New A9a System User Instruction Manual Innovative Health Concepts LLC

New A9a System
User Instruction Manual
Manual rev: 2014-08
For assistance please contact:
Innovative Health Concepts LLC
Chicago, Illinois, USA
www.allevawave.com
Table of Contents
User Assistance Information .................................................................... 2
Warnings and Precautions ...................................................................... 2
Purpose of the A9a System ..................................................................... 2
Description of the A9a System ................................................................. 3
Environmental Conditions that Affect Use .................................................... 4
System Setup ...................................................................................... 5
Operating Instructions ........................................................................... 6
Coil Placement Examples ........................................................................ 9
Frequency and Duration of Use ............................................................. ...10
Cleaning and Maintenance ................................................................... .. 10
Servicing ........................................................................................ .. 10
Storage .......................................................................................... .. 10
FAQ .............................................................................................. .. 11
Recommendations…………………………………………………………………………………………………………………….13
Expected Battery Life ......................................................................... .. 14
www.allevawave.com
1 of 14
A9a User Instruction Manual
User Assistance Information
If you have questions or require assistance you may contact us at:
Innovative Health Concepts LLC
Chicago, Illinois, USA
www.allevawave.com mike@allevawave.com
Warnings and Precautions
Meanings of Warnings, Cautions and Notes
This manual contains specific warnings, cautions and notes which you should read and follow carefully. They are displayed as
shown below and have the following meanings:
WARNING
If a specific action or procedure is not followed, you may suffer personal injury or the device may
be damaged.
Caution
If a specific action or procedure is not followed, the device may be damaged.
Note
If a specific action or procedure is not followed, the device may not function properly.
General Precautions
This device emits electromagnetic energy by design. This type of energy goes by various names, including RF energy, radio
frequency energy, EMF, EMP PEMF, ICES and various other designations. It is a very low power device. The average power
output is approximately 1/3 watt, which is much less than the power output of a cellular telephone. All of this energy first
passes through the upper green LED, which then charges the system for each pulse. The total system generates only as much
power as you can visually see illuminating the upper green LED.
You should not use the device on or around the brain, ears, and eyes, unless you are under specific instructions from a medical
professional. If you are pregnant you should not use the device near the fetus. If you feel pain during the use of the device you
should stop using it and consult your physician.
If you experience increased pain or discomfort, you should consider using a lower power setting if available, changing the location of the
coils, or discontinuing use of the device altogether if the problem persists. If the application of ICES seems to make your problem worse,
we recommend that you immediately discontinue use of the device, and promptly talk to your physician.
There is no valid scientific reason to believe it is harmful to use this device. However, if you have concerns about exposure to
electromagnetic energy you must not use this device.
WARNING
You should not use this device near the brain, ears and eyes. You should not use the device near the fetus if
you are pregnant. If you feel pain while using the device you should stop using it and consult your
physician. And if you have concerns about exposure to electromagnetic energy then you must not use the
device.
Purpose of the Alleva-Wave™ A9a System
The Micro-Pulse Alleva-Wave ™system is intended to safely assist in recovery from injury or damage in hard and soft tissue, and to
reduce or eliminate swelling and pain in most tissues. This includes bones, tendons, ligaments, muscles, and cartilage. The AllevaWave system operates by applying a precisely controlled pulsed electro-magnetic field (PEMF )and inducing an electrical pulse .
The pulsed electromagnetic fields operate at a cellular level to promote repair and regeneration of structural tissues; it is because
of this cellular level action that the (ICES) system can positively affect such a wide range of tissue types.
www.allevawave.com
2 of 14
A9a User Instruction Manual
More specifically, the Alleva-Wave™ (ICES) Inductively Coupled Electrical Stimulation can be used to accelerate tissue
recovery following surgical procedures, promote faster tissue recovery following injuries (from sports, daily activities, work,
etc.), and reduce chronic swelling in tissues. An important application is the reduction of joint pain, whether chronic or from
injury.
Description of the Alleva-Wave™ A9a System
System Components
The A9a system has the following major components:
1
Pulse generator
Creates precisely controlled electrical pulses that power the coils.
2
9-volt battery
Supplies power to the pulse generator. Any type of 9v will work
3
Coils
Transmits electromagnetic pulses into the tissue being
stimulated. Each of the two coils has a bumpy side and a smooth
side, for easy orientation.
4
1
2
Battery charger (not shown) optional
Recharges the 9-volt battery when the battery is not being used
in the pulse generator.
Pulse Generator Details
The pulse generator has the following components. Many of these are explained in more detail in the following manual sections:
1 Connector for coils
Receives the plug for the cabling to the coils.
2
2 Upper green LED
1
Indicates the system is generating pulses in Stimulation Mode during
3
operation.
2 Lower red LED
Indicates there is a problem with the system that is not serviceable by
the user.
4
3 Cycle profile switch
Allows you to select from among four cycle profiles, which define how
the pulse generator outputs the stimulation pulses. The cycle profiles are indicated by L M H X
L= very low intensity setting 70 Tesla/S, small animals, sensitive subjects
M = standard intensity fields are generated, 100 Tesla/S, suggested initial
treatment protocol. Same as setting 1 on earlier models.
H = higher intensity fields are generated, 160 Tesla/S, for deep tissue injuries
X = 35% higher intensity fields are generated continuously, 200+ Tesla/S. Higher than setting 3 for older model
For experienced users only, 500 milliwatts of electro-magnetic energy in RF spectrum.
4 Socket for 9-volt battery with polarity indicators
Holds the 9-volt battery. The positive (+) and negative (-) polarity markings on the socket indicate how to insert
the battery to properly align its positive and negative terminals. Unit starts upon insertion.
www.allevawave.com
3 of 14
A9a User Instruction Manual
Theory of Operation
Pulse generator
The Alleva-Wave ™ ICES system promotes tissue recovery through the application of precisely generated magnetic and electrical
stimulation pulses which are created by the pulse generator. The stimulation pulses are designed with proprietary combinations of
frequencies, amplitudes, field directions, and wave shapes to activate tissue recovery at a cellular level
Four levels of ICES magnetic field intensity
The Alleva-Wave™ can generate either Low intensity magnetic fields for very mild or sensitive tissue stimulation, or M standard
intensity fields for mild ,standard tissue stimulation. You can move the cycle profile switch to position H to generate vigorous
intensity fields for deep tissue injuries and stimulation or you can select position X to generate 35% higher intensity fields.
When first using the Alleva-Wave ™it is recommended that you set the cycle profile to M (standard intensity) on the first day or
two as you become familiar with the device. Thereafter you can select cycle profile H (higher intensity) or X (higher intensity, only
experienced users). *only use when clear lower settings not effective, Do not try to accelerate healing by applying excessive power.
Higher intensity fields are recommended when the injured tissue is far from the coils. It is also recommended after you feel
comfortable with low intensity fields and you feel that more stimulation intensity would be beneficial for your tissue.
Cycle profile switch
The Alleva-Wave™ A9a operates on about a 40 minute cycle that repeats continuously and automatically. Using the cycle profile
switch, you can select from among four predefined cycle profiles.
Switch
Magnetic
Mode for
position
field intensity
Operation
L
M
Low
Standard
Stimulation (flashing green)
Stimulation (flashing green)
H
Higher
Stimulation (flashing green)
X
Higher
Stimulation (flashing green)
LED indicators
The LED's indicate various conditions and activities:
All LED's are OFF =
battery is dead or removed
Upper green LED is flashing =
system is operating and is in Stimulation Mode
Lower red LED is flashing =
system has a malfunction, requiring you to contact the supplier
Coils
The ICES pulses are transmitted into the tissue between or adjacent to the two coils. Having two coils provides flexibility in shaping
and directing the magnetic fields. Configuring the coils properly is described later in this manual. During stimulation you may hear
a high-frequency clicking sound coming from the coils. This is normal and is caused by the rapid change of the magnetic field
produced by the coils. Listening for the clicking sound is a good way to assure the system is working properly.
Environmental Conditions that Affect Use
The Alleva-Wave™ ICES system can be used during most normal daily activities. However, it cannot be used or worn when bathing,
swimming or during any activity that would expose the device to a wet environment. Do not submerge the device in water or any
liquid.
Caution
Do not expose the Alleva-Wave™ A9a system to a wet environment, or submerge the A9a system under
water, or clean the device using water. Doing so may permanently damage its electronics
www.allevawave.com
4 of 14
A9a User Instructions
Connect the Coils
Plug the coils into the pulse generator. Be sure to always connect and disconnect the coils with a battery removed. Do not
remove coils while unit is in operation. This is to avoid possible damage to the internal electronics.
Caution
Always remove battery before inserting or removing the coil connector. This will prevent possible
damage to the pulse generator electronics.
From the very first use of this system the coils should always remain plugged into the pulse generator and should never be
unplugged during use. Unplug the coils only when they need to be replaced. Though it is unlikely, the system can be damaged if it
is operated with a partially-inserted coil plug, or if the plug is removed during operation.
Insert the Battery
The battery bay has a spring loaded quick-connect mechanism to make it easy to insert and remove batteries during operation.
This allows you to keep the system in place while changing batteries. No tools are needed to change batteries. The system will
operate as long as the battery remains inserted, until the battery is fully drained.
You can use any standard 9-volt battery. Either non-rechargeable batteries (typically
Alkaline) or rechargeable batteries (Lithium or NiMH) can be used. 7.2/8.4/9.6 volts. It is
recommended to use rechargeable batteries to minimize your operating cost. A very
common approach is to have two rechargeable batteries and a quick-charging battery
charger. This allows you to use the Alleva-Wave ™system with one battery while the other
charges, so that you can switch between the two batteries for continuous stimulation.
Insert a fully charged battery into the pulse generator. Confirm that its terminals are
correctly aligned to the positive (+) and negative (-) markings on the pulse generator.
Unit will NOT operate if battery is inserted in reversed polarity.
A rechargeable NiMH battery will normally last four to eight hours during operation. For
more detailed battery life information please refer to the "Expected Battery Life" section
of this manual.
It is okay to run the Alleva-Wave ™system until the battery is completely drained.
Simply recharge or replace the battery to continue stimulation.
WARNING
Do not charge Alkaline and other types of non-rechargeable batteries. They can explode and cause injury.
Check the System
Before using the Alleva-Wave ™A9a ICES system you should:
1)
Ensure the connector for the coils is completely inserted into the pulse generator.
2)
Inspect the coils and the cabling to ensure there are no exposed or broken wires.
3)
Ensure the pulse generator has power from the 9-volt battery. Do this by moving the cycle profile switch to
position L, M or H. Then the upper green LED should begin flashing, which indicates the system is operating in
Stimulation Mode. If there is no LED activity or if the LED's are dim, then the battery is drained and needs to be
recharged or replaced. Red LED indicates system failure. Contact manufacturer.
www.allevawave.com
5 of 14
A9a User Instruction Manual
Operating Instructions
Record the Injury or Pain History
It is very useful to put into writing the history and progression of the injury or pain before using the Alleva-Wave™ ICES system
and then record your response throughout the stimulation period. Keeping a written record will help you objectively monitor
your pets progress and make better decisions regarding coil placement strategies (explained later in this section).
Understand Coil Configurations and Stimulation Zones
The Alleva-Wave™ ICES system has two specially- designed coils. The two coils allow you to control how the pulsed magnetic field
is shaped into different stimulation zones so that you can better focus the ICES on the injury site. This maximizes the stimulation
to the injured tissue while minimizing unnecessary stimulation to surrounding tissues. This ability to control stimulation zones is a
unique advantage of the A9a system.
In addition, each coil is flexible to allow you to bend and shape the coils to conform them to the surface of the body, limb,
neck, face, jaw, digits, or whatever area is being stimulated.
The two coils are always placed with the bumpy side away from the skin. They are positioned in one of two basic configurations
depending on the area of the body to be stimulated:
Opposite-Side - The Opposite- Side configuration is generally used to stimulate thinner body sections and joints, such as
arms, elbows, knees, ankles, etc. In such cases the two coils are located on opposite sides of the injured or painful tissue. The
Opposite-Side configuration is preferred over Side-by-Side when possible, as it produces stronger and better focused
electromagnetic fields in the tissue being stimulated.
Side-by- Side - The Side-by-Side configuration can be effective for large, thick parts of the body when the injury is relatively
close to the surface, within 1 to 2 inches (25 to 50 mm) of the skin. In these cases the coils are situated on the same side of
the injury area, adjacent to each other.
The coil configurations result in the two different stimulation zones illustrated below. The darker regions of the stimulation zones
indicate where the magnetic fields are strongest. The stimulation zone of the opposite-side coil configuration is narrower and
deeper, whereas the stimulation zone of the side-by-side configuration is wider and more shallow.
ICE Stimulation Zone of the Opposite-Side Configuration
ICE Stimulation Zone of the Side-by-Side Configuration
www.Alleva-Wave.com
6 of 14
A9a User Instruction Manual
Sometimes it may not be obvious which configuration should be used. For example, a shoulder injury might best be stimulated
with the Opposite-Side configuration or the Side-by-Side. Or, it may be stimulated by alternating between the two configurations.
You are encouraged to experiment with configurations and coil placements to determine which works best for you. The coils can
also be stacked with bumps facing bumps to focus wave on single injury point similar to acupuncture or laser treatment.
Place the Coils in Position
The orientation of the coils to each other and to the injury is important for optimal benefit from the ICES fields. There are
three guidelines when locating the coils on the body:
1) Put the bumpy side out of each coil, away from the skin
2) Place the coils as close together as possible (Side-by-Side configuration)
3) Position the coils so the injured tissue is between the coils (Opposite-Side configuration)
4 ) Stacked with bumps facing each other.
Note
Follow the three coil placement guidelines to make sure you are receiving the full benefit from
the magnetic fields.
Visualizing the ICES field can be helpful when considering how to place the coils. When the coils are placed in an Opposite-Side
configuration, they will generate a magnetic/electrical field that fills the space between them. The closer together you place the
coils, the stronger the field will be, and the more stimulation you will be able to focus onto the injured tissue. When the coils are
placed in a Side-by-Side configuration, they will generate a ICES field that is wider across the area. The closer the two adjacent coils
are, the stronger the field will be. It is therefore recommended that the coils touch each other in the Side-by-Side configuration.
As stated in the guidelines, place the bumpy side of each coil away from the skin. This applies to both the Opposite-Side and Sideby-Side configurations. This ensures the ICES fields are lined up properly. If the bumpy sides are not out, then the ICES fields may
oppose and cancel one another, and will render the system ineffective.
It is okay if the wound is larger than can be completely covered by the coils. The stimulation has a beneficial effect on tissues in
the general area of stimulation, so coil placements do not need to be precise.
It may also be helpful to move and reposition the coils occasionally, especially if the exact site of the injury is not clear.
Repositioning will change the orientation of the magnetic field, which is often beneficial in the recovery process. For example, on
the first day or two you might place the coils on a limb on the lateral (outside) and medial (inside) surfaces. Then for the following
day or two you might choose to reposition the coils on the dorsal (back) and ventral (front) surfaces around the injury. You might
later decide to place them obliquely (at an angle).
The coils can be stacked (bumpy side to bumpy with smooth side against skin) this stacked configuration is used when the
location of an injury is pinpointed. This type of treatment is similar to acupuncture, where the needle is precisely located on
known meridians for pain erasure.
To the extent possible, it is best to try several options to determine what seems to work best. The preferred placement of the coils
might ultimately be driven mainly by practical concerns such as ease of securing or bandaging the coils or placement of the pulse
generator.
You should not use the device on or around the brain, ears, and eyes, unless you are under specific instructions from a
medical professional.
WARNING
Do not use the Alleva-Wave™ on or around the brain, ears, and eyes unless under specific instructions
from a medical professional. Always use on low setting if advised.
www.allevawave.com
7 of 14
A9a User Instruction Manual
Secure the Coils and the Pulse Generator
The two coils are normally placed directly against the skin and secured in
place using adhesive medical tape or medical bandages. Loose clothing can be
worn over the coils, if their location permits.
Alternatively, the coils can be located outside the wraps and fixed in place using
tape or elastic bands. In this case the wraps and coils at the location should be
held close to the skin for maximum magnetic field penetration into the injured
area.
The ICES fields will penetrate clothing, tape, bandages, leather and other nonmetallic materials with no problems. However, the magnetic fields can be
distorted or weakened by the presence of metallic objects. These can include
watches, metal zippers, belt buckles, coins, cell phones, and other electronic
devices. You should keep metallic objects at least 8 inches (200 mm) away from
the coils.
Note
Keep metallic objects at least 8 inches (200 mm) away from the coils to ensure the magnetic fields
are properly focused.
The pulse generator can similarly be secured by adhesive medical tape or medical bandages. It can also be placed in a pocket if one
is available. Some users fashion a special simple garment with extra pockets or pouches to hold the coils and pulse generator in
position during stimulation, for example, using an old t-shirt and iron-on pockets or stitched-on pouches. This works well because
each individual case is different.
Coil Life and Coil Test Chip
The Alleva-Wave coils are very durable but they tend to see a lot of mechanical wear and tear depending on amount of use. It is
possible that due to continuous daily usage, customers should expect to have to replace coils about once per month (675-1500
hrs). The average user will get 3 to 6 months of use out of their coils, but you have to keep in mind the coils are like a toothbrush:
more use = shorter life expectancy. We provide 2 sets of coils with each new unit order at no extra cost, to provide a backup so
that no interruption of treatment exists. We suggest you should reorder and replace worn out coils immediately. If you can hear
the coils clicking, they are working. Also avoid any cross contamination by avoiding sharing coils between animals.
The coil test chip can also be used to visually verify that the coils are properly working. Simply place the coil over the tester chip. If
the coil test chip’s LED blinks green, then the coil is operational.
Monitor the Pulse Generator LED's
As you use the Alleva-Wave™, occasionally look at its upper green LED is blinking. If the upper green LED is flashing then the A9a is
in Stimulation Mode during its cycle.
If the LED is not lit then likely the battery is drained.
Occasionally Change Coil Locations
One very good strategy is to occasionally change the placement of the coils. This helps direct the ICES fields in different directions
toward the injured area, and thus aids the recovery process. Changing coil locations regularly works particularly well on long-term
programs.
For example, for lower back stimulation you might change the positions of the coils on your back several times during the day,
assuming they are in a Side-by-Side configuration. Then every two or three days switch from the Side-by -Side configuration to
an Opposite-Side configuration where one coil is located on the lower abdomen and the other on the lower back (both coils with
bumps out).
www.allevawave.com
8 of 14
A9a User Instruction Manual
You can try many different coil placement strategies to see which ones work best for you. Changing coil locations, and therefore
changing the tissue stimulation pattern, is similar to occasionally changing the way you exercise. In both cases your body
experiences longer term health gains as it's subjected to a variety of different external stimulations.
Another reason to change coil locations is the effect of "accommodation", where tissues of the body often learn to ignore repeated
external signals. You can see this in many ways. One way is where certain exercises or workouts that used to work well no longer
seem to give the same benefit. Another is where pain drugs can have slightly less effectiveness through repeated use. And still
another is where very spicy foods that once were intolerable can taste good after building resistance to their chemical burn.
In the same way body tissues over time will tend to accommodate the Alleva-Wave’s™ Inductively Coupled Electrical Stimulation
(ICES)- if it is kept the same - and will therefore reduce the effectiveness of the stimulation. To reduce this accommodation effect
the brain and nervous system need to be kept "guessing" by periodically changing the PEMF stimulation orientation. You are
encouraged to try different coil location and configuration strategies to reduce the effects of tissue accommodation.
Remove Battery When Done
Please remove the battery when you are not using the Alleva-Wave™. This helps to preserve battery energy. If a rechargeable 9volt battery is being used and if it is drained, then it can be placed in an appropriate battery charger in preparation for next use. It
is advised to remove battery when not in use to avoid battery drain. Unit will continue to operate with charged battery inserted.
Continue Recording the Injury or Pain Response
As you progress through the stimulations continue to keep a daily written record of your response. This will help you make
adjustments to coil placement locations and assist you in monitoring your progress.
Coil Placement Examples
Below are examples of how coils can be located on different parts of the body in either Side-by-Side or Opposite -Side
configurations. Two or more coil placement alternatives are shown for each injury area below to emphasize the importance of
periodically moving the coils into different orientations. Many other areas of the body can be stimulated with the Alleva-Wave™
system beyond those shown.
The illustrations below show the coils in their "flat" geometry. However, you are encouraged to flex and shape the coils to your
pets’ body contours for both comfort and greater stimulation effectiveness.
Please Note: DO NOT use any type of creams, liniments, oils etc. with the coils. It may cause irritations. It is also not recommended
to treat multiple horses or pets with the same set of coils. This will avoid any cross contamination. Additional coils are available for
purchase.
www.allevawave.com
9 of 14
A9a User Instruction Manual
Frequency and Duration of Use
The frequency and duration of use varies with the person and situation. A first time user with a mild injury or mild symptoms might
use the Alleva-Wave™ only once per day for 4 hours at a standard stimulation intensity setting, for example, setting M. A more
experienced user with a more severe injury or severe symptoms might use the device 2 or 3 times each day for 4 to 6 hours each
time at a higher stimulation setting. Between these two extremes there are many possible regimens you can experiment with. A
very typical regimen is 6 to 8 hours per day, either during the day or while sleeping at night.
Rest periods are as important as stimulation periods, to allow tissues to recover. Simply not using the device for certain periods of
time are advised.
Each person's daily pattern of use will be unique to their injury or symptom, their daily routine, and their response to the
stimulation. You are encouraged to experiment with usage patterns and coil locations to find what works best for you. Also,
keeping a daily written record of your response will help you make adjustments to usage patterns and coil locations.
Output Intensity
Coils configuration determines the intensity of the ICES field. Separately, each coil together will produce a specific intensity (for the
A9a model, 70 to 90 *Gauss on setting L). When placed near each other (coils next to each other on the back, for example), or
opposite each other (one coil on each side of a knee, for example), the intensity increases (for the A9a model, up to 120 Gauss on
setting M). Setting H output is up to 180 Gauss. When the coils are stacked (nubby side to nubby side), intensity increases further
(for the A9a model, up to 200 Gauss.) Setting X output is consistently in excess of 200+ Gauss regardless of coil configuration.
* Gauss is standard measurement of magnetic wave strength. 1 Tesla=10,000 Gauss
Cleaning and Maintenance
The Alleva-Wave™ system requires very little cleaning. If the coils and cable become dirty with use then you can clean them
with common rubbing alcohol (isopropyl alcohol).
The Alleva-Wave ™system requires no maintenance or calibration.
Servicing
The Alleva-Wave ™system has no parts that you can service or repair.
If the coils become damaged please contact your supplier for advice on replacement options. www.alleva-wave.com
If the lower red LED is flashing, then there is an electrical fault in the device and it will not operate properly. Please contact your
supplier for advice on repair and replacement options.
Storage
When storing the system for more than one month you should remove the 9-volt battery to prevent corrosion. Recharge batteries
before using the system again.
Caution
Do not store the system with the 9-volt battery installed, as this may result in damage to the
pulse generator from battery corrosion. The unit will trickle drain a new battery.
www.allevawave.com
10 of 14
A9a User Instruction Manual
FAQ
Q: How do I know if my device is working properly?
First, plug in your coils and make sure they are plugged in all the way. Then insert a fresh, fully charged 9-volt battery. Then
move the cycle profile switch to either position L, M, H or X. You should see the center LED flash three times, and then see the
upper green LED flash as the device operates. Finally, put each coil close to your ear and listen for its "click-click -click" sound.
Or test with chip, If you see the green LED flashing and if you hear the clicking sound, then the Alleva-Wave ™system is
functioning properly.
Q: How will I know when the battery needs charging?
The LED's will become dim or will go out entirely, and you will no longer be able to hear the coils clicking. Refer to the
"Expected Battery Life" section in this manual to determine the approximate battery life that applies to your battery
and usage.
Q: Can I use any battery charger to recharge the battery?
Any commercial battery charger will work as long as you have rechargeable batteries. Be sure to match rechargeable batteries to
the correct type of charger or you could easily damage or destroy your batteries and charger. Almost all types of rechargeable
batteries require a different specific charger: NiCad, NiMH, Li-Ion, Li-Polymer, Hybrid, etc.7.2/8.4/9.6 volt will work.
Q: The device makes a clicking or humming noise. Is this normal?
Yes, this is normal. The device is humming for the same reason that power transformers on telephone poles sometimes hum
and vibrate. The phenomenon is known as magnetostriction, and it is normal and harmless. Unfortunately it is an
unavoidable side effect of the physical operation of the device. We are working to reduce the noise in future designs but this
may not be possible due to the physics of the device operation. When you hear these sounds in most cases it means the
device is operating properly. To diminish noise, try covering unit with a towel or wrap.
Q: One of the coils does not seem to be working (no clicking sound) but the other one still works. Will
the device still be effective?
The device will not be as effective with a non-functional coil. Both coils are required for effective stimulation and we
recommend that you order replacement coils.
Q: I would like to locate the pulse generator farther from the coils when I use the device. Are longer coil
cables available? Or can I add an extension cable to the coil cable?
We offer 3 lengths of coil cable. 10”,20” & 47”. We do this because the cable length of the Alleva-Wave™ is designed to
transmit maximum energy to the coils. Longer cable lengths decrease the transmitted energy due to electrical losses in the
cable.
If you feel it necessary to add an extension cable, it should conform to the following guidelines:
a) cable length: 24 inches (600 mm) or shorter; shorter is better
b) cable type: high-performance audio cable; higher quality cable provides better performance
c) connector style: 3-conductor audio, 1/8" diameter audio plug and socket
d) connector material: gold-plated; wipe contacts clean before first use
Depending on various factors, adding an extension cable will create an approximate 20% reduction in magnetic energy at
the coils. This will reduce stimulation to the tissue and will reduce the device's effectiveness.
Q: If the pulse generator is damaged in any way, is there a chance the device can cause a fire?
NEVER use a damaged device.
Although the risk of creating fire with the device is very remote, we recommend that you do not take any risks. The units are
internally sealed against moisture and tampering, so it is not possible to repair a pulse generator once it has sustained
damage. We recommend you contact your reseller to order a replacement unit.
www.allevawave.com
11 of 14
A2 User Instruction Manual
FAQ continued
Q: I dropped the pulse generator in water. What do I need to do? Will it ever work again?
There is no guarantee it will work again. However, in some cases, if it is dried quickly and thoroughly enough, it might function
again. First, IMMEDIATELY get the pulse generator out of the water and disconnect the battery. Then disconnect the coils.
Second, shake the device vigorously to remove any excess water and wipe the exterior dry. Third, put about 0.5 inch (12 mm)
of dry white rice in the bottom of a jar or bowl and place the pulse generator on top of the rice. Fourth, seal the top of the jar
or bowl, and let sit overnight. The dry rice will absorb moisture in the container, including (hopefully) the water in the pulse
generator. Discard the rice after use. If available, it is preferable to use desiccant instead since it will absorb moisture better
than rice. DO NOT place the pulse generator in an oven or microwave to dry.
Q: Why don't I feel the current when the system is in operation? Will it give me a shock?
The Alleva-Wave’s™ output is a magnetic field and inductively coupled electrical field NOT a direct electric current like a TENS
unit. The device is very low power. In fact, all of the energy that goes into the magnetic field first passes through the upper green
LED, so the total amount of output power is very low, only 1/3 watt. This is much less than the power of a typical cell phone. The
voltage and amperage is too low to give a painful shock. A slight tingling may be felt if the coil wires become exposed through
damage, but no harm to the skin will result.
Q: I feel a tingling sensation when I use the device. Is this normal?
Yes, it is a common occurrence and generally does not indicate any type of problem. This happens most often when
stimulating tissues where the coils can be placed on opposite sides and very close together, such as across the wrist, hand,
ankle, or foot. If the sensation causes discomfort you could try lowering the intensity setting or placing the coils slightly
farther apart. Usually this sensation is not unpleasant and after some use it will gradually fade until you no longer notice it.
Q: When I use the device it causes the pain to intensify. What should I do?
Generally the device should not cause pain or discomfort. If this happens you might try shifting the position of the coils and
reducing the intensity setting on the device if possible. If the problem persists for any length of time we strongly recommend
that you discontinue use of the device and seek medical advice. Based on our experience this type of reaction often indicates
that you need a physician or specialist to take a detailed look at the area with the increased pain. Increased pain indicates a
need for close medical attention and advice.
Q: How soon should I expect a reduction in pain?
Each person's response to the ICES stimulation is different and we cannot guarantee results. However, pain reduction often
occurs within the first 3 to 4 days of use, and permanent reduction of pain often occurs within 2 to 3 weeks of correct use.
.Q: Does it matter which way the battery or optional 9v adapter is inserted?
The proper polarity +/- must be used for the unit to operate with a 9v battery. The optional 9v adapter must also be
inserted with the correct polarity prior to being plugged into a wall outlet. Failure to do so could result in a electrical fire or
shock and damage to the internal components of the unit.
Please Note: DO NOT use any type of creams, liniments, oils etc. with the coils. It may cause irritations. It is also not
recommended to treat multiple horses, pets or people with the same set of coils. This will avoid any cross
contamination. Additional coils are available for purchase.
www.allevawave.com
12 of 14
A9a User Instruction Manual
General Recommendations for ICES Setting and Usage
L = LOW. This setting is about half the power of the minimum setting (#1) used in earlier ICES models. Nonetheless, the power
setting "L" has been demonstrated by independent laboratory testing to be very effective at reducing acute inflammation in
controlled laboratory animal testing. RECOMMENDED USE: For very sensitive animals such as horses and very small animals. We
recommend beginning any treatment of small or very sensitive animals using the setting "L". The setting may be increased without
limit if the animal seems to tolerate the device well and if more powerful stimulation is thought to be likely to be more effective.
M = MEDIUM. This setting has equivalent peak power to the original ICES setting #1, but there is no rest period on this or any of
the new ICES power settings (L M H and X). RECOMMENDED USE: We recommend beginning all treatments on the setting "M"
unless otherwise recommended for the setting "L" above.
H = HIGH. This setting has equivalent peak power to the original ICES settings #2 and #3, but there is no rest period on this or any
of the new ICES power settings (L M H and X). So the setting "H" is exactly equivalent to the original power setting #3, which also
had no rest period. RECOMMENDED USE: For experienced users who have experience with the way injuries respond to ICES
technology, it is OK to begin at power setting "H", especially for deep tissue injuries, acute recent and severe injuries, and for
injuries that are not responsive to treatment for at least two weeks on a lower setting ("L" or "M").
X = Extra-high power. This setting delivers more power than was available from earlier ICES systems. It delivers about 35% more
power than the highest previous setting (#3). RECOMMENDED USE: For experienced users only. Please use this setting only when
it is absolutely clear that lower power will not yield acceptable results. Please do not try to accelerate the effects by applying
excessive power. On this power setting the ICES system delivers about 500 mW (500 milli-Watts, which is the same as 1/2 Watt) of
electro-magnetic energy in the RF spectrum, which is non-ionizing radio-frequency radiation. At that power level, which is
equivalent to about half the maximum RF power radiation of a typical cellular phone, there is no reliable data that we are aware of,
nor any scientific reason that we know of, to suggest safety concerns. However, we believe it is always good practice to minimize
the total exposure to radiated energy in any form. And finally, based upon independent laboratory testing the data clearly indicate
that lower power settings cab be very effective at reducing inflammation, with somewhat diminishing returns for higher power
settings. We believe it is best to reserve the "X" setting only for immediate use after severe injury, or when lower power settings
after careful and prolonged use do not yield adequate relief.
GENERAL ADVICE:
Reduce the power setting if any discomfort is evident. Discontinue use if discomfort persists or worsens.
If symptoms do not improve within the first few days of use, before trying higher power settings try different placements of the
coils around and near the injury. Often placing the coils closer to the center of the body than the site of the injury works well,
especially for injures involving nerve tissue or the lower spine.
Always place the coils correctly: bumps away from the skin.
Allow at least three weeks for beneficial effects. Then continue use for a more permanent effect (see below)
Replace/recharge batteries as needed. Expected battery life depends mostly upon the power setting (L M H or X). Using a 500
mAh energy capacity Lithium-polymer rechargeable 9 volt battery, the expected battery life between charges is as follows:
power
setting hours of continuous use
L
generally more than 24
M
about 20-24
H
about 10-12
X
about 8 hours maximum
Don't stop treatment as soon as a benefit is seen. Continue treatment for at least three weeks after symptoms have subsided, or
for at least as long a period as was required to get symptomatic relief, whichever is longer, to get the best chance for a permanent
improvement. Otherwise symptoms may return after a period of time when use of the ICES system has been prematurely
discontinued.
www.allevawave.com
13 of 14
A9a User Instruction Manual
Expected Battery Life
The hours of battery operation you can expect depends on two factors: the battery type and the cycle profile setting selected (1, 2
or 3). Certain makes and types of batteries have more available energy than others. For a given battery, as the cycle profile setting
is increased, the expected battery life decreases due to larger power requirements.
Tests were run on a sampling of rechargeable and non-rechargeable batteries to give you an idea of the approximate battery life
in hours of continuous operation for each switch setting. The results are shown in the summary table below. The table assumes a
battery is still "good" if its output power remains at or above 80% of its initial full power value.
There are many good batteries available for you to buy and this is not a recommendation for the brands included in this test.
Rather, this test data is provided to give you an estimation of the range of operating battery life likely depending on battery
type and cycle profile setting. With this data you can estimate the duration and cost of operating the device and therefore
select the best battery for your overall needs.
Battery Life Estimates Summary Table
Hours of operation until battery output drops below 80% full power
Setting L
Setting M
Setting H Setting X
+ Li-Ion MaxPower 550 mAh
34
24
12
8
+
Li-Ion Marsell 350 mAh
24
16
9
4
+
NiMH Tenergy 250 mAh
11
6
4
2
+
NiMH Tenergy 200 mAh
13
8
7
6
*
Alkaline Energizer MAX
30
22
12
6
*
Alkaline DuraCell
28
24
12
5
*
Alkaline Generic
25
17
10
4
+ Rechargeable batteries. Can be used many times.
Non-rechargeable batteries. Can be used only once.
* Never recharge a non rechargeable battery
Alleva-Wave™ is designed and assembled in USA by Micro-Pulse LLC www.pulse-pet.com
www.allevawave.com
14 of 14
A2a User Instruction Manual