Gas outlets OR range
The terminal units of a medical gas pipeline system are
devices used to make connections and disconnections in
order to provide specific medical gases to medical devices,
pulmonary ventilators or other appliances.
An incorrect connection puts to hazard the life of patients,
therefore the terminal units have been designed following
different technical standards so to grant a non-interchange-
ability among different gases and prevent from any possible
mistake.
The OR range socket can be fixed to the base block in only
one position.
The profile of the installed socket is always in the correct
position.
Base block OR range
This component is the part of the terminal unit connected to the
medical gas pipeline distribution system (suction in case of vacuum).
The base blocks are designed complying with the standard
EN ISO 9170-1 for design and manufacture requirements.
All base blocks, except those for vacuum, are supplied with a
maintenance valve to be fitted after the plumbing is completed.
Materials:
Brass EN 12164 - CuZn39Pb3
NBR/EPDM
AISI
Vertical inlet base block OR range
P/N
ØB
9
ZARMOR-O
ZARMOR-V
ZARMOR-A4
ZARMOR-P
ZARMOR-A8
ZARMOR-C
ZARMOR-OP
ZARMOR10-O
ZARMOR10-V
ZARMOR10-A4
ZARMOR10-P
ZARMOR10-A8
ZARMOR10-C
ZARMOR10-OP
36
45
A
A
øB
Gas
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
8
8
8
8
8
8
8
10
10
10
10
10
10
10
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen/Nitrous oxide
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen/Nitrous oxide
A
øB
Gas
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
8
8
8
8
8
8
8
10
10
10
10
10
10
10
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen/Nitrous oxide
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen/Nitrous oxide
c/w plug for test.
45° side inlet base block OR range
P/N
A
36
45
ØB
ZARMOR-45-O
ZARMOR-45-V
ZARMOR-45-A4
ZARMOR-45-P
ZARMOR-45-A8
ZARMOR-45-C
ZARMOR-45-OP
ZARMOR10-45-O
ZARMOR10-45-V
ZARMOR10-45-A4
ZARMOR10-45-P
ZARMOR10-45-A8
ZARMOR10-45-C
ZARMOR10-45-OP
c/w plug for test.
P RE S BLO C K s . p . a .
3 Medical
Gas OutletS
Vertical inlet base block OR range c/w copper pipe
ØB
P/N
150
ZARMOR-T-O
ZARMOR-T-V
ZARMOR-T-A4
ZARMOR-T-P
ZARMOR-T-A8
ZARMOR-T-C
ZARMOR-T-OP
ZARMOR10-T-O
ZARMOR10-T-V
ZARMOR10-T-A4
ZARMOR10-T-P
ZARMOR10-T-A8
ZARMOR10-T-C
ZARMOR10-T-N
ZARMOR10-T-OP
A
36
45
A
øB
Gas
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
8
8
8
8
8
8
8
10
10
10
10
10
10
10
10
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen/Nitrous oxide
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
Oxygen/Nitrous oxide
c/w plug for test.
45° side inlet base block OR range c/w copper pipe
P/N
36
45
ZARMOR-45T-O/1
ZARMOR-45T-V/1
ZARMOR-45T-A4/1
ZARMOR-45T-P/1
ZARMOR-45T-A8/1
ZARMOR-45T-C/1
ZARMOR-45T-OP/1
ZARMOR1045T-O/1
ZARMOR1045T-V/1
ZARMOR1045TA4/1
ZARMOR1045T-P/1
ZARMOR1045TA8/1
ZARMOR1045T-C/1
ZARMOR1045TOP/1
A
ØB
A
øB
Gas
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
18,7
8
8
8
8
8
8
8
10
10
10
10
10
10
10
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen/Nitrous oxide
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen/Nitrous oxide
c/w plug for test.
Base block OR range c/w back copper pipe
P/N
150
ØB
ZARMORD-T-O
ZARMORD-T-V
ZARMORD-T-A4
ZARMORD-T-P
ZARMORD-T-A8
ZARMORD-T-C
ZARMORD-T-N
øB
Gas
8
8
8
8
8
8
8
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
A
øB
Gas
18,7
18,7
18,7
18,7
18,7
18,7
18,7
6
8
6
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
c/w plug for test.
A
36
45
A
23,7
23,7
23,7
23,7
23,7
23,7
23,7
Base block OR range for flexible hose
ØB
A
36
45
P/N
ZARMORD-O-PG6
ZARMORD-V-PG8
ZARMORD-A4-PG6
ZARMORD-P-PG6
ZARMORD-A8-PG6
ZARMORD-C-PG6
ZARMORD-N-PG6
c/w plug for test.
50
P RE S BLO C K s . p . a .
4 Medical
Gas OutletS
Base block OR range for flexible hose
P/N
ØB
A
øB
Gas
23,7
23,7
23,7
23,7
23,7
23,7
23,7
6
8
6
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
P/N
A
øB
Gas
ZARMORG-PG6-O
ZARMORG-PG8-V
ZARMORG-PG6-A4
ZARMORG-PG6-P
ZARMORG-PG6-A8
ZARMORG-PG6-C
ZARMORG-PG6-N
30
30
30
30
30
30
30
6
8
6
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
69
ZARMOR-PG6-O
ZARMOR-PG8-V
ZARMOR-PG6-A4
ZARMOR-PG6-P
ZARMOR-PG6-A8
ZARMOR-PG6-C
ZARMOR-PG6-N
c/w plug for test.
36
45
A
Rotating base block OR range c/w hosetail
34
ØB
c/w plug for test.
A
Base block OR range c/w NIST connector
P/N
ZARMORD-O-MNIST
ZARMORD-V-MNIST
ZARMORD-A4-MNIST
ZARMORD-P-MNIST
ZARMORD-A8-MNIST
ZARMORD-C-MNIST
ZARMORD-N-MNIST
65
A
Gas
23,7
23,7
23,7
23,7
23,7
23,7
23,7
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
c/w plug for test.
A
P RE S BLO C K s . p . a .
5 Medical
Gas OutletS
Sockets OR range
The socket is the female part of the terminal unit connected to the
base block by a gas specific interface and which contains the gasspecific connecting point.
These sockets have been designed in accordance with the following standards:
BS 5682:1998, DIN 13260-2, NF S 90-116, FD S 90-119, SS
875 24 30, JIS T 7101, AS 2896 and ENV 737-6 for the dimensions of linking point to the probe;
EN ISO 9170-1 for design and manufacture requirements.
Materials:
Brass EN 12164 - CuZn39Pb3 nickel plated
NBR/EPDM
AISI
ABS/POM
DIN 13260-2
DIN ORPHEUS
P/N
Ø 53
43
Gas
VGTDEP-OR-O
VGTDEP-OR-V
VGTDEP-OR-A4
VGTDEP-OR-P
VGTDEP-OR-C
VGTDEP-OR-N
Oxygen
Vacuum
Air
Nitrous oxide
Carbon dioxide
Nitrogen
Supplied with releaser sleeve that can be custom made on request.
Connecting backside and sleeve made of plastic.
DIN ORPHEUS Plus
P/N
43
Gas
Ø 53
VGTDE-OR-O
VGTDE-OR-V
VGTDE-OR-A4
VGTDE-OR-P
VGTDE-OR-C
VGTDE-OR-N
Oxygen
Vacuum
Air
Nitrous oxide
Carbon dioxide
Nitrogen
Supplied with releaser sleeve that can be custom made on request.
Connecting backside made of metal and sleeve of plastic.
plus
DIN ORPHEUS Premium
P/N
Ø 53
43
Gas
VGTDEM-OR-O
VGTDEM-OR-V
VGTDEM-OR-A4
VGTDEM-OR-P
VGTDEM-OR-C
VGTDEM-OR-N
Oxygen
Vacuum
Air
Nitrous oxide
Carbon dioxide
Nitrogen
Supplied with releaser sleeve that can be custom made on request.
Connecting backside and sleeve made of metal.
premium
P RE S BLO C K s . p . a .
6 Medical
Gas OutletS
DIN 13260-2 socket OR range
P/N
45
Gas
Ø45
VGTDM-OR-O
VGTDM-OR-V
VGTDM-OR-A4
VGTDM-OR-P
VGTDM-OR-C
VGTDM-OR-N
Oxygen
Vacuum
Air
Nitrous oxide
Carbon dioxide
Nitrogen
Supplied with releaser sleeve that can be custom made on request.
DIN 13260-2 horizontal socket OR range
P/N
45
Gas
VGTDMHO-OR-O
VGTDMHO-OR-C
VGTDMHO-OR-N
Oxygen
Carbon dioxide
Nitrogen
Ø45
Supplied with releaser sleeve that can be custom made on request.
BS 5682:1998
BS 5682:1998 socket OR range
P/N
44,5
Gas
Ø50
VGTBSM-OR-O
VGTBSM-OR-V
VGTBSM-OR-A4
VGTBSM-OR-P
VGTBSM-OR-A8
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Supplied with releaser sleeve that can be custom made on request.
BS 5682:1998 horizontal socket OR range
P/N
44,5
Gas
Ø50
VGTBSMHO-OR-O
VGTBSMHO-OR-V
VGTBSMHO-OR-A4
VGTBSMHO-OR-P
VGTBSMHO-OR-A8
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Supplied with releaser sleeve that can be custom made on request.
P RE S BLO C K s . p . a .
7 Medical
Gas OutletS
BS 5682:1998 plastic socket OR range
P/N
43
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Supplied with releaser sleeve that can be custom made on request.
Ø50
BS 8532:2011
Gas
VGTBSP-OR-O
VGTBSP-OR-V
VGTBSP-OR-A4
VGTBSP-OR-P
VGTBSP-OR-A8
AGSS BS 8532:2011 c/w plastic box
Ø 12
P/N
VEGS-BS
VEGS-BS/1
150
without box
AGSS BS 8532:2011 c/w NIST connector
P/N
VEGS-BS-MNIST
NF S 90-116 and FD S 90-119
AFNOR socket OR range
P/N
M27x1
42
Gas
VGTFM-OR-O
VGTFM-OR-V
VGTFM-OR-A4
VGTFM-OR-P
VGTFM-OR-A8
VGTFM-OR-C
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
12
24
P RE S BLO C K s . p . a .
8 Medical
Gas OutletS
Bulkhead nut
P/N
3
A
B
Colour
18,7
36,6
grey
B
M27x1
PGTF/1
A
Gas identification ring
A
B
SS 875 24 30
P/N
A
B
Disting.colour
Gas
ZGTF-O
ZGTF-V
ZGTF-A4
ZGTF-P
ZGTF-A8
ZGTF-C
ZGTFD70-O
ZGTFD70-V
ZGTFD70-A4
ZGTFD70-P
ZGTFD70-A8
ZGTFD70-C
38
38
38
38
38
38
38
38
38
38
38
38
55
55
55
55
55
55
70
70
70
70
70
70
white
yellow
black-white
blue
white-black
grey
white
yellow
black-white
blue
white-black
grey
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
SS 875 24 30 socket OR range
P/N
43
Gas
Ø45
VGTAM-OR-O
VGTAM-OR-V
VGTAM-OR-A4
VGTAM-OR-P
VGTAM-OR-A8
VGTAM-OR-C
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Supplied with releaser sleeve that can be custom made on request.
SS 875 24 30 horizontal socket OR range
P/N
Ø45
43
Gas
VGTAMHO-OR-O
VGTAMHO-OR-V
VGTAMHO-OR-A4
VGTAMHO-OR-P
VGTAMHO-OR-A8
VGTAMHO-OR-C
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Supplied with releaser sleeve that can be custom made on request.
P RE S BLO C K s . p . a .
9 Medical
Gas OutletS
ENV 737-6
ENV 737-6 socket OR range
P/N
52
Ø53
JIS T 7101
Gas
VGTEM-OR-O
VGTEM-OR-V
VGTEM-OR-A4
VGTEM-OR-P
VGTEM-OR-A8
VGTEM-OR-C
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Supplied with releaser sleeve that can be custom made on request.
JIS T 7101 socket OR range
P/N
45
Gas
VGTJSM-OR-O
VGTJSM-OR-V
VGTJSM-OR-A4
VGTJSM-OR-P
Ø50
Oxygen
Vacuum
Air
Nitrous oxide
AS 2896
AS 2896 outlet
ØB
P/N
Ø50
150
VGTAS-O
VGTAS-V
VGTAS-A4
VGTAS-P
VGTAS-A8
VGTAS-C
VGTAS-N
øB
Gas
1/4’’
5/16’’
1/4’’
1/4’’
1/4’’
1/4’’
1/4’’
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
48
AS 2896 outlet c/w NIST connector
M22x1,5
P/N
Gas
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Nitrogen
Ø 50
200
VGTAS-O-MNIST
VGTAS-V-MNIST
VGTAS-A4-MNIST
VGTAS-P-MNIST
VGTAS-A8-MNIST
VGTAS-C-MNIST
VGTAS-N-MNIST
49
P RE S BLO C K s . p . a .
10 M e d i c a l
Gas OutletS
AS 2896 Anaesthetic Gas Scavenge
ØB
P/N
øB
VEGS-AS
8
50
AS 2896 Anaesthetic Gas Scavenge c/w NIST connector
P/N
M22x1,5
200
VEGS-AS-MNIST
50
P RE S BLO CK s . p . a .
11 M e d i c a l
Gas OutletS
Plastic boxes for terminal units OR
Box without lid
P/N
99
B
ZBOX-GT-OR
ZBOX-GT-OR/01
ZBOX-GT-OR/02
79
A
B
Profile
56,8
56,8
58,8
46,2
52,5
55
suitable for DIN 13260-2 and SS 875 24 30 standards
suitable for BS 5682:1998 and JIS T 7101 standards
suitable for ENV 737-6 standard and DIN ORPHEUS
A
Suitable to be either flush-mounted or wall-mounted.
19
Box c/w lid for AFNOR outlet
P/N
65
ZBOXGTF
ZBOXGTF/1
c/w locking nut
67
65
46
Lid for BHU Afnor outlet
P/N
65
BOXGTF
BOXGTF-1
15
65
c/w locking nut
56
P RE S BLO C K s . p . a .
12 M e d i c a l
Gas OutletS
UNI 9507 Gas outlets
The terminal units of a medical gas pipeline system are
devices used to make connections and disconnections in
order to provide specific medical gases to medical devices,
pulmonary ventilators or other appliances.
An incorrect connection puts to hazard the life of patients,
therefore the terminal units have been designed following
different technical standards that ensure the accordance to
the European directive 93/42/EEC so to grant a non-interchangeability among different gases and prevent from any
possible mistake.
Base blocks
This component is the part of the terminal unit connected to the
medical gas pipeline distribution system (suction in case of vacuum). The base blocks are designed in accordance with the standard EN ISO 9170-1 for design and manufacture requirements.
All base blocks, except those for vacuum, are supplied with a
maintenance valve to be fitted after the plumbing is completed.
Materials:
Brass EN 12164 - CuZn39Pb3
NBR/EPDM
AISI
Base block c/w connector to weld
32
P/N
Gas
ZARM60-O/PB
ZARM60-V/PB
ZARM60-A4/PB
ZARM45-P/PB
ZARM60-A8/PB
ZARM60-C/PB
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
8
Hexagonal head plug for base block
P/N
AGT-O/TPP
AGT-V/TPP
AGT-A4/TPP
AGT-P/TPP
AGT-A8/TPP
AGT-C/TPP
17
5
Ch 7
A
Disting.colour
Gas
M22x1,5D
M24x1,5D
M20x1,5S
M16x1,5D
M18x1,5D
M20x1,5D
white
yellow
black
blue
black
grey
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
7
c/w O-ring
A
Ø36
P RE S BLO CK s . p . a .
13 M e d i c a l
Gas OutletS
Sockets
The socket is the female part of the terminal unit connected to the
base block by a gas specific interface and which contains the gasspecific connecting point.
These sockets have been designed in accordance with the following standards:
UNI 9507;
EN ISO 9170-1 for design and manufacture requirements.
UNI 9507
Materials:
Brass EN 12164 - CuZn39Pb3 nickel plated
NBR/EPDM
AISI
ABS/POM
UNI 9507 socket
P/N
9
45
A
7
Ø53
5,5
VGTX60-O
VGT60-V
VGT60-A4
VGT45-P
VGT60-A7
VGT60-C
A
Disting.colour
Gas
M22x1,5D
M24x1,5D
M20x1,5S
M16x1,5D
M18x1,5D
M20x1,5D
white
yellow
black-white
blue
white-black
grey
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
The sleeve can be custom made on request.
Plastic boxes
Box without lid suitable for UNI 9507 outlets
P/N
BOX-GT
Suitable to be either flush-mounted or surface-mounted
85
20
Supplied with:
- service cap;
- components designed to link the boxes to form a series.
69
36,5
85
107
P RE S BLO C K s . p . a .
14 M e d i c a l
Gas OutletS
Probes
All these male connectors to be coupled to the sockets are
non-interchangeable and are designed in accordance with
the following standards:
UNI 9507, NF S 90-116, FD S 90-119, DIN 13260-2, BS
5682:1998, SS 875 24 30 and ENV 737-6 for the dimensions of linking point to the socket;
EN ISO 9170-1 for design and manufacture requirements.
UNI 9507
Materials:
Brass EN 12164 - CuZn39Pb3 nickel plated
AISI
NBR/EPDM
PE/POM/ABS
Elbow barb end probe
P/N
62
IGTG60-PG6-O
IGTG60-PG8-V
IGTG60-PG6-A4
IGTG45-PG6-P
IGTG60-PG6-A7
IGTG60-PG6-C
55
35
Disting.colour
H
Gas
white
yellow
black-white
blue
white-black
grey
6
8
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Supplied with sticky label that can be custom made on request.
H
NF S 90-116 and FD S 90-119
Straight barb end probe
P/N
51
H
IGTF-PG6-O
IGTF-PG8-V
IGTF-PG6-A4
IGTF-PG6-P
IGTF-PG6-A8
IGTF-PG6-C
Disting.colour
H
Gas
white
yellow
black-white
blue
white-black
grey
6
8
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Supplied with sleeve that can be custom made on request.
Elbow barb end probe
P/N
48
IGTGF-PG6-O
IGTGF-PG8-V
IGTGF-PG6-A4
IGTGF-PG6-P
IGTGF-PG6-A8
IGTGF-PG6-C
Disting.colour
H
Gas
white
yellow
black-white
blue
white-black
grey
6
8
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Supplied with the sleeve that can be custom made on request.
H
P RE S BLO CK s . p . a .
15 M e d i c a l
Gas OutletS
Straight swivel male threaded probe
P/N
IGTF-OR-M12-O
IGTF-OR-M12-V
IGTF-OR-M12-A4
IGTF-OR-M12-P
IGTF-OR-M12-C
IGTF-OR-M1/8-O
IGTF-OR-M1/8-V
IGTF-OR-M1/8-A4
IGTF-OR-M1/4-O
IGTF-OR-M1/4-V
B
46
Ch 20
41,5
B
Disting.colour
Gas
M12x1
M12x1
M12x1
M12x1
M12x1
1/8’’
1/8’’
1/8’’
1/4’’
1/4’’
white
yellow
white-black
blue
grey
white
yellow
white-black
white
yellow
Oxygen
Vacuum
Air
Nitrous oxide
Carbon dioxide
Oxygen
Vacuum
Air
Oxygen
Vacuum
Supplied with the sleeve that can be custom made on request.
DIN 13260-2
Elbow barb end probe
P/N
55
IGTGD-PG6-O
IGTGD-PG8-V
IGTGD-PG6-A4
IGTGD-PG6-P
IGTGD-PG6-C
IGTGD-PG6-N
22
Disting.colour
H
Gas
white
yellow
black-white
blue
grey
black
6
8
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Carbon dioxide
Nitrogen
Supplied with sticky label that can be custom made on request.
28
H
64
Elbow barb end probe with check valve
P/N
55
IIGTGD-PG6-O
IIGTGD-PG6-A4
IIGTGD-PG6-P
IIGTGD-PG6-C
22
Disting.colour
H
Gas
white
black-white
blue
grey
6
6
6
6
Oxygen
Air
Nitrous oxide
Carbon dioxide
Supplied with sticky label that can be custom made on request.
28
H
64
ENV 737-6
Elbow barb end probe
P/N
60
55
41
IGTGE-PG6-O
IGTGE-PG8-V
IGTGE-PG6-A4
IGTGE-PG6-P
IGTGE-PG6-A8
IGTGE-PG6-C
Disting.colour
H
Gas
white
yellow
black-white
blue
white-black
grey
6
8
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
Supplied with sticky label that can be custom made on request.
H
P RE S BLO C K s . p . a .
16 M e d i c a l
Gas OutletS
BS 5682:1998
Straight barb end probe
H
P/N
66
SS 875 24 30
IGTBS-PG6-O
IGTBS-PG8-V
IGTBS-PG6-A4
IGTBS-PG6-P
IGTBS-PG6-A8
H
Gas
6,5
8
6,5
6,5
6,5
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Made of nickel plated brass.
Straight barb end probe
P/N
H
Gas
IGTA-PG6-O
IGTA-PG10-V
IGTA-PG8-A4
IGTA-PG6-P
IGTA-PG8-A8
IGTA-PG6-C
6
10
8
6
8
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Carbon dioxide
P RE S BLO CK s . p . a .
17 M e d i c a l
Gas OutletS
NIST connectors
Range of male and female components intended to maintain gas-specificity by allocation of a set of different diameters
and screw thread to mating components to each gas. These
connectors are suitable for medical gases when the pressure
range is 300 kPa - 1400 kPa and for vacuum when the pressure range is 10 kPa - 100 kPa.
These components are designed in accordance with the
standard EN 15908
Materials:
Brass EN 12164 - CuZn39Pb3 nickel plated
NBR/EPDM
AISI
1st fix Nist connector c/w copper pipe welded
A
Ch 29
M22x1,5
P/N
VGTN-PL-O-T1/2
VGTN-PL-A4-T1/2
VGTN-PL-V-T1/2
VGTN-PL-P-T1/2
VGTN-PL-C-T1/2
VGTN-PL-A8-T1/2
120
161,5
A
Gas
1/2’’
1/2’’
1/2’’
1/2’’
1/2’’
1/2’’
Oxygen
Air
Vacuum
Nitrous oxide
Carbon dioxide
Air-800
PL = free flow
A
Ch 32
M22x1,5
1st fix Nist connector c/w check valve and copper pipe welded
150
200
P/N
A
Gas
VGTN-O-T10
VGTN-A4-T10
VGTN-V-T12
VGTN-P-T10
VGTN-C-T10
VGTN-A8-T10
VGTN-AGSS-T15*
10
10
12
10
10
10
15
Oxygen
Air
Vacuum
Nitrous oxide
Carbon dioxide
Air-800
AGSS
*No check valve
1st fix Nist connector c/w hosetail & check valve
A
20
M22x1,5
Ch22
P/N
A
Gas
VGTN-O-PG6
VGTN-A4-PG6
VGTN-V-PG8
VGTN-P-PG6
VGTN-C-PG6
VGTN-A8-PG6
6
6
8
6
6
6
Oxygen
Air
Vacuum
Nitrous oxide
Carbon dioxide
Air-800
21
70
1st fix Nist connector c/w hosetail
A
M22x1,5
Ch17
21
P/N
A
Gas
VGTN-PL-O-PG6
VGTN-PL-A4-PG6
VGTN-PL-V-PG8
VGTN-PL-P-PG6
VGTN-PL-A8-PG6
VGTN-PL-C-PG6
6
6
8
6
6
6
Oxygen
Air
Vacuum
Nitrous oxide
Air-800
Carbon dioxide
PL= free flow
59
P RE S BLO C K s . p . a .
18 M e d i c a l
Gas OutletS
A
M22x1,5
NIST straight barb end probe
P/N
A
Gas
IGTN-PG6-O
IGTN-PG8-V
IGTN-PG6-A4
IGTN-PG6-P
IGTN-PG6-A8
IGTN-PG6-OA
6
8
6
6
6
6
Oxygen
Vacuum
Air
Nitrous oxide
Air-800
Air/Oxygen
25
60,5
st
1 fix Nist connector c/w M1/4’’ thread
P/N
A
M22x1,5
Ch 23
VGTN-O-M1/4
VGTN-V-M1/4*
VGTN-A4-1/4
A
Gas
1/4’’
1/4’’
1/4’’
Oxygen
Vacuum
Air
* Free flow
11
60
P RE S BLO CK s . p . a .
19 M e d i c a l
Gas OutletS
Anaesthetic Gas Scavenging System
(AGSS) and Air Motor
They are the part of system to which the operator makes
connection and disconnection for exhausting the medical
gases coming from anaesthetic and breathing systems.
To grant the gas-specifity and prevent any mistake, these
terminal units have been designed in accordance to the european standard EN ISO 9170-2 that ensure the conformity
to the european directive 93/42/EEC, taken as reference.
EN ISO 9170-2 AGSS sockets
These are the female parts of a terminal unit attached to the base
block by a gas-specific interface which contain the gas-specific
connection point.
Materials:
Brass EN 12164 - CuZn39Pb3
Aluminium series 2000 or 6000
Steel
NBR
AISI
PA/POM
AGSS type 1L outlet for centralized system
P/N
Ø12
Disting.colour
VEGS120/EN1
magenta
Ø50
Supplied with metal sleeve that can be custom made on request.
31,5
47,5
AGSS type 1L active
P/N
Ø50
ØA
VEGA120/EN1
Disting.colour
øA
øB
magenta
10 (supply)
16 (evacuation)
Supplied with:
- adjusting screw for setting following the standard EN ISO 7396-2
.- green suction indicator.
- sleeve that can be custom made on request.
ØB
- manifold for connecting the exhaust copper pipe.
Ø65
P RE S BLO C K s . p . a .
20 M e d i c a l
Gas OutletS
AGSS type 1L active c/w bulkhead plate
P/N
Ø50
VEGA120/EN2
ØA
Disting.colour
øA
øB
magenta
10 (supply)
16 (evacuation)
Supplied with:
- adjusting screw for setting following the standard EN ISO 7396-2.
- green suction indicator.
- sleeve that can be custom made on request.
ØB
- manifold for connecting the exhaust copper pipe.
Ø65
AGSS type 1L active for ceiling pendants
P/N
Ø50
VEGA120/EN3
ØA
20
Disting.colour
øA
magenta
10 (supply )
Supplied with:
15
- adjusting screw for setting following the standard EN ISO 7396-2.
- green suction indicator.
- sleeve that can be custom made on request.
- adapter for connecting the exhaust copper pipe or flexible hose.
Ø65
Box for AGSS active outlet
P/N
BOX-AGSS
20
Made of aluminium.
157
125
52
120
212
Ø13
252
60
P RE S BLO CK s . p . a .
21 M e d i c a l
Gas OutletS
EN ISO 91790-2 AGSS probes
These are the male non-interchangeable parts to be connected to
the outlets.
Materials:
Aluminium series 2000 or 6000 anodized
60° Barb end type 1L
P/N
Disting.colour
D
øG
E
magenta
30
22
63
30
D
E
Ø
G
60°
IEGG120-PG22/EN
Air Motor ENV 737-6
This terminal unit integrates a medical tool drive outlet with a gas
scavenging terminal unit.
Materials:
Brass EN 12164 - CuZn39Pb3 nickel plated
Steel
NBR
Anodized Aluminium
ENV 737-6 outlet for ceiling pendants
P/N
Ø15
Ø9
Ø70
VAM737-PG6
82
Ø64
P RE S BLO C K s . p . a .
22 M e d i c a l
Gas OutletS
ENV 737-6 outlet c/w copper pipes
P/N
68
60
55,5
Ø70
Ø64
VAM737-TL
A Ø8
Ø15
ENV 737-6 barb end probe
P/N
10
IAM737-PG6
20
Ø6
163
P RE S BLO CK s . p . a .
23 M e d i c a l
Gas OutletS
Accessories
Adapters for ambulances
Materials:
Brass EN 12164 - CuZn39Pb3
NBR/EPDM
AISI
ABS/POM
Aluminium
Adapter with one AFNOR NF S 90-116 socket and barb end inlet
P/N
A
Gas
AGTF-1V-PG6-V
6
Vacuum
20 26
24
A
60
Adapter with two AFNOR NF S 90-116 sockets and barb end inlet
24
P/N
A
Gas
AGTF-2V-PG6-O
AGTF-2V-PG6-V
6
6
Oxygen
Vacuum
20
26
62
A
110
Adapter with three AFNOR NF S 90-116 sockets and barb end inlet
19
P/N
A
Gas
AGTF-3V-PG6-O
6
Oxygen
20
26
40
A
117
Adapter with four AFNOR NF S 90-116 sockets and barb end inlet
P/N
A
Gas
AGTF-4V-PG6-O
6
Oxygen
A
20
26
40 18
156
P RE S BLO C K s . p . a .
24 M e d i c a l
Gas OutletS
Adapter with one SS 875 24 30 socket and barb end inlet
P/N
A
Gas
AGTA-1V-O
6
Oxygen
54
A
21
42
Adapter with two SS 875 24 30 sockets and barb end inlet
21
P/N
A
Gas
AGTA-2V-O
6
Oxygen
54
A
68
110
Adapter with two DIN 13260-2 sockets and barb end inlet
P/N
A
Gas
AGTD-2V-O
AGTD-2V-A4
6
6
Oxygen
Air
A
64,5
68
20,5
110
Adapter with four DIN 13260-2 sockets and barb end inlet
20,5
A
Gas
AGTD-4V-O
AGTD-4V-A4
6
6
Oxygen
Air
A
64,5
68
P/N
190
P RE S BLO CK s . p . a .
25 M e d i c a l
Gas OutletS
Quick couplers for flowmeter and regulator
DIN mini socket male threaded
P/N
41,5
VGTD-O-M1/4
VGTD-A4-M1/4
VGTD-V-M1/4
VGTD-P-M1/4
VGTD-C-M1/4
VGTD-O-M3/8
ØA
Ch27
10
øA
Gas
1/4’’
1/4’’
1/4’’
1/4’’
1/4’’
3/8’’
Oxygen
Air
Vacuum
Nitrous oxide
Carbon dioxide
Oxygen
26
DIN mini socket with barb end
A
50,5
P/N
A
Gas
VGTD-O-PG6
VGTD-A4-PG6
6
6
Oxygen
Air
Ch 27
SS mini socket male threaded
P/N
41
A
VGTA-O-M10X1
VGTA-A4-M10X1
VGTA-P-M10X1
VGTA-C-M10X1
VGTA-O-M1/4
VGTA-A4-M1/4
A
Gas
M10x1
M10x1
M10x1
M10x1
1/4’’
1/4’’
Oxygen
Air
Nitrous oxide
Carbon dioxide
Oxygen
Air
Ch 24
AFNOR mini socket male threaded
P/N
39
VGTF-O-M1/4
VGTF-A4-M1/4
VGTF-O-M3/8
VGTF-A4-M3/8
VGTF-V-M3/8
øA
Gas
1/4’’
1/4’’
3/8’’
3/8’’
3/8’’ Oxygen
Air
Oxygen
Air
Vacuum
25
ØA
B
B
6,5
6,5
10
10
10
Ch23
BS mini socket male threaded
P/N
VGTBS-O-M1/4/V
A
L
Gas
1/4’’
60
Oxygen
A
L
P RE S BLO C K s . p . a .
26 M e d i c a l
Gas OutletS
JIS mini socket male threaded
P/N
L
A
L
Gas
1/4’’
54,5
Oxygen
A
VGTJS-O-M1/4
AS mini socket male threaded
P/N
VGTAS-O-M1/4
A
L
Gas
1/4’’
47
Oxygen
A
L
UNI mini socket male threaded
P/N
VGTI60-O-M1/4
VGTI60-O-M3/8
A
L
Gas
1/4’’
3/8’’
56
41
Oxygen
Oxygen
A
L
Flexible hoses for medical gas
Flexible hose for EN ISO 5359 assemblies
P/N
NTPB6x12-O
NTPB8x14-V
NTPB6x12-A4
NTPB6x12-P
NTPB6x12-C
NTPB6x12-NC
Disting.colour
Gas
white
yellow
white-black
blue
grey
black
Oxygen
Vacuum
Air
Nitrous oxide
Carbon dioxide
-
50 meters coil.
Without DEHP.
P RE S BLO CK s . p . a .
27 M e d i c a l
Gas OutletS
PRES BLOCK S.p.A. - Via Alpignano, 151-155
I - 10040 CASELETTE (TO) Tel. +39 011 9688055 (r.a.) - Fax +39 011 9688668
http://www.presblock.com
e-mail: export@presblock.com
sales@presblock.com
The products described herein, including without limitation, product features, specifications,
designs are subject to change by Pres Block at any time without notice.
PBCTGTGB
plus