TLC-S, Thermostated Laminar Flow Cabinet, Class II
Biological Safety Cabinet
DESCRIPTION
Thermostated Laminar Flow Cabinet TLC-S is a Class II Biological
Safety Cabinet (BSC), specially designed for cell culture
bioprocessing. Additionally, when connected to exhaust, TLC-S
protects from exposure to vapours, gases, fumes and particulates.
Besides HEPA-filtered sterile air flow the TLC-S has 2 additional
options:
1. Temperature controlled (at 50°C), sterilized laminar air flow.
2. UV-air recirculation systems for DNA/RNA inactivation in the
recirculated air.
1. Thermostated Air Flow
Due to the recirculation of the preheated air we guarantee not only
temperature controlled working surface (at 50°C), but also
uniformity of temperature inside the whole working space.
As a result, instruments (e.g. rollers, rockers, vortexes and
pipettes), materials (e.g. racks and pipette tips), reagents (e.g. kits
for cell line expansion and differentiation, culture media and
supplements), and cell lines or cell culture inoculum — all kept at
the same temperature.
The preheated air flow protects cell lines from possible thermal
shock and prevents condensation of liquid on the upper side of
culture vessels, due to the even temperature distribution inside
the cabinet.
2. Physical Inactivation of DNA/RNA Inside Air Flow
The TLC-S cabinet is equipped with the UV air recirculation system
— using 4 UV lamps installed inside the rear wall which disinfect
the laminar air flow. UV- air recirculation systems - as been shown
earlier by BioSan — are highly effective against infectious
biomaterials that are much smaller than HEPA filter’s pores (less
than 0.3 μm), such as viruses, and DNA/RNA fragments and
amplicons.
SPECIFICATION
Biological Class Safety (EN
12469)
HEPA Filter Seal Type EN 18221:2009
Standard inflow velocity
Illumination of the working area
Class II
Н14
0.35 m/s ±20%
1000 Lux
Temperature setting range
+25°C ... +50°C
Temperature control range
+10°C above ambient ... +50°C
Recirculation of air inside the
cabinet
Noise level, not more
Integrated socket
Power consumption
Maximum permissible outlet
70 %
57 dB
1
1100 W
600 W
CAT. NR.
TLC-S incl. power cord
BS-040114-AAA
230VAC 50/60Hz Euro plug
Biosan SIA., Ratsupites 7, build. 2, Riga, LV-1067, Latvia, Phone: +371 674 261 37, Fax: +371 674 281 01,
E-mail: marketing@biosan.lv, http://www.biosan.lv
TLC-S, Thermostated Laminar Flow Cabinet, Class II
Biological Safety Cabinet
Daylight lamp (for working area
illumination)
2 × 24 W
Bactericidal UV lamp in the
working chamber, TUV25WG13
UV-C
1 × 25 W
Bactericidal UV lamp inside the
recirculator, TUV25WG13 UV-C
4 × 25 W
Working chamber material
Dimensions of the inner
chamber
Overall dimensions
Weight
Stainless steel
645 × 580 × 485 mm
700 × 705 × 1025 mm
98 kg