Phys. Med. Biol. 59 (2014) R349–R369
Physics in Medicine & Biology
doi:10.1088/0031-9155/59/21/R349
Topical Review
MR guidance in radiotherapy
Jan J W Lagendijk, Bas W Raaymakers, Cornelis A T Van den Berg, Marinus A Moerland, Marielle E Philippens
and Marco van Vulpen
Department of Radiotherapy, University Medical Centre Utrecht, Heidelberglaan 100,
Utrecht, The Netherlands
Received 7 February 2014, revised 20 May 2014
Accepted for publication 28 July 2014
Published 16 October 2014
Keywords: MRI, radiotherapy, adaptive, imaging
(Some figures may appear in colour only in the online journal)
1. Introduction
The use of modern MRI technology in radiotherapy is fast growing. MRI is entering all facets
of radiotherapy. This review will start with a very short overview of the newest developments
in MRI technology after which it will cover, in separate paragraphs, the various radiotherapy
related topics:
•Developments MRI technology
•Tumour visualisation, characterization and response assessment for Radiotherapy
•Treatment simulation
•MRI guidance for brachytherapy
•MRI guidance for external beam radiotherapy
This review will serve as a short introduction to this wide field. The focus will lie on the
new development of MRI guidance of the actual therapy.
2. Developments MRI technology
Magnetic resonance imaging (MRI) has become, over the last 30 years, one of the main pillars of modern diagnostic imaging, which is widely used for many clinical problems. One of
the reasons for its success is the high level of technical innovation which has turned modernday MRI into a versatile medical imaging modality (Haacke et al 1999). MRI stands out in
terms of soft tissue visualization. MRI does not require ionizing radiation. In addition, it provides several functional imaging modalities for measuring biological function and physiology.
Throughout its relatively short history the strive for higher magnetic field strength has been
a constant factor from 0.5 T MR scanners in the eighties to 1.5 and 3 T MR scanners in current clinics. A higher magnetic field strength is favourable as it leads to higher signal-to-noise
0031-9155/14/21R349+22$33.00 © 2014 Institute of Physics and Engineering in Medicine Printed in the UK & the USA
R349
Topical Review
Phys. Med. Biol. 59 (2014) R349
ratios that can be traded off for higher spatial resolution or imaging speed. In addition, various functional imaging contrasts such as Blood Oxygen Level Dependence (BOLD) are augmented by the higher magnetic field strength (Ogawa et al 1993, Donahue et al 2011). At the
moment, 7 T MRI forms the forefront of this quest for higher field strength and recently 7 T
neuro-imaging has produced remarkable results, which provide new insight into brain anatomy and function (van Veluw et al 2013).
Another technological aspect that has played a key role in the advancement of MRI is the
use of multi-detector receive. While in the early days, a single surface loop was used for signal
reception, a modern MR scanner employs 32 or even 64 receive channels connected to dedicated
multi element detectors (receive coil arrays) geometrically tailored to the scanned body part. A
phased array operation of such a multi-channel detector allows for maximum signal reception
and minimization of noise contamination of the signal (Roemer et al 1990). A very recent trend
is digitalization of the signals on the detector, which are then passed to the image reconstructor
by optical fibres minimizing electronic distortions and RF loss (Possanzini et al 2011). Vendors
have integrated this digital technology into new generation 1.5 and 3 T MR scanners with greater
bore sizes going from 60–70 cm bore diameter. This is relevant for radiotherapy purposes and the
extra space allows scanning patients in radiotherapy treatment postures.
Although MRI cannot yet compete with multi-detector CT imaging, in terms of acquisition
speed, it has made vast progress in image acceleration. The above mentioned development
of parallel imaging technology was a key step and allows for sub-Nyquist undersampling of
k-space by exploiting the different receive sensitivity patterns of the receive channels to unfold
the otherwise aliased images of the different channels into one unfolded image (Sodickson et
al 2005, Pruessmann 2006). A more recent trend to accelerate image acquisition is the use of
compressed sensing methodology, which has been very successful when applied to various MR
imaging techniques (Lustig et al 2007). Today, very rapid advancements are being achieved in
functional imaging of the brain where new multi-slice, multi-band imaging technology allows
high resolution, whole brain images in a few seconds (Larkman et al 2001, Breuer et al 2005).
Neuro-science is the main driver behind the development of various new MR imaging
technologies. MR imaging of the brain is relativity easy due to minimal motion compared to
thoracic, abdominal and pelvic organ sites. Abdominal and thoracic MR imaging faces more
severe challenges due to cardiac and respiratory induced motion that can result in a variety of
artefacts. Especially cardiac MR imaging, which played a pioneering role in the development
of strategies to mitigate or correct patient motion (Lanzer et al 1984). Navigator echo techniques were developed to reduce such artefacts allowing for the monitoring and correcting of
motion during MR acquisitions by tracking the movement of objects in real time (Ehman and
Felmlee 1989, Stam et al 2012). Other acquisition strategies to address patient motion include
the use of breath holds and gated acquisitions (Runge et al 1984). By gating the MR acquisition
on the ECG or respiratory signal, artefact free MR images can be made of several phases of the
cardiac and respiratory cycle (Ehman et al 1984, Lewis et al 1986). Such sequences allow for
new detailed imaging of body locations like the oesophagus, pancreas and kidneys (figure 1).
3. Tumour visualisation, characterization and response assessment for
radiotherapy
In radiology the diagnostic performance of an imaging method is leading. In contrast, for
radiotherapy treatment planning accurate tumour delineation is the most important factor.
Therefore, first robust imaging with high resolution and high contrast between tumour and
surrounding normal tissue is required.
R350
Topical Review
Phys. Med. Biol. 59 (2014) R349
Figure 1. Example MRI oesophagus tumour, improvement image quality using cardiac
triggering and breath hold (reproduced with permission from van Rossum et al 2013).
Since all directions, left–right, anterior–posterior and cranial–caudal are equally important
for treatment planning and delivery, spatial resolution is important and preferably isotropic,
3-dimensional (3D) MR images should be used. An additional advantage of 3D MRI is the
improved signal to noise ratio. However, this is generally at the cost of increased motion
sensitivity and increased acquisition time. Steady state free precession (SSFP) acquisition
methods are most appropriate for fast, high resolution acquisition and are geometrically robust
(Patz and Hawkens 1986). The disadvantage is that these sequences show a mixed T1 and
T2 contrast and are susceptible to magnetic field (B0) inhomogeneities, which may lead to
banding artefacts (Hargreaves 2012). Fast spin echo methods show a better defined contrast,
are intrinsically corrected for B0 inhomogeneities, but are more time consuming and sensitive to blurring due to long decaying echo trains (Reichenbach et al 1997, Markl and Leupold
2012). A special acquisition method is the echo planar imaging (EPI) read out. This acquisition method is the workhorse for fMRI brain studies and diffusion weighted imaging, allowing
very fast data collection with excellent SNR performance (Bammer 2003). The application in
radiotherapy, however, is far from ideal due to its extreme sensitivity to B0 inhomogeneities,
causing large geometric distortions, especially in areas with air-tissue contrasts, such as head
and neck (Le Bihan et al 2006, Schakel et al 2013).
For abdominal and thoracic organs, respiratory induced motion deteriorates MR images
and imaging motion artefacts have to be resolved for accurate tumour visualization. The simplest method is to apply breath-holds. However, breath-holds elongate measurement time and
restrict image resolution due to the limited acquisition time available per breath hold (Low
2007, Wielpütz and Kauczor 2012). Further, breath-hold positions are poorly reproduced,
which will lead to discontinuities in the image. A more patient friendly approach is to apply
respiratory triggering either using MR navigators or with an external respiratory pressure belt
(Oechsner et al 2009). This enables data acquisition in a consistent phase of the breathing
cycle and prevents motion artefacts in the images. Using these techniques, tumours in the pancreas, liver and stomach are imaged successfully (Marin et al 2010). A comparable procedure,
using cardiac triggering, can be applied to correct for cardiac motion while measuring thoracic
lesions, like the oesophageal cancer or mediastinal lymph nodes (Riddell et al 2007)
Besides high resolution, high tumour-background contrast is also needed to accurately
delineate tumour volumes. Figure 2 gives a visualization example of a Gd-enhanced
T1-weighted MRI of a hypopharyngeal carcinoma. The strength of MRI is that different
R351
Topical Review
Phys. Med. Biol. 59 (2014) R349
Figure 2. Computed tomography image (left) of a patient with a T2N2bM0 hy-
popharyngeal carcinoma, compared with a Gd-enhanced T1-weighted MR image
(right) at the same location (reproduced with permission from Verduijn et al 2009).
endogenous contrasts can be created by manipulation of the pulse sequences parameters
(Plewes and Kucharczyk 2012). The most important independent contrast mechanisms are
T1 and T2 relaxation and water diffusion (Bottomley et al 1987). Both native T1 and T2
contrasts are useful to distinguish solid tumours from normal tissue, depicting water content and chemical environment. The additional contrast is apparent between water and fat
containing tissues, which helps to define tumour boundaries. In the head and neck, squamous cell carcinoma appear slightly darker on a T1 weighted image (Becker et al 2008a).
For oesophageal and rectal tumours, T2 weighted images are vry useful to distinguish the
tumour from normal rectal, oesophageal tissue surrounding fat (Beets-Tan and Beets 2012).
Tumours can be surrounded by oedema due to peritumoral inflammation or due to restricted
venous and lymph drainage (Becker et al 2008b). This effect is apparent in head and neck
tumours, where the area with high signal on T2 is, for the most part, largely overestimating the gross tumour volume (Caldas-Magalhaes et al 2012). Another endogenous contrast
is T2*, which reflects the induced (microscopic) field inhomogeneities due to e.g. deoxyhemoglobin. BOLD imaging is based on this mechanism where deoxyhemoglobin causes a
decrease in T2*. Since the amount of deoxyhemoglobin is dependent on the amount of blood
and the oxygenation of blood, BOLD MRI might indicate the oxygenation status of a tumour
(Hoskin et al 2007). Improved estimation of the oxygenation status can be achieved using
D(C)O2E MRI which requires respiratory challenges of CO2 and O2 (Rijpkema et al 2002).
An important physiological characteristic of most tumours is the impaired vasculature due to
imbalanced angiogenesis. To measure altered micro-vascular parameters, an exogenous contrast agent is injected and its tissue uptake monitored by (repeated) imaging, which results
in information on blood volume and flow. Commonly, gadolinium chelates contrast agents
are used. These contrast agents decrease the T1 relaxation of water protons and thereby
affect T1 based contrast in blood and in tissue due to contrast agent leakage (Padhani and
Dzik-Jurasz 2004). An alternative technique used to assess blood volume and flow is Arterial
Spin Labelling (ASL), which is conceptually based on the flow of magnetically labelled
blood. This has a subtle effect on the signal intensity of the tissue where the blood arrives.
The main advantage of this technique is that it does not require any contrast agent injection
(Petersen et al 2006). However, its sensitivity compared to contrast enhanced techniques is
considerably lower.
R352
Topical Review
Phys. Med. Biol. 59 (2014) R349
Figure 3. An example of a prostate cancer patient. (a) T2w image, (b) ADC map and
(c) Ktrans map of a prostate cancer patient (d) GTV areas by threshold combinations of
the ADC and Ktrans maps. The overlap volume increases for increasing ADC thresholds and decreasing Ktrans thresholds. However, a complete overlap between thresholded ADC and thresholded Ktrans regions was not found for any of the threshold
combinations (reproduced with permission from Groenendaal et al 2010b).
Despite the improved contrast, accurate tumour definition is still difficult and generally
tumour volume is largely overestimated due to tumour induced tissue alterations around the
gross tumour volume (Daisne et al 2004). Diffusion tensor imaging has been proposed as a
novel tool for improved distinction of the CTV (Price et al 2006).
Poor tumour delineation is a major source of inaccuracies in radiotherapy treatment. To
improve this, (semi)automatic segmentations based on multi-parameter quantitative MRI
might be helpful. Dynamic enhanced MRI (DCE-MRI) and diffusion weighted MRI (DWI)
are two widely used quantitative MRI methods (van der Heide et al 2012). DCE MRI provides information on micro-vascular properties like vessel permeability, blood flow, and
extra-cellular space which are altered in tumour tissue due to presence of angiogenesis (Cao
et al 2011). The application of pharmo-kinetic analysis on signal enhancement curves on
a voxel basis, allows the generation of quantitative parameter maps e.g. so-called Ktrans
maps are frequently used for GTV delineation in prostate cancer reflecting a mix of blood
flow and vessel permeability (Groenendaal et al 2010b). The apparent diffusion coefficient
(ADC) derived from DWI reflects the microscopic water mobility, which is related to the
structural microenvironment (Padhani et al 2009). For automatic assignment of tumour voxels, a high specificity per voxel is needed. Different parameters generally do not indicate
that the same voxels are tumours (Groenendaal et al 2010b). Therefore, probability mapping
using combined parameter classifiers is applied, as is shown in prostate cancer (figure 3)
(Groenendaal et al 2012). A drawback of multiparameter mapping is the time consuming
imaging acquisition. Recently, Ma et al introduced MRI fingerprinting, which is a conceptually new, quantitative image acquisition and processing method (Ma et al 2013) which
promises fast, repeated, quantitative MRI.
Recently, the need for better lymph node imaging has been emphasized. As local tumour
treatment improves, recurrences in lymph nodes become apparent (Nomden et al 2013).
Conventionally lymph node imaging is based on T1 contrast. The identification of metastasized lymph nodes is a challenging task. Generally they are diagnosed based on size criteria,
but both false positive and false negative lymph nodes are frequently found. In head and neck,
DWI showed a high sensitivity and specificity for sub-centimetre positive lymph nodes to
help in this distinction (Vandecaveye et al 2009). However, in the pelvic area, compared to
conventional MRI, DWI did not add any additional diagnostic value (Klerkx et al 2012). Most
promising results were achieved using ultra-small super-paramagnetic iron oxide (USPIO)
R353
Topical Review
Phys. Med. Biol. 59 (2014) R349
Figure 4. Example dedicated diffusion weighted T2-FFE scan with fat, blood and mus-
cle suppression for MRI lymph node detection in the neck area (reproduced with permission from Philippens et al 2012).
in breast and prostate cancer (Harisinghani et al 2006, Barentsz et al 2007). USPIO particles
collect in negative lymph nodes, because they are incorporated by macrophages. Due to the
iron loading, the MR signal of normal lymph nodes will disappear about 24 h after contrast
administration, while a minimal MR signal change will occur in metastatic nodes due to much
lower iron uptake. Despite the promising results and the need for accurate lymph node diagnosis, no approval for use of USPIO from the Food and Drug Agency has been released yet.
Apart from the diagnosis of positive lymph nodes, elective radiotherapy of lymph nodes in
high risk areas is common practice. CT is the conventional modality for the definition of these
regions, but the smaller lymph nodes are not visible on CT. Therefore, anatomical boundaries
are defined where most of the lymph nodes will be found. MRI might visualize lymph nodes
with a higher sensitivity due to the improved contrast. To distinguish lymph nodes from the
surrounding tissue suppression of fat, blood and muscle tissue is required. This is achieved
by designing a pulse sequence with intrinsic diffusion weighting in a T2 weighted water only
scan, as is used in neurography (Zhang et al 2008, Philippens et al 2012) (figure 4). Moreover,
USPIO enhanced MRI has shown that a substantial number of positive lymph nodes in prostate cancer were not located in the routine surgical areas or in the axillary region defined by
delineation guidelines (Heesakkers et al 2009, MacDonald et al 2010).
The above mentioned MR imaging procedures can also be used for treatment response
assessment. As an example, the response of chemoradiotherapy for rectal cancer must be
quantified to decide in which patients surgery may be omitted (Solanki et al 2013). The difference in ADC seems to be a powerful tool for predicting pathologic complete response early
in the treatment course (Lambrecht et al 2012, Bäuerle et al 2013, Intven et al 2013). Only
FDG-PET seems insufficient for prediction response in clinical practice, e.g. in the treatment
of patients with rectum cancer (Guillem et al 2013). Not all tumours show a response during
treatment, which can lead to a decision to differentiate or change dose. In prostate cancer,
changes in ADC are not observed during treatment (Foltz et al 2013). However, diffusion
weighted imaging can be used to define the exact location of recurrences of prostate tumours
(van Vulpen et al 2009), providing essential feedback on treatment choices. In prostate cancer
a higher dose is required in areas with more aggressive tumour cells and with more tumour
cells (van Vulpen et al 2009).
R354
Topical Review
Phys. Med. Biol. 59 (2014) R349
4. Treatment simulation
The definition of irradiation fields requires a geometrically accurate dataset, acquired with the
patient in a fixed radiotherapy position. Both conditions need attention since MRI is prone
to geometry and intensity distortions (Moerland et al 1995) and current MR scanners with
hollow couches and different types of RF coils attached to the patient’s body, may conflict
with the radiotherapy position (Devic 2012). Instead of using MRI alone for radiotherapy
planning, the use of registered MRI and CT datasets is a possible approach. The advantages of
this approach is that CT provides geometric accuracy with a large field of view and delivery
of tissue electron densities for dose calculation, while the MRI may zoom in on the tumour
volume and provides detailed delineation and characterization information.
Whatever approach is chosen, insight into the origin and impact of MR image distortions
is needed before introducing MRI in radiotherapy treatment planning.
MR images can be acquired with a variety of pulse sequences. The occurrence and
severity of MR image distortions depends on the type of pulse sequence and on the object.
Investigations on MR image distortions date back from the 80s when Lerski (1988) observed
machine dependent geometric distortions up to 13 mm in a European multi-centre investigation. Geometric distortion in MR imaging predominantly arises from the inhomogeneity of
the static magnetic field and the nonlinearity of the gradients. Bakker et al (1992) investigated
system related imperfections by systematic variation of the strength, direction, and polarity
of the read-out gradient in imaging experiments on a grid of cylindrical sample tubes. For the
1.5 T Philips Gyroscan system used in this study, static field related errors up to 7 mm and
gradient related errors up to 4 mm were observed (midcoronal plane, FOV 400 mm, read out
gradient between 0.5 and 3.0 mT m−1). It was also shown that in spin echo and gradient echo
imaging, both gradient errors and static field inhomogeneity can lead to image distortions
in the direction of the frequency encoding gradient, whereas in the phase encoding gradient image distortions are solely caused by gradient errors. System related distortions caused
by imperfections in the static magnetic field and gradients can be readily corrected for and
nowadays this is available on most modern MR scanners as a post-processing tool (Doran et al
2005). Gradient and static field distortions grow toward the periphery of the imaging field. Old
MRI radiotherapy simulation systems, like the Panorama 0.23 T (Philips), have such a small
field of view that peripheral distortions are uncontrolled (Mah et al 2002). In most clinical
situations the tumour will be positioned in close proximity to the isocentrec and system related
distortions will be minimised, in modern MRI systems, to the sub-millimetre range.
B0 magnetic field errors mainly stem from imperfections of the MR scanner, which can
be corrected as indicated above, but B0 magnetic field errors also originate from the patient.
Since the human body susceptibility distribution is a major factor perturbing the magnetic
field, those geometrical distortions become patient anatomy dependent. Due to the relatively
large susceptibility difference between tissue and air, the image distortions are expected to be
largest at tissue–air interfaces. Magnetic field distributions and image distortions for simple
objects, such as cylinders and spheres, have been solved analytically (Chu et al 1990). For the
magnetic field in and around the patient, numerical methods have to be applied. Bhagwandien
et al (1994) developed a method to calculate the magnetic field distribution in and around a
3D object using a technique based on the finite difference method and applied it to the human
body. Indeed, field perturbations were largest at the tissue–air interfaces, between −5–6 ppm
outside the head and between −6 and 5 ppm inside the head near the sinuses. Resulting image
distortions depend on the strength of the read out gradient, e.g. between −5 and 5 mm for a
typical read out gradient strength (1.5 mT m−1 at 1.5 T). Those susceptibility related errors are
inversely proportional to gradient strength, whereas gradient related errors turned out to be
R355
Topical Review
Phys. Med. Biol. 59 (2014) R349
independent of gradient strength. It is therefore recommended to apply the strongest available
read-out gradients when geometric fidelity is given preference to signal-to-noise considerations, making 1.5 T systems easier to handle compared to 3.0 T systems. Undistorted images
may be acquired with pulse sequences that only apply phase encoding of the MR signal using
so-called single point imaging (SPI) (Gruwel et al 2007, Crijns et al 2012a). However, phase
encoding is time consuming and usually not applied for clinical MR imaging. Compressed
sensing can accelerate SPI and make the method suitable for QA purposes (Crijns et al
2012a). Another approach is to measure the magnetic field error and use it for correcting the
MR image (Baldwin et al 2009, Crijns et al 2011a). For all normal imaging applications,
susceptibility related errors can be minimised to sub-mm level (Stanescu et al 2012). There is
an exception for special DWI sequences used for functional imaging (Embleton et al 2010).
DWI is typically based on EPI sequences, which are highly sensitive for susceptibility errors
due to the low pixel-bandwidth in the phase encode direction, with errors running up to the
cm scale. Careful mapping of the magnetic field must be applied to correct these distortions.
For the moment the University Medical Centre Utrecht has chosen to use CT as a geometry
master for Radiotherapy Treatment Planning (RTP) and to register the high resolution and
often zoomed MR images, used for delineation, with this CT. Scan protocols generally contain T2 weighted 2D multiple slice turbo spin echo (TSE) sequences and high resolution 3D
steady state free precession (SSFP) sequences with relatively high bandwidth and sequences
for functional imaging (DCE-MRI and DWI). Tumour delineation is saved on the high resolution SSFP image set, which is registered with the CT data used for setting up the irradiation
beams. Image co-registration is possible in many ways and is becoming a separate profession
(Devic 2012). It is essential that the delineation from the MRI is transferred from the MRI to
the CT. Mis-registration may introduce a systemic error. To minimise the introduction of new
errors it is essential that the structures and/or fiducial markers used in the position verification process are registered correctly. Focus on the registration of tumour contours may be
considered to be an erroneous approach. Using only the MRI for delineation, without the CT
information, may prevent this error. The fundamental problem of registering different images
will only be solved if the fiducials used for position verification are clearly visible on the MRI
(Kapanen et al 2012) or if direct MRI is available at the treatment itself. Those systems will
be discussed in the chapter on MRI guidance for external beam radiotherapy.
Like 4D CT, MRI can supply information on organ movements (McBain et al 2009, Chan
et al 2008). Cine MRI can easily supply breathing related information at a sub-second scale
without any radiation dose involved. These conventional techniques use binning techniques
and thus provide 4D information by post-processing. Modern techniques try to acquire MRI
data real time in 2D (Bauer et al 2013) and recently even in 3D (Kim et al 2013, Niebergall
et al 2013) and thus provide the data needed for margin assessment (Paulson et al 2011, see
example in figure 5) and treatment guidance (Stemkens et al 2013).
A new development is the use of MRI-only for generating the Hounsfield data required
for treatment planning. This area of generating synthetic CT data from MRI is stimulated by
the introduction of integrated MRI/PET systems. Hounsfield data can be obtained via image
segmentation from the MRI or via direct visualization of the bony structures using UTE
sequences (Hsu et al 2013, Korhonen et al 2014), figure 6).
5. MRI guidance for brachytherapy
The value of MRI guidance for brachytherapy has been proven for patients with cervix cancer (Jürgenliemk-Schulz et al 2009, Pötter et al 2011, Nomden et al 2013). Guidelines were
R356
Topical Review
Phys. Med. Biol. 59 (2014) R349
Figure 5. An example of a 2D FLASH cine MRI (3 Hz) to measure GTV displacement
(red contours) and corresponding percent change in GTV area (b,d) during one resting
(a,b) and one swallowing event (c,d) for a H/N patient with a supraglottic tumor
(reproduced with permission from Paulson et al 2011).
Figure 6. An example of a pseudo-CT image (b) generated from T1/T2* MR images
by using a HU conversion technique with the original MR image (a) from which the
pseudo-CT image was constructed. (c) shows the corresponding transverse standard CT
slice (reproduced with permission from Korhonen et al 2014).
developed for the imaging and delineation of cervix cancer (GTV, high risk CTV, intermediate
risk CTV) and for dose prescription (Haie-Meder et al 2005). It was observed that standard
tandem/ovoid and tandem/ring applicator dose distributions from the radiograph/CT era did
not cover the target in many instances and special applicators were developed to increase
target coverage and to increase local control of the tumour (Jürgenliemk-Schulz et al 2009).
R357
Topical Review
Phys. Med. Biol. 59 (2014) R349
At the University Medical Centre Utrecht delineation of tumour volumes and OAR’s as well
as the reconstruction of applicators is performed on axial, coronal and sagittal T2 weighted
images. The accuracy of applicator reconstruction is in the order of pixel size, usually
1–2 mm, for plastic and titanium applicators at 1.5 T (De Leeuw et al 2009, Haack et al 2009).
Titanium applicators and needles may cause susceptibility and RF artefacts, especially in 3 T
MR images (Kim et al 2011). At higher field strengths, a potential hazard for interventional
applications is the heating of devices, due to resonating electromagnetic RF waves along the
conducting structure (Yeung et al 2007, van den Bosch et al 2010a).
Other applications of MRI guided brachytherapy under development are focal HDR
brachytherapy for patients with localized prostate cancer or recurrent localized prostate
cancer, where tumour localization is based on multi-parametric imaging using T2w, DCEMRI, DWI and biopsy findings (Groenendaal et al 2010a, Moman et al 2010). In current
MRI guided prostate implantation procedures the patient is moved in and out of the MR
scanner between needle insertion, imaging and irradiation due to restricted access to the
MR scanner and lack of MRI compatible brachytherapy afterloaders (Susil et al 2004).
Several groups are developing MR compatible robotic devices for prostate implant procedures inside a cylindrical bore MR scanner and to reduce prostate deformations, due to the
intervention the needle is inserted by tapping (Lagerburg et al 2006, van den Bosch et al
2010b) or shooting (Muntener et al 2008). Prostate brachytherapy is developing towards
single-fraction HDR monotherapy and if early data are confirmed with longer follow-up, it
may become the treatment of choice for many men with localised prostate cancer (Morton
and Hoskin 2013). To deliver single fraction HDR monotherapy in a safe manner, the use of
an MRI compatible afterloader and the monitoring of the source position in relation to the
target and organs at risk may be of great value. De Leeuw et al (2013) have shown that with
special MR sequences and reconstruction algorithms HDR source position verification on
MRI is feasible in real time.
6. MRI guidance for external beam radiotherapy
The excellent soft-tissue contrast of MRI, combined with its real-time/cine capacities, while
using harmless radiofrequency instead of ionizing radiation, theoretically allows direct target
visualization for treatment guidance. It allows patient set-up based on the actual position,
shape and motion characteristics, but also actual motion tracking during irradiation. Motion
tracking can be used for real-time treatment guidance and adaptation, but also for recording
anatomical changes for dose reconstruction and dose accumulation, thus providing unique
positioning and dose certainty in the treatment process. To achieve this hybrid MRI, a radiotherapy technology is required, developed and under varying stages of construction.
A first, moderate step, which does not require integration of the MRI and radiotherapy
system, is presented by Karlsson et al (2009). They presented a next door MRI configuration
where the patient is scanned prior to irradiation and then brought on the same table top and a
dedicated trolley to the radiotherapy treatment room. No technical modifications are required
to either system to facilitate the use of MRI for the determination of the actual anatomy,
however, clearly no intra-fraction motion can be detected, and the patient transport introduces
potential shifts in anatomy. The IMRIS system (Varian press release Oct 5 2010) offers a similar solution; it consists of a radiotherapy accelerator and an MRI that can be brought into the
treatment room for scanning prior to treatment. This reduces the patient transport, but comes
at the price of moving the MRI system on rails in and out of the room. A prototype system is
currently under construction in Princess Margaret Hospital, Canada.
R358
Topical Review
Phys. Med. Biol. 59 (2014) R349
For an MRI during treatment itself, a more integrated solution is required. UMC Utrecht,
the Netherlands, have together with Elekta AB, Stockholm, Sweden and Philips, Best, The
Netherlands, pioneered such integration (Lagendijk et al 2002, Lagendijk et al 2008) with a
1.5 T MRI and a 6 MV accelerator. In the Cross Cancer Institute a prototype integrated 0.2 T
MRI and a 6 MV radiotherapy accelerator has been developed and built (Fallone et al 2009).
A commercial alternative is the combined 0.35 T MRI and 60Cobalt radiotherapy system
being developed by Viewray, OH, U.S.A.
6.1. UMC Utrecht MRI accelerator
The present prototype system comprises of a modified 1.5 T Achieva Philips MRI and an Elekta
6 MV accelerator in a ring in the mid-transversal plane around the MRI. The three main technical hurdles to allow both systems to operate unhampered are the magnetic interference, the
radio-frequency interference and the beam transmission through the MRI. The active magnetic
shielding of the MRI is modified to create a low field toroid in the transversal mid-plane to magnetically decouple the MRI and the accelerator (Overweg et al 2009), this way the accelerator still
operates, although it is in the vicinity of the MRI and the MRI is not perturbed by the presence of
ferrous accelerator components. The radio-frequency (RF) interference between the accelerator
components and the MRI acquisition system cope because of the redesign of the Faraday cage.
Conventionally the MRI is placed inside a Faraday cage to prevent interference with the outer
world. For the hybrid MRI accelerator, the aluminium cryostat wall is part of the Faraday cage
and therefore, the accelerator is positioned outside the cage. Another modification is for the beam
transmission, to minimize beam absorption and heterogeneity a homogeneous, circumferential
beam portal in the mid plane is made to allow beam passage (Overweg et al 2009).
The proof of concept of the system for a static configuration was given in 2009 (Raaymakers
et al 2009). Simultaneous irradiation and MRI was shown to be feasible. As designed, both
systems operate independent of each other. Since then the system is being upgraded with a
ring mounted accelerator which allows continuous rotation in either direction (figure 7). At the
same time the system is equipped with a MLC.
This technical prototype has been used to demonstrate MRI based gating and tracking on
moving phantoms (Crijns et al 2012b). 1D MRI navigators where used to follow the moving
phantom and with a latency of less than 200 ms the radiation could be gated or the beam could
be moved along with the phantom. Additionally, the MRI could be used to reconstruct the dose
on the moving phantom (Crijns et al 2012b), which is crucial for establishing precise dose
response relationships.
6.2. Renaissance system of Viewray
Viewray, OH, USA is a start-up company from the University of Florida and has developed
and built a hybrid 0.35 T MRI and Cobalt-60 radiotherapy system, the Renaissance, see1
­(figure 8). Three Cobalt-60 units equipped with MLC are positioned in a ring in the midtransversal plane of the double donut, so open, MRI system and can rotate 120° to allow
radiation delivery from each gantry position. Obviously, the radio-active sources are not
affected by the magnetic stray field and the open MRI design facilitates beam transmission,
also RF interference is limited due to the use of radio-active sources. The potential impact
of the ring mounted radiation units on the MRI performance is not reported on yet. The first
non-clinical systems are installed and imaging and dosimetric studies are underway.
1
www.viewray.com
R359
Topical Review
Phys. Med. Biol. 59 (2014) R349
Figure 7. Photograph of the MRI linac experimental prototype system UMC Utrecht.
Figure 8. Schematic drawing ViewRay renaissance system (courtesy ViewRay).
6.3. The Cross Cancer Institute system
The system at the Cross Cancer Institute is a bi-planar, 0.2 T MRI system with a 27 cm gap
integrated with a 6 MV accelerator (Fallone et al 2009). The accelerator is positioned laterally to the MRI, so the beam enters through the open side of the magnet, facilitating beam
transmission. The magnetic interference in this system is solved by passively shielding the
accelerator. This passive shielding on its turn perturbs the main magnetic field; therefore the
main magnetic field is shimmed for the presence of the passive accelerator shielding. As a
consequence, the geometry of accelerator and MRI must be fixed since shimming is done
R360
Topical Review
Phys. Med. Biol. 59 (2014) R349
Figure 9. Phase II linac-MR system (6 MV linac, 0.6 T MR system, and 60 cm bore)
at the Cross Cancer Institute, University of Alberta, Edmonton2. The magnet is a high
temperature superconductor. Currently building Phase III with 85 cm bore.
Figure 10. Schematic drawing of the MRI linac design by Paul Keall for investigating
the magnetic field orientation according to the beam (reproduced with permission from
Constantin et al 2011).
specifically for one position of the accelerator. To allow irradiation from multiple beam angles
the accelerator has to rotate together with the MRI around the target in order to maintain the
same geometrical relation between MRI and accelerator.
Fallone et al demonstrated simultaneous MRI and irradiation using this system in 2008
(Fallone et al 2009). A human sized prototype is under construction comprising a 0.6 T MRI
and a 6 MV accelerator. In addition, the orientation of the accelerator and the MRI can be
changed in that system. The accelerator can be positioned either laterally to the MRI, so the
beam enters through the gap, or on top of the MRI, where a dedicated hole in the MRI allows
the beam to enter from the top (Keyvanloo et al 2012) (figure 9). The reason for doing so is
that in this orientation the central radiation beam line is parallel with the magnetic field, reducing the impact of the magnetic field on the dose distribution, this will be discussed in more
detail in section ‘Impact of the magnetic field on the dose distribution’. To investigate the
orientation of the beam according to the magnetic field Keall and co-workers are developing a
comparable system (Constantin et al 2011, Oborn et al 2012) (figure 10).
2
www.mp.med.ualberta.ca/linac-mr/photo_gallery.html
R361
Topical Review
Phys. Med. Biol. 59 (2014) R349
6.4. Consequences of magnetic field strengths
The B0 magnetic field strength for the various designs ranges from 0.2 T–1.5 T. A higher
field strength yields a higher signal to noise ratio (SNR) and in MRI SNR it can be traded off
against temporal or spatial resolution and improve contrast. So higher field strengths yield
more versatile MRI and with that improved image guidance scenarios.
However, an increased B0 strength also yields increased susceptibility artefacts, which
much be countered with increased gradient strengths. These artefacts need to be resolved
before MRI can be used to guide radiation delivery as discussed above in section ‘MRI
simulation’. So specifically for hybrid MRI radiotherapy systems dedicated MRI correction
schemes are investigated to obtain geometrically correct MRI during treatment (Baldwin et al
2009, Crijns et al 2011b, 2012b).
More related to the hybrid MRI radiotherapy systems, the main magnetic field strength is
affecting the radiation dose distribution as well as radiation detectors.
6.5. Impact of the magnetic field on the dose distribution
All systems treat with a photon beam, and a photon beam is unaffected by a magnetic field.
However, secondary electrons are affected. The magnitude of this effect is determined by the
magnetic field strength and the orientation of the incident beam relative to the magnetic field.
For a transverse magnetic field relative to the beam a decreased built-up distance and
­a-symmetrical penumbra is seen, while the magnitude of this effect depends on the magnetic
field strength (Raaymakers et al 2004, Raaijmakers et al 2008). Most striking is the impact at a
tissue–air interface; without a magnetic field electrons would scatter away from such interface
along with the incident radiation beam, with a magnetic field, these electrons will return, due
to the Lorentz force, to the existing surface yielding a local dose increase (Raaijmakers et al
2005). This so called Electron Return Effect (ERE) manifests itself in a different way for different radiation field sizes, magnetic field strengths and surface orientations (Raaijmakers et al
2008). Opposing beams do, in the first order, cancel the ERE effect as shown by Raaijmakers
et al (2005) and when using multi-beam (so approximately multiple opposing beams) IMRT
the effect can be entirely compensated (Raaijmakers et al 2007). To do so, the magnetic field
has to be accounted for in the dose calculations, which at this stage means a Monte Carlo
approach is required.
The impact of the magnetic field is also dependent on the orientation relative to the incident
radiation beam. For a parallel configuration with the beam along the magnetic field lines, the
impact for tissue–air interfaces is smaller (St Aubin et al 2010). However, in such a configuration, the contamination electrons from the radiation source and air column travelled, are
captured in the photon beam by the parallel magnetic field and potentially increase the dose at
the entrance side of the beam (Oborn et al 2012). For realistic MRI stray fields this effect may
be limited (Keyvanloo et al 2012) but have to be investigated further.
6.6. Radiation dosimetry in magnetic fields
Beside an altered patient dose distribution in the presence of a magnetic field, also radiation detectors can be affected. This is clear particularly for ionization chambers, which are
basically an air cavity in a water-like environment and thus the response will be affected.
Similar to the ERE discussed above, the response is likely to be dependent on the orientation
of the chamber relative to the magnetic field. This was demonstrated experimentally and by
R362
Topical Review
Phys. Med. Biol. 59 (2014) R349
Monte Carlo simulations for the NE2571 Farmer-type chamber (Meijsing et al 2009). The
response could vary up to 11% depending on the orientation for a 1.5 T magnetic field.
6.7. Impact of MRI radiofrequency coils
In MRI the signal is typically picked up by a RF antenna positioned on the patient. The optimal position for this antenna is inevitably in the radiation field. Fortunately, the receiving
antenna can be made to be thin enough to yield a negligible attenuation to the beam. This
was demonstrated already for a hybrid 0.5 T MRI kV fluoroscopy system (Fahrig et al 2001),
where the MRI coil was optimised for beam transmission and made transparent for kV beams.
For MV radiation therapy beams, even non-optimised RF coils can be suitable as shown by
Hoogcarspel et al (2013).
Another impact of the RF coil is the potential image degradation due to the irradiation of
the antenna loop. Because the copper loop has a different density to the surrounding tissue,
an electron dis-equilibrium can occur in the RF antenna, with a Radiation Induced Current as
a consequence (Burke et al 2012). This effect has been demonstrated experimentally for low
magnetic field strength and thus a relatively low RF frequency of 9 MHz (Burke et al 2012).
At 1.5 T (64 MHz), the impact of irradiating the RF coil did not result in the degradation of
SNR (Hoogcarspel et al 2013).
6.8. MRI guided proton therapy
The promise of real-time anatomical feedback is appealing for photon radiotherapy, but even
more so for proton therapy. There, the strong dose gradient from the Bragg peak demands high
precision image guidance to fully exploit the proton beam advantages. At present the range uncertainty is countered by additional margins and limited IMRT modulation (McGowan et al 2013).
So far, investigations into such hybrid system are limited to the impact on dosimetry. As the proton beam itself consists of charged particles, the beam trajectory will be affected by the magnetic
field. This deflection can be analytically calculated (Wolf et al 2012). The impact of the magnetic
field on the secondary electrons is negligible (Raaymakers et al 2008). This is due to the low
energy of the secondary electrons in proton therapy relative to those with photon therapy.
7. Conclusion
It may be clear from this review that the impact of modern MRI on the radiotherapy process
is growing fast. MRI based treatment simulation combined with tumour characterization is
becoming part of the daily routine in clinics. The impact of MRI on brachytherapy cannot be
underestimated; MRI is making brachytherapy a quantitative therapy and reduces the operator dependency. Seeing what you treat, bringing certainty and quantification in the treatment
process, is essential for brachytherapy, but will also have a huge impact on external beam
radiotherapy, making radiotherapy more of an interventional radiology process.
References
Bakker C J, Moerland M A, Bhagwandien R and Beersma R 1992 Analysis of machine-dependent and
object-induced geometric distortion in 2DFT MR imaging Magn. Reson. Imag. 10 597–608
Baldwin L N, Wachowicz K and Fallone B G 2009 A two-step scheme for distortion rectification of
magnetic resonance images Med. Phys. 36 3917–26
R363
Topical Review
Phys. Med. Biol. 59 (2014) R349
Bammer R 2003 Basic principles of diffusion-weighted imaging Eur. J. Radiol. 45 169–84
Barentsz J O, Fütterer J J and Takahashi S 2007 Use of ultrasmall superparamagnetic iron oxide in lymph
node MR imaging in prostate cancer patients Eur. J. Radiol. 63 369–72
Bauer R W, Radtke I, Block K T, Larson M C, Kerl J M, Hammerstingl R, Graf T G, Vogl T J and
Zhang S 2013 True real-time cardiac MRI in free breathing without ECG synchronization using a
novel sequence with radial k-space sampling and balanced SSFP contrast mode Int. J. Cardiovasc.
Imag. 29 1059–67
Bäuerle T et al 2013 Diffusion-weighted imaging in rectal carcinoma patients without and after
chemoradiotherapy: a comparative study with histology Eur. J. Radiol. 82 444–52
Becker M, Burkhardt K, Dulguerov P and Allal A 2008a Imaging of the larynx and hypopharynx Eur.
J. Radiol. 66 460–79
Becker M, Zbären P, Casselman J W, Kohler R, Dulguerov P and Becker C D 2008b Neoplastic invasion
of laryngeal cartilage: reassessment of criteria for diagnosis at MR imaging Radiology 249 551–9
Beets-Tan R G and Beets G L 2011 Local staging of rectal cancer: a review of imaging J. Magn. Reson.
Imag. 33 1012–9
Bhagwandien R, Moerland M A, Bakker C J G, Beersma R and Lagendijk J J W 1994 Numerical analysis
of the magnetic field for arbitrary magnetic susceptibility distributions in 3D Magn. Reson. Imag.
12 101–7
Bottomley P A, Hardy C J, Argersinger R E and Allen-Moore G 1987 A review of 1H nuclear magnetic
resonance relaxation in pathology: are T1 and T2 diagnostic? Med. Phys. 14 1–37
Breuer F A, Blaimer M, Heidemann R M, Mueller M F, Griswold M A and Jakob P M 2005 Controlled
aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging
Magn. Reson. Med. 53 684–91
Burke B, Wachowicz K, Fallone B G and Rathee S 2012 Effect of radiation induced current on the
quality of MR images in an integrated linac-MR system Med. Phys. 39 6139–47
Caldas-Magalhaes J, Kasperts N, Kooij N, van den Berg C A, Terhaard C H, Raaijmakers C P and
Philippens M E 2012 Validation of imaging with pathology in laryngeal cancer: accuracy of the
registration methodology Int. J. Radiat. Oncol. Biol. Phys. 82 e289–98
Cao Y 2011 The promise of dynamic contrast-enhanced imaging in radiation therapy Semin. Radiat.
Oncol. 21 147–56
Chan P, Dinniwell R, Haider M A, Cho Y B, Jaffray D, Lockwood G, Levin W, Manchul L, Fyles A and
Milosevic M 2008 Inter—and intrafractional tumor and organ movement in patients with cervical
cancer undergoing radiotherapy: a cinematic-MRI point-of-interest study Int. J. Radiat. Oncol.
Biol. Phys. 70 1507–15
Chu S C, Xu Y, Balschi J A and Springer C S Jr 1990 Bulk magnetic susceptibility shifts in NMR studies
of compartmentalized samples: use of paramagnetic reagents Magn. Reson. Med. 13 239–62
Constantin D E, Fahrig R and Keall P J 2011 A study of the effect of in-line and perpendicular
magnetic fields on beam characteristics of electron guns in medical linear accelerators Med. Phys.
38 4174–85
Crijns S P M, Kok J G, Lagendijk J J W and Raaymakers B W 2011a Towards MRI-guided linear
accelerator control: gating on an MRI accelerator Phys. Med. Biol. 56 4815–25
Crijns S P, Raaymakers B W and Lagendijk J J 2011b Real-time correction of magnetic field
inhomogeneity-induced image distortions for MRI-guided conventional and proton radiotherapy
Phys. Med. Biol. 56 289–97
Crijns S P M, Bakker C J G, Seevinck P R, de Leeuw H, Lagendijk J J W and Raaymakers B W
2012a Towards inherently distortion-free MR images for image-guided radiotherapy on an MRI
accelerator Phys. Med. Biol. 57 1349–58
Crijns S P M, Raaymakers B W and Lagendijk J J W 2012b Proof of concept of MRI-guided tracked
radiation delivery: tracking 1D motion Phys. Med. Biol. 57 7863–72
Daisne J F, Duprez T, Weynand B, Lonneux M, Hamoir M, Reychler H and Grégoire V 2004 Tumor
volume in pharyngolaryngeal squamous cell carcinoma: comparison at CT, MR imaging, and FDG
PET and validation with surgical specimen Radiology 233 93–100
De Leeuw A A, Moerland M A, Nomden C, Tersteeg R H, Roesink J M and Jürgenliemk-Schulz I M
2009 Applicator reconstruction and applicator shifts in 3D MR-based PDR brachytherapy of
cervical cancer Radiother. Oncol. 93 341–6
de Leeuw H, Moerland M A, van Vulpen M, Seevinck P R and Bakker C J 2013 A dual-plane co-RASOR
technique for accurate and rapid tracking and position verification of an Ir-192 source for single
fraction HDR brachytherapy Phys. Med. Biol. 58 7829–39
R364
Topical Review
Phys. Med. Biol. 59 (2014) R349
Devic S 2012 MRI simulation for radiotherapy treatment planning Med. Phys. 39 6701–11
Donahue M J, Hoogduin H, van Zijl P C, Jezzard P, Luijten P R and Hendrikse J 2011 Blood oxygenation
level-dependent (BOLD) total and extravascular signal changes and ΔR2* in human visual cortex at
1.5, 3.0 and 7.0 T NMR Biomed. 24 25–34
Doran S J, Charles-Edwards L, Reinsberg S A and Leach M O 2005 A complete distortion correction for
MR images: I. Gradient warp correction Phys. Med. Biol. 50 1343–61
Ehman R L and Felmlee J P 1989 Adaptive technique for high definition MR imaging of moving
structures Radiology 173 255–63
Ehman R L, McNamara M T, Pallack M, Hricak H and Higgins C B 1984 Magnetic resonance imaging
with respiratory gating—techniques and advantages Am. J. Radiol. 143 1175–82
Embleton K V, Haroon H A, Morris D M, Ralph M A and Parker G J 2010 Distortion correction for
diffusion-weighted MRI tractography and fMRI in the temporal lobes Hum. Brain Mapp.
31 1570–87
Fahrig R, Butts K, Rowlands J A, Saunders R, Stanton J, Stevens G M, Daniel B L, Wen Z, Ergun D L
and Pelc N J 2001 A truly hybrid interventional MR/x-ray system: feasibility demonstration
J. Magn. Reson. Imag. 13 294–300
Fallone B G, Murray B, Rathee S, Stanescu T, Steciw S, Vidakovic S, Blosser E and Tymofichuk D
2009 First MR images obtained during megavoltage photon irradiation from a prototype integrated
linac-MR system Med. Phys. 36 2084–8
Foltz W D et al 2013 Changes in apparent diffusion coefficient and T2 relaxation during radiotherapy for
prostate cancer J. Magn. Reson. Imag. 37 909–16
Groenendaal G, Borren A, Moman M R, Monninkhof E, van Diest P J, Philippens M E, van Vulpen M
and van der Heide U A 2012 Pathologic validation of a model based on diffusion-weighted imaging
and dynamic contrast-enhanced magnetic resonance imaging for tumor delineation in the prostate
peripheral zone Int. J. Radiat. Oncol. Biol. Phys. 82 e537–44
Groenendaal G, Moman M R, Korporaal J G, van Diest P J, van Vulpen M, Philippens M E P and
van der Heide U A 2010a Validation of functional imaging with pathology for tumor delineation in
the prostate Radiother. Oncol. 94 145–50
Groenendaal G, van den Berg C A, Korporaal J G, Philippens M E, Luijten P R, van Vulpen M and van
der Heide U A 2010b Simultaneous MRI diffusion and perfusion imaging for tumor delineation in
prostate cancer patients Radiother. Oncol. 95 185–90
Gruwel M L H, Latta P, Tanasiewicz M, Volotovskyy V, Sramek M and Tomanek B 2007 MR imaging
of teeth using a silent single point imaging technique Appl. Phys. A 88 763–7
Guillem J G et al 2013 Neither FDG-PET nor CT can distinguish between a pathological complete
response and an incomplete response after neoadjuvant chemoradiation in locally advanced rectal
cancer: a prospective study Ann. Surg. 258 289–95 763–7
Haack S, Nielsen S K, Lindegaard J C, Gelineck J and Tanderup K 2009 Applicator reconstruction
in MRI 3D image-based dose planning of brachytherapy for cervical cancer Radiother. Oncol.
91 187–93
Haacke E M, Brown R W, Thompson M R and Venkatesan R 1999 Magnetic Resonance Imaging:
Physical Principles and Sequence Design (New York: Wiley-Liss)
Haie-Meder C et al 2005 Recommendations from Gynaecological (GYN) GEC ESTRO Working Group
(I): concepts and terms in 3D image-based 3D treatment planning in cervix cancer brachytherapy
with emphasis on MRI assessment of GTV and CTV Radiother. Oncol. 74 235–45
Hargreaves B A 2012 Rapid gradient-echo imaging J. Magn. Reson. Imag. 36 1300–13
Harisinghani M G, Saksena M A, Hahn P F, King B, Kim J, Torabi M T and Weissleder R 2006
Ferumoxtran-10-enhanced MR lymphangiography: does contrast-enhanced imaging alone suffice
for accurate lymph node characterization? AJR Am. J. Roentgenol. 186 144–8
Heesakkers R A, Jager G J, Hövels A M, de Hoop B, van den Bosch H C, Raat F, Witjes J A,
Mulders P F, van der Kaa C H and Barentsz J O 2009 Prostate cancer: detection of lymph
node metastases outside the routine surgical area with ferumoxtran-10-enhanced MR imaging
Radiology 251 408–14
Hoogcarspel S J, Crijns S P, Lagendijk J J, van Vulpen M and Raaymakers B W 2013 The feasibility of
using a conventional flexible RF coil for an online MR-guided radiotherapy treatment Phys. Med.
Biol. 58 1925–32
Hoskin P J et al 2007 Hypoxia in prostate cancer: correlation of BOLD-MRI with pimonidazole
immunohistochemistry—initial observations Int. J. Radiat. Oncol. Biol. Phys. 68 1065–71
R365
Topical Review
Phys. Med. Biol. 59 (2014) R349
Hsu S H Cao Y, Huang K, Feng M and Balter J M 2013 Investigation of a method for generating
synthetic CT models from MRI scans of the head and neck for radiation therapy Phys. Med. Biol.
58 8419–35
Intven M, Reerink O and Philippens M E 2013 Diffusion-weighted MRI in locally advanced rectal
cancer: pathological response prediction after neo-adjuvant radiochemotherapy Strahlenther.
Onkol. 189 117–22
Jürgenliemk-Schulz I M, Tersteeg R J, Roesink J M, Bijmolt S, Nomden C N, Moerland M A and
de Leeuw A A 2009 MRI-guided treatment-planning optimisation in intracavitary or combined
intracavitary/interstitial PDR brachytherapy using tandem ovoid applicators in locally advanced
cervical cancer Radiother. Oncol. 93 322–30
Kapanen M, Collan J, Beule A, Seppälä T, Saarilahti K and Tenhunen M 2012 Commissioning of MRIonly based treatment planning procedure for external beam radiotherapy of prostate Magn. Reson.
Med. Ahead of print
Karlsson M, Karlsson M G, Nyholm T, Amies C and Zackrisson B 2009 Dedicated magnetic resonance
imaging in the radiotherapy clinic Int. J. Radiat. Oncol. Biol. Phys. 74 644–51
Keyvanloo A, Burke B, Warkentin B, Tadic T, Rathee S, Kirkby C, Santos D M and Fallone B G 2012
Skin dose in longitudinal and transverse linac-MRIs using Monte Carlo and realistic 3D MRI field
models Med. Phys. 39 6509–21
Kim Y C, Lebel R M, Wu Z, Ward S L, Khoo M C and Nayak K S 2013 Real-time 3D magnetic
resonance imaging of the pharyngeal airway in sleep apnea Magn. Reson. Med. 71 1501–10
Kim Y C, Muruganandham M, Modrick J M and Bayouth J E 2011 Evaluation of artefacts and
distortions of titanium applicators on 3.0 T MRI: feasibility of titanium applicators in MRI-guided
brachytherapy for gynecological cancer Int. J. Radiat. Oncol. Biol. Phys. 80 947–55
Klerkx W M et al 2012 The value of 3.0 T diffusion-weighted MRI for pelvic nodal staging in patients
with early stage cervical cancer Eur. J. Cancer 48 3414–21
Korhonen J, Kapanen M, Keyrilainen J, Seppala T and Tenhunen M 2014 A dual model HU conversion
from MRI intensity values within and outside of bone segment for MRI-based radiotherapy
treatment planning of prostate cancer Med. Phys. 41 011704
Lagendijk J J W, Raaymakers B W, Raaijmakers A J E, Overweg J, Brown K J, Kerkhof E M, van
der Put R W, Hårdemark B, van Vulpen M and van der Heide U A 2008 MRI/linac integration
Radiother. Oncol. 86 25–9
Lagendijk J J W, Raaymakers B W, Van Der Heide U A, Topolnjak R, Dehnad H, Hofman P,
Nederveen A J, Schulz I M, Welleweerd J and Bakker C J G 2002 MRI guided radiotherapy: MRI
as position verification system for IMRT Radiother. Oncol. 64 224 (suppl 1) (Czech Republic
European Society for Therapeutic Radiation Oncology)
Lagerburg V, Moerland M A, van Vulpen M and Lagendijk J J W 2006 A new robotic needle insertion
method to minimise attendant prostate motion Radiother. Oncol. 80 73–7
Lambrecht M, Vandecaveye V, De Keyzer F, Roels S, Penninckx F, Van Cutsem E, Filip C and
Haustermans K 2012 Value of diffusion-weighted magnetic resonance imaging for prediction
and early assessment of response to neoadjuvant radiochemotherapy in rectal cancer: preliminary
results Int. J. Radiat. Oncol. Biol. Phys. 82 863–70
Lanzer P et al 1984 Cardiac imaging using gated magnetic resonance Radiology 150 121–7
Larkman D J, Hajnal J V, Herlihy A H, Coutts G A, Young I R and Ehnholm G 2001 Use of multicoil
arrays for separation of signal from multiple slices simultaneously excited J. Magn. Reson. Imag.
13 313–7
Le Bihan D, Poupon C, Amadon A and Lethimonnier F 2006 Artifacts and pitfalls in diffusion MRI
J. Magn. Reson. Imag. 24 478–88
Lerski R A, McRobbie D W, Straughan K, Walker P M, de Certaines J D and Bernard A M 1988 Multicenter trial with protocols and prototype test objects for the assessment of MRI equipment. EEC
concerted research project Magn. Reson. Imag. 6 201–14
Lewis C E, Prato F S, Drost D J and Nicholson R L 1986 Comparison of respiratory triggering and gating
techniques for removal of respiratory artifacts in MR imaging Radiology 160 803–10
Low R N 2007 Abdominal MRI advances in the detection of liver tumours and characterisation Lancet
Oncol. 8 525–35
Lustig M, Donoho D and Pauly J M 2007 Sparse MRI: the application of compressed sensing for rapid
MR imaging Magn. Reson. Med. 58 1182–95
Ma D J, Gulani V, Seiberlich N, Liu K, Sunshine J L, Duerk J L and Griswold M A 2013 Magnetic
resonance fingerprinting Nature 495 187–92
R366
Topical Review
Phys. Med. Biol. 59 (2014) R349
MacDonald S M, Harisinghani M G, Katkar A, Napolitano B, Wolfgang J and Taghian A G 2010
Nanoparticle-enhanced MRI to evaluate radiation delivery to the regional lymphatics for patients
with breast cancer Int. J. Radiat. Oncol. Biol. Phys. 77 1098–104
Mah D, Steckner M, Palacio E, Mitra R, Richardson T and Hanks G E 2002 Characteristics and quality
assurance of a dedicated open 0.23 T MRI for radiation therapy simulation Med Phys. 29 2541–7
Marin D, Husarik D B, Boll D T and Merkle E M 2010 Abdominal magnetic resonance imaging at 3 T:
oncological applications Top. Magn. Reson. Imag. 21 149–56
Markl M and Leupold J 2012 Gradient echo imaging J. Magn. Reson. Imag. 35 1274–89
McBain C A, Khoo V S, Buckley D L, Sykes J S, Green M M, Cowan R A, Hutchinson C E, Moore C J
and Price P M 2009 Assessment of bladder motion for clinical radiotherapy practice using cinemagnetic resonance imaging Int. J. Radiat. Oncol. Biol. Phys. 75 664–71
McGowan S E, Burnet N G and Lomax A J 2013 Treatment planning optimisation in proton therapy
Br. J. Radiol. 86 20120288
Meijsing I, Raaymakers B W, Raaijmakers A J E, Kok J G, Hogeweg L, Liu B and Lagendijk J J W 2009
Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer
NE2571 ionization chamber Phys. Med. Biol. 54 2993–3002
Moerland M A, Beersma R, Bhagwandien R, Wijrdeman H K and Bakker C J G 1995 Analysis and
correction of geometric distortions in 1.5 T magnetic resonance images for use in radiotherapy
treatment planning Phys. Med. Biol. 40 1651–4
Moman M R, van den Berg C A, Boeken Kruger A E, Battermann J J, Moerland M A, van der Heide U A
and van Vulpen M 2010 Focal salvage guided by T2-weighted and dynamic contrast-enhanced
magnetic resonance imaging for prostate cancer recurrences Int. J. Radiat. Oncol. Biol. Phys.
76 741–6
Morton G C and Hoskin P J 2013 Brachytherapy: current status and future strategies—can high dose rate
replace low dose rate and external beam radiotherapy? Clin. Oncol. (R. Coll. Radiol.) 25 474–82
Muntener M, Patriciu A, Petrisor D, Schär M, Ursu D, Song D Y and Stoianovici D 2008 Transperineal
prostate intervention: robot for fully automated MR imaging—system description and proof of
principle in a canine model. Radiology 247 543–9
Niebergall A, Zhang S, Kunay E, Keydana G, Job M, Uecker M and Frahm J 2013 Real-time MRI
of speaking at a resolution of 33 ms: undersampled radial FLASH with nonlinear inverse
reconstruction Magn. Reson. Med.69 477–85
Nomden C N, de Leeuw A A, Roesink J M, Tersteeg R J, Moerland M A, Witteveen P O, Schreuder H W,
van Dorst E B and Jürgenliemk-Schulz I M 2013 Clinical outcome and dosimetric parameters of
chemo-radiation including MRI guided adaptive brachytherapy with tandem-ovoid applicators for
cervical cancer patients: a single institution experience Radiother. Oncol. 107 69–74
Oborn B M, Metcalfe P E, Butson M J, Rosenfeld A B and Keall P J 2012 Electron contamination
modeling and skin dose in 6 MV longitudinal field MRIgRT: impact of the MRI and MRI fringe
field Med. Phys. 39 874–90
Oechsner M, Pracht E D, Staeb D, Arnold J F, Köstler H, Hahn D, Beer M and Jakob P M 2009 Lung
imaging under free-breathing conditions Magn. Reson. Med. 61 723–7
Ogawa S, Menon R S, Tank D W, Kim S G, Merkle H, Ellermann J M and Ugurbil K 1993 Functional
brain mapping by blood oxygenation level-dependent contrast magnetic resonance imaging.
A comparison of signal characteristics with a biophysical model Biophys. J. 64 803–12
Overweg J, Raaymakers B W, Lagendijk J J W and Brown K 2009 System for MRI guided radiotherapy
Proc. Int. Soc. Mag. Reson. Med. 17 594
Padhani A R and Dzik-Jurasz A 2004 Perfusion MR imaging of extracranial tumor angiogenesis Top.
Magn. Reson. Imag. 15 41–57
Padhani A R et al 2009 Diffusion-weighted magnetic resonance imaging as a cancer biomarker:
consensus and recommendations Neoplasia 11 102–25
Patz S and Hawkes R C 1986 The application of steady-state free precession to the study of very slow
fluid flow Magn. Reson. Med. 3 140–5
Paulson E S, Bradley J A, Wang D, Ahunbay E E, Schultz C and Li X A 2011 Internal margin assessment
using cine MRI analysis of deglutition in head and neck cancer radiotherapy Med. Phys. 38 1740–7
Petersen E T, Zimine I, Ho Y C and Golay X 2006 Non-invasive measurement of perfusion: a critical
review of arterial spin labelling techniques Br. J. Radiol. 79 688–701
Philippens M, Bongard H, Asselen B, van Ulpen M and Lagendijk J 2012 New high resolution T2-FFE
MR imaging of axillary lymph nodes for real time radiation treatment guidance Med. Phys. 39 3889
Plewes D B and Kucharczyk W 2012 Physics of MRI: a primer J. Magn. Reson. Imag. 35 1038–54
R367
Topical Review
Phys. Med. Biol. 59 (2014) R349
Pötter R et al 2011 Clinical outcome of protocol based image (MRI) guided adaptive brachytherapy
combined with 3D conformal radiotherapy with or without chemotherapy in patients with locally
advanced cervical cancer Radiother. Oncol. 100 116–23
Possanzini C, Van Liere P, Roeven H, Den Boef J, Saylor C, Van Eggermond J, Harvey P and Moore E
2011 Scalability and channel independency of the digital broadband dStream architecture Proc. of
Annual meeting of the International Society on Magnetic Resonance in Medicine. abstract 5103
(Montreal, Canada)
Price S J, Jena R, Burnet N G, Hutchinson P J, Dean A F, Peña A, Pickard J D, Carpenter T A and
Gillard J H 2006 Improved delineation of glioma margins and regions of infiltration with the use
of diffusion tensor imaging: an image-guided biopsy study AJNR Am. J. Neuroradiol. 27 1969–74
Pruessmann K P 2006 Encoding and reconstruction in parallel MRI NMR Biomed. 19 288–99
Raaijmakers A J, Hårdemark B, Raaymakers B W, Raaijmakers C P and Lagendijk J J 2007 Dose
optimization for the MRI-accelerator: IMRT in the presence of a magnetic field Phys. Med. Biol.
52 7045–54
Raaijmakers A J E, Raaymakers B W and Lagendijk J J W 2005 Integrating a MRI scanner with a 6 MV
radiotherapy accelerator: dose increase at tissue-air interfaces in a lateral magnetic field due to
returning electrons Phys. Med. Biol. 50 1363–76
Raaijmakers A J E, Raaymakers B W and Lagendijk J J W 2008 Magnetic-field-induced dose effects
in MR-guided radiotherapy systems: dependence on the magnetic field strength Phys. Med. Biol.
53 909–23
Raaymakers B W et al 2009 Integrating a 1.5 T MRI scanner with a 6 MV accelerator: proof of concept
Phys. Med. Biol. 54 N229–37
Raaymakers B W, Raaijmakers A J E, Kotte A N T J, Jette D and Lagendijk J J W 2004 Integrating a
MRI scanner with a 6 MV radiotherapy accelerator: dose deposition in a transverse magnetic field
Phys. Med. Biol. 49 4109–18
Raaymakers B W, Raaijmakers A J E and Lagendijk J J W 2008 Feasibility of MRI guided proton
therapy: magnetic field dose effects Phys. Med. Biol. 53 5615–22
Reichenbach J R, Venkatesan R, Yablonskiy D A, Thompson M R, Lai S and Haacke E M 1997 Theory
and application of static field inhomogeneity effects in gradient-echo imaging J. Magn. Reson.
Imag. 7 266–79
Riddell A M, Davies D C, Allum W H, Wotherspoon A C, Richardson C and Brown G 2007 Highresolution MRI in evaluation of the surgical anatomy of the esophagus and posterior mediastinum
AJR Am. J. Roentgenol. 188 W37–43
Rijpkema M, Kaanders J H, Joosten F B, van der Kogel A J and Heerschap A 2002 Effects of
breathing a hyperoxic hypercapnic gas mixture on blood oxygenation and vascularity of headand-neck tumors as measured by magnetic resonance imaging Int. J. Radiat. Oncol. Biol. Phys.
53 1185–91
Roemer P B, Edelstein W A, Hayes C E, Souza S P and Mueller O M 1990 The NMR phased array
Magn. Reson. Med. 16 192–225
Runge V M, Clanton J A, Partain C L and James A E Jr 1984 Respiratory gating in magnetic resonance
imaging at 0.5 T Radiology 151 521–3
Schakel T, Hoogduin J M, Terhaard C H, Philippens M E 2013 Diffusion weighted MRI in head-andneck cancer: geometrical accuracy Radiother. Oncol. 109 394–7
Sodickson D K et al 2005 Rapid volumetric MRI using parallel imaging with order-of-magnitude
accelerations and a 32-element RF coil array: feasibility and implications Acad. Radiol. 12 626–35
Solanki A A, Chang D T and Liauw S L 2013 Future directions in combined modality therapy for rectal
cancer: reevaluating the role of total mesorectal excision after chemoradiotherapy Onco Targets
Ther. 6 1097–110
St Aubin J, Santos D M, Steciw S and Fallone B G 2010 Effect of longitudinal magnetic fields on a
simulated in-line 6 MV linac Med. Phys. 37 4916–23
Stam M K, Crijns S P M, Zonnenberg B A, Barendrecht M M, van Vulpen M, Lagendijk J J W and
Raaymakers B W 2012 Navigators for motion detection during real-time MRI-guided radiotherapy
Phys. Med. Biol. 57 6797–805
Stanescu T, Wachowicz K and Jaffray D A 2012 Characterization of tissue magnetic susceptibilityinduced distortions for MRIgRT Med. Phys. 39 7185–93
Stemkens B, Tijssen R H, Van den Berg C A T, Lagendijk J J W, Moonen C T W, Ries M and Denis de
Senneville B 2013 Optical flow analysis on undersampled radial acquisitions for real-time tracking
of the pancreas in MR guided radiotherapy Proc. ISMRM (Salt lake City)
R368
Topical Review
Phys. Med. Biol. 59 (2014) R349
Susil R C, Camphausen K, Choyke P, McVeigh E R, Gustafson G S, Ning H, Miller R W, Atalar E,
Coleman C N and Ménard C 2004 System for prostate brachytherapy and biopsy in a standard 1.5 T
MRI scanner Magn. Reson. Med. 52 683–7
Vandecaveye V, De Keyzer F, Vander Poorten V, Dirix P, Verbeken E, Nuyts S and Hermans R 2009 Head
and neck squamous cell carcinoma: value of diffusion-weighted MR imaging for nodal staging
Radiology 251 134–46
van den Bosch M R, Moerland M A, Lagendijk J J W, Bartels L W and van den Berg C A T 2010a New
method to monitor RF safety in MRI-guided interventions based on RF induced image artefacts
Med. Phys. 37 814–21
van den Bosch M R, Moman M R, van Vulpen M, Battermann J J, Duiveman E, van Schelven L J, de
Leeuw H, Lagendijk J J W and Moerland M A 2010b MRI-guided robotic system for transperineal
prostate interventions: proof of principle Phys. Med. Biol. 55 N133–40
van der Heide U A, Houweling A C, Groenendaal G, Beets-Tan R G and Lambin P 2012 Functional MRI
for radiotherapy dose painting Magn. Reson. Imag. 30 1216–23
van Rossum P S, van Hillegersberg R, Lever F M, Lips I M, van Lier A L, Meijer G J, van Leeuwen M S,
van Vulpen M and Ruurda J P 2013 Imaging strategies in the management of oesophageal cancer:
what’s the role of MRI? Eur. Radiol. 23 1753–65
van Veluw S J, Zwanenburg J J, Engelen-Lee J, Spliet W G, Hendrikse J, Luijten P R and Biessels G J
2013 In vivo detection of cerebral cortical microinfarcts with high-resolution 7 T MRI J. Cereb.
Blood Flow Metab. 33 322–9
van Vulpen M, van den Berg C A, Moman M R and van der Heide U A 2009 Difficulties and potential
of correlating local recurrences in prostate cancer with the delivered local dose Radiother. Oncol.
93 180–4
Verduijn G M, Bartels L W, Raaijmakers C P, Terhaard C H, Pameijer F A and van den Berg C A
2009 Magnetic resonance imaging protocol optimization for delineation of gross tumor volume in
hypopharyngeal and laryngeal tumors Int. J. Radiat. Oncol. Biol. Phys. 74 630–6
Wielpütz M and Kauczor H U 2012 MRI of the lung: state of the art Diagn. Interv. Radiol. 18 344–53
Wolf R and Bortfeld T 2012 An analytical solution to proton Bragg peak deflection in a magnetic field
Phys. Med. Biol. 57 N329–37
Yeung C J, Karmarkar P and McVeigh E R 2007 Minimizing RF heating of conducting wires in MRI
Magn. Reson. Med. 58 1028–34
Zhang Z, Meng Q, Chen Y, Li Z, Luo B, Yang Z, Mao L and Lin E 2008 3 T imaging of the cranial nerves
using 3D reversed FISP with diffusion-weighted MR sequence J. Magn. Reson. Imag. 27 454–8
R369