PACKED BED REACTOR OPERATING MANUAL REVISION: FALL 2014 – C EDITOR: HAROLD J. TOUPS INTRODUCTION The study of reaction kinetics is at the heart of chemical engineering. Coupled with separations (i.e., mass transfer) and process control, these three elements are chief distinguishing characteristics separating chemical engineering from, say, mechanical engineering. Similarly, knowledge and skill in these three disciplines distinguish chemical engineers from the rest of engineers. Laboratory and pilot scale experiments in kinetics – properly performed – provide the necessary data and insights needed to understand reaction mechanisms; evaluate the effects of composition, temperature, and catalysts on the rates of reaction; and ultimately allow chemical engineers to design commercial scale reactors for the manufacture of valuable products. BACKGROUND ON THE ETHYLENE HYDROGENATION REACTION The hydrogenation of C2H4 has been studied as a model reduction reaction in characterizing new metal catalysts. Beeck, in 1950, was one of the first to systematically study the reaction[1]. He found that nickel, while not the most active metal catalyst for this reaction, is active enough that the hydrogenation reaction can take place at < 200°C. A S H 2 2S AS HS HS Section: Introduction The reaction typically involves adsorbed, dissociated H2 reacting with adsorbed C2H4. In other words, both hydrogen atoms and C2H4 molecules form bonds with a metal site (here denoted as S). The strong bonding of C2H4 with S weakens the double bond sufficiently to allow hydrogen atoms to add to C2H4, forming ethane, which is not adsorbed. A mechanism for the reaction has been suggested by Butt (letting A = C2H4, E = C2H6 and S = a metal site) and can be written as follows[2]: (1) (2) 1 AS HS AHS S (3) AHS HS E S S (4) If we assume the third reaction is the rate-limiting step, and that the total amount of S sites is constant (So) we can use an approximate mass balance: (S0 ) (S ) ( AS ) ( HS ), (5) and the quasi-equilibrium assumption of the first two reaction steps to obtain a theoretical kinetics expression: r k ( H 2 )1/2 ( A)(S0 )2 [1 K1 ( A) K21/2 ( H 2 )1/2 ]2 (6) Note that in the approximate mass balance we assume that (S ), ( HS ) and ( AS ) ( AHS ) (7) Also note that (So) is a constant as long as the total number of metal sites remains the same. When the number of metal sites decreases with respect to time, we say the catalyst deactivates; when it increases, the catalyst activates. In this reaction, deactivation can be caused by a side reaction with this stoichiometry: aC2 H 4 (CH )2 a aH 2 (8) Section: Background on the Ethylene Hydrogenation Reaction The component (CH)2a – called coke – is too heavy (i.e., a is large) to desorb from the metal sites and so these metal sites are effectively removed from the catalysis[3]. However, subsequent reaction conditions may cause the coke to break down, in effect re-activating the catalyst. For the kinetics described above, it is evident that for low concentrations of C2H4 the rate is first-order in C2H4 while for high concentrations of C2H4 it is -1 order. Rate expressions of this type –also common in enzyme-catalyzed reactions – are called Langmuir-Hinshelwood rate expressions. Since the equilibrium constant K1 is temperature-dependent, this rate expression tells us that the order m for C2H4 in a power law rate expression of the type r k ( A)m ( H 2 )n (9) will change with temperature. Most rate expressions regressed from experimental data are of the power law type. The catalyst used in the packed bed reactor employs Ni as the active component, but contains SiO2 in significant quantity as well. The SiO2 serves as a support for the Ni; its only purpose is to provide a large surface area for the Ni to cover. The SiO2 is chemically inert at the conditions used here. Another inert material, SiC, is used to fill up the rest of the reactor. General reaction kinetics references abound. Fogler’s book is particularly good[4]. 2 PACKED BED REACTOR – EXXON CATALYTIC REACTOR PILOT UNIT Section: Packed Bed Reactor – Exxon Catalytic Reactor Pilot Unit Donated by the former Exxon Research and Development Laboratories here in Baton Rouge, the CAT unit – as it is sometimes called – is capable of many different catalyst experimental testing strategies. Currently, it is equipped to facilitate heterogeneous packed bed reactor studies testing nickel and other metal-on-support hydrogenation catalysts in the hydrogenation of ethylene – C2H4. Figure 1 is a simplified schematic of the reactor system. Figure 1 – Exxon Catalytic Reactor Pilot Unit schematic. The reactor itself (detailed in Figure 2) is a steel tube contained within a sandbath. The sandbath is fluidized using air and is heated by metal resistance heaters. Large quantities of heat can be transferred rapidly to this pilot plant reactor. Catalyst cylinders are located in a narrow layer at a point roughly 4 inches above the lowest movable thermocouple position. The precise location and extent can be determined by detection of the reaction exotherm (see Longitudinal Temperature Variation section later in this document). Precautions have been taken to make this a safe system. There are relief valves on the system, a high temperature shutdown, and only diluted H2 (see the specification information on/next to the cylinder) is used. However, with any reacting system, strict adherence to safety procedures 3 is necessary. The lower flammability limit of H2 in air is 4 vol. %[5], therefore it is important to ensure that the reactor is not leaking H2 into the surrounding sandbath. Section: Packed Bed Reactor – Exxon Catalytic Reactor Pilot Unit Key measurement components of this pilot unit are instrumented through the Emerson Process Management DeltaVTM control system. Figure 2 – Reactor mechanical drawing denoting thermowells and inlet preheat coil. 4 OPERATING INSTRUCTIONS The following section is devoted to instructions on the typical operation of the pilot unit as currently configured. LOGGING INTO AND USING THE DELTAV™ SYSTEM The DeltaV system can be accessed in virtual fashion from any of the computers in the control room. Though there are several DeltaV virtual machines that allow such access, the che-uolabdv3 virtual machine is reserved for, and set up properly for access to the CAT Unit. Accessing this virtual machine is done through VMware’s Horizon View Client, as outlined in the multistep process given below. To access the che-uolab-dv3.lsu.edu DeltaV virtual machine, follow the steps outlined here: Section: Operating Instructions 1. Logon to Windows 7 using a valid LSU ID and password 2. Double-click the VM Horizon View Client icon on the Desktop. The app will open and a screen such as that show in the figure below will appear: 3. If an icon for che-uolab-dv3.lsu.edu is already showing in the gray area, skip to Step 8 in this process. If NO icon for che-uolab-dv3.lsu.edu is showing in the gray area (as depicted in the figure above), then double-click the Add Server icon. A new popup will appear, as shown in the figure below: 5 Section: Operating Instructions 4. Here, enter the desired server’s name: che-uolab-dv3.lsu.edu and click the Connect button. At this point, another popup will appear; one that looks like the following figure: 5. Here, enter administrator as the User name and deltav as the Password. Click the Login button. 6. The dv3 virtual machine should appear next and another DeltaV Logon menu displays. Type deltav in the Password: field. Click OK. 7. A Flexlock menu window appears next. Click the Windows Desktop button. Minimize this menu window. At this point, skip to Step 13. 8. You are here because an icon for che-uolab-dv3.lsu.edu is showing in the gray area, appearing like that show in the figure below: 6 Section: Operating Instructions 9. Double-click the che-uolab-dv3.lsu.edu icon, at which point the following popup will appear: 10. Here, the proper User name has already been loaded but you must enter deltav in the Password field. Click the Login button. 7 11. The dv3 virtual machine should appear next and another DeltaV Logon menu displays. Type deltav in the Password: field. Click OK. 12. A Flexlock menu window appears next. Click the Windows Desktop button. Minimize this menu window. 13. You have logged in successfully to the che-uolab-dv3.lsu.edu virtual machine. At this point, the DeltaV virtualized station of choice is up and operating. From here you can access the DeltaV Operate Run program and the Process History View program (and Control Studio from che-uolab-dv1) from the Windows Start menu. You should also be able to access the apps drive and a connected jump drive (if installed) from Windows Explorer on the DeltaV virtual machine. To start up the control schematic navigate to Start > DeltaV > Operator > DeltaV Operate Run. The UOLAB_Overview display should come up. Click on the hot-linked photo of the Exxon Catalytic Reactor unit. Doing this should automatically open the CAT1 display. If a dialog box appears indicating an error, click Skip All on the dialog box. CHANGING CONTROLLER PARAMETERS FROM CONTROLLER FACEPLATES On the larger operations schematic, the appearance of a small icon – identifiable by a gray box with three vertical lines – signals the presence of an automatic controller. To the left of each controller icon are the process value (PV, the measurement) in yellow, the setpoint (SP) in white, and the output (OP) in cyan. To change a parameter on a controller, click on this icon to bring up a faceplate. From the faceplate, controller mode can be changed by clicking on the desired mode button on the left side of the faceplate. Section: Operating Instructions Manually changing the controller output (OP) is only possible in MANual mode. To change the controller output, click on the MANual value field at the top of the faceplate and enter a new value. Click-dragging the large cyan pointer – present only when the controller is in MANual mode – to a new position also changes the output of the controller. The smaller cyan pointers are output limit indicators and cannot be changed from the controller faceplate. When the mode is not MAN, the controller uses the process value (PV), setpoint (SP) and tuning constants to calculate a new output (OP) every processing pass. Manually changing the controller setpoint (SP) is only possible in AUTO mode. To change the controller setpoint, click on the setpoint value field on the right side of the faceplate and enter a new value. Click-dragging the large white pointer – present ONLY when the controller is in AUTO – to a new position also changes the setpoint of the controller. The smaller white pointers setpoint limit indicators and cannot be changed from the controller faceplate. 8 ACCESSING ADDITIONAL CONTROLLER DETAILS Across the bottom of the faceplate are six icons that call up other displays with more details about this controller. The two most useful ones for this experiment are the first one from the left, which provides access to controller parameters; and the second one from the right, which calls up the historical trend for this controller. ACCESSING REAL-TIME PROCESS HISTORY To start up the real-time process history view at any time, simply navigate to Start > DeltaV > Operator > Process History View, and then open CATUnitOverview (if it does not open automatically). Chart scales can be compressed or expanded by clicking those buttons on the menu bar. Scales can be shifted up or down by click dragging on the scale of interest. STARTUP 1. Be certain the air to the sandbath is on. The rotameter should read ≈ 5 or higher and be held constant from run to run to provide consistent heating. 2. Enable main power to the sandbath heater by pushing the black-colored START button on the CAT unit’s panel board. 3. Set automatic temperature controller (TIC-10, for the sandbath) to desired initial set point, if the reactor is to be heated. Set mode to AUTO unless you know the exact output desired. (You can also turn auxiliary heater on (output ≤ 90%) if you desire a rapid heatup, but be sure to it turn off when the skin (heater) temperature nears your desired reactor T; the auxiliary heater is not controlled at this time and cooling (≈10 °C/hr) takes much longer than heating.) – Not typically recommended – See Instructor for guidance. Section: Operating Instructions The reactor temperature can lag the heater temperature significantly. You may want to change the selector switch for TIC-10 to the skin thermocouple (this tells the controller to use the skin T/C as its input), and raise the set point to compensate for the difference in temperature between skin and reactor temperature (anywhere from 2 to 5°C is typically, depending on the desired reactor temperature). You can also change the selector switch later. The highest temperature to be used in any experiment is 200oC. The DeltaVTM computer has been programmed to reduce the input heating power by half at 200oC and to shut off the heaters at 220oC. A "CRITICAL" warning light will alert you to the high temperature. Both the sandbath and reactor temperatures are monitored, and both are programmed to alarm and shut off the heaters. 9 4. Be sure the breaker for the sandbath main heater (#10) is on, as well as breakers for instrument power (#4), tape heater and GC (#5), and control (#7). 5. If the ammeter for the sandbath heaters is not showing current, press the START switch. Steps 1 through 5 can be rapidly performed the morning of the experiment. Then you can leave the experiment unattended with NO GAS FLOWS (unless otherwise negotiated with your instructor). The catalyst is stable in a no flow condition and, most typically, you can start the flows during the normal afternoon lab period. Special Item: If experimental guidance from your instructor calls for, or you believe you will need to and have obtained permission to introduce gas flows AND leave the unit unattended during the morning of an experiment, complete and post the Unattended Operations Placard provided in the Appendix of this manual, following the instructions given therein. Do NOT leave the reactor unattended AND at elevated pressure. Use 0 psig during the pretreating period. The temperatures can be monitored on the control schematic chart or – even better – using Process History View. The reactor effluent flow rate should be checked periodically. If problems (loss of flow, runaway temperature) develop, SHUT OFF BOTH REACTANT FLOWS USING APPROPRIATE SHUTOFF VALVES AND TURN ALL HEATERS OFF. LEAVE SANDBATH AIR ON. Should the experimental program suggest or call for the use of elevated pressure reaction conditions, this can be accomplished by loading the dome of the reactor back pressure regulator (V-8) with nitrogen pressure. Unless otherwise advised by your instructor, limit reactor pressures to 30 psig. (Note: Tests in the fall of 2009 indicated that the flow meters yield the same flow calibration at elevated pressure that they do at 0 psig.) SETTING FLOW RATES Set flow rates prior to turning on heaters on Day 1. Note that the main operating screen has a digital value designating whether the computer or the panel board will be used for this. Set this value to 1 for computer use. The H2/N2 mixture is set first, using mass flow controller FIC-301. FIC-301 is a linear response device, but only operates between roughly 20% and 100% of rated flow. To minimize uncertainty in rate and other calculations which may be based on this flow rate, a precise calibration of this meter is needed, relating raw input (Field Value % in DeltaVTM terminology) to the actual flow as measured by the soap bubble-o-meter. Section: Operating Instructions The reactor temperature will typically show some oscillation around the controller set point. This is normal. However, it is not normal for the sandbath T’s to greatly exceed the reactor temperature by more than 10 °C after a prolonged period. This condition may indicate a blocked filter on top of the sandbath. Remove and clean it; you can do this while the unit is running. 10 To build this calibration, a) Bypass the reactor; and b) Determine the flow rates over a range of field value, or output, percent by changing the controller’s valve opening in MAN (manual) mode. Wait a minute prior to taking readings to ensure that a steady flow is achieved. The results from this work can be used as a stand-alone calibration for FIC-301 and, if desired, the linear calibration coefficients can be entered into the DeltaVTM system to avoid the need for the stand-alone tool. If the calculated zero (0% of input) is far from zero flow and/or the calibrated span (100% of input) is far from the estimated maximum flow rate of this meter, there may be a problem with the instrument requiring the attention of the Lab Coordinator. When finished checking this flow meter calibration, re-route flow through the reactor. Check a flow rate again at a high output %. It should be close to your previous reading; if well below (say > 10%), there may be a serious leak in the reactor; notify the Lab Coordinator. Calibrate the ethylene mass flow controller, FIC-302, in a similar manner by bypassing the reactor again, and shutting off the N2/H2 flow. Be sure to close the shutoff valve just before the mixing tee. Set C2H4 flow rates using FIC-302 in MAN mode. This is also a linear response device, but it too only operates between roughly 20% and 100% of rated flow. Again use the bubble-o-meter to determine (using two or more points along the span) the calibration slope between flow rate and input (Field Value %), and then the calculated zero and span. To start a reaction, route (or re-route) the combined H2/N2 mixture and C2H4 flows to the reactor. If the chromatographic method in use is capable of total feed analysis, it is probably wise to take at least one sample with the feed through the bypass (a blank sample) to determine the feed composition. It's OK to change flow rates on the fly without bypassing the reactor. Again, if GC method allows for it, you can verify that your gas chromatograph is set-up properly by running the H2/N2 and C2H4 separately, and checking the results versus the composition noted on the cylinder labels. Special Item: LONGITUDINAL TEMPERATURE VARIATION As conversion of ethylene increases, even the small quantity of catalyst and the large amount of carborundum diluent present in the reactor may not prevent a longitudinal increase in temperature from occurring, resulting in a non-isothermal kinetic impact on reaction rate. If this is of consequence to your experimental program, the longitudinal temperature variation can be measured and its effect accounted for in post-run analysis. The following steps illustrate how to measure longitudinal temperature variation whilst carrying out the ethylene conversion reaction: Section: Operating Instructions N.B.: When either calibrating meters or checking exit flow rate during reaction – indeed, anytime the bubble-o-meter is being actively used to measure flow – the Agilent GC sample line should be blocked out. This must be done to avoid having gases from the GC contribute erroneously to measured flow rate through the bubble-o-meter. 11 1. With the reactor thermocouple bottomed-out to its full extent in the reactor centerline thermowell, collect sufficient data (using the method outlined later in this manual) to identify an average temperature from this thermocouple. Note: If a sandbath or skin couple is being used for control, the reactor couple can be moved with impunity. If, however, the reactor thermocouple is being used to control temperature, then moving the reactor couple will affect temperature control undesirably. To avoid this, temporarily place the temperature controller in MANual mode while performing this procedure, returning it to AUTO mode after completion of the procedure. 2. Lifting this thermocouple a small and known distance up, again collect sufficient data to identify an average temperature from this thermocouple. As the couple may be hot, use appropriate gloves or a cloth to protect your fingers from the lower portion of the couple (see Mr. Perkins for any needed materials). 3. Repeat step 2 until the thermocouple no longer shows a rise in reactor temperature. A chart of these average values will reveal the start and end of the catalyst bed temperature rise, as well as the quantitative data necessary to deal with longitudinal temperature variation resulting from this exothermic reaction. 4. After gathering all necessary data, return the reactor thermocouple to its bottomed-out position, being careful to guide the couple down without crimping it. As the couple may be hot, use appropriate gloves or a cloth to protect your fingers from the lower portion of the couple (see Mr. Perkins for any needed materials). CHROMATOGRAPH OPERATING INSTRUCTIONS (AGILENT 3000A MICRO GC) USING THE AGILENT 3000A 1. Check to make sure the carrier gases (i.e., helium and argon cylinders located behind the GC itself) are open to the GC and that pressure levels are sufficiently high (i.e., supply pressures of roughly 80-to-90 psig and cylinder pressures of 200+ psig). If the levels look low, contact the Lab Coordinator. 2. Check to ensure the direct connect tubing line to the GC is connected to the source to be analyzed; i.e., the standard gas cylinder if performing a calibration or the CAT unit itself if analyzing process gas. If this tubing line needs to be moved from one source to the other, contact the Lab Coordinator. 3. Log into a suitable UO Lab Control Room computer using your LSU ID and password. 4. Open the Remote Desktop Connection application. In the destination line, type the virtual computer’s name: che-uolab-gc3. Hit Enter on the keyboard. You will be asked to Section: Operating Instructions Use of the Agilent 3000A Micro GC requires the following: a) the GC is powered up and on, b) carrier gases are available and aligned, c) the gas source to be analyzed is aligned, and d) software to operate the GC and analyze results is accessible. Instructions for meeting these requirement are outlined below. 12 log in again, using your LSU ID and password. This will take you to the che-uolab-gc3 virtual computer desktop. 5. On the desktop of that virtual computer, click on the icon for MicroGC 3000 (online), as shown in the image below: 9. Performing an analysis: a. When you are ready to perform a run, click on Control (same top line as File) and then Single Run. A popup menu will appear. Section: Operating Instructions This will open the GC operating and analysis software application. 6. At this point you will see a navigation window, a top menu with icon buttons, and a grey primary screen. 7. Load the method “PH-CAT4”. To do so, click File → Open → Method and choose the method titled “PH-CAT4”. The method is now loaded onto the system and ready to be used. The method name should be displayed at the very top of the window. It may be loaded already; check the top line of the MicroGC program to see if it is already enabled. 8. Display the Instrument Status by clicking on Control → Instrument Status. It may take some time for the method settings to equilibrate the GC. You cannot perform a run until all parameters in Instrument Status are green (even though we are not using Channels C or D). 13 b. Your run will need both a Sample ID as well as a Result Name (group name or name and run number and date-time, for example). Once you have those lines complete, click Start in the popup window. Four plots will pop up (one for each detector). c. It may take a minute or two for the software to begin plotting. You will know that the run setup was successful by the changing of the status bar (bottom right-hand corner) from green to blue. Section: Operating Instructions 10. For the compounds being tested, your results will come from Channel A and Channel B. Hydrogen and nitrogen will be detected on channel A, while methane, ethane, and ethylene will be detected on channel B. 11. To view the report for your run, click Reports → View → Normalization. A report with the known compounds will pop up. 14 12. If prompted to save Method, click NO. Clicking Yes would create a new method file and could potentially change the CAT-PH settings. 13. When finished with the GC, log out of the Remote Desktop Connection application using the Start menu (do NOT exit via the X at the top of the blue Remote Desktop Connection bar). Logging off not using the Start button would leave you logged into che-uolab-gc3, preventing the other groups/instructors from logging in. 14. Leave the carrier gases on. 15. When completely finished (both the GC and DeltaV), log off of the main desktop via the Start button again. REACTOR SHUTDOWN The following steps constitute a safe and orderly shutdown of the Exxon Catalytic Reactor Unit: 1. Shutdown main power to the sandbath heater by pushing the red EMERG. STOP button on the CAT unit’s panel board. 2. Set automatic controller temperature to 0ºC (AUTO) or 0% output (MAN). 3. Set auxiliary heater output to 0%, if on. 4. Set C2H4 flow to 0%. 5. Shut off C2H4 block valve at the mixing tee. 6. Shut off C2H4 main cylinder valve. 7. Shut off air flow to fluidized bed. 8. Wait ~ 2 minutes; then shut off H2/N2 mixture flow to the reactor by setting flow to 0%. (Maintaining flow for those 2 minutes leaves a blanket of this mixture in the reactor, prolonging the life of the catalyst.) 9. Shut off H2/N2 mixture main cylinder valve. 11. All other valves should be left open. EMERGENCY HEATER SHUTDOWN If the reactor heater needs to be shut down under emergency conditions, press the red Emergency Stop (EMERG. STOP) button on the front panel of the unit IF CONDITIONS PERMIT THIS TO BE DONE SAFELY. Section: Operating Instructions 10. Shut off N2 cylinder main cylinder valve, if open; then, relieve pressure on the reactor back pressure regulator IF it has been set above 0 psig. 15 HISTORICAL DATA ACCESS USING DELTAV CONT INUOUS HISTORIAN A continuous historical record of all relevant temperatures, flows, and levels is kept by the DeltaV system on its hard disk(s). Selected portions of these data can be imported into an Excel spreadsheet for analysis. This spreadsheet must be saved to a flash drive or a personal directory if it is to be used on any other personal computer than a dedicated DeltaV workstation as files stored on the DeltaV network are not accessible from the general LSU network. To import data from DeltaV history to Excel, Emerson has provided an Excel add-in called the DeltaV Continuous Historian. It appears under the Add-Ins menu in Excel 2007 when this program is opened on a DeltaV workstation. Any process variable that is enabled in History Collection can be imported. Most of these variables are collected every 10 seconds, 30 seconds, or 1 minute, so it doesn't make sense to try to read the data any faster. Though the menu features of the DeltaV Continuous Historian can be used for their intended purpose, requesting an ad hoc retrieval of data for more than one or two tags is tedious and time-consuming. To avoid this labor, an Excel 2007 template file is available for TDU data retrieval. This template is preloaded to request ALL the historically trended TDU tags. The only information that the user must supply is the starting date and time, the ending date and time, and the time interval between data values. To import data, open Excel 2010 on the DeltaV workstation and follow these steps: 1. Put Excel calculations in Manual mode before opening the template or attempting to change the starting and ending dates. 2. Open the data retrieval template file for CAT. It is located on the Desktop, in the DeltaV_Excel_Data_Collector_Files folder under the name CAT_Data_Retrieval.xlsx For example, data starting October 5th, 2011 at 9AM would be entered as 10/5/2011 9:00 4. Enter (or modify) the ending date and time for the data of interest as mm/dd/yyyy hh:mm into cell A6. For example, data ending October 5th, 2011 at 3PM would be entered as 10/5/2011 3:00PM or 10/5/2011 15:00 5. Enter (or modify) the desired time interval between data values. Section: Operating Instructions 3. Enter (or modify) the starting date and time for the data of interest as mm/dd/yyyy hh:mm into cell A4. 16 For example, to request data values every 10 seconds, enter =”10seconds” into cell A8. For data values every 2 minutes, enter =”2minutes” into cell A8. Any value of seconds, minutes, or hours may be used, but some values make more sense than others. Using values faster than the fastest data originally collected makes no sense. So, values below 10 seconds only make sense if data collection on some variables has been set to time intervals below 10 seconds. Data will be interpolated where there are no values. A maximum of 2161 readings (2160 intervals) are available in any one data retrieval operation. This will allow the retrieval of 6 hours of 10 second data, 36 hours of 1 minute data, or any other combination that results in 2160 (or fewer) intervals. If more data are required, two separate requests must be made and concatenated manually. Of course, if only a few tags are needed, one can bypass the use of the template and retrieve data ad hoc using DeltaV features. 6. After entering the last of these user inputs, click the Add-Ins menu tab, click the DeltaV drop-down tab, roll over the Continuous Historian option, and select Refresh. All values in the spreadsheet should update, signaling successful retrieval of data. Select the entire worksheet (by clicking in the upper left hand corner of the worksheet adjacent to the A1 cell. Copy this selection (e.g., ctrl-c). Select the 2nd worksheet tab. Using Paste Values, paste the values copied from the Data Collection tab into the 2nd worksheet. These values are now static and will be available even when the spreadsheet is relocated to another computer. The values in the first tab can be deleted at this point IF the file will not be used to retrieve additional data from the DeltaV. 8. Save this file with an appropriate name to a device or drive separate from the DeltaV for use at a later time. (Remember that calculation mode for that file is Manual.) 9. The units for the data are the same as in DeltaV. If the word shutdown appears instead, then either: History collection was not enabled for that variable, or No history collection because the system was shut down, or No history collection because the variable did not change more than its deviation value set in history collection (i.e., nothing's changing!) PROCESS CONTROL BACKGROUND AND THE DELTAV SYSTEM The control of reaction processes is critical in the chemical, polymer, pharmaceutical, electronics, and food processing industries. The control problems can be difficult due to the different demands the reaction imposes on the process – rapid startups and shutdowns, Section: Process Control Background and the DeltaV System 7. To save these results, these data must be turned into static values, as follows: 17 possible runaway temperature, narrow operating range, etc. Also, it may be difficult to determine optimal control parameters for any control loop, because different operations may require markedly different control strategies. Equally important is the setting of operator alarms and interlocks (an interlock is a logical statement which, when true, locks out the operation of specified equipment). The logic involved in such operations can be complex even for simple units. For the catalytic reactor, you may also be assigned a study of its control/operations programming in order to make improvements to the control system. The control platform is DeltaVTM, an Emerson Process Management system that is compatible with several ISA Instrument Society of America) protocols. It is a modular system; i.e., modules built in a program called “Control Studio” perform different types of tasks. The main programs and the modules that have been constructed are listed below: Control Studio – creates control loop and sequential function chart modules. A sequential function chart (SFC) is a set of commands for performing a subset of plant operations in a sequence. Operate (Configure) – creates process and instrumentation diagrams (P&IDs), with special links for operation of equipment and control loops Operate (Run) – general plant operations – setting basic control loop parameters, starting sequential function charts, directly operating equipment from the P&IDs. Process History View – examines historical data graphically, if the system has been programmed to record such data. The operation of Operate (Run) and Process History View will be demonstrated by the instructor. The other three programs are for configuration only; you will not need them to operate the system and they are not available on the computer near the experiment. Instructions for retrieving data from the control system into Excel are given in “retrievedata.txt” (in d:\uolab). There are extensive on-line instructional manuals and help files (right click usually gives you a choice for help) to aid in understanding how to use DeltaVTM. Basically, if you are familiar with VBA, or any other object-oriented programming language, then you already understand a lot about Control Studio. If you are familiar with Windows-based drawing programs, you could pick up Operate (Configure) easily. If you are not familiar with these, you may consider going back to the 19th century. The only schematic now provided is CAT1. Start the program DeltaVTM Operate. Open the file CAT1. Now let’s work with CAT1. You see icons that can be operated in the schematic. For example, you can click on a faceplate and get the temperature controller TIC-10. This is a PID controller manipulating the main heater output (OUT) in order to control reactor T. The reactor T is the measured (process) variable, PV. However, you can click on the input TSELEC and change the PV that the controller uses. Leaving this loop in manual, you can increase the output of the heater from 0 to 100% (use the left control). You see the measured variable in the right indicator. There is also an input box for the set point (SP) – the desired process variable value Section: Process Control Background and the DeltaV System Explorer - configure the system, including the I/O. 18 when the controller is in AUTO. Set OUT back to 0%; change to AUTO; and try a SP of 80°C. You should see OUT gradually increase. Clicking on the bottom left icon gives you the module detail which contains parameters you would need to tune a controller. Such tuning can be done either here or in Control Studio. However, Control Studio can only be opened with write access on the DV1 console (or the DeltaV server itself). Now examine a control module in Control Studio on a console with write access. Open the DeltaVTM Explorer, select module TIC-10 in area CAT, and open it in Control Studio. This module contains the temperature controller, the analog output devices (heaters) and analog input devices (thermocouples). This will allow you to operate this subsystem from Control Studio. Click on menu item “VIEW” then “ONLINE”. You will see numbers appear on the screen – you are operating the entire module in real time. Click on the PID1 box. This also gives you access to the controller. Set “OUT” to 30% using the “Parameter List” in the bottom left corner. This will set the main heater output to 30%, if some conditions are met. Following the arrows (logic flow), determine what these conditions are (click in the appropriate boxes; then click on “Expressions”). Conditional expressions are logical statement(s) which must be satisfied before the program performs a subsequent action. Expressions evaluate as 0 = false, 1 = true. This completes a brief tour of DeltaVTM. In order to complete your assignment you will have to learn how to use Operate (Run) and Process History View. If problems, you can consult the Books Online and the extensive online Help files. However, any direct question on a particular topic will be answered. Some additional information on process control background is provided below; if you have already completed a process control lab, you can probably skip this information. Equipment specifications are in the figures below. The reaction system is equipped with thermocouples, gauge pressure transmitters, thyristors (also called SCRs, these are power controllers) for the heaters, electronic flow control valves, and a motor controller for the recycle pump. The following is some background on the operation of such instruments and their control. For more detailed descriptions you can read Ch. 8 and 9 of Process Dynamics and Control by Seborg et al. or Ch. 5 of Smith & Corripio, Principles and Practice of Automatic Process Control, or a similar book on process instrumentation and control. Sensor/Transmitter: An example differential pressure transmitter is the one installed to measure the level in the reactor. The transmitter sends an electric signal in the range of 4 to 20 mA (converted to 1-5 V by a resistor) to the controller; this signal is proportional to the level in the tank. The controller converts this signal into a percent of range reading according to the following formula: PV 100 (h h0 ) hs (10) Section: Process Control Background and the DeltaV System Similarly, open FIC-301B in Control Studio. This contains the transmitter and controller for N2/H2 flow. Locate where the zero and span for the transmitter output would be set. Examine which alarms have been configured, and how they have been configured. 19 where PV is the process variable (level) in percent of transmitter range, h is the level, h0 is the zero of the transmitter range (lowest level that it can measure), hs is the span of the transmitter range, that is, the difference between the maximum measurable level and the zero level. The DeltaVTM controller can then convert this reading into any engineering unit desired – this is done in Control Studio. Power Controllers and Control Valves: The controller outputs (AOs) are transmitted as 4 to 20 mA corresponding to 0-100% range. When a control valve is actuated by air pressure, a currentto-pressure (I/P) transducer is used to convert the current signal into air pressure (3-15 psig range). The control valve can be air-to-open (fails closed), i.e., it opens as the controller output signal increases; it can also be air-to-close, which is just the opposite. The flow through a control valve is usually measured by a d/p transmitter (with orifice), a turbine meter, a Coriolis (spiral tube) flow meter, or the resistance of a heated metal filament (anemometer). In this experiment, electronic controllers with anemometer flow sensors are used. An important characteristic of a feedback controller is the direction of its action, i.e., which way should it change its output when the measured variable rises. If an increase in the output signal opens a valve on the inlet flow to a tank, which way should the controller output to the valve actuator move (increase or decrease) when the tank level (measured variable) rises? If your answer is "increase" we would have a direct acting controller, and if your answer is "decrease", it would be a reverse acting controller. Which action is correct for the controller in TIC-10? Why? The controller calculates the error or difference between the SP and the PV: E SP PV (11) The ideal PID controller output signal is given by: OUT OUT0 Kc [ E 1 dE Edt Td ] Ti dt (12) where the controller output in the range of 0 to 100%, OUT0 is the initial (time zero) output, and KC, Ti, and Td, are, respectively, the proportional gain, reset (or integral) time, and rate (or derivative) time. These adjustable parameters can be set on the controller using the module, either in Control Studio or Operator Interface. Real PID controllers are more complex, but the above is a useful approximate model. The three terms in the bracket of the equation represent the proportional (P), integral (I), and derivative (D) modes of the feedback controller (PID). As you can see, the proportional term makes the controller output proportional to the error and acts immediately upon a change in the process variable or in the set point. The second term, integral, is slower because it acts over time; its purpose is to force the error to zero. The derivative term acts on the rate of change of Section: Process Control Background and the DeltaV System Feedback Controllers: These regulate a measured process variable (e.g., temperature) by adjusting an OUT signal to vary a manipulated variable (e.g., power output). Control is accomplished by adjusting the SP of the controller (e.g., TIC-10 in Operator Interface), and by changing the mode of the controller to AUTO. 20 the error; it is the fastest of the three terms. Flow rate controllers usually do not require the derivative term, so Td may be set to zero for these. TUNING CONTROLLERS Figure 3. Controller tuning map for tuning of PI controllers. All subplots are the response of PV versus time after step change in set point. (Practical Process Control by Cooper, 2004) There are two general approaches to controller tuning: trial-and-error and model-based (an approach covered extensively in ChE 4198 – Process Control). Because flow rate dynamics are relatively fast, a trial-and-error method is preferable in establishing optimal Kc and Ti for flow rate loops. But for the more complex loops controlling temperature or pressure, model-based approaches are preferred. A standard model for the open-loop (controller in MANUAL mode) Section: Process Control Background and the DeltaV System Consider a simple flow controller. The position of a flow control valve or a variable pump speed is the manipulated or controller output, and the flow rate is the measured or process variable. The adjustment of the controller parameters to obtain a fast stable response of the variable being variable is known as tuning the controller. Figure 3 (Controller Tuning Map) shows, qualitatively, criteria for evaluating controller performance. The “detail” screen of the module faceplate in Operator Interface would be used to enter candidate Kc, Ti and Td values, as well as other controller parameters. 21 dynamics of many systems is first-order-plus-dead-time, or FOPDT. The open-loop dynamics are assumed to follow the form (when put into dimensionless form): dY Y K X H (t ) dt (13) The measured variable is Y (dimensionless, measured as fraction of span), the output or disturbance variable is X (dimensionless, same units) and the H-operator is the unit step function, which models the pure time delay of the process. The adjustable parameters are the steady-state gain (K), the system time constant (), and the time delay (). Normally, a test of the system in open-loop is performed as follows. The system is allowed to come to a steady state, then a disturbance is introduced (X = M). A plot of Y vs. t is constructed – known as a process reaction curve – and from this, values of K, , and are estimated, from which candidate Kc, Ti and Td values of the controller parameters can then be generated. Estimating the process characterization parameters (K, , and ) can be as simple as using graphical interpretation of the resulting process reaction curve (e.g., the Cohen-Coon open-loop method[6]) or as complex as performing an optimization procedure to minimize the error between the entire set of process reaction curve data and the presumed first-order-plus-deadtime model[7]. Candidate tuning parameters are then estimated using an appropriate choice from the myriad methods available, such as Direct Synthesis (DS)[8], Internal Model Control (IMC)[9], and Cohen-Coon (CC)[6]. [1] O. Beeck, "Hydrogenation catalysts," Discussions of the Faraday Society, vol. 8, pp. 118128, 1950. [2] J. B. Butt, "Progress toward the a priori determination of catalytic properties," AIChE Journal, vol. 22, pp. 1-26, 1976. [3] J. B. Peri, "Infra-red studies of carbon monoxide and hydrocarbons adsorbed on silicasupported nickel," Discussions of the Faraday Society, vol. 41, pp. 121-134, 1966. [4] H. S. Fogler, Elements of chemical reaction engineering, 4th ed. Upper Saddle River, NJ: Prentice Hall PTR, 2006. [5] (2005, August 18, 2010). Lees' Loss Prevention in the Process Industries: Hazard Identification, Assessment and Control (3rd ed.). [6] J. A. Romagnoli and A. Palazoglu, Introduction to Process Control. Boca Raton: Taylor & Francis, 2006. [7] D. J. Cooper. (2008, August 18). Modeling Gravity Drained Tanks Data Using Software. Section: References REFERENCES 22 [8] D. A. Mellencamp, et al. (August 18). Process Dynamics and Control, Chapter 12 slides. Available: http://www.che.utexas.edu/course/che360/lecture_notes/chapter_12.ppt [9] D. A. Cooper. (2006, August 18). PID Control of the Heat Exchanger. APPENDIX Section: Appendix Unattended Operations Placard (See next page.) 23 Exxon Catalytic Reactor Unattended Operations Unattended Use1 by: ______________ between _______ and2 _______ on __________ Hydrogen/Nitrogen Flow Rate (cc/min): _______________________________________ Ethylene Flow Rate (cc/min): ________________________________________________ Heating to Reactor Temperature, °C ____________ Pressure, psig __________________ Email Notification to Lab Coordinator (pdoring@lsu.edu) at: _________ (clock time) If necessary, hydrogen/nitrogen flow can be stopped immediately by blocking out the supply cylinder behind the unit. It is the far right cylinder as you face the back of the unit. Simply turn the main cylinder valve clockwise until closed. 1 2 If necessary, ethylene flow can be stopped immediately by blocking out the supply cylinder at floor level on the unit skid’s left side. It is the only cylinder at that location. Simply turn the main cylinder valve clockwise until closed. Hang this form on the reactor sandbath after completing accurately with needed information. Make sure that the exit gas valve is directed to vent and NOT the bubble meter. Here, post the times between which the unit will be unattended by you, and the date.
© Copyright 2024