Concentration of spinbath in Viscose (Rayon)
fiber and filament production - a key for perfect
fiber and filament quality.
Our TALI 200 measures the sulphuric acid and
zinc sulphate concentration of the spinbath and
controls the process by means of a continuous
on-line analysis. In connection with a density
measurement, also the amount of Natrium
Sulphate can be determined.
Reduced deviations of spinbath concentrations
stabilize the product quality. The on-line
measurement and process control reduce
costs for labor, input of chemicals and avoid
human errors.
The benefits of TALI 200 have been proven
for more than 20 years in numerous Viscose
productions worldwide.
PROCESS CONTROL
TALI 200
TITRATING APPARATUS
Scope:
Titrating apparatus for continuous
on-line analysis of spinbath in
Viscose (Rayon) fiber and
filament productions to determine
the concentrations of sulfuric acid
and zinc sulfate and to control the
process using industrially
standardized signals.
Closed control loop:
The output signals can be
used for automatic dosing and
control of H2SO4 and ZnSO4,
for keeping the concentration
automatically within certain
limits
Measuring ranges:
H2SO4
up to 250 g/l
ZnSO4
up to 250 g/l
Na2SO4* up to 250 g/l
Accuracy:
H2SO4
ZnSO4
Na2SO4*
TALI 200 consists of a fully
stainless steel housing with
splash proof protection IP65,
analysing unit with two storage
containers, electrical unit with free
programmable control system
and operator panel with LCD
display
<±1.0 g/l
<±1.0 g/l
<±1.0 g/l
Electrical mains supply:
220 V, 50 - 60 Hz, single phase
with earth contact, max. 400 VA
(other voltages on request)
Measurement frequency:
Programmable, max. 6
measurements/hour
Output signal:
Analogous 0 - 20 mA or
4 - 20 mA for H2SO4 and ZnSO4
concentrations
Analogous 0 - 20 mA or
4 - 20 mA for Na2SO4*
concentration
Ambient conditions:
Temperature max. 35°C
Relative humidity max. 90%,
not condensing
Measuring principle:
Automatic pH titration with NaOH,
Zn titration with complex former
and NaOH.
Dimensions:
Height:
Width:
Depth:
Weight:
1600 mm
1040 mm
720 mm
approx. 175kg
Alarm levels:
Can be specified individually
due to requirements
Optional:
Serial interface for data output
to a printer, process control
system
Limit signal (relay contacts)
Example of process improvement through TALI 200
Concentration after
installation of TALI200 in
combination with closed
control loop
Technical data and pictures are subject to change!