Patient Simulator MiniSim 1000 Basic D

MiniSim 1000 Basic
The MiniSim 1000, Arrhythmia Simulator is a
powerful, comprehensive but basic patient simulator in
a compact case. It is designed for testing the
performance of basic patient monitoring instrumentation
quickly and easily.
The microprocessor-based instrument is menu driven via
its tactile feel keypad and easy to follow LCD display.
The MiniSim 1000 provides full 12 lead ECG
simulation with 14 user selectable rates from 30 to 350
BPM and 14 user selectable amplitudes from 0.15 to 5
mV. Sine, Square, and Triangle performance
waveforms are simulated with 14 user selectable rates
from 0.1 to 100 Hz.
The MiniSim also simulates 45 different arrhythmias
including Atrial, ventricular, tall T wave, ST elevation,
ST depression, myocardial infraction, blocks, pulse, and
pacer waveforms.
A 9-volt alkaline battery with a typical life of 40 continuous hours powers the MiniSim 1000 multiparameter patient simulator.
Operating Features:
Features:
• Easy to Use
• Small, Compact Size
• Menu driven operation
• Large LCD Display
• Tactile Keypad
Testing Features:
• 12 Lead ECG
• 45 Arrhythmias
• ECG Performance waveforms
Specifications:
„
„
ECG Normal Sinus
Rate: 30, 60, 70, 80, 90, 100,120, 150, 180, 210, 240, 270, 300, 350 BPM
Amplitude: 0.15 to 5mV, High Level Output: 0.5 Volt/mV of Low Level (Lead II)
Test Waveforms
Sine, Triangle, Square Rate: 0.1 to 100 Hz
0.1 to 1 Hz in 0.1 Hz increments
1 to 10 Hz in 1 Hz increments, 10 to 100 Hz in 10 Hz increments
110 Toledo Street, Farmingdale, NY 11735, Phone: 631-531-0100, Fax: 631-531-0101
www.Netech.org
1-800-547-6557 (US)
Defibrillator / Pacemaker
Patient Simulator
MiniSim 1000
The MiniSim 1000, Arrhythmia Simulator is a powerful,
comprehensive patient simulator in a compact case. It is
designed for testing the performance of patient
monitoring instrumentation quickly and easily.
The microprocessor based instrument is menu driven via its
tactile feel keypad and easy to follow LCD display. The
MiniSim 1000 provides full 12 lead ECG simulation with 14
user selectable rates from 30 to 350 BPM and 14 user
selectable amplitudes from 0.15 to 5 mV. Sine, Square, and
Triangle performance waveforms are simulated with 14 user
selectable rates from 0.1 to 100 Hz.
The MiniSim also simulates 45 different arrhythmias including
Atrial, ventricular, tall T wave, ST elevation, ST depression,
myocardial infraction, blocks, pulse, and pacer waveforms.
In addition, two invasive blood pressures are simulated with choice of 12 static pressures and 6 dynamic
waveforms. Respiration is simulated with 4 rates. 4 base line impedance, and 4 delta impedance selections.
Apnea may be off, continuous, and with either 12 or 32 second apnea bradycardia. Finally, temperature is
available at either 25 or 37 degrees Centigrade. A 9-volt alkaline battery with a typical life of 40
continuous hours powers the MiniSim 1000 multi-parameter patient simulator. The instrument provides
unmatched choices of parameter simulation in a truly compact instrument.
Features:
Operating Features:
Testing Features:
• 12 Lead ECG
• 45 Arrhythmias
• ECG Performance waveforms
• 2 Blood Pressure
• Respiration / Apnea
• Temperature
• Easy to Use
• Small, Compact Size
• Menu driven operation
• Large LCD Display
• Tactile Keypad
Specifications:
„
„
ECG Normal Sinus
Rate: 30, 60, 70, 80, 90, 100,120, 150, 180, 210, 240, 270, 300, 350 BPM
Amplitude: 0.15 to 5mV, High Level Output: 0.5 Volt/mV of Low Level (Lead II)
Test Waveforms
Sine, Triangle, Square Rate: 0.1 to 100 Hz
0.1 to 1 Hz in 0.1 Hz increments
1 to 10 Hz in 1 Hz increments, 10 to 100 Hz in 10 Hz increments
110 Toledo Street, Farmingdale, NY 11735, Phone: 631-531-0100, Fax: 631-531-0101
www.Netech.org
1-800-547-6557 (US)
MiniSim 1000 Page 1 of 2
Defibrillator / Pacemaker Analtzer
Patient Simulator
MiniSim 1000
Specifications:
„
„
„
„
„
„
„
Tall T Wave
Test Waveform@ 70 BPM, Atrial Pacer, Ventricular Pacer*, Paced Rhythms - Variable
from 0.1to 2mS, R Wave detection - Variable form 10 to 120 ms
Pulse: 2mS Pulse of 1mV amp @ 4 sec intervals
Arrhythmias
ST Segment Analysis, ST Elevation and Depression, ST Depression Positive and
Negative Slope, Myocardial Infarction, Atrial, Ventricular, Blocks
Respiration
Normal Physiological Simulation: Base Line Impedance: 250, 500, 750, 100 Ohms
Delta Impedance: 0.1, 0.5, 1.0, 1.5 Ohms, Rates: 15, 30, 60, 120 BPM
Apnea: Off, Continuous, 12 Seconds and 32 Seconds Apnea-Bradycardia
Accuracy: Delta Impedance 5% of selection, Rate: 1% of selection
Blood Pressure: Impedance: 350 Ohms, E x c i t a t i o n : 2 to 16 Volts
Sensitivity: 5uV/V/mmHg
Static: 0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, 300 mmHg.
Dynamic: 50/10, 60/20, 70/30, 80/40, 100/60, 120/80 mmHg.
Accuracy: (±2% FS. ±1 mmHg
Synchronized with all normal sinus rates
Physiologically track all arrhythmia selections.
Temperature
Compatibility: YSI 400/700, 25/37 Degree Centigrade
Accuracy: ±1% of reading
General
Display: 16 Character 2 Line LCD, Key Pad: 8 tactile keys Ð F1, F2, F3, F4, >, <, ON/
OFF, Menu, Output Connectors ECG: 10 Banana Jacks
High Level: 3.5mm Phono Jack, Pressure: 5 contact circular Din
Temperature: 3.5mm Stereo Jack
Size: 5½ x 3½ x 1 ½ inches (13.97 x 8.9 x 3.8 cm)
Case: High Impact Plastic
Weight: 10 Oz (0.28 kg)
Power Requirements: 9V Alkaline Battery. Optional AC Adapter
Environmental:
Operating Temperature: 59° to 95° F (15° to 35° C)
Storage Temperature: 32° to 131° F (0° to 55° C)
Ordering Information
Part #
300
302
303
304
Description
------- M i n i S i m 1000, Arrhythmia Simulator
------- A C Adapter
------- P r e s s u r e Cable (Unterminated)
------- T e m p e r a t u r e Cable
For all other Accessories Please call 1-800-547-6557 or E-mail sales@netech.org
110 Toledo Street, Farmingdale, NY 11735, Phone: 631-531-0100, Fax: 631-531-0101
www.Netech.org
1-800-547-6557 (US)
MiniSim 1000 Page2 of 2
Defibrillator
ibrillator / Pacemaker Analtzer
Patient Simulator
MiniSim - EEG
EEG SIMULATOR
The MiniSim 330, EEG Simulator is designed to test
EEG instruments including recorders and Sleep Study
Monitors. the compact, microcontroller based
instrument has five separate floating outputs and
simulates Alpha, Beta Rhythm (ABR), Sine, Square
and Triangle waveforms with selectable frequencies
and amplitudes.
Patented EEG
cup electrode
Connector
The easy to use MiniSim 330 is menu operated via 8
tactile feel keys and a large LCD display. High
accuracy is provided by its advanced microprocessor
with a crystal oscillator and precision voltage
reference.
Features:
The MiniSim 330 EEG Simulator is a unique test
instrument that provides quick and easy performance
testing of all EEG instrumentation.
Patented EEG cup electrode Connector
5 Channel Indipendent Outputs.
Simulates ABr, SIN, Triangle
Clean Microvolt Signals
Selectable Amplitudes and
Frequencies
Large LCD Display
Compact, Rugged Case
Excellent Value.
Battery or AC Operation
Ordering Information
Part #
Description
330 ---- MiniSim EEG Simulator
301 ---- Carrying Case
302 ---- AC Adapter
Specifications:
Waveforms
ABR Waveforms at 1 kHz, 0.64 µV.
Spike waveform, Sine, Square, and Triangle:
Frequencies: 0.1, 0.5, 2, 50, and 60 Hz.
Amplitude: 10, 30, 50, 100, 500 µV, 1,2, and 2.5 mV
Key Pad:8 tactile membrane Keypad
Display:2 line 16 character LCD Display
Case: High Impact Plastic
Environmental
Operating Range:
59° to 95° F (15° to 35° C)
Power Requirements
9V Alkaline Battery.
Storage Range:
32° to 122° F (0° to 55° C)
Physical Dimensions
110 Toledo Street, Farmingdale, NY 11735, Phone: 631-531-0100, Fax: 631-531-0101
www.Netech.org
1-800-547-6557 (US)
Defibrillator / Pacemaker Analtzer