA
Zero Calibration (In Progress), Fault 2
2mA ±0.3mA
Fault 1 (non-Critical)
2.5mA ± 0.3mA
Misalignment fault
3mA ±0.3mA
“Maintenance Call”
4mA±0.5mA
No gas present
4-20mA
Continuous measuring of gas concentration at a
range between 0 and full scale. For methane and
propane
This translates to 3.2mA per LEL.m and for ethylene
to 2mA per LEL.m.
21mA
Concentration is over the range limit (more than fullscale concentration).
RS-485 Interface
The RS-485 input/output sends complete data information to a PC and
receives data or control commands from the PC. The protocol is Modbus
compatible. The communication with the PC that is operated through the
interface is executed only when used with appropriate host software.
3.4
System Setup
This section includes the following topics:
3.4.1

Detection Function Programming, page 16

Detection Setup Function, page 17

Detector Default Setup, page 18
Detection Function Programming
The SafEye Quasar 900 incorporates several functions that can be set by the
customer using:
16

Host software. Refer to manual TM 888050 for programming
instructions.

Handheld unit (P/N 799810). Refer to manual TM 799060 for
programming instructions. The connection of the handheld to the
detector is fast and intrinsically safe and allows function change with no
need to open the detector.
System Setup
TM 888200 Rev (I), January 2014
3.4.2
Detection Setup Function
See Detector Default Setup, page 18 for default settings.
Setup includes the following options:
3.4.2.1

Gas Calibration, page 17

Zero Calibration, page 17

Address Setup, page 17

Heated Optic Operation, page 17
Gas Calibration
Three gas types can be selected for maximum compatibility to the required
measured gas(es).
Gas Types:

Methane – full scale 5 LEL.m.

Propane – full scale 5 LEL.m.

Ethylene – full scale 8 LEL.m.
These three calibrations are standard calibrations.
3.4.2.2
3.4.2.3
Zero Calibration

Enable – Zero calibration is performed according to background

Disable – The Detectors are not updated due to change of background
Address Setup
The Detector provides up to 247 addresses that can be used in RS-485
communication link.
3.4.2.4
Heated Optic Operation
The heated optics for the Detector unit can be defined as one of the
following modes:

OFF -Not operated

ON -Operated continuously

AUTO - On, per temperature change (default)
When operated per temperature change, the user can define the start
temperature below which the window will be heated between 0 to 50°C. The
Detector will stop heating the window when the temperature will be 15°C
above the defined temperature.
This feature relates to the Detector only.
The Source heated optic must be defined with the order in two options:

Heated continuously.

Start heating below 41°F (5°C). (default)
System Setup
17
SafEyeTM Quasar 900 Gas Detector
3.4.3
Detector Default Setup
The Detector has 8 functions that can be programmed according to the
customer requirement at factory or at customer facility using a software
Host or a handheld unit. The standard set up is as follows:
Table 6: Detector Default Set Up
Function
Setup
Gas Type
1
BG Zero Calibration
Enabled
0-20mA
Continuous
Heat mode
Auto
Heat on
5
Table 7: Source Default Set Up
Function
Setup
Heat mode
Auto
Heat on
5
The Source default can be changed with the same Host.
18
System Setup
4
Technical Specifications
➣ In this chapter…
4.1
General Specifications
page 19
Electrical Specifications
page 20
Mechanical Specifications
page 21
Environmental Specifications
page 22
General Specifications
Detected Gases:
Simultaneous detection of C1-C8
flammable gases
Detection Distance range:
Table 8
Table 8: Detection Distance Range
Model
No.
Detector
Source
Min.
Installation
Distance
Max.
Installation
Distance
901
QR-X-11X
QT-X-11X
23 ft (7 m)
66 ft (20 m)
902
QR-X-11X
QT-X-21X
50 ft (15 m)
132 ft (40 m)
903
QR-X-11X
QT-X-31X
115 ft (35 m)
330 ft (100m)
904
QR-X-11X
QT-X-41X
265 ft (80 m)
660 ft (200m)
Response Time:
3 sec to T90
Spectral Response:
2.0 - 3.0 micron
Sensitivity Range:
Full
Scale
LEL.m.
Warning
LEL.m.
Alarm
LEL.m.
Gas 1
Methane
5
1
3
Gas 2
Propane
5
1
3
Gas 3
Ethylene
8
1.6
4.8
Field of View:
Line of Sight
Alignment
Tolerance:
± 0.5°
Drift:
± 7.5% of the reading or ±4% of the full scale
(whichever is greater)
Minimum Detectable
Level:
0.15 LEL.m
Temperature Range:
-67°F (-55°C) to +149°F (+65°C)
General Specifications
19
SafEyeTM Quasar 900 Gas Detector
Immunity to false
alarm:
4.2
Does not produce false alarm and is not
influenced by Solar Radiation, Hydrocarbon
Flames and other external IR Radiation Sources.
Electrical Specifications
Operating Voltage:
4.2.1
18-32 VDC
Power Consumption
Table 9: Detector and Source Maximum Power Consumption
Without Heated
Optic (Max.)
4.2.2
With Heated
Optic (Max.)
Detector
200mA
250mA
Source
200mA
250mA
Electrical input protection
The input circuit is protected against voltage-reversed polarity, voltage
transients, surges and spikes according to EN50270.
4.2.3
Electrical outputs
4.2.3.1
0-20mA Current Output
The 0-20mA is isolated sink option. This output can be configured as Source
(see Appendix A).
The maximum permitted load resistance is 600 ohm.
4.2.3.2
Communication Network
The Detector is equipped with an RS 485-communication link that can be
used in installations with computerized controllers.
Communication is compatible with the Modbus protocol:
4.2.3.3

This protocol is a standard and is widely used.

It enables continuous communication between a single standard Modbus
controller (Master device) and a serial Network of up to 247 Detectors.

It enables connection between different types of Spectrex Detectors or
other Modbus devices to the same Network.
HART Protocol
The HART Protocol is a digital communication signal at low level on top of
the 0-20mA.
This is a bi-directional field communications protocol used to communicate
between intelligent field instruments and the Host System.
20
Electrical Specifications
TM 888200 Rev (I), January 2014
Through the HART Protocol the detector can:
4.3

Display set-up

Reconfigure set-up

Display detector status and definition

Perform detector diagnostic

Troubleshooting
Mechanical Specifications
Enclosure:
The Detector, Source and tilt mount are St. St. 316
Electro chemical and passivized coating.
Explosion Proof:
ATEX &
IECEx
EX II 2 GD,
FM/FMC
Class I Div. 1 Groups B, C, and D
Ex d e ib [ib Gb] IIB+H2 T4 Gb
Ex tb IIIC T135°C Db IP66
Class II/III Div. 1 Groups E, F, and G
T6 -58°F (-50°C) ≤ Ta ≤ 149°F (65°C)
Functional
Testing:
Approved per FM 6325
Tested by FM per EN60079-29-4
Water and Dust
Tight:
IP 66 and IP 68
IP68 is rated for 2 meter depth for 45 minutes.
NEMA 250 type 6p
Electrical
Modules:
Conformal coated
Electrical
Connection:
(Two options - specified at time of order)
2 X M25 (ISO)
2 X 3/4” - 14NPT conduits
Dimensions:
Weight:
Detector
10.5 x 5.1 x 5.1 in
(267 x 130 x 130mm)
Source
10.5 x 5.1 x 5.1 in
(267 x 130 x 130mm)
Tilt Mount 4.7 x 4.7 x 5.5 in
(120 x 120 x 40mm)
Detector
11 lb
(5 kg)
Source
11 lb
(5 kg)
Tilt Mount 4.2 lb
Mechanical Specifications
(1.9 kg)
21
SafEyeTM Quasar 900 Gas Detector
4.4
Environmental Specifications
The SafEye system is designed to withstand harsh environmental conditions.
The source and detector units compensate for adverse conditions while
maintaining accuracy.
4.4.1
High Temperature
Designed to meet MIL-STD-810C, method 501.1 procedure II.
4.4.2
Operating Temperature:
+149 ºF (+65 ºC)
Storage Temperature:
+149 ºF (+65 ºC)
Low Temperature
Designed to meet MIL-STD-810C, method 502.1, procedure I
4.4.3
Operating Temperature:
-4 ºF (-20 ºC)
Storage Temperature:
-40 ºF (-40 ºC)
Humidity
Designed to meet MIL-STD-810C, method 507.1, procedure IV
relative humidity of up to 95% for the operational temperature range.
4.4.4
Salt and Fog
Designed to meet MIL-STD-810C, method 509.1 procedure I.
Exposure to a 5% salt solution for 48 hours.
4.4.5
4.4.6
22
Water and Dust

IP67 per EN60529

IP66 per EN60529
Dust:
Totally protected against dust
Liquids:
Protected against immersion between 15
cm and 1m in depth. Protected against
all water jets from all directions.
Shock and Vibration
Vibration:
Designed to meet MIL-STD-810C,
method 514.2, procedure VIII.
Mechanical Shock:
Designed to meet MIL-STD-810C,
method 516.1, procedure I.
Environmental Specifications
TM 888200 Rev (I), January 2014
4.4.7
Electromagnetic Compatibility (EMC)
This product is in conformance with EMC per EN50270:
Radiated Emission:
EN55022
Conducted Emission:
EN55022
Radiated Immunity:
EN61000-4-3
Conducted Immunity:
EN61000-4-6
ESD:
EN61000-4-2
Burst:
EN61000-4-4
Surge:
EN61000-4-5
Magnetic Field:
EN61000-4-8
Environmental Specifications
23
SafEyeTM Quasar 900 Gas Detector
24
Environmental Specifications
5
Installation Instructions
➣ In this chapter…
5.1
Introduction
page 25
General Considerations
page 25
Preparations for Installation
page 26
Certification Instructions
page 28
Conduit/Cable Installation
page 29
Detector/Source Mounting
page 29
Detector Wiring
page 30
Detector Terminal Wiring
page 34
Flash Source Wiring
page 34
Introduction
The Detector and Flash Source units can be installed and maintained with
the use of general-purpose common tools and equipment. The installation
procedure has to be performed by suitably qualified personnel.
This section does not attempt to cover all of the standard practices and
codes of installation. Rather, it emphasizes specific points of consideration
and provides some general rules for suitably qualified personnel. Special
safety precautions are stressed wherever applicable.
5.2
General Considerations
5.2.1
Personnel
Only suitably qualified personnel, familiar with the local codes and practices,
trained for gas detection maintenance should be employed. Wiring should
only be performed or supervised by someone with knowledge of electronics
and in particular wiring installation.
5.2.2
Tools Required
The SafEye system requires the following tools:

Set of screwdrivers

Set of hex keys/Allan wrenches (supply with commissioning kit)

Voltage Multimeter
Introduction
25
SafEyeTM Quasar 900 Gas Detector
5.2.3
Site Requirements
The installation position of the SafEye system must take into account if the
gas being monitored is heavier or lighter than air, and the individual site
requirements. The site selected must give the Detector a direct view to the
Source. The mounting point for each item should be secure and stable with
minimal vibrations. Equipment should be either mounted in a position where
it cannot be knocked out of alignment, or guarded from physical impact.
5.2.4
The Source and Detector
The appropriate Detector should be selected for the length of open path to
be monitored. To allow for ageing of the Source and a reduction of the IR
signal due to adverse weather we recommend using a Detector that is not at
the limit of its operating range.
The open path between the Source and Detector and the immediate
surroundings should be kept clear of obscuration that might hinder the free
movement of air in the protected area or block the infrared beam.
5.2.5
Guidance Tips for Gas Detector Locations
Guidance Tips for Gas Detector locations in order to provide the best
detection coverage:
5.3

Below potential leak source for heavier than air gases

Above potential leak sources for lighter than air gases

Near to leak sources along the expected leak trajectory, taking into
account prevailing wind directions

Between leak source and potential ignition sources

In area with expected heavy fog, rain or snow, consider the effect of
long range installation and install the Detector at shorter range with the
maximum intensity model available.
Preparations for Installation
Installation should comply with local, national and international regulations
and norms as applicable to gas Detectors and approved electrical devices
installed in hazardous areas. The Detectors can be installed with generalpurpose common tools and equipment.
The System should include the following (in addition to this manual):

Detector unit - QR-X-11X (See Models and Types, page 8)

Source unit - QT-X-X1X (See Models and Types, page 8)

Two Tilt Mount Bases - P/N 888270

26

1 base is used for the Detector

1 base is used for the Flash Source
Commissioning Kit - P/N 888247
Preparations for Installation
TM 888200 Rev (I), January 2014
The Commissioning kit includes Function Check Filter and the Telescope
Kit, which is used during each SafEye installation and then removed.
They can be reused for all other SafEye installations on the site.
Therefore, only one set is provided for several detectors.

Telescope Kit - P/N 888210

Function Check Filters - P/N 888260-1, -2

Handheld Unit - P/N 799810

Other accessories are available (per the customer request):

Pole Mount (U-Bolt 5 inch) - P/N 799225

HART Hand-Held Diagnostic Unit - P/N 888810

HART Hand-Held Harness Kit - P/N 888815

USB/RS485 Harness Converter Kit - P/N 794079-8

Mini Laptop Kit - P/N 777820

Sunshade - P/N 888263
See details in Appendix B.
The following procedure is used for preparing the installation of the 900 Gas
Detector:
1
Verify the appropriate Purchase Order. Record the Part No. and the
Serial No. of the detectors and source units and the installation date in
the appropriate log book.
2
Open the container package immediately prior to detector installation
and visually inspect the detectors; sources and accessories.
3
Verify that all components required for the detector installation are
readily available before commencing the installation. In the event that
the installation is not completed in a single session, secure and seal
detectors and conduits.
4
For wiring, use color-coded conductors or suitable wire markings or
labels. Cross section of wire must be between 1mm² to 2.5mm²
(18- 14AWG). The selection of wire gauge should be based on the
number of detectors used on the same loop and the distance from the
control unit. Max. Wire connection in the same terminal is two wire
cross section of each wire is 1mm².
Preparations for Installation
27
SafEyeTM Quasar 900 Gas Detector
5.4
Certification Instructions
Warning: Do not open the detector, even when isolated, when
flammable atmosphere is present.
Use the following certification instructions:

The cable entry point may exceed 182°F (83°C) suitable precautions
should be taken when selecting the cable.

The equipment may be used with flammable gases and vapors with
apparatus groups IIA and IIB +H2 T4 in the ambient temperature range
-67°F (-50°C) to +149°F (+65°C).

Installation shall be carried out by suitably trained personnel in
accordance with the applicable code of practice e.g. EN 60079-14:1997.

Inspection and maintenance of this equipment shall be carried out by
suitably trained personnel in accordance with the applicable code of
practice e.g. EN 60079-17.

Repair of this equipment shall be carried out by suitably trained
personnel in accordance with the applicable code of practice e.g. EN
60079-19.

The certification of this equipment relies upon the following materials
used in its construction:


28

Enclosure: 316L Stainless Steel

Window: Sapphire Glass

Seals: EPDM
If the equipment is likely to come into contact with aggressive
substances, then it is the responsibility of the user to take suitable
precautions that prevent it from being adversely affected, thus ensuring
that the type of protection provided by the equipment is not
compromised.

Aggressive substances: For example, acidic liquids or gases
that may attack metals, or solvents that may affect polymeric
materials.

Suitable precautions: For example, regular checks as part of
routine inspections or establishing from the material’s data
sheets that it is resistant to specific chemicals.
Special Conditions for safe use: The Quasar 900 Open Path Gas
Detectors & Flash Source Units shall not be used as safety related
devices in accordance with directive 94/9/EC.
Certification Instructions
TM 888200 Rev (I), January 2014
5.5
Conduit/Cable Installation
The conduit and cable installation must comply with the following guidelines:
5.6
1
To avoid water condensation in the detector, it should be installed with
the conduits/cable entries facing downwards.
2
Use flexible conduits/cables for the last portion connecting to the
detector.
3
When pulling the cables through the conduits, ensure that they are not
tangled or stressed. Extend the cables about 12 inches (30cm) beyond
the Detector location to accommodate wiring after installation.
4
After the conductor cables have been pulled through the conduits,
perform a continuity test.
Detector/Source Mounting
The Detector Source should be mounted with the Tilt Mount Kit, Model
888270. The Tilt enables the Detector to be rotated up to 45º in all
directions and fine alignment of up to 3º.
5.6.1
Tilt Kit P/N 888270
The following contents are included with the Tilt Mount Kit:
Table 10: Tilt Mount Kit
Item
5.6.2
Qty
Type/Model
Tilt Mount
1
888269
Screw
1
M10 x 1.5
Spring Washer
1
No. 10
Detector/Source Installation
(Figure 4 and Figure 5)
The detector and the source can be installed in two ways with the same Tilt
Mount.
➣ To install the Detector/Source:
1
Place the Tilt Mount holding plate (item 1) in its designated location and
secure it with (4) fasteners through four (4) holes dia. 8.5mm.
Notes:

Skip this step if the Tilt Mount is already installed.

Detector removal for maintenance purpose does not require
Tilt Mount removal.
Conduit/Cable Installation
29
SafEyeTM Quasar 900 Gas Detector
5.7
2
Place the Detector, with its conduit/cable inlets pointing downwards on
the Detector holding plate of the Tilt Mount (item 2). Secure the
Detector with M10 x 1.5 screw with No. M10 spring washers (9,10).
Secure the detector to the Tilt Mount use Hex Key No. 7 for M10 x 1.5
screw (item 9).
3
Repeat steps 1 – 2 to install the Source.
Detector Wiring
➣ To install the Detector wiring:
30
1
Release the back cover secure bolt (item 15, Figure 5), and open the
detector back cover (item 14, Figure 5). The chamber is now revealed.
2
Remove the protective plug mounted on the Detector conduit/cable
entry inlet; pull the wires through the Detector inlet (item 4, Figure 6).
Use a 3/4” – 14NPT or M25x1.5 explosion-proof conduit connection /
cable gland to assemble the cable / explosion-proof conduit to the
Detector.
3
Connect the wires to the required terminals (item 2, Figure 6) according
to the wiring diagram. See paragraph 5.8 and figures 8, 10, 11, and 12
Appendix A.
4
Connect the grounding wire to the ground screw outside Detector
(item 3. Figure 6). The Detector must be well grounded to Earth
Ground.
5
Place and secure the Detector back cover by screwing the cover and
securing it using the secure bolt (item 15, Figure 5).
Detector Wiring
TM 888200 Rev (I), January 2014
1
Tilt Mount Holding Plate
5
Vertical Fine Alignment
Tightening Screw
Vertical Crude Alignment
Tightening Screw
2
Detector/Source Holding
Plate
6
3
Horizontal Crude Alignment
Tightening Screw
7
4
Horizontal Fine Alignment
Tightening Screw
8
Vertical Find Alignment Screw
Horizontal Fine Alignment
Screw
Figure 4: Tilt Mount
Detector Wiring
31
SafEyeTM Quasar 900 Gas Detector
1
Tilt Mount Holding Plate
9
2
Detector/Source Holding
Plate
10
3
Horizontal Crude Alignment
Tightening Screw
11
4
Horizontal Fine Alignment
Tightening Screw
12
5
Vertical Fine Alignment
Tightening Screw
13
6
Vertical Crude Alignment
Tightening Screw
14
7
Vertical Fine Alignment
Screw
15
8
Horizontal Fine Alignment
Screw
Detector Tightening Screw
Detector Tightening Washer
Detector
Telescope
Telescope Tightening Bolt
Detector Back Cover
Detector Back Cover Secure
Bolt
Figure 5: Detector and Tilt Mount Assembly
32
Detector Wiring
TM 888200 Rev (I), January 2014
1
Housing
5
Internal Earth Connection
2
Terminal Board
6
Connection to Handheld Unit
3
Earth Terminal
7
Detector Holding Plate
4
Inlet Conduit
Figure 6: Detector with Cover Removed
Detector Wiring
33
SafEyeTM Quasar 900 Gas Detector
5.8
Detector Terminal Wiring
The Detector has six wiring terminals.
The following describes the function of each electrical terminal of the
detectors.
Table 11: Wiring Options
Terminal
No.
Function
1
Power +24V DC
2
Return +24V DC
3
0-20mA (Input)
4
0-20mA (Output)
5
RS-485 (+)
6
RS-485 (-)
5.9
Flash Source Wiring
5.9.1
Wiring
➣ To install the wiring:
34
1
Release the back screw bolt (item 15, Figure 5), and open the source
back cover (item 14, fig. 4). The chamber is now revealed.
2
Remove the protective plug mounted on the source conduit/cable entry
inlet; pull the wires through the source inlet (item 4, Figure 7). Use a
3/4” – 14NPT or M25x1.5 explosion-proof conduit connection / cable
gland to assemble the cable / explosion-proof conduit to the Detector.
3
Connect the wires to the required terminals (item 2, Figure 7) according
to the wiring diagram. See paragraph Terminal Wiring, page 35 and
Figure 11 Appendix A.
4
Connect the grounding wire to the ground screw outside Detector
(item 3, Figure 7). The source must be well grounded to Earth Ground.
5
Place and secure the source back cover by screwing the cover and
secure the back screw bolt.
Detector Terminal Wiring
TM 888200 Rev (I), January 2014
5.9.2
Terminal Wiring
The Flash Source contains six wiring terminals.
Table 12: Flash Source Wiring Options
Terminal
No.
Function
1
Power +24V DC
2
Return +24V DC
3
Spare
4
Spare
5
RS-485 (+)
6
RS-485 (-)
1
Housing
5
Internal Earth Connection
2
Terminal Board
6
N/A
3
Earth Terminal
7
Detector Holding Plate
4
Inlet Conduit
Figure 7: Source with Cover Removed
Flash Source Wiring
35
SafEyeTM Quasar 900 Gas Detector
36
Flash Source Wiring
6
Operating Instructions
➣ In this chapter…
6.1
SafEye Operation
page 37
Alignment of Unit
page 37
Powering up the System
page 38
Safety Precautions
page 39
Signal Verification
page 39
Zero Calibration
page 40
Functional Check of Unit
page 41
SafEye Operation
Once the system is positioned, it will monitor for possible specified gases
automatically sending signals to a standard control panel or a PC. This
section describes the alignment, calibration and operation.
Important! Accurate Alignment is essential for proper operation of the
SafEye system.
6.2
Alignment of Unit
Using the Telescope performs full alignment.
Alignment procedure should be performed in two stages: Crude alignment
and fine adjustment.
The Telescope includes a Periscope that consists of a prism and an ocular
located vertical to the Telescope assembly. This allows the user to observe
into the opposite unit perpendicularly to the alignment when access from
the rear of the unit is impossible. For installations where rear access is
possible the Periscope is not necessary. In this case it can be removed by
releasing the Periscope Fastening Screw.
Important Notes:
1
Prior To Telescope Installation, verify that the Telescope and Its Sight
Mounting are free from any dirt to ensure proper alignment according to
factory calibration.
2
Do not attempt to change any factory calibration at the Telescope or its
mounting. This may prevent optimal alignment.
➣ To align the unit (see Figure 4):
1
Make sure that the Detector and the Flash Source are installed properly.
Installation instructions are described in section 5 of this document.
2
Remove the front shield using the two captive screws.
SafEye Operation
37
SafEyeTM Quasar 900 Gas Detector
6.3
3
Install the Telescope Assembly (12) on the detector/source front. Fasten
the Telescope with Fastening Screws (13).
4
Crude Alignment
a
Use ¼” Allen screw driver for all alignment screws
b
Loosen screws 5, 6
c
Approximately aim the source horizontally toward the detector
d
Tighten screw 6
e
Loosen screws 3, 4
f
Approximately aim the source vertically toward the detector
g
Tighten screw 3
5
Repeat step 4 for the Detector
6
Fine Alignment
a
Aim the Source to the Detector within horizontal axis using screws
7. Aim the cross to the detector or source center of the front
window (see figure 1 & 2 item 8).
b
Tighten screw 5
c
Aim within the vertical axis using screw 8
d
Tighten screw 4
e
Make sure the telescope cross is pointing to the detector and
source center of the window
7
Repeat step 6 for the Detector alignment
8
Remove the telescope. Install the front shield.
Powering up the System
Important! Prior to any operation or maintenance, follow the Safety
Precautions, page 39.
➣ To power up the system:
1
Make sure that the Source and Detector are connected to power.
2
Make sure that the 4-20mA meter is connected to the Detector.
3
Power up the system 18-32VDC.
4
After 60 seconds, the current meter will indicate 4mA.
Note: Zero calibration should be performed after powering up the system
(see Zero Calibration, page 40).
38
Powering up the System
TM 888200 Rev (I), January 2014
6.4
Safety Precautions
After powering up, the Detector requires minimal attention in order to
function properly, but the following should be noted:
6.5
1
Follow the instructions in the Manual; refer to the drawings and
specifications issued by the manufacturer.
2
Do not open the detector/source housing while power is supplied.
3
Disconnect external devices, such as automatic extinguishing systems
before carrying out any maintenance task warranty.
Signal Verification
The signal verification can be done through Host software supplied by
Spectrex. Refer to Manual TM 888050 or by Handheld unit refer to Manual
TM 799060.
6.5.1
Signal Values Limitation
Table 13 describes the maintenance data channels limitation limits.
Table 13: Maintenance Channels Limit Values
Installation Distance
Channel
Min
Med
Max
Maintenance
Reference
1V Gain1
1V Gain2
1V Gain4
The minimum signal
allowed is 2V at Gain9
Signal 1
1V Gain1
1V Gain2
1V Gain4
The minimum signal
allowed is 2V at Gain9
Ratio 0
0.6-1.4
0.6-1.4
0.6-1.4
0.5-3
NQRat 0
LEL
0.95 – 1.05
0 LEL x m
Must be 0.95 – 1.05
0 LEL x m
Temp.
Up to 25° C beyond room temp.
Up to 25° C beyond
room temp.
Voltage
32VDC > V > 18VDC
32VDC > V > 18VDC
Note: The installation information refers to the installation distance.

Min: The minimum distance as defined on the model no.

Med: Half of the maximum distance as defined on the model no.

Max: The maximum distance defined on the model no.
Safety Precautions
39
SafEyeTM Quasar 900 Gas Detector
6.6
Zero Calibration
Zero calibration must be done after:

Installation.

Realignment.

Window Cleaning.

Any change in Detector or Source position.
Precise alignment must be performed prior to the zero calibration procedure.
Zero calibration should be performed in good weather conditions with
insignificant gas concentrations in the surrounding environment or indoors.
➣ To perform the zero calibration procedure:
1
Switch from Normal to Alignment mode indication.
2
Switch from Alignment to Standby mode.
3
Switch from Standby to Zero Calibration mode.
The 0-20mA output should now be at 1mA.
4
Wait up to 60 seconds until it switches to Normal mode. The Detector
reading is now set to normal. The 0-20mA output should now indicate
4mA.
Switching from each position (above first three instructions) can be done
by HOST HART (refer to TM888030) or RS485 (refer to TM888050 or
TM799060), or by moving the magnetic mode selector above the
magnetic switch (see Figure 8).
Figure 8: Magnetic Mode Selector
40
Zero Calibration
TM 888200 Rev (I), January 2014
6.7
Functional Check of Unit
The SafEye system has been calibrated at the factory for the User's specific
gas or vapor detection requirements. The following procedure validates the
functional operation of the system. The Functional Check Filter is a
convenient operational check used to confirm that response has not changed
from previous readings. The filter is not used for calibration, which is
unnecessary, nor does it equate to a particular quantity of gas.
Caution: Automatic activation or any external device that should not be
activated during the calibration check should be disconnected.
Notes:
1
This functional verification procedure is for a standard 0-20mA output.
2
Prior to starting the functional check, verify that the power to the units
is on and that the current of the 0-20mA channel is stable. Record the
reading.
➣ To perform the functional check:
1
Position the Functional Check filter in front of the SafEye Detector.
2
The functional check filter's window must be centered over the viewing
window of the Detector.
3
Wait 20 seconds.
4
Read the 0-20mA current. Determine the difference between the
reading taken with and without the Functional Check Filter. This
difference is the 0-20mA current variance.
5
Record the 0-20mA current variance in the Maintenance Logbook. If the
variance is more than a 30% change when compared to the previous
check (see delivery form), repeat the alignment.
Functional Check of Unit
41
SafEyeTM Quasar 900 Gas Detector
42
Functional Check of Unit
7
Maintenance Instructions
➣ In this chapter…
7.1
General Maintenance
page 43
Periodic Maintenance
page 43
General Maintenance
The SafEye Quasar 900 requires only simple periodic maintenance to
provide satisfactory service and achieve maximum performance. The
Detector and Source units can be maintained with the use of common tools
and equipment. Record the periodic test results in Maintenance Logbook
with a copy of the delivery form inside.
7.2
Periodic Maintenance
Recommended cleaning of optical surfaces to be performed periodically.
Note: The frequency of cleaning operations is ultimately dependent upon
the existing environmental conditions and the applications used.
➣ To perform periodic maintenance:
1
Proper maintenance will allow the SafEye system to retain maximum
performance and reliability.
2
The optical surfaces of the Source and Detector viewing windows should
be kept as clean as possible as it is an active device.
3
Alignment procedures must be performed each time that the Source or
the Detector unit has been opened or moved for any reason.
4
The Signal Verification Check corroborates the current signals from the
Flash Source compared to that of previous alignments. It is
recommended to perform this check every 6-12 months. The signal
should be checked according to threshold levels (see Signal Verification,
page 39).
5
The functional check should be performed every 6 months (see
Functional Check of Unit, page 41)
6
Alignment procedure should be performed only if the signals are below
Threshold value (see Signal Verification, page 39)
7
Zero calibration (see Zero Calibration, page 40) must be done every
time the Detector or Source is realigned or windows cleaned.
General Maintenance
43
SafEyeTM Quasar 900 Gas Detector
7.2.1
Routine Optical Surface Cleaning
The SafEye system, being an optical device, must be kept as clean as
possible. The optical surfaces concerned are the Source and Detector
viewing windows.
➣ To clean the optical window, proceed as follows:
1
Turn off the power to the SafEye detector and source.
2
In places where dust or dirt has accumulated on the optical surface,
clean the surface with a small, soft-bristle brush.
3
The surfaces must then be washed thoroughly with water and a mild
non-abrasive detergent.
4
Thoroughly rinse the glass surface with clean water; ensuring no residue
is left behind.
5
Dry the glass with a clean dry soft cloth.
6
Enter in the Maintenance Logbook: Date, name of person and company
who performed the maintenance service.
7
Turn on power to the SafEye detector and source.
8
Perform signal verification (see Signal Verification, page 39).
9
Perform zero calibration (see Zero Calibration, page 40).
10 Perform functional check (see Functional Check of Unit, page 41)
7.2.2
Signal Verification
The signal verification check determines the proper operation of the open
path. It checks the alignment and cleanliness of the window or any problem
in Source or Detector. Use the PC Host software to measure the signal
verification.
Refer to Manual TM 888050 or use I.S handheld unit. Refer to Manual
TM 799060 for non I.S Handheld Unit.
7.2.3
Function Check of Unit
The SafEye Quasar has been calibrated at the factory per the user’s specific
gas or vapor detection requirements. This procedure validates the functional
operation. The functional check must be done periodically. Refer to 6.7 for
instructions.
Caution: Automatic activation or any external device that should not be
activated during the calibration check should be disconnected.
44
Periodic Maintenance
8
Troubleshooting
Table 14: Troubleshooting
Problem
Cause
Solution
“Maintenance call”
status or R and S are
below 2V DC at Gain
9
Poor alignment
Perform alignment
Dirt on the
window
Clean the window
Poor light source
Replace the light source
Detector fault
Replace/repair Detector
NQRat below the
permitted limit
Gas in the path
Make sure that the path is
clean and the weather
conditions are good
NQRat above the
permitted limit
Poor alignment
Perform alignment
Temperature higher
than 25° C beyond
the room
temperature
Electronic
problem
Replace/repair Detector
Ratio1 and Ratio2 out
of the limit
Poor alignment
Perform alignment
Dirt on the
window
Clean the window
Detector fault
Replace/repair Detector
Low input
voltage
Check the power supply and
installation
Voltage less than 16V
DC. The Detector at
“V” fault
Periodic Maintenance
45
SafEyeTM Quasar 900 Gas Detector
46
Periodic Maintenance
Appendices
Appendices
47
SafEyeTM Quasar 900 Gas Detector
48
Appendices
A
Wiring Configurations
➣ In this appendix…
RS-485 Communication Network
page 53
Figure 9: Detector Wiring Terminal
RS-485 Communication Network
49
SafEyeTM Quasar 900 Gas Detector
Figure 10: Source Wiring Terminal
50
RS-485 Communication Network
TM 888200 Rev (I), January 2014
Figure 11: 0-20mA Sink 4-Wire
Figure 12: 0-20mA Non-isolated Sink 3-Wire
RS-485 Communication Network
51
SafEyeTM Quasar 900 Gas Detector
Figure 13: 0-20mA Source 3-Wire
52
RS-485 Communication Network
TM 888200 Rev (I), January 2014
A.1
RS-485 Communication Network
By using the RS-485 network capability of the SafEye Quasar 900 detector
and additional software, it is possible to connect up to 32 Detectors in an
addressable system with four (4) wires only (2 for power and 2 for
communication). Using repeaters, the number of Detectors can be much
larger (32 Detectors for each repeater) up to 247 on the same four (4)
wires. When using the RS-485 network, it is possible to read each Detector
status (FAULT, WARNING, and ALARM).
For more details, consult the factory.
Figure 14: RS-485 Networking for Wiring Option 3
RS-485 Communication Network
53
SafEyeTM Quasar 900 Gas Detector
54
RS-485 Communication Network
B
Accessories
➣ In this appendix…
B.1
Tilt Mount P/N 888270
page 55
Pole Mount (U-Bolt 5 inch) P/N 799225
page 55
Commissioning Kit P/N 888247
page 55
RS485 Hand-Held Diagnostic Unit P/N
799810
page 55
HART Hand-Held Diagnostic Unit P/N
888810
page 56
HART Hand-Held Harness Kit P/N 888815
page 56
USB/RS485 Harness Converter Kit P/N
794079-8
page 56
Mini Laptop Kit P/N 777820
page 56
Sunshade P/N 888263
page 56
Tilt Mount P/N 888270
The Tilt Mounting Brackets allow accurate alignment of the detector/source
for proper operation of the open path. The brackets give crude alignment of
±30° and fine alignment of ±5°.
B.2
Pole Mount (U-Bolt 5 inch) P/N 799225
The U-bolt mount is available to facilitate 5” pipe mounting.
B.3
Commissioning Kit P/N 888247
The Commissioning/Alignment Kit unit is required for commissioning and
future maintenance checks. Only one kit is required per site.
The kit includes an Alignment Telescope P/N 888210, a Magnetic Mode
Selector P/N 790285, two Function Check Filter P/N 888260-1, -2 for system
installation and periodical functional testing along with socket keys for
access to units.
B.4
RS485 Hand-Held Diagnostic Unit P/N 799810
RS485 Hand-Held diagnostic unit, fitted with harness and converter to the
quick plug connection which provides easy and economical connection, will
provide verification, status and instructions for correcting the detector’s
parameters.
RS485 Hand-Held Diagnostic Unit P/N 799810
55
SafEyeTM Quasar 900 Gas Detector
This is an I.S. approved Hand-Held with I.S. harness and converter to suit
the detector and programmed with a special host for maintenance and
commissioning.
B.5
HART Hand-Held Diagnostic Unit P/N 888810
HART Hand-Held diagnostic unit fitted with harness to the quick plug
connection which provides easy and economical connection. The HART
Hand-Held will provide verification, status and instructions for correcting the
detector’s parameters.
This is an I.S. approved Hand-Held with a special harness to suit the
detector and a host for maintenance and commissioning.
B.6
HART Hand-Held Harness Kit P/N 888815
A quick plug connection with harness which is connected on one side to a
standard HART Hand-Held unit and includes Spectrex host software that can
be uploaded to an existing Hand-Held (does not include the Hand-Held).
B.7
USB/RS485 Harness Converter Kit P/N
794079-8
The USB RS485 Harness Kit with RS485/USB converter, used with Spectrex
Host software, enables the user to connect to any available PC or laptop to
re-configure settings or perform diagnostics on the Quasar 900 gas
detector.
Refer to manual TM 888050 for programming instructions when using the
USB RS485 Harness Kit.
B.8
Mini Laptop Kit P/N 777820
The mini laptop, pre-loaded with Spectrex software, enables the user to reconfigure settings or perform diagnostics on all flame and gas detectors
series.
Refer to manual TM777070 for programming instructions when using the
Mini Laptop Kit. The kit includes cable harness with RS485/USB converter.
The mini laptop is programmed with maintenance winhost for all detectors'
type.
B.9
Sunshade P/N 888263
The Sunshade is designed to protect the detector from the heat of the sun.
56
Sunshade P/N 888263
C
SIL-2 Features
➣ In this appendix…
Safety Relevant Parameters for the Quasar
900 Open Path Gas Detector
page 57
General Conditions for Safe Use
page 58
This appendix details the special conditions to comply with the requirements
of EN 61508 for SIL-2.
The SafEye Quasar 900 IR Open-Path Gas Detector can be used in low and
high demand mode applications, see IEC 61508-4:2010, Chapter 3.5.16.
C.1
Safety Relevant Parameters for the Quasar
900 Open Path Gas Detector
Type:
B
Structure:
1oo1
HFT:
0
Main time to repair:
72 h
Ambient temperature:
max. 65°C
Proof-Test-Interval:
52 weeks
 S  2056.1fit

 D  1976.1fit

 DU  114.8fit

 SD  1933.4fit

 DD  1861.4fit

SFF  97 %
DC = 94%
PFDavg = 6.45x 10-4
PFH = 1.15 x 10
-7
1/h
PFD%_SIL2 = 6,4 %
PFH%_SIL2 = 11,5 %
Safety Relevant Parameters for the Quasar 900 Open Path Gas Detector
57
SafEyeTM Quasar 900 Gas Detector
C.2
58
General Conditions for Safe Use

The SafEye Quasar 900 IR Open-Path Gas Detector shall consist only of
the approved hardware and software modules.

The application advice and the limitations of the manual must be taken
into consideration. For calibration and maintenance, the regional and
national regulations have to be taken into consideration.

The 24V power supply must fulfill the requirements for PELV / SELV of
EN 60950.

The HART and RS 485 interfaces are not allowed to be used for the
transmission of safety related data.

The alert conditions according to SIL-2 can be implemented by an Alert
signal via the 20mA current loop.

After installation and configuration the set-up parameters must be
verified and the function of the SafEye Quasar 900 IR Open-Path Gas
Detector must be checked completely.

The alarm conditions of the transmitter must be checked periodically
together with the typical gas calibration checks. The SafEye Quasar 900
IR Open-Path Gas Detector must be switched OFF and ON.

The connected controller must monitor the 4-20mA signal current for
values below 4mA and above 20mA.

A defect on the SafEye Quasar 900 IR Open-Path Gas Detector must be
repaired within 72 hours.
General Conditions for Safe Use
Technical Support
For all technical assistance or support, contact:
218 Little Falls Road
Cedar Grove, NJ 07009, USA
Tel: +1 (973) 239 8398
Fax: +1 (973) 239 7614
Email: spectrex@spectrex.net
Web-site: www.spectrex.net
Your Local Office:
SPECTREX INC.
USA (Texas)
Mr. Jay Cooley, Regional Sales Manager
16203 Park Row, Suite 150
Houston, Texas 77084
USA
Phone: +1 (832) 321 5229
Email: jay@spectrex.net
Europe
Mr. Ian Buchanan, Regional Manager
6 Applecross Road
Glasgow G66 3TJ
United Kingdom
Phone: +44 (0) 141 578 0693
Email: ian@spectrex.net
Far East
Mr. Deryk Walker, Regional Sales Manager
59 Fen Ji Hu, Danshui
Taipei County 25163
Taiwan (ROC)
Phone: +886 2 8626 2893
Mobile: +886 926 664 232
Email: deryk@spectrex.net
Technical Support
59
SafEyeTM Quasar 900 Gas Detector
60
Technical Support